QUANTEX C3
Device Facts
| Record ID | K962199 |
|---|---|
| Device Name | QUANTEX C3 |
| Applicant | Instrumentation Laboratory CO |
| Product Code | CZW · Immunology |
| Decision Date | Sep 25, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.5240 |
| Device Class | Class 2 |
Indications for Use
This product permits the quantitative in vitro diagnostic determination of Complement 3 in serum and plasma on the ILab Clinical Chemistry System by turbidimetric immunoassay method. C3 is one of a group of serum proteins which destroy infectious agents. Measurement of this protein aids in the diagnosis of immunologic disorders, especially those associated with deficiencies of C3.
Device Story
Turbidimetric immunoassay for quantitative C3 determination in serum/plasma; utilizes anti-human C3 reagent and buffer on ILab Clinical Chemistry System. Used in clinical laboratory settings by trained personnel. Output is quantitative C3 concentration; aids clinicians in diagnosing immunologic disorders and C3 deficiencies.
Clinical Evidence
Bench-only; method comparison study of 50 serum samples comparing quantex C3 on ILab system vs. IL Test™ C3 on Monarch system. Correlation coefficient (r) = 0.9842.
Technological Characteristics
Turbidimetric immunoassay; reagents include anti-human C3 and buffer. Designed for use on ILab Clinical Chemistry System. No specific materials, energy sources, or software algorithms described.
Indications for Use
Indicated for quantitative in vitro diagnostic determination of Complement 3 (C3) in serum and plasma to aid in the diagnosis of immunologic disorders associated with C3 deficiencies.
Regulatory Classification
Identification
A complement components immunological test system is a device that consists of the reagents used to measure by immunochemical techniques complement components C1q , C1r , C1s , C2 , C3 , C4 , C5 , C6 , C7 , C8 , and C9 , in serum, other body fluids, and tissues. Complement is a group of serum proteins which destroy infectious agents. Measurements of these proteins aids in the diagnosis of immunologic disorders, especially those associated with deficiencies of complement components.
Predicate Devices
- IL Test™ C3 (K920007/B)
Related Devices
- K962198 — QUANTEX C4 · Instrumentation Laboratory CO · Sep 26, 1996
- K964298 — N-ASSAY TIA MULTI V-NL · Crestat Diagnostics, Inc. · Jul 14, 1997
- K983441 — C3 · Abbott Laboratories · Nov 4, 1998
- K012361 — TINA-QUANT COMPLEMENT C3C TEST SYSTEM · Roche Diagnostics Corp. · Nov 29, 2001
- K993437 — K-ASSAY C3 · Kamiya Biomedical Co. · Nov 24, 1999