← Product Code [LYZ](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ) · K051393

# DISPOSABLE POWDER FREE VINYL EXAM GLOVE YELLOW COLOR (K051393)

_Ever Global Enterprise Corporation · LYZ · Jul 27, 2005 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ/K051393

## Device Facts

- **Applicant:** Ever Global Enterprise Corporation
- **Product Code:** [LYZ](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ.md)
- **Decision Date:** Jul 27, 2005
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A PATIENT EXAMINATION GLOVES IS A DISPOSABLE DEVICE INTENDED FOR MEDICAL PURPOSE THAT IS WORN ON THE EXAMINER'S HAND OR FINGERS TO PREVENT CONTAMINATION BETWEEN PATIENT AND EXAMINER.

## Device Story

Disposable powder-free vinyl examination glove; yellow color. Worn by healthcare examiners on hands/fingers to provide barrier protection against contamination between patient and examiner. Manufactured to meet ASTM D5250-00E4 standards. Device is non-sterile and intended for single use.

## Clinical Evidence

No clinical data; bench testing only. Testing included dimensions, watertight integrity (ASTM D-5151-99, AQL 2.5), sensitization/irritation testing, and residual powder testing (ASTM D6124-01) to confirm powder-free status.

## Technological Characteristics

Vinyl material; powder-free. Dimensions and physical properties conform to ASTM D5250-00E4. Watertight integrity tested per ASTM D-5151-99. Residual powder content verified per ASTM D6124-01 (≤ 2 mg per glove).

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Safehealth Medical Supply Corp Vinyl Examination Glove, Powder Free ([K955091](/device/K955091.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### JUL 2 7 2005

## 510(K) SUMMARY

This summary of 510(K) safety and effectiveness information is being submitted in Accordance with requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(K) number is:

## 1. Submitter's Identification:

Mr. Nick Bien Ever Global Enterprise Corporation Long Thanh Industrial Zone Taman Village, Long Thanh District Dong Nai Province, Vietnam

Date Summary Prepared: April 5, 2005

## 2. Name of the Device:

Ever Global Enterprise Corporation Disposable Powder Free Vinyl Exam Glove Yellow Color

## 3. Predicate Device Information:

Safehealth Medical Supply Corp Vinyl Examination Glove,Powder Free (K955091)

#### 4. Device description:

Classified by FDA's General and Plastic Surgery Device panel as Class 1, 21 CFR 880.6250, Powder-Free Vinyl Patient Examination Glove, 80LYZ, and meets all requirements of ASTM Standard D5250-00E4.

#### 5. Intended Use:

A patient examination glove is a disposable device intended for medical purposes . I parceller in ...
that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

{1}------------------------------------------------

## 6.Comparison to Predicate Devices:

Comparison to I realleries Disposable Powder Free Vinyl Exam Glove,
Ever Global Enterprise Corporation's Disposable Powder Free Schoolth Evel Cloudi Linerprise Corporation
Yellow Color is substantially equivalent to the device manufactured by Safehealth Yenow Color is submansany -11
Medical Supply Corp. (K955091) except for the yellow color additive being added to this glove.

# 7. Discussion of Non-Clinical tests Performed for Determination of Substantial Equivalence are as follows:

Substantial Data. In Ever Global Enterprise Corporation's glove production are
The standards used for Ever Global Enterprise Corporation of Picture The standards about 11 Dimensions Testing conducted on gloves, Inspection level S-2, AQL 4.0.

The FDA 1000 ml Watertight Test based on ASTM D-5151-99 was also conducted with samplings of AQL 2.5, Inspection level I, meeting these requirements. Primary With samplings of Pressitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.

There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels.

A residual Powder test that based on ASTM D6124-01 for Starch at finished A residual I oware 1997 ...
inspection is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per gloves)

# 8. Discussion of Clinic Tests Performed:

Not applicable – There is no Hypoallergenic Claim.

#### 9. Conclusions:

Collectusions. I:VEl Olooal Linerprise Carp .
Yellow Color conform fully to ASTM D-5250-00E4 standard as well as applicable Tenow Octor of the same
21CFR references, and, meets pinhole FDA requirements, biocompatibility zTCPA references, and labeling claims as shown by data in Section 7. There are no requirements and november of the "substantial equivalence" products cited.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is surrounded by the words "U.S. Department of Health and Human Services, Inc." in a circular arrangement.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 27 2005

Mr. Nick Bien Manager Ever Global Enterprise Corporation Long Thanh Industrial Zone, Taman Village, Long Thanh District, Dong Nai Province, VIETNAM

Re: K051393

Trade/Device Name: Disposable Powder Free Vinyl Exam Glove, Yellow Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: July 6, 2005 Received: July 20, 2005

Dear Mr. Bein:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Bein

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements mount that I I any Federal statutes and regulations administered by other Federal agencies. or are For of any with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and if moments as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Switte y. Michael Omd

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### Indications for Use

#### EVER GLOBAL ENTERPRISE CORPORATION Applicant:

### 510(k) Number (if known): K051393

Device Name: Disposable Powder Free Vinyl Exam Glove Yellow Color

Indications for Use:

A PATIENT EXAMINATION GLOVES IS A DISPOSABLE DEVICE INTENDED FOR MEDICAL PURPOSE THAT IS WORN ON THE EXAMINER'S HAND OR FINGERS TO PREVENT CONTAMINATION BETWEEN PATIENT AND EXAMINER.

| Prescription Use            |  |
|-----------------------------|--|
| (Part 21 CFR 801 Subpart D) |  |

AND/OR

| Over-The-Counter Use   | X |
|------------------------|---|
| (21 CFR 801 Subpart C) |   |

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER

PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
PAGE ADD 6-1
510(k) Number: K051393

7/26/05

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ/K051393](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ/K051393)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
