← Product Code [LYY](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYY) · K222693

# Single-use medical latex examination gloves (LG100) (K222693)

_Fitone Latex Products Co., Ltd. Guangdong · LYY · Sep 24, 2022 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYY/K222693

## Device Facts

- **Applicant:** Fitone Latex Products Co., Ltd. Guangdong
- **Product Code:** [LYY](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYY.md)
- **Decision Date:** Sep 24, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

The Single-use medical latex examination gloves is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## Device Story

Single-use medical latex examination gloves (LG100) are disposable, non-sterile natural rubber latex gloves. Used by clinicians in medical settings to provide a physical barrier between the examiner's hand and the patient, preventing cross-contamination. The device is worn on the hand or finger during examinations. Effectiveness is established through physical barrier testing, biocompatibility, and adherence to ASTM standards. Benefits include protection against contamination during routine medical procedures.

## Clinical Evidence

No clinical data. Bench testing only. Testing included physical dimensions, physical properties (tensile strength, elongation), water leak test (ASTM D5151-19), residual powder content (ASTM D6124-06), protein content (ASTM D5712-15), and biocompatibility (ISO 10993 series).

## Technological Characteristics

Material: Natural rubber latex. Form factor: Non-sterile, powder-free, five sizes (XS-XL). Standards: ASTM D3578-19 (physical properties/dimensions), ASTM D5151-19 (barrier), ASTM D6124-06 (powder), ASTM D5712-15 (protein). Biocompatibility: ISO 10993-1, 10, 11. Shelf life: 3 years.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Examination gloves-Type A (Latex gloves) ([K214017](/device/K214017.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 24, 2022

Fitone Latex Products Co., Ltd. Guangdong % Stuart Situ Director Landlink Healthcare Technology (Shanghai) Co., Ltd. Room1308,Baohua International Plaza,West Guangzhong Road 555 Jingan District Shanghai. 200072 China

Re: K222693

Trade/Device Name: Single-use medical latex examination gloves (LG100) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LYY Dated: August 30, 2022 Received: September 6, 2022

Dear Stuart Situ:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian.M.D.,Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

Submission Number (if known)

K222693

Device Name

Single-use medical latex examination gloves (LG100)

Indications for Use (Describe)

The Single-use medical latex examination gloves is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary - K222693

### I. Submitter

Fitone Latex Products Co., Ltd. Guangdong No.5 Huitong road, Lingbei Industrial Zone, Suixi, 524338 Zhanjiang, Guangdong, China

Contact person: Christine Ou Position: Manager Tel.: 0759-7905808 E-mail: market-intl@fitonelatex.com Preparation

date: September 20, 2022

### Submission Correspondent

Ms. Stuart Situ Landlink Healthcare Technology (Shanghai) Co., Ltd. E-mail: stuart.situ@landlink-healthcare.com

### US Agent

Qihui Zhang ZYPPEL LLC 1337 Massachusetts Avenue #158 Arlington MA, MA US 02476

### II. Proposed Device

| Device Trade Name: | Single-use medical latex examination gloves |
|--------------------|---------------------------------------------|
| Model:             | LG100                                       |
| Common name:       | Latex Patient Examination Glove             |
| Regulation Number: | 21 CFR 880.6250                             |
| Regulatory Class:  | Class I                                     |
| Product code:      | LYY                                         |
| Review Panel:      | General Hospital                            |

### III. Predicate Devices

| 510(k) Number: | K214017                                  |
|----------------|------------------------------------------|
| Trade name:    | Examination gloves-Type A (Latex gloves) |
| Common name:   | Latex Patient Examination Glove          |

{4}------------------------------------------------

| Classification: | Class I                                                          |
|-----------------|------------------------------------------------------------------|
| Product Code:   | LYY                                                              |
| Manufacturer    | Jiangxi Kemei Medical Apparatus & Instruments Group<br>Co., Ltd. |

## IV. Device description

The proposed device. The Single-use medical latex examination qloves is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

The proposed device is made of natural rubber latex and meet all the current specifications listed under the ASTM Specification D3578-19, Standard Specification for Rubber Examination Gloves. The principal operation and mechanism of this device is to prevent contamination between patient and examiner and this principle is achieved through testing of barrier, physical properties and other testing stated in the performance data. This device is provided non-sterile, in sizes XS, S, M, L and XL, natural white color, powder free, and the shelf life is 3 years.

### V. Indication for use

The Single-use medical latex examination gloves is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

| Item                   | Proposed device<br>(K222693)                                                                                                                                                                                                   | Predicate device<br>(K214017)                                                                                                                                                                                                   | Discussion |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product<br>name        | Single-use medical latex<br>examination gloves                                                                                                                                                                                 | Examination gloves-Type<br>A(Latex gloves)                                                                                                                                                                                      | -          |
| Indications<br>For Use | The Single-use medical<br>latex examination gloves is a<br>disposable device intended<br>for medical purposes that is<br>worn on the examiner's hand<br>or finger to prevent<br>contamination between<br>patient and examiner. | The Examination gloves-<br>Type A (Latex gloves) is<br>a disposable device<br>intended for medical<br>purposes that is worn on<br>the examiner's hand or<br>finger to prevent<br>contamination between<br>patient and examiner. | Same       |
| Product<br>Code        | LYY                                                                                                                                                                                                                            | LYY                                                                                                                                                                                                                             | Same       |

#### VI. Comparison of technological characteristics with the predicate devices Table 1 General Comparison of Single-use medical latex examination gloves

{5}------------------------------------------------

| Regulation No.               | 21 CFR 880.6250                                                                                                  | 21 CFR 880.6250                                                        | Same                                                       |
|------------------------------|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|------------------------------------------------------------|
| Class                        | I                                                                                                                | I                                                                      | Same                                                       |
| Powdered or<br>Powdered free | Powdered free                                                                                                    | Powdered free                                                          | Same                                                       |
| Material                     | Natural Rubber Latex                                                                                             | Natural Rubber Latex                                                   | Same                                                       |
| Color                        | No colour pigment added.<br>Natural White                                                                        | No colour pigment added.<br>Natural White                              | Same                                                       |
| Dimensions                   | Length:<br>6(XS): 238-241 mm<br>61/2(S):240-245mm<br>71/2(M): 240-245mm<br>81/2(L):246-250mm<br>9(XL): 246-252mm | Length:<br>S: 240-244mm<br>M: 241-245mm<br>L: 242-245mm                | Similar<br>Meeting<br>requirem<br>ent of<br>ASTM D<br>3578 |
|                              | Width:<br>6(XS): 74-76 mm<br>61/2 (S):80-84mm<br>71/2 (M): 94-97mm<br>81/2 (L): 105-108mm<br>9(XL): 114-120mm    | Width:<br>S: 84-86mm<br>M: 96-98mm<br>L: 105-108mm                     | Similar<br>Meeting<br>requirem<br>ent of<br>ASTM D<br>3578 |
|                              | Palm Thickness: 0.11-<br>0.13mm<br>Finger Thickness: 0.13-<br>0.16mm                                             | Palm Thickness: 0.11-<br>0.13mm<br>Finger Thickness: 0.13-<br>0.14mm   | Similar<br>Meeting<br>requirem<br>ent of<br>ASTM D<br>3578 |
| Biocompatibility             | Non-sensitizing<br>Non-irritating<br>Non-systemic toxicity                                                       | Non-sensitizing<br>Non-irritating<br>Non-systemic toxicity             | Same                                                       |
| Tensile<br>strength          | Before Aging $24.4$ MPa,<br>min<br>Stress at<br>500 %<br>Elongation                                              | Before<br>Aging $20.6$ MPa,<br>min<br>Stress at<br>500 %<br>Elongation | Similar<br>Meeting<br>requirem<br>ent of<br>ASTM D         |
|                              |                                                                                                                  | $4.2$ Mpa,<br>max                                                      | $5.1$ Mpa,<br>max                                          |

{6}------------------------------------------------

|                     | After Aging                          | 20.5 MPa, min | After Aging                          | 18.9MPa, min | 3578     |
|---------------------|--------------------------------------|---------------|--------------------------------------|--------------|----------|
| Ultimate Elongation | Before Aging                         | 650%, min     | Before Aging                         | 683%, min    |          |
|                     | After Aging                          | 524%, min     | After Aging                          | 623%, min    |          |
| Freedom from Holes  | Meets ASTM D5151-19 AQL=2.5          |               | Meets ASTM D5151-19 AQL=2.5          |              | Same     |
| Protein Content     | Meets ASTM D5712-2015                |               | Meets ASTM D5712-2015                |              | Same     |
| Powdered residue    | Meets ASTM D6124-06(Reapproved 2017) |               | Meets ASTM D6124-06(Reapproved 2017) |              | Same     |
| Sterility           | Non-sterile                          |               | Non-sterile                          |              | Same     |
| Shelf life          | 3 Years                              |               | 5 Years                              |              | Similar¹ |
| Single Use          | Yes                                  |               | Yes                                  |              | Same     |
| Label and labeling  | Meet FDA's Requirement               |               | Meet FDA's Requirement               |              | Same     |

1 The Shelf life of proposed device and predicate device is different; However, the product performance after 3 years has been determined based on accelerated aging study. The results can demonstrate that the performance of proposed device meets the ASTM D3578 requirements after 3 years accelerated aging.

# VII. Non-Clinical Testing

Non clinical tests were conducted in accordance with following standards to verify that the proposed device met all design specifications.

- . ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves
- . ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves
- . ASTM D3578-19 Standard Specification for Rubber Examination Gloves
- . ASTM D5712-15 Standard Test Method for Analysis of Aqueous Extractable Protein in Natural Rubber in Latex, Natural Rubber, and Elastomeric Products Using the Modified Lowry Method
- . ASTM D7160-05 Standard Practice for Determination of Expiration Dating for medical Gloves
- ISO 10993-1: 2009 Biological evaluation of medical devices Part1: Evaluation and .

{7}------------------------------------------------

testing within a risk management process

- ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for • irritation and skin sensitization
- . ISO 10993-11:2017 Biological Evaluation of Medical Devices - Part 5: Tests For systemic toxicity

| Test Methodology        | Purpose                                   | Acceptance Criteria                                                                                                                                                                                                                                                                                                                                                            | Results                                        |
|-------------------------|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| ASTM D3578-19           | Physical Dimension                        | The actual measured<br>dimension of the<br>gloves shall be meet<br>the stated tolerance<br>specified in Table 2 of<br>the ASTM D3578-19<br>Width ± 10mm<br>Length XS-S 220 min<br>and M-XL 230 min<br>Thickness (finger and<br>Palm): 0.08 min                                                                                                                                 | Meet the<br>requirement                        |
| ASTM D3578-19           | Determination of<br>Physical Properties   | Before and after<br>accelerated aging, the<br>gloves shall conform<br>to the physical<br>requirements in the<br>Table 3 of ASTM<br>D3578-19<br>Before Aging:<br>Tensile Strength: 18<br>MPa min<br>Stress Elongation: 5.5<br>MPa max<br>Ultimate Elongation:<br>650% min<br>After Accelerated<br>Aging:<br>Tensile Strength: 14<br>MPa min<br>Ultimate Elongation:<br>500% min | Meet the<br>requirement                        |
| ASTM D5151-19           | Water Leak Test for<br>Detection of Holes | The gloves shall be<br>free from hole when<br>tested in accordance<br>with the method given<br>in ASTM D5151<br>AQL=2.5                                                                                                                                                                                                                                                        | Meet the<br>requirement                        |
| ASTM D6124-<br>06(2017) | Residual Powder<br>Content Test           | The powder residue<br>content shall be not<br>more than 2mg per<br>glove.                                                                                                                                                                                                                                                                                                      | Meet the<br>requirement                        |
| ASTM D3578-19           | Protein Content                           | $\leq$ 200 ug/dm²                                                                                                                                                                                                                                                                                                                                                              | Meet the<br>requirement                        |
| ISO 10993-<br>5:2009    | In vitro Cytotoxicity                     | Non-cytotoxicity                                                                                                                                                                                                                                                                                                                                                               | Failed<br>Potential toxicity<br>Grade (3 or 4) |
| ISO 10993-10:<br>2010   | Skin Sensitization                        | Non-sensitizing                                                                                                                                                                                                                                                                                                                                                                | Meet the<br>requirement                        |
| ISO 10993-10:<br>2010   | Skin irritation                           | Non-irritating                                                                                                                                                                                                                                                                                                                                                                 | Meet the<br>requirement                        |
| ISO 10993-11:<br>2017   | Systemic toxicity                         | Non-systemic toxicity                                                                                                                                                                                                                                                                                                                                                          | Meet the<br>requirement                        |

{8}------------------------------------------------

# VIII. Clinical Testing

No clinical study is included in this submission.

### IX. Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, the Single-use medical latex examination gloves is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K214017.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYY/K222693](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYY/K222693)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
