N Antisera to Human Immunoglobulins (IgG, IgA, and IgM)
Applicant
Siemens Healthcare Diagnostics Products GmbH
Product Code
CFN · Immunology
Decision Date
Dec 13, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 866.5510
Device Class
Class 2
Indications for Use
In-vitro diagnostic reagents for the quantitative determination of immunoglobulins (IgG, IgA and IgM) in human serum, heparinized and EDTA plasma, and IgG in human urine and cerebrospinal fluid (CSF) by means of immunonephelometry on the BN II and BN ProSpec® System. Measurements of IgG aid in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.
Device Story
In-vitro diagnostic reagent kit for quantitative immunonephelometric analysis of immunoglobulins. Input: human serum, heparinized/EDTA plasma, urine, or CSF samples. Principle: proteins in sample form immune complexes with rabbit-derived anti-human IgG antibodies; complexes scatter light beam; scattered light intensity proportional to protein concentration. Output: quantitative concentration result compared against known standard. Used in clinical laboratories; operated by trained technicians on BN II or BN ProSpec systems. Provides clinicians with quantitative data to aid diagnosis of abnormal protein metabolism and immune system function. Modification adds High Dose Hook (HDH) effect claim for CSF samples (up to 1130 mg/L) to ensure accurate reporting at high concentrations.
Clinical Evidence
Bench testing only. Verification and validation activities performed based on risk analysis to confirm the high dose hook effect claim for CSF samples. No clinical data required for this labeling modification.
Technological Characteristics
Liquid reagent containing rabbit-derived anti-human IgG antibodies and sodium azide preservative. Principle: immunonephelometry. Analyzers: BN II and BN ProSpec systems. Traceability: ERM-DA470k/IFCC. Storage: 2-8°C. No changes to formulation, instrument parameters, or fundamental technology.
Indications for Use
Indicated for the qualitative or quantitative determination of human immunoglobulins (IgG, IgA, and IgM) in human serum, plasma, or cerebrospinal fluid (CSF) for the assessment of immune status or disease states.
Regulatory Classification
Identification
An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.
Predicate Devices
N Antisera to Human Immunoglobulins (IgG, IgA, and IgM) (K233663 - previously cleared version)
Related Devices
K042735 — N ANTISERA TO HUMAN IMMUNOGLOBULINS (IGG, IGA AND IGM) · Dade Behring, Inc. · Feb 15, 2005
K963868 — IMMAGE IMMUNOCHEMISTRY SYSTEM IMMUNOGLOBULIN G (IGG), A (IGA) AND M (IGM) · Beckman Instruments, Inc. · Feb 10, 1997
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# SPECIAL 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information:
A 510(k) Number
K233663
B Applicant
Siemens Healthcare Diagnostics Products GmbH
C Proprietary and Established Names
N Antisera to Human Immunoglobulins (IgG, IgA, and IgM)
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| CFN | Class II | 21 CFR 866.5510 - Immunoglobulins A, G, M, D, And E Immunological Test System | IM - Immunology |
## II Review Summary:
This 510(k) submission contains information/data on modifications made to the submitter's own CLASS II device requiring 510(k). The following items are present and acceptable.
1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.)
2. Submitter's statement that the INDICATIONS FOR USE/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
This change was for a labeling modification to include a new high dose hook effect claim for N Antiserum to Human Immunoglobulins IgG for cerebrospinal fluid (CSF) samples.
4. Comparison Information (i.e., similarities and differences) to the submitter's legally marketed predicate device including, labeling, intended use, and physical characteristics.
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis.
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied.
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
K233663 - Page 2 of 2
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.