← Product Code [MLR](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/MLR) · K151725

# STERRAD NX Sterilizer, STERRAD 100NX Sterilizer (K151725)

_Advanced Sterilization Products · MLR · Jan 13, 2016 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/MLR/K151725

## Device Facts

- **Applicant:** Advanced Sterilization Products
- **Product Code:** [MLR](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/MLR.md)
- **Decision Date:** Jan 13, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6860
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The STERRAD Sterilizers are designed for sterilization of both metal and nonmetal medical devices at low temperatures. Because the cycle operates within a dry environment and at low temperatures, it is especially suitable for instruments sensitive to heat and moisture.

## Device Story

STERRAD NX and 100NX are stand-alone hydrogen peroxide gas plasma sterilizers. Process: aqueous hydrogen peroxide injected into vaporizer, heated, introduced into chamber under sub-ambient pressure, and transformed into gas plasma via electrical energy. System monitors hydrogen peroxide concentration; cancels cycle if specs not met. Operated by hospital staff in clinical settings. Software enables network connectivity (LAN) via TCP/IP for transferring cycle parameters (PDF/CSV) to servers or instrument tracking systems. Connectivity features include DHCP configuration, diagnostic tools, and transmission failure notifications. Benefits: low-temperature sterilization of heat/moisture-sensitive instruments without toxic residue.

## Clinical Evidence

Bench testing only. No clinical data. Performance validated through electrical safety (IEC/EN 61010-1, IEC/EN 61010-2-040), electromagnetic compatibility (CISPR 11, EN 60601-1-2), and software verification/validation per FDA guidance.

## Technological Characteristics

Hydrogen peroxide gas plasma sterilization system. Chamber/door: aluminum. Sterilant: 59% nominal hydrogen peroxide in disposable cassettes. Connectivity: LAN via TCP/IP, DHCP. Software: onboard microprocessor-controlled. Sterilization cycles: Standard, Advanced, Flex, EXPRESS, DUO. Standards: CAN/CSA-C22.2 No. 61010-1, UL 61010-1, IEC/EN 61010-1, IEC/EN 61010-2-040, IEC/EN 60601-1-2, EN 55011.

## Regulatory Identification

An ethylene gas sterilizer is a nonportable device intended for use by a health care provider that uses ethylene oxide (ETO) to sterilize medical products.

## Predicate Devices

- STERRAD NX Sterilizer and STERRAD 100NX Sterilizer ([K142454](/device/K142454.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure, with three faces overlapping to create a sense of depth and connection. The logo is simple, yet recognizable, and represents the department's mission to protect the health of all Americans.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 13, 2016

Advanced Sterilization Products Ms. Sun Choi Regulatory Affairs Specialist IV 33 Technology Drive Irvine, CA 92618

Re: K151725

Trade/Device Name: STERRAD® NX® Sterilizer STERRAD® 100NX® Sterilizer Regulation Number: 21 CFR 880.6860 Regulation Name: Ethylene Oxide Gas Sterilizer Regulatory Class: Class II Product Code: MLR Dated: December 14, 2015 Received: December 17, 2015

Dear Ms. Sun Choi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{1}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K151725

#### Device Name STERRAD® NX® Sterilizer

Indications for Use (Describe)

Page 1 of 3

The STERRAD® NX® Sterilizer is designed for sterilization of both metal medical devices at low temperatures. The STERRAD sterilization process is a multiphase sterilization process that utilizes a combination of exposure to hydrogen peroxide vapor and plasma to affect sterilization. The STERRAD NX Sterilize instruments which have diffusion-restructed spaces, such as the hinged portion of forceps and scissors.

Medical devices with the following materials and dimensions can be processed in the STERRAD NX Sterilizer Standard cycle:

Single channel stainless steel lumens with

· An inside diameter of 1 mm or larger and a length of 150 mm or shorter*

- An inside diameter of 2 mm or larger and a length of 400 mm or shorter*
Medical devices, including most flexible endoscopes, with the following materials and dimensions can be processed in the STERRAD NX Sterilizer Advanced cycle:

Single channel stainless steel lumens with

- An inside diameter of 1 mm or larger and a length of 500 mm or shorter*
Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscope with • An inside diameter of 1 mm or larger and length of 850 mm or shorter*

* The validation testing for this lumen size was conducted using a maximum of 10 lumens per load.

- Hospital loads should not exceed the maximum number of lumens validated by this testing.
- *Only one flexible endoscope can be processed per sterilization cycle with or without a silicone mat. No additional load.

Note: With the exception of the 1 x 850 mm flexible endoscopes, the validation studies were performed using a validation load consisting of one instrument tray weighing 10.7 lbs. The 1 x 850 mm flexible without any additional load.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|× Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K151725

Device Name STERRAD® 100NX® Sterilizer

Indications for Use (Describe)

Page 2 of 3

The STERRAD® 100NX® Sterilizer is designed for sterilization of both metal medical devices at low temperatures. The STERRAD sterilization process is a multiphase sterlization process that utilizes a combination of exposure to hydrogen peroxide vapor and plasma to safely sterilize medical instruments and materials without leaving toxic residue.

The STERRAD 100NX Sterilizer can sterilize instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

Medical devices with the following materials and dimensions can be processed in the STERRAD 100NX Sterilizer Standard cycle:

· Single channel stainless steel lumens with an inside diameter of 0.7 mm or larger and a length of 500 mm or shorter*

Medical devices, including most flexible endoscopes, with the following materials and dimensions can be processed in the STERRAD 100NX Sterilizer Flex Scope cycle:

• Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and length of 850 mm or shorter**

Note: With the exception of the 1 x 850 mm flexible endoscopes, the validation studies were performed using a validation load consisting of two instrument trays each weighing 10.7 bs. The 1 x 850 mm flexible endoscopes were validated without any additional load.

* A maximum of ten single channel stainless steel lumens, five per tray per sterilization cycle.

** A maximum of two flexible endoscopes, one per tray per sterilization cycle. No additional load.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|× Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known) K151725

Device Name STERRAD® 100NX® Sterilizer

Indications for Use (Describe)

Page 3 of 3

The STERRAD 100NX EXPRESS Cycle is an additional optional cycle designed for surface sterilization of both metal medical devices at low temperatures.

• It can sterlize instrument surfaces and instruments having diffusion-restricted spaces, such as the hinged portion of forceps and scissors

· It can sterilize rigid and semi-rigid endoscopes without lumens

Note: The validation studies for EXPRESS Cycle were performed using a validation load consisting of a single instrument tray weighing 10.7 lbs placed on the bottom shelf.

The STERRAD 100NX DUO Cycle is an additional optional cycle designed for sterlization of medical devices including most flexible endoscopes, with the following materials and dimensions:

· Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and a length of 875 mm or shorter

· Accessory devices that are normally connected to a flexible endoscope during use

· Flexible endoscopes without lumens

Note: The validation studies for DUO Cycle were performed using a validation load consisting of two flexible endoscopes with their accessory devices weighing a total of 13.2 lbs.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|× Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{5}------------------------------------------------

# 510(k) Summary K151725

#### SUBMITTER I.

Advanced Sterilization Products 33 Technology Drive Irvine, CA 92618

Contact Person: Sun Choi Regulatory Affairs Specialist IV Telephone: (949) 453-6378 Fax: (949) 798-3900

Date Prepared: January 7, 2016

#### II. DEVICE

Name of Device: STERRAD® NX® Sterilizer and STERRAD® 100NX® Sterilizer Common or Usual Name: Hydrogen Peroxide Gas Plasma Sterilization System Classification Name: Ethylene Oxide Gas Sterilizer (21 CFR 880.6860) Regulatory Class: II Product Code: MLR

# III. PREDICATE DEVICE

STERRAD NX Sterilizer and STERRAD 100NX Sterilizer, K142454

The subject and predicate devices are manufactured by Advanced Sterilization Products.

# IV. DEVICE DESCRIPTION

### STERRAD NX Sterilizer

The STERRAD NX Sterilizer is a self-contained stand-alone system of hardware and software designed to sterilize medical instruments and devices using a patented hydrogen peroxide gas plasma process. Hydrogen peroxide vapor is generated by injecting aqueous hydrogen peroxide into the vaporizer where the solution is heated and vaporized, introducing 

{6}------------------------------------------------

the vapor into the chamber under sub-ambient pressure and transforming the vapor into a gas plasma using electrical energy. The STERRAD NX Sterilizer has two cleared sterilization cycles, the Standard and Advanced Cycles.

The sterilizer uses a disposable sterilant cassette that contains the 59% nominal hydrogen peroxide solution in a plastic cell pack and cassette shells. The hydrogen peroxide is concentrated before introducing into the sterilizer chamber and its concentration is monitored during the cycle. The sterilizer cancels the cvcle if the hydrogen peroxide monitor data does not meet the specification.

The hardware for the STERRAD NX Sterilizer consists of a sterilizer chamber, constructed with aluminum, and a variety of instruments and components which are housed in a covered frame. The sterilizer also uses accessories such as reusable instrument trays, printer paper, and an optional movable cart. The STERRAD NX Sterilizer can be placed directly on a table, counter top, or on the movable cart.

# STERRAD 100NX Sterilizer

The STERRAD 100NX Sterilizer is a self-contained stand-alone system of hardware and software designed to sterilize medical instruments and devices using a patented hydrogen peroxide gas plasma process. Hydrogen peroxide vapor is generated by injecting aqueous hydrogen peroxide into the vaporizer where the solution is heated and vaporized, introducing the vapor into the chamber under sub-ambient pressure and transforming the vapor into a gas plasma using electrical energy. The STERRAD 100NX Sterilizer has four cleared sterilization cycles (Standard, Flex, EXPRESS, and DUO Cycles).

The sterilizer uses a disposable sterilant cassette that contains the 59% nominal hydrogen peroxide solution in a plastic cell pack and cassette shells. For the Standard and Flex Cycles, the hydrogen peroxide is concentrated before into the sterilizer chamber. For the EXPRESS and DUO Cycles, the concentration process is not used. For all four cycles, the hydrogen peroxide concentration is monitored during the cycle and the sterilizer cancels the cycle if the hydrogen peroxide monitor data does not meet the specification.

The hardware for the STERRAD 100NX Sterilizer consists of a sterilizer chamber, constructed with aluminum, and a variety of instruments and components which are housed in a covered frame. The sterilizer also uses accessories such as reusable instrument trays and printer paper.

### Network Connectivitv

The STERRAD NX and 100NX are controlled by software running on an onboard microprocessor. The software is designed to control the sterilizer and provide an interface for user interaction with the sterilizer.

{7}------------------------------------------------

The software has the network connectivity feature that allows the Hospital IT Department to connect the STERRAD NX or STERRAD 100NX to a Hospital Local Area Network (LAN) for transfer of cycle parameters to a server and then, if desired, to an Instrument Tracking System. The software has been designed for ease of configuration using Dynamic Host Configuration Protocol (DHCP). The cycle information will be available in Portable Document Format (PDF) and Comma Separated Values (CSV) formats and transmitted using Transmission Control Protocol/Internet Protocol (TCP/IP). The network digital information will be identical to the existing cycle information printed out by the devices after each cycle (PDF file) and the existing electronic delimited data (CSV file) that can be downloaded through the USB port.

The basis for this 510(k) premarket notification is to offer the software revision that enables the network connectivity for the STERRAD NX and 100NX Sterilizers that were already placed on the market with the previous version of microprocessor. This proposed software revision includes the same software codes which were cleared for the systems under K142454 and has the same performance features as compared to the predicate devices as follows:

- It can be configured automatically or manually;
- It will be activated only upon the request of the user; ●
- Once it is activated, the user can turn it off and back on; ●
- It will have diagnostic tools for network troubleshooting; and ●
- It will notify the user of an unsuccessful transmission.

#### v. INDICATIONS FOR USE

# STERRAD NX Sterilizer

The STERRAD NX Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. The STERRAD sterilization process is a multiphase sterilization process that utilizes a combination of exposure to hydrogen peroxide vapor and plasma to affect sterilization. The STERRAD NX Sterilizer can sterilize instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

Medical devices with the following materials and dimensions can be processed in the STERRAD NX Sterilizer Standard cvcle:

Single channel stainless steel lumens with

- An inside diameter of 1 mm or larger and a length of 150 mm or shorter™ ●
- An inside diameter of 2 mm or larger and a length of 400 mm or shorter™ .

Medical devices, including most flexible endoscopes, with the following materials and

{8}------------------------------------------------

dimensions can be processed in the STERRAD NX Sterilizer Advanced cycle:

Single channel stainless steel lumens with

- An inside diameter of 1 mm or larger and a length of 500 mm or shorter" ●
Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscope with

- An inside diameter of 1 mm or larger and length of 850 mm or shorter* ●
TThe validation testing for this lumen size was conducted using a maximum of 10 lumens per load.

Hospital loads should not exceed the maximum number of lumens validated by this testing.

*Only one flexible endoscope can be processed per sterilization cycle with or without a silicone mat. No additional load.

Note: With the exception of the 1 x 850 mm flexible endoscopes, the validation studies were performed using a validation load consisting of one instrument tray weighing 10.7 lbs. The 1 x 850 mm flexible endoscope was validated without any additional load.

# STERRAD 100NX Sterilizer

The STERRAD 100NX Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. The STERRAD sterilization process is a multiphase sterilization process that utilizes a combination of exposure to hydrogen peroxide vapor and plasma to safely sterilize medical instruments and materials without leaving toxic residue.

The STERRAD 100NX Sterilizer can sterilize instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

Medical devices with the following materials and dimensions can be processed in the STERRAD 100NX Sterilizer Standard cycle:

- Single channel stainless steel lumens with an inside diameter of 0.7 mm or larger and ● a length of 500 mm or shorter*
Medical devices, including most flexible endoscopes, with the following materials and dimensions can be processed in the STERRAD 100NX Sterilizer Flex Scope cycle:

- Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and length of 850 mm or shorter**
Note: With the exception of the 1 x 850 mm flexible endoscopes, the validation studies were performed using a validation load consisting of two instrument trays each weighing 10.7 lbs. The 1 x 850 mm flexible endoscopes were validated without any additional load.

* A maximum of ten single channel stainless steel lumens, five per tray per sterilization cycle.

{9}------------------------------------------------

** A maximum of two flexible endoscopes, one per tray per sterilization cycle. No additional load.

The STERRAD 100NX EXPRESS Cycle is an additional cycle designed for surface sterilization of both metal and nonmetal medical devices at low temperatures.

- It can sterilize instrument surfaces and instruments having diffusion-restricted spaces, . such as the hinged portion of forceps and scissors
- . It can sterilize rigid and semi-rigid endoscopes without lumens

Note: The validation studies for EXPRESS Cycle were performed using a validation load consisting of a single instrument tray weighing 10.7 lbs placed on the bottom shelf.

The STERRAD 100NX DUO Cycle is an additional optional cycle designed for sterilization of medical devices including most flexible endoscopes, with the following materials and dimensions:

- . Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and a length of 875 mm or shorter
- Accessory devices that are normally connected to a flexible endoscope during use
- Flexible endoscopes without lumens

Note: The validation studies for DUO Cycle were performed using a validation load consisting of two flexible endoscopes with their accessory devices weighing a total of 13.2 lbs.

{10}------------------------------------------------

# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The subject devices have the same intended use, indications for use, materials, design features, and sterilizer functions as the predicate devices. Both the subject and predicate devices share the same technological characteristics with the exception of different versions of the microprocessor module.

The predicate devices use a later version microprocessor module due to the obsolescence of the previous version by the supplier. ASP plans to offer the software upgrade that enables the network connectivity for the subject devices, i.e., the STERRAD NX and STERRAD 100NX Sterilizers that were manufactured and distributed with the previous version of the microprocessor module.

Refer to Table 1 and Table 2 for comparison between subject and predicate devices for the STERRAD NX and STERRAD 100NX, respectively.

| Table 1: STERRAD NX - Comparison Table for Subject Device vs. Predicate Device |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                           |
|--------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
|                                                                                | Predicate Device:<br>STERRAD NX Sterilizer<br>(K142454)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Subject Device:<br>STERRAD NX Sterilizer<br>(K151725)                                     |
| Intended Use                                                                   | The STERRAD Sterilizers are designed for<br>sterilization of both metal and nonmetal medical<br>devices at low temperatures. Because the cycle<br>operates within a dry environment and at low<br>temperatures, it is especially suitable for instruments<br>sensitive to heat and moisture.                                                                                                                                                                                                                                                                                                                                                                                                                  | Same as predicate                                                                         |
| Indications for<br>Use                                                         | Refer to Section V, Indications for Use of this<br>510(k) Summary.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same as predicate                                                                         |
| Technological<br>Characteristics                                               | The STERRAD NX is a hydrogen peroxide gas<br>plasma sterilizer designed to sterilize medical<br>instruments and devices using a hydrogen peroxide<br>process. Hydrogen peroxide vapor is generated<br>by injecting aqueous hydrogen peroxide into the<br>vaporizer where the solution is heated and<br>vaporized, introducing the vapor into the chamber<br>and transforming the vapor into a gas plasma<br>using electrical energy.<br>The network connectivity software allows the<br>hospital IT department to connect the sterilizer to a<br>hospital local area network for transfer of cycle<br>parameters to a server and then, if desired, to an<br>instrument tracking system. The software has been | Same as predicate with the<br>exception of an earlier<br>microprocessor module<br>version |

### Table 1: STERRAD NX - Comparison Table for Subiect Device vs. Predicate Device

{11}------------------------------------------------

|                         | Predicate Device:<br>STERRAD NX Sterilizer<br>(K142454)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Subject Device:<br>STERRAD NX Sterilizer<br>(K151725) |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| Materials               | designed for ease of configuration using Dynamic<br>Host Configuration Protocol (DHCP) for network<br>connectivity.                                                                                                                                                                                                                                                                                                                                                                                                                         | Same as predicate                                     |
|                         | The chamber and door are constructed with<br>aluminum.<br>The cassette contains the 59% nominal hydrogen<br>peroxide solution in the plastic cell pack and cassette<br>shells.                                                                                                                                                                                                                                                                                                                                                              |                                                       |
| Design<br>Features      | The sterilizer concentrates the hydrogen peroxide.<br>The hydrogen peroxide concentration is monitored<br>during the cycle and the sterilizer cancels the cycle<br>if the hydrogen peroxide monitor data does not meet<br>the specification.                                                                                                                                                                                                                                                                                                | Same as predicate                                     |
| Sterilizer<br>Functions | Logout is used when the current operator is finished<br>using the sterilizer. When Logout is selected, login<br>is required before using the sterilizer.<br>System Summary displays the System Summary file<br>and allows operator to print a copy.<br>Cycle History displays the Select Cycle History<br>screen. This screen allows operator to select a cycle<br>history file and view or print it.<br>Additional Utilities is available only to operators<br>with Supervisor-level access. It displays the<br>Additional Utilities Menu. | Same as predicate                                     |

# Table 2: STERRAD 100NX - Comparison Table for Subject Device vs. Predicate Device

|                                  | Predicate Devices:<br>STERRAD 100NX Sterilizer<br>(K142454)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Subject Device:<br>STERRAD 100NX Sterilizer<br>(K151725)                                  |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| Intended Use                     | The STERRAD Sterilizers are designed for<br>sterilization of both metal and nonmetal medical<br>devices at low temperatures. Because the cycle<br>operates within a dry environment and at low<br>temperatures, it is especially suitable for instruments<br>sensitive to heat and moisture.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same as predicate                                                                         |
| Indications for<br>Use           | Refer to Section V, Indications for Use of this<br>510(k) Summary.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same as predicate                                                                         |
|                                  | Predicate Devices:<br>STERRAD 100NX Sterilizer<br>(K142454)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Subject Device:<br>STERRAD 100NX Sterilizer<br>(K151725)                                  |
| Technological<br>Characteristics | The STERRAD 100NX is a hydrogen peroxide gas<br>plasma sterilizer designed to sterilize medical<br>instruments and devices using a hydrogen peroxide<br>process. Hydrogen peroxide vapor is generated<br>by injecting aqueous hydrogen peroxide into the<br>vaporizer where the solution is heated and<br>vaporized, introducing the vapor into the chamber<br>and transforming the vapor into a gas plasma<br>using electrical energy.<br>The network connectivity software allows the<br>hospital IT department to connect the sterilizer to a<br>hospital local area network for transfer of cycle<br>parameters to a server and then, if desired, to an<br>instrument tracking system. The software has been<br>designed for ease of configuration using Dynamic<br>Host Configuration Protocol (DHCP) for network<br>connectivity. | Same as predicate with the<br>exception of an earlier<br>microprocessor module<br>version |
|                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                           |
| Materials                        | The chamber and door are constructed with<br>aluminum.<br>The cassette contains the 59% nominal hydrogen<br>peroxide solution in the plastic cell pack and cassette<br>shells.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same as predicate                                                                         |
| Design<br>Features               | The sterilizer concentrates the hydrogen peroxide for<br>Standard and Flex Cycles. The hydrogen peroxide<br>concentration is monitored during the cycle and the<br>sterilizer cancels the cycle if the hydrogen peroxide<br>monitor data does not meet the specification.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same as predicate                                                                         |
| Sterilizer<br>Functions          | Logout is used when the current operator is finished<br>using the sterilizer and the option is enabled. When<br>Logout is selected, the operator must re-login to use<br>the sterilizer.<br>Cycle History displays the Select Cycle History<br>screen. This screen allows the operator to select a<br>cycle history file and view or print it.<br>Utilities are available only to operators with<br>Supervisor-level access. It displays the Additional<br>Utilities Menu.<br>Door Open opens the active door.                                                                                                                                                                                                                                                                                                                          | Same as predicate                                                                         |
|                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                           |
|                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                           |
|                                  | Door Close closes the active door.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                           |

{12}------------------------------------------------

{13}------------------------------------------------

# VII. PERFORMANCE DATA

The subject devices have the same intended use, indications for use, materials, design features, technological characteristics, and sterilizer functions as the predicate devices. The following studies were performed in support of the substantial equivalence determination between the subject and predicate devices.

# Electrical safety and electromagnetic compatibility (EMC)

The subject devices were tested for Radiated and Conducted Emissions according to the Standards shown in the table below. All test results passed the requirements of the Standards (Table 3).

Table 3: Testing of STERRAD NX and 100NX Sterilizers for Radiated and Conducted Emissions

| Test Description    | Standard                       | Pass/Fail |
|---------------------|--------------------------------|-----------|
| Radiated Emissions  | CISPR 11:2009 +A1:2010 Class A | Pass      |
| Radiated Emissions  | EN 60601-1-2:2007 Class A      | Pass      |
| Conducted Emissions | CISPR 11:2009 +A1:2010 Class A | Pass      |
| Conducted Emissions | EN 60601-1-2:2007 Class A      | Pass      |

The STERRAD NX and 100NX Sterilizers were evaluated for safety under the standards listed below. The testing, conducted to the set of standards, provided a standardized level of assurance that the system is electrically and mechanically safe when operated and maintained in accordance with the STERRAD User Guide.

- CAN/CSA-C22.2 No.: 61010-1 (R2009); Safety Requirements for Electrical Equipment . for Measurement, Control, and Laboratory Use.
- UL 61010-1 (R2008); Standard for Safety for Electrical Equipment for Laboratory Use. .
- . IEC/EN 61010-1:2001; Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use.
- IEC/EN 61010-2-040:2005, Particular requirements for sterilizers and washer-. disinfectors used to treat medical materials, First Ed.
- IEC/EN 60601-1-2:2007, CLASS A; Medical Electrical Equipment Part 1-2: General . Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests.
- . EN 55011:2009 +A1:2010. Group I Class A limits, based on CISPR 11:2009. Group I Class A limits (subset of EN 60601-1-2), Industrial, Scientific and Medical Equipment -Radio-frequency Disturbance Characteristics – Limits and Methods of Measurement.

{14}------------------------------------------------

#### Software Verification and Validation Testing

Software verification and validation testing was conducted as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." All testing met the predefined acceptance criteria. The software for this device was considered as a "moderate" level of concern since prior to mitigation of hazards a failure of the software could result in Minor Injury, either to a patient or to a user of the device.

# VIII. CONCLUSIONS

The subject devices have the same intended use and technological characteristics as the predicate devices; additionally, non-clinical performance testing demonstrates equivalent performance between the subject and predicate devices. Therefore, based on the intended uses, technological characteristics, and non-clinical performance data, the subject devices (K151725) are substantially equivalent to the predicate devices cleared under K142454.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/MLR/K151725](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/MLR/K151725)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
