BECKMAN COULTER IMMAGE IMMUNOCHEMISTRY SYSTEM LOW CONCENTRATION IMMUNOGLOBULIN, BECKMAN COULTER

K130122 · Beckman Coulter, Inc. · CFN · Jan 7, 2014 · Immunology

Device Facts

Record IDK130122
Device NameBECKMAN COULTER IMMAGE IMMUNOCHEMISTRY SYSTEM LOW CONCENTRATION IMMUNOGLOBULIN, BECKMAN COULTER
ApplicantBeckman Coulter, Inc.
Product CodeCFN · Immunology
Decision DateJan 7, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5510
Device ClassClass 2

Indications for Use

IMMAGE® Immunochemistry Systems IGMLC Reagent, when used in conjunction with IMMAGE® Immunochemistry Systems and Cerebrospinal Fluid Protein Calibrator, is intended for quantitative determination of immunoglobulin M in human serum or cerebrospinal fluid (CSF) by rate nephelometry.

Device Story

IMMAGE Immunochemistry System uses rate nephelometry to measure human IgM concentration in serum and CSF. System automatically dilutes and delivers samples and reagents to reaction cuvettes. Particle-bound anti-IgM antibodies bind to IgM, forming insoluble complexes that increase light scatter. 670 nm laser light source; detector at 90° angle measures scatter intensity increase. System calculates rate of change of scatter signal, which is directly proportional to IgM concentration. Used in clinical laboratories by trained personnel. Output provides quantitative IgM levels to assist clinicians in diagnosing CNS conditions related to blood-CSF barrier permeability and intrathecal synthesis.

Clinical Evidence

Bench testing only. Precision evaluated per CLSI EP5-A2 (n=80 per sample pool); total %CV ranged 2.85-15.64% for CSF and 2.85-10.84% for serum. Linearity demonstrated per CLSI EP06-A (r² ≥ 0.997). Method comparison (n=80) against predicate showed high correlation (R² ≥ 0.994 for CSF, R² ≥ 0.995 for serum). LoB, LoD, and LoQ established per CLSI EP17-A.

Technological Characteristics

Rate nephelometry; 670 nm laser light source; 90-degree detector angle. Automated random-access analyzer. Reagent formulation includes particle-bound anti-IgM antibodies. Traceable to ERM-DA-470(k)-IFCC standard. Analytical range: 0.15-10 mg/L (CSF) and 32.4-2160 mg/L (Serum).

Indications for Use

Indicated for quantitative determination of Immunoglobulin M (IgM) in human serum or cerebrospinal fluid (CSF) for clinical laboratory use. Prescription use only.

Regulatory Classification

Identification

An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k130122 B. Purpose for Submission: Modification to cleared device (k993547) C. Measurand: Immunoglobulin M D. Type of Test: Quantitative, Nephelometric E. Applicant: Beckman Coulter, Inc. F. Proprietary and Established Names: IMMAGE® Immunochemistry System Low Concentration Immunoglobulin M (IGMLC) Reagent IMMAGE® Immunochemistry Systems CSF-CAL Cerebrospinal Fluid Protein Calibrator G. Regulatory Information: 1. Regulation section: 21 CFR §866.5510, Immunoglobulins A, G, M, D, and E immunological test system 21 CFR §862.1150, Calibrator 2. Classification: Class 2 3. Product code: CFN – Method, Nephelometric, Immunoglobulins (G, A, M) JIX – Calibrator, Multi-Analyte Mixture {1} 4. Panel: Immunology (82) Clinical Chemistry (75) H. Intended Use: 1. Intended use(s): IMMAGE® Immunochemistry System Low Concentration Immunoglobulin M (IGMLC) Reagent: IMMAGE® Immunochemistry Systems IGMLC Reagent, when used in conjunction with IMMAGE® Immunochemistry Systems and Cerebrospinal Fluid Protein Calibrator, is intended for quantitative determination of Immunoglobulin M (IGMLC) in human serum or cerebrospinal fluid (CSF) by rate nephelometry. IMMAGE® Immunochemistry Systems CSF-CAL Cerebrospinal Fluid Protein Calibrator: CSF CAL (Cerebrospinal Fluid Protein Calibrator), when used in conjunction with Beckman Coulter Low Concentration Immunoglobulin A (IGALC) and Low Concentration Immunoglobulin M (IGMLC) reagents is intended for use on IMMAGE for the calibration of these reagents. 2. Indication(s) for use: Same as above 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: IMMAGE Immunochemistry Systems I. Device Description: The IMMAGE® Immunochemistry System Low Concentration Immunoglobulin M (IGMLC) Reagent kit includes (a) one IGMLC Cartridge containing IGMLC Particle Reagent (particle bound anti-IgM), Antigen Excess Solution (processed diluted human serum), sodium azide (used as a preservative) and bovine serum albumin and non-reactive chemicals, (b) two Evaporation Caps, and (c) one IGMLC Reagent Bar Code Card. 2 {2} The Beckman Coulter Cerebrospinal Fluid Protein Calibrator (CSF CAL) components include (a) Cerebrospinal Fluid Protein Calibrator, (b) Cerebrospinal Fluid Protein Calibrator Bar Code Card, (c) Cerebrospinal Fluid Protein Calibrator Bar Code Strips, and (d) Value Assignment Sheet. # J. Substantial Equivalence Information: 1. Predicate device name(s) and 510(k) number(s): Beckman Coulter IMMAGE Immunochemistry System Low Concentration Immunoglobulin M (IGMLC) Reagent and Beckman Coulter Cerebrospinal Fluid Protein Calibrator (CSF CAL); (k993547) 2. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device IGMLC Reagent and Calibrator (k130122) | Predicate IGMLC Reagent and Calibrator (k993547) | | Intended Use: | IGMLC reagent is intended for the quantitative determination of immunoglobulin M (IGMLC) in human serum or cerebrospinal fluid (CSF) by rate nephelometry. CSF CAL is intended for use on IMMAGE for the calibration of these reagents. | Same | | Instrument | IMMAGE Immunochemistry Systems | Same | | Technology | Rate nephelometric method | Same | | Result Output | Quantitative result based on calibration curve | Same | | Calibrator Formulation | Prepared from human urine to which immunoglobulin A and immunoglobulin M have been added | Same | | Specimen types | Serum, CSF | Same | {3} | Differences | | | | --- | --- | --- | | Item | Device IGMLC Reagent and Calibrator (k130122) | Predicate IGMLC Reagent and Calibrator (k993547) | | Analytical Sensitivity | CSF: 0.15 mg/L Serum: 32.4 mg/L | CSF: 0.3 mg/L Serum: 64.8 mg/L | | Calibrator traceability | IFCC Reference preparation ERM-DA470(k)/IFCC | IFCC Reference preparation CRM 470 | | Analytical Range | CSF 0.15 - 10 mg/L Serum 32.4 - 2160 mg/L | CSF 0.3 - 10 mg/L Serum 64.8 - 2160 mg/L | | Equivalency | Comparison to the Predicate assay | Concordance to IFE method | | Internal Standards | Two additional internal standard levels were added to the internal standard curve at the low end | | ## K. Standard/Guidance Document Referenced (if applicable): EP05-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline, Second Edition EP06-A, Evaluation of Linearity of Quantitative Measurement, Approved Guideline, Second Edition EP09-A2, Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline - Second Edition EP17-A, Protocols for Determination of Limits of Detection and Limits of Quantification, Approved Guideline EN ISO 17511: In vitro diagnostic medical devices. Measurement of quantities in biological samples. Metrological traceability of values assigned to calibrators and control materials ## L. Test Principle: The IGMLC test measures the rate of increase in light scattered from particles suspended in solution as a result of complexes formed during an antigen-antibody reaction. A beam of light is passed through the cuvette and, as the antigen-antibody reaction proceeds, the light passing through the cuvette is scattered increasingly as insoluble immune complexes are formed. The light scatter is monitored by measuring the light intensity at an angle away from incident light. The IGMLC Reagent and CSF Calibrator are designed for optimal performance on the IMMAGE Immunochemistry Systems. A curve fit model is applied to the assay result to create parameters that allow each reagent lot to have its own unique stored curve. For a CSF sample, the instrument will use a $50~\mu \mathrm{L}$ sample size and concentration will {4} read directly from the stored curve. For a serum sample, the instrument will make 1:39 dilution and pick up $8\mu \mathrm{L}$ of sample and then multiply result by 216. # M. Performance Characteristics (if/when applicable): # 1. Analytical performance: # a. Precision/Reproducibility: Precision testing was performed in accordance with CLSI EP5-A2. Three separate serum sample pools representing low (32.4 - 43.4 mg/L; Serum-1), mid (432 - 1296 mg/L; Serum-2) and high (1728 - 2160 mg/L; Serum-3) range, and CSF samples pools representing low (0.15 - 0.20 mg/L; CSF-1), mid (2 - 6 mg/L; CSF-2) and high (8 - 10 mg/L; CSF-3) range were included in the study. The experimental design utilized duplicate sample analysis, twice daily, over the course of twenty working days $(n = 80)$ . The data were analyzed for within-run, between-run, between-day, and total precision and the mean U/mL and percent coefficient of variation (%CV) are summarized below. | Sample | N | Mean) (mg/L) | Within-Run | | Between Run | | Between Day | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | CSF-1 | 80 | 0.196 | 0.03 | 16.55% | 0 | 0% | 0.01 | 4.57% | 0.03 | 15.64% | | CSF-2 | 80 | 2.49 | 0.05 | 2.07% | 0.06 | 2.22% | 0 | 0% | 0.07 | 2.85% | | CSF-3 | 80 | 8.81 | 0.29 | 3.25% | 0.29 | 3.32% | 0 | 0% | 0.38 | 4.30% | | Serum-1 | 80 | 38.5 | 3.84 | 9.99% | 0.69 | 1.80% | 1.46 | 3.81% | 4.17 | 10.84% | | Serum-2 | 80 | 1250.1 | 28.87 | 2.31% | 21.33 | 1.71% | 0 | 0% | 35.59 | 2.85% | | Serum-3 | 80 | 1935.8 | 55.23 | 2.85% | 29.58 | 1.53% | 10.23 | 0.53% | 63.48 | 3.28% | # b. Linearity/assay reportable range: Linearity studies were conducted in accordance with CLSI Guideline EP06-A, Evaluation of Linearity of Quantitative Measurement, Approved Guideline, Second Edition. Two sample sources, including internal standards traceable to ERM-DA470k/IFCC and patient sample pools, were used to demonstrate linearity. Each sample and dilution was evaluated in replicates of five. The linear regression parameters (slope, intercept and $\mathbf{r}^2$ ) of the observed values vs. predicted values are shown below. | Sample type | Sample | Dilution range (mg/L) | Slope | Intercept | r2 | | --- | --- | --- | --- | --- | --- | | Serum | Low-1 | 29.70 – 297.01 | 0.977 | 6.918 | 0.999 | | | Low-2 | 29.72 – 297.20 | 0.999 | 2.340 | 0.997 | | | Low-3 | 30.51 – 305.12 | 0.970 | 5.538 | 0.999 | | | High-1 | 341.40 – 1707.02 | 1.033 | 34.837 | 0.998 | {5} | Sample type | Sample | Dilution range (mg/L) | Slope | Intercept | r2 | | --- | --- | --- | --- | --- | --- | | | High-2 | 354.68 – 1773.39 | 1.019 | 22.365 | 0.999 | | | High-3 | 386.35 – 1931.75 | 1.034 | 58.827 | 0.999 | | CSF | Low-1 | 0.13 – 1.33 | 1.073 | -0.110 | 0.997 | | | Low-2 | 0.15 – 1.52 | 1.012 | -0.030 | 0.999 | | | Low-3 | 0.15 – 1.50 | 1.006 | -0.011 | 0.999 | | | High-1 | 1.74 – 8.70 | 0.986 | -0.002 | 0.998 | | | High-2 | 1.73 – 8.63 | 0.985 | 0.088 | 0.999 | | | High-3 | 1.74 – 8.72 | 0.992 | 0.033 | 0.999 | The claimed reportable Analytical Measuring range is shown in the table below: | IgM Assay | Assay reportable range | | --- | --- | | Serum | 32.4 – 2160 mg/L | | CSF | 0.15 – 10 mg/L | c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability: The IMMAGE system uses a stored response curve that is created during manufacture by diluting single-level internal standard which is traceable to the IFCC reference preparation to plasma protein, material ERM-DA470k/IFCC. The traceability process is based on EN ISO 17511. The curve is established using multiple instruments. A curve fit model is applied to the assay result to create parameters that allow each reagent lot to have its own unique stored curve. The single point calibrator is also value assigned using the same internal standard that is used for the stored curve. Reagent Stability: No change (referred to k993547). d. Detection limit: Limit of blank (LoB), limit of detection (LoD), and limit of quantitation (LoQ) were determined in accordance with CLSI EP17-A. LoB: For serum, four patient sample pools created from IgM negative sera were assayed daily in replicates of 5 for a period of three days (total 60 data points) on two instruments. For CSF, CSF diluent was assayed daily in replicates of 5 for a period of three days (total 15 data points) on two instruments. Datasets were ranked from lowest to highest for these blank readings and the 95th percentile reading was determined from this ranking. LoD: For serum, four pools created from a serum patient sample and IgM depleted human serum at a mean value $\leq 32.4\mathrm{mg / L}$ and assayed daily in replicates of 5 for a {6} period of three days (total 60 data points) on two instruments. For CSF, four pools patient sample pools at a mean value $\leq 0.15\mathrm{mg / L}$ created from a CSF patient sample pool and CSF diluent were assayed daily in replicates of 5 for a period of three days (total 60 data points) on two instruments. LoQ: In order to establish LoQ for serum and CSF, three low-concentration IgM pools were run over 2 reagents lots and 2 instruments. For serum, there were a total of 45 replicates, 3 reps per pool, 3 pools, and 5 runs. For CSF, there were a total of 40 replicates from 3 pools: 3 reps for two of the pools and 2 reps for the third pool, and 5 total runs. The LoQ samples were created from an internal standard with a known value. | IgM assay | LoB (mg/L) | LoD (mg/L) | LoQ (mg/L) | | --- | --- | --- | --- | | Serum | 7.46 | 18.05 | 32.4 | | CSF | 0 | 0.0586 | 0.13 | e. Analytical specificity: Established in k993547. f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: Method comparison and bias estimation experiments were designed using CLSI guideline EP09-A2. The testing compared IGMLC pre-design change lots and IGMLC design changed lots. Eighty (80) CSF and serum specimens with IgM concentrations spanning the analytical ranges of the two sample types were obtained from commercial sources, including paired CSF and serum patient samples from clinical laboratories and commercial sources. About $5\%$ CSF samples (4 of 80 samples) and $12.5\%$ of the serum samples (10 of 80 samples) were contrived. The method comparison samples were tested internally over multiple days; one run per day with eight of each of the sample types (CSF and Serum) per day, over a period of 10 days. The IgM values in CSF samples ranged from 0.150 to $8.982\mathrm{mg / mL}$ while IgM values in serum samples ranged from 34.73 to $2,156.19\mathrm{mg / mL}$ (by the new method). Deming regression analysis was performed to obtain slope, $\mathbf{R}^2$, intercept and a scatter plot of data. Results are summarized in tables below. {7} CSF samples: | n | Slope (95% CI) | Y-Intercept (95% CI) | R² | | --- | --- | --- | --- | | 80 | 0.963 (0.946 to 0.980) | -0.133 (-0.190 to -0.076) | 0.994 | Serum samples: | n | Slope (95% CI) | Y-Intercept (95% CI) | R² | | --- | --- | --- | --- | | 80 | 1.048 (1.031 to 1.065) | 5.695 (-12.7 to 24.117) | 0.995 | b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: The reference interval values of 429 to 1893 mg/L for serum IGMLC were established using a population of 130 apparently healthy male and female adults from California. The reference interval of 0.19 – 0.29 mg/L for CSF was taken from published literature [Burtis, C. A., Ashwood, E. R., Tietz, Texbook of Clinical Chemistry, 3rd Edition, W. B. Saunders, Philadelphia, PA (1999)]. {8} N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 9
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...