QUANTEX IGA

K962200 · Instrumentation Laboratory CO · CZP · Sep 25, 1996 · Immunology

Device Facts

Record IDK962200
Device NameQUANTEX IGA
ApplicantInstrumentation Laboratory CO
Product CodeCZP · Immunology
Decision DateSep 25, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5510
Device ClassClass 2

Indications for Use

This product permits the quantitative in vitro diagnostic determination of Immunoglobulin A in serum and plasma on the ILab Clinical Chemistry System by turbidimetric immunoassay method. Measurement of IgA aids in the diagnosis of abnormal protein metabolism and the body’s lack of ability to resist infectious agents.

Device Story

Turbidimetric immunoassay for quantitative IgA determination in serum/plasma; utilizes ILab Clinical Chemistry System; reagent kit includes anti-human IgA and buffer; used in clinical laboratory settings by trained personnel; provides quantitative IgA concentration values; aids clinicians in assessing protein metabolism and immune function.

Clinical Evidence

Bench testing only. Method comparison study evaluated 45 serum samples comparing quantex IgA on ILab system vs. IL Test™ IgA on Monarch system. Correlation coefficient (r) = 0.9905.

Technological Characteristics

Turbidimetric immunoassay; reagents include anti-human IgA and buffer; designed for use on ILab Clinical Chemistry System; quantitative measurement principle.

Indications for Use

Indicated for quantitative in vitro diagnostic determination of Immunoglobulin A in serum and plasma to aid in diagnosis of abnormal protein metabolism and immune resistance deficiencies.

Regulatory Classification

Identification

An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} # SECTION 3 ## quantex IgA - 510(k) SUMMARY ### (Summary of Safety and Effectiveness) **Submitted by:** Betty Lane Director, Regulatory Affairs Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, MA 02173 phone (617) 861-4182 fax (617) 861-4464 *K962200* **SEP 26 1996** **Contact Person:** Betty Lane phone (617) 861-4182 **Summary Prepared:** May 29, 1996 **Name of the device:** quantex IgA: 8 x 6 mL anti-human IgA P/N 3000-22133 2 x 100 mL Buffer P/N 3000-22130 **Classification name(s):** 866.5510 Immunoglobulins A, G, M, D and E immunological test systems Class II **Identification of predicate device(s):** IL Test™ IgA #K920014/B **Description of the device/intended use(s):** This product permits the quantitative *in vitro* diagnostic determination of Immunoglobulin A in serum and plasma on the ILab Clinical Chemistry System by turbidimetric immunoassay method. Measurement of IgA aids in the diagnosis of abnormal protein metabolism and the body’s lack of ability to resist infectious agents. **Statement of how the Technological Characteristics of the Device compare to the Predicate device:** Biokit’s quantex IgA is substantially equivalent in performance, intended use, safety and effectiveness to the predicate, IL Test™ IgA -- 510(k) #K920014/B. **Summary of Performance Data:** A comparative performance study was performed with IL Test™ IgA on a Monarch Clinical Chemistry System and Biokit’s quantex IgA on an ILab Clinical Chemistry System. Method comparison studies evaluated IgA levels of forty-five serum samples on the respective instruments. The correlation $(r)$ in these studies was 0.9905. Section 3 quantex IgA 510(k) Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...