← Product Code [MLR](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/MLR) · K212200

# STERLINKTM FPS-15s Plus Sterilizer with STERLOADTM Cassette (K212200)

_Plasmapp Co,., Ltd. · MLR · Oct 22, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/MLR/K212200

## Device Facts

- **Applicant:** Plasmapp Co,., Ltd.
- **Product Code:** [MLR](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/MLR.md)
- **Decision Date:** Oct 22, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6860
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The STERLINK™ FPS-15s Plus sterilizer with STERLOAD™ cassette is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed, and dried reusable metal and nonmetal medical devices used in healthcare facilities. A pre-programmed sterilization lumen cycle operates at low pressure and is thus suitable for processing medical devices sensitive to heat and moisture. The STERLINK™ FPS-15s Plus can sterilize: - Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors - Medical devices with a single stainless-steel lumen with: o An inside diameter of 2.4 mm or larger and a length of 280 mm or shorter *The validation testing for all lumen sizes was conducted using a maximum of five (5) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load with a total weight of 3.97 lbs.

## Device Story

Low-temperature vaporized hydrogen peroxide sterilizer; uses STERLOAD™ cassette (59% H2O2) as sterilant. Input: cleaned, rinsed, dried medical devices. Operation: automated sterilization cycle (vacuum, sub-atmospheric pressure, 60°C) in 14L chamber; fixed-dose sterilant delivery. Output: sterilized instruments. Used in healthcare facilities; operated by staff. Provides terminal sterilization for heat/moisture-sensitive devices; benefits patient by reducing infection risk from contaminated instruments.

## Clinical Evidence

Bench testing only. Validated per ISO 14937, ISO 11737, and AOAC 966.04. Testing included sporicidal activity, lumen/surface/mated-surface sterilization, simulated-use, and material compatibility. Electrical safety, EMC, and software validation (IEC 62304) confirmed. No clinical data.

## Technological Characteristics

Low-temperature vaporized H2O2 sterilizer; 14L chamber; 1000W max power. Uses 59% aqueous H2O2 unit-dose cassettes. Control system: Embedded Linux; chamber/vaporizer thermocouples; pressure transducers. Connectivity: Standalone. Materials/Standards: ASTM D638, E8/E8M, D790, E290, D256, E23, E1164, E313, D3985, F1249. Sterilization: 36-minute cycle at 60°C, sub-atmospheric pressure (down to 3 Torr).

## Regulatory Identification

An ethylene gas sterilizer is a nonportable device intended for use by a health care provider that uses ethylene oxide (ETO) to sterilize medical products.

## Predicate Devices

- PSD-85 ([K140464](/device/K140464.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 22, 2021

Plasmapp Co.,Ltd % Candace Cederman Consultant CardioMed Device Consultants LLC 1783 Forest Drive. Suite 254 Annapolis, Maryland 21401

Re: K212200

Trade/Device Name: STERLINKTM FPS-15s Plus, STERLOAD™ Cassette Regulation Number: 21 CFR 880.6860 Regulation Name: Ethylene oxide gas sterilizer Regulatory Class: Class II Product Code: MLR Dated: September 20, 2021 Received: September 21, 2021

Dear Candace Cederman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212200

Device Name

STERLINKTM FPS-15s Plus Sterilizer with STERLOAD™ Cassette

Indications for Use (Describe)

The STERLINK™ FPS-15s Plus Sterilizer with STERLOAD™ Cassette is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed, and dried reusable metal medical devices used in healthcare facilities. A pre-programmed sterilization lumen cycle operates at low temperature and is thus suitable for processing medical devices sensitive to heat and moisture.

The STERLINK™ FPS-15s Plus can sterilize*:

- · Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
- · Medical devices with a single stainless steel lumen with:
- o An inside diameter of 2.4 mm or larger and a length of 280 mm or shorter

* The validation testing for all lumen sizes was conducted using a maximum of five (5) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load with a total weight of 3.97 lbs.

| Type of Use (Select one or both, as applicable) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------|---------------------------------------------|
|-------------------------------------------------|---------------------------------------------|

_ | Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 1. General Information

| Applicant/Submitter:        | Plasmapp Co., Ltd.<br>BVC-111, 125, Gwahak-ro,<br>Yuseong-gu, Daejeon, 34141,<br>Rep. of Korea (South Korea)<br>Tel: +82 (0)42 716 2115 |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:<br>Address: | Candace Cederman<br>CardioMed Device Consultants LLC<br>1783 Forest Drive<br>Suite 254<br>Annapolis, MD 21401<br>Tel: +1 410 674 2060   |
| Preparation Date:           | October 12, 2021                                                                                                                        |

## 2. Device Name and Code

| Device Trade Name:   | STERLINK™ FPS-15s Plus Sterilizer with STERLOAD™ Cassette |
|----------------------|-----------------------------------------------------------|
| Common Name:         | Vapor Phase Hydrogen Peroxide Sterilization System        |
| Classification Name: | Ethylene Oxide Gas Sterilizer                             |
| Product Code:        | MLR                                                       |
| Regulation Number:   | 880.6860                                                  |
| Classification:      | Class II                                                  |
| Review Panel:        | General Hospital                                          |

# 3. Device Description

The STERLINK™ FPS-15s Plus sterilizer with STERLOAD™ cassette is a low temperature sterilizer which uses the STERLINK™ process to inactivate microorganisms on a broad range of 

{4}------------------------------------------------

medical devices and instruments. This sterilizer offers an effective, safe, fast, economical, easy to use, reliable, and flexible sterilization method.

This system consists of a main device and cassette which are called the STERLINK™ and STERLOAD™, respectively. The STERLOAD™ cassette contains 59% (weight concentration) of hydrogen peroxide (H2O2) which is utilized as the sterilant.

# 4. Indications / Intended Use

The STERLINK™ FPS-15s Plus sterilizer with STERLOAD™ cassette is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed, and dried reusable metal and nonmetal medical devices used in healthcare facilities. A pre-programmed sterilization lumen cycle operates at low pressure and is thus suitable for processing medical devices sensitive to heat and moisture.

The STERLINK™ FPS-15s Plus can sterilize :

- Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors ●
- . Medical devices with a single stainless-steel lumen with:
	- o An inside diameter of 2.4 mm or larger and a length of 280 mm or shorter

*The validation testing for all lumen sizes was conducted using a maximum of five (5) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load with a total weight of 3.97 lbs.

## 5. Technical Characteristics in Comparison to Predicate Devices

A summary of the technical characteristics of the subject device and predicate device can be found in the table below.

|                                                    | Subject Device                                                                                                                                                                                                                                                                                                           | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                    | Plasmapp Co., Ltd.                                                                                                                                                                                                                                                                                                       | Sterilucent, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device Name                                        | FPS-15s Plus                                                                                                                                                                                                                                                                                                             | PSD-85                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| 510(k) Number                                      | K212200                                                                                                                                                                                                                                                                                                                  | K140464                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Device Classification<br>Name                      | Ethylene oxide gas sterilizer                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Classification Product<br>Code                     | MLR                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Regulation Number                                  | 21 CFR 880.6860                                                                                                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                                    | Subject Device                                                                                                                                                                                                                                                                                                           | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                                    | Plasmapp Co., Ltd.                                                                                                                                                                                                                                                                                                       | Sterilucent, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Intended Use                                       | The STERLINKTM FPS-15s Plus sterilizer<br>with STERLOADTM cassette is a vaporized<br>hydrogen peroxide sterilizer intended for use<br>in the terminal sterilization of cleaned,<br>rinsed, and dried reusable metal and<br>nonmetal medical devices used in healthcare<br>facilities. A pre-programmed sterilization<br> | The PSD-85 is a vaporized hydrogen<br>peroxide sterilizer intended for use in the<br>terminal sterilization of cleaned, rinsed and<br>dried reusable metal and nonmetal medical<br>devices used in healthcare facilities.<br>The two pre-programmed sterilization<br>cycles, the Lumen and the Non-Lumen<br>Cycles, operate at low pressure and low<br>temperature and are thus suitable for<br>processing medical devices sensitive to heat<br>and moisture.<br><br>The PSD-85 Lumen Cycle can sterilize*:<br>• Instruments with diffusion-restricted<br>spaces such as the hinged portion of forceps<br>and scissors<br>• Medical devices with a single stainless steel<br>lumen with:<br>o An inside diameter of 1 mm or larger and<br>a length of 60 mm or shorter<br>o An inside diameter of 2 mm or larger and<br>a length of 250 mm or shorter<br>o An inside diameter of 3 mm or larger and<br>a length of 350 mm or shorter<br><br>*The validation testing for all lumen sizes<br>was conducted using a maximum of ten (10)<br>lumens per load. Hospital loads should not<br>exceed the maximum number of lumens<br>validated by this testing. The validation<br>studies were performed using a validation<br>load consisting of one instrument tray with a<br>total weight of 10.4 lbs.<br><br>The PSD-85 Non-Lumen Cycle can<br>sterilize**:<br>• Non-lumen instruments including non-<br>lumen instruments with stainless steel<br>diffusion-restricted areas such as the hinged<br>portion of forceps or scissors.<br><br>**The validation studies were conducted<br>using a validation load consisting of one<br>instrument tray with a total weight of 25.3<br>lbs. |
| Physical Characteristic                            | Self-contained, stand-alone device                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Design and<br>Construction                         | Welded frame onto which is mounted the<br>sterilization chamber along with a variety of<br>instruments and components, controls,<br>piping, and vacuum pump, all of which is<br>housed in covered frame                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                                    | Subject Device                                                                                                                                                                                                                                                                                                           | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                                    | Plasmapp Co., Ltd.                                                                                                                                                                                                                                                                                                       | Sterilucent, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Chamber Volume                                     | 14 L                                                                                                                                                                                                                                                                                                                     | 85 L                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Weight                                             | 147 lbs (67 kg)                                                                                                                                                                                                                                                                                                          | 440 lbs (200 kg)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Max Power                                          | 1000 W                                                                                                                                                                                                                                                                                                                   | 1650 W                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Control System                                     | Embedded Linux                                                                                                                                                                                                                                                                                                           | Atmel AT91SAM7 Microprocessor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                                    | Internal process monitor                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Temperature                                        | Chamber and vaporizer thermocouple                                                                                                                                                                                                                                                                                       | Chamber and vaporizer thermistors                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Pressure                                           | Chamber pressure transducers                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Sterilant Concentration                            | None                                                                                                                                                                                                                                                                                                                     | Real-time hydrogen peroxide vapor monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Operational Principle                              | Sterilization of the intended devices by<br>exposure under controlled conditions of<br>pressure, temperature, and time                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Operational<br>Parameters                          | Low pressure (vacuum; sub-atmospheric<br>down to 3 Torr) and temperature (60°C)                                                                                                                                                                                                                                          | Low pressure (vacuum; sub-atmospheric<br>down to 0.4 Torr) and temperature (55°C)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Pre-processing<br>Requirements                     | Cleaned, rinsed, and dried devices                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Devices                                            | Reusable metal and non-metal medical<br>devices that are used in healthcare facilities,<br>including those that are sensitive to heat and<br>moisture                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Sterilization Cycles                               | One (1) pre-programmed; approximately 36<br>minutes                                                                                                                                                                                                                                                                      | Two (2) pre-programmed;<br>Lumen (approximately 72 minutes) and<br>Non-Lumen (approximately 40 minutes)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                    | Sterilant                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Model Name                                         | STERLOAD™                                                                                                                                                                                                                                                                                                                | Sterilant Disc                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Type                                               | Cassette type (unit dose)                                                                                                                                                                                                                                                                                                | Disk type                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Sterilant                                          | 59% aqueous solution of hydrogen peroxide                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                                    | Monitoring accessories                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Biological Indicator                               | Self-contained biological indicator,<br>Geobacillus stearothermophilus                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Process Challenge<br>Device /<br>Routine Test Pack | Self-contained biological indicator,<br>Geobacillus stearothermophilus                                                                                                                                                                                                                                                   | Same…

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/MLR/K212200](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/MLR/K212200)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
