NON-STERILE, POWDER FREE NITRILE EXAMINATION GLOVE, BLUE COLOR, POLYMER COATED, TESTED WITH CHEMOTHERAPY DRUGS

K082957 · Siam Sempermed Corp., Ltd. · LZA · May 29, 2009 · General Hospital

Device Facts

Record IDK082957
Device NameNON-STERILE, POWDER FREE NITRILE EXAMINATION GLOVE, BLUE COLOR, POLYMER COATED, TESTED WITH CHEMOTHERAPY DRUGS
ApplicantSiam Sempermed Corp., Ltd.
Product CodeLZA · General Hospital
Decision DateMay 29, 2009
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

A patient examination glove is a medical device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Story

Non-sterile, powder-free nitrile examination glove; blue color; polymer-coated. Designed for use by healthcare professionals during patient examinations to provide a barrier against contamination. Tested for use with chemotherapy drugs. Device functions as a physical barrier worn on the hand. No electronic or software components.

Clinical Evidence

Bench testing only. Device performance verified against ASTM D 6319-00a-05 for dimensions, physical properties, and freedom from pinholes; ASTM D 6124-06 for powder-free residue; and biocompatibility standards (primary skin irritation and guinea pig sensitization).

Technological Characteristics

Nitrile material; powder-free; polymer-coated; blue color. Complies with ASTM D 6319-00a-05 (dimensions, physical properties, pinholes) and ASTM D 6124-06 (powder residue). Biocompatibility confirmed via rabbit skin irritation and guinea pig sensitization tests.

Indications for Use

Indicated for use as a patient examination glove to prevent cross-contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K082957 ## 510 (k) Summary As Required by 21 section 807.92 (c) MAY 2 9 2009 - Submitter Name: Siam Sempermed Corp., Ltd 1. - 352 Kanjanavanit Road. Pahtong Hatyai 2. Address: Songkhla. Thailand 90230 - (+66) 74 291 471 3. Phone: - (+66) 74 471 337 4. Fax: Contract Person: Mrs. Parawan Paiyasan(Quality System Manager) ഗ് Date summary prepared: September 22, 2008 ર. 7. Official Correspondent: Sempermed USA Inc. - 13900 49th Street North 8. Address: Clearwater, USA , FL 33762 - 727 787 7250 9. Phone: - 727 787 7558 10. Fax: - Mr. William E. Harris 11. Contact person: 12. Device Trade or Proprietary Name: Non sterile, Powder free Nitrile Examination Glove, Blue with Polymer coating Tested for use with chemotherapy drugs. 13. Device Common or usual name: Examination glove 14. Device Classification Name: Nitrile Patient Examination Glove (Powder Free(Polymer coated),Blue color) - 15. Description of the Device: Non sterile, Powder free Nitrile Examination Glove, Blue with Polymer coated, Tested for use with chemotherapy drugs. 16. Indications for se of the device: Based upon 21 CFR Prt 880.6250: "Patient examination glove." A patient res mination glove is a medical device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. ## 17. Summary of The Technological Characteristics of The devices: amary of Title Technological Characteristics of Trio decised for use with Polymer coated, Tested for use with Non seenle, Fowder Inc. Forme Exammation Over, Story, Free Pological characteristics compared to ASTM or equivalent standards. | CHARACTERISTICS | STANDARDS | DEVICE PERFORMANCE | |-----------------------|------------------------------------|--------------------| | Dimensions | ASTM D 6319-00a-05 | Meets | | Physical Properties | ASTM D 6319-00a-05 | Meets | | Freedom from pinholes | ASTM D 6319-00a-05 | Meets | | Powder Free Residue | ASTM D 6124-06 | Meets | | Biocompatibility | Primary Skin Irritation in Rabbits | Passes | | | Guinea Pig Sensitization | Passes | {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES.USA" is arranged in a circular fashion around the caduceus. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Siam Sempermed Corporation C/O Mr. William E. Harris President & CEO Sempermed USA, Incorporated 13900 49th Street North Clearwater, Florida 33762 MAY 2 9 2009 Re: K082957 Trade/Device Name: Non-Sterile, Powder-Free Nitrile Examination Glove, Blue with Polymer Coating, Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: May21, 2009 Received: May 26, 2009 Dear Mr. Harris: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ ## Page 2- Mr. Harris Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Anthony D. Watson for Susan Runner, D.D.S., MA Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use Applicant: Siam Sempermed Corp. Ltd. 510(k) Number: K082957 Use Device Name: Non-Sterile, Powder-Free Nitrile Examination Glove, Blue with Polymer Coating, Tested for Use with Chemotherapy Drugs Indications for : Based upon 21 CFR Part 880.6250; "Patient examination glove" . > A patient examination glove is a medical device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Shiela A. Murphy (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K082457 Page 1 of __
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...