K212438 · Thai Rubber Industry Company Limited · LYY · Dec 7, 2021 · General Hospital
Device Facts
Record ID
K212438
Device Name
Comfortpro
Applicant
Thai Rubber Industry Company Limited
Product Code
LYY · General Hospital
Decision Date
Dec 7, 2021
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
Comfortpro latex examination powder free gloves is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
Device Story
Comfortpro is a disposable, non-sterile, powder-free natural rubber latex examination glove. It is worn by healthcare examiners to provide a barrier against contamination between the patient and the examiner. The device is used in clinical or medical settings. It functions as a physical barrier. The healthcare provider dons the glove manually before patient interaction. The device benefits the patient and provider by reducing the risk of cross-contamination during medical examinations.
Clinical Evidence
No clinical data. Bench testing only. Performance verified against ASTM D3578-19 (dimensions, physical properties), ASTM D5151-19 (water-tightness/holes), ASTM D6124-06 (residual powder), and ASTM D5712-15 (extractable protein). Biocompatibility testing performed per ISO 10993-10 (irritation/sensitization), ISO 10993-5 (cytotoxicity), and ISO 10993-11 (acute systemic toxicity).
Technological Characteristics
Material: Natural rubber latex. Form factor: Disposable, non-sterile, powder-free examination glove. Dimensions and physical properties (tensile strength, elongation) meet ASTM D3578-19. Biocompatibility: Complies with ISO 10993-5, 10993-10, and 10993-11. No electronic components or software.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand to prevent contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Palm Care Latex Examination Powder Free Gloves (K202377)
K982872 — TG MEDICAL POWDER-FREE LATEX EXAMINATION GLOVE PROTEIN LABELING CLAIM (50 MICROGRAMS OR LESS) · Tg Medical Sdn. Bhd. · Oct 15, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 7, 2021
Thai Rubber Industry Company Limited % Manoj Zacharias Consultant Liberty Management Group Ltd. 75 Executive Dr. STE 114, Aurora, Illinois 60504
Re: K212438
Trade/Device Name: Comfortpro Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LYY Dated: November 4, 2021 Received: November 4, 2021
Dear Manoj Zacharias:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K212438
Device Name Comfortpro
Indications for Use (Describe)
Comfortpro latex examination powder free gloves is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### A. APPLICANT INFORMATION
| 510(K) Owner's Name | THAI RUBBER INDUSTRY COMPANY LIMITED |
|---------------------|--------------------------------------------------------------|
| Address | 738 MOO 5, MANAM KOO,<br>PLUAKDAENG<br>RAYONG 21140 THAILAND |
| Phone | +66-81-6298773, +66-85-6000373 |
| Fax | - |
| E-mail | chawalit@thairubberindustry.com |
| Contact Person | Mr. Chawalit Tiyadechachai |
| Designation | Owner and Managing Director |
| Contact Number | +66-81-6298773, +66-85-6000373 |
| Contact Email | chawalit@thairubberindustry.com,<br>Chawalit.tiya@gmail.com |
| Date Submitted | July 15, 2021 |
#### B. DEVICE IDENTIFICATION
| Name of the device | COMFORTPRO |
|-----------------------------------|--------------------------------------|
| Product proprietary or trade name | COMFORTPRO |
| Common or usualname | Latex examination powder free gloves |
| Classification name | Latex Patient Examination Glove |
| Device Classification | Class-1 |
| Product Code | LYY |
| Regulation Number | 21 CFR 880.6250 |
| Review Panel | General Hospital |
#### C. PREDICATE DEVICE
| Predicate Device | Hi-Care Thai Gloves Co. Ltd. |
|------------------|------------------------------|
| 510(k) Number | K202377 |
| Regulatory Class | Class 1 |
| Product code | LYY |
### D. DESCRIPTION OF THE DEVICE:
COMFORTPRO Latex Examination Powder Free Gloves are manufactured to meet all the current specifications listed under the ASTM Specification D3578-19, Standard Specification for Rubber Examination Gloves. They are made from Natural Rubber Latex. These gloves are Natural in color (No color added) and are powder free.
{4}------------------------------------------------
#### E. INDICATIONS FOR USE OF THE DEVICE:
COMFORTPRO latex examination powder free gloves is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
#### F. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE
| CHARACTERSTICS | STANDARDS | DEVICE PERFORMANCE | | Comparison | |
|------------------------------------------|--------------|-----------------------------------------------------------------------------------------------|----------------------------------------------------|-----------------|---------|
| | | PREDICATE | CURRENT | | |
| 510(k) Number | -- | K202377 | K212438 | | |
| Name of device | -- | Palm Care Latex<br>Examination Powder<br>Free Gloves | ComfortPro latex Examination<br>Powder Free Gloves | -- | |
| Dimensions- Length | ASTMD3578-19 | Length > 230 mm | Length > 230 mm | Similar | |
| | | | Size | Average | |
| | | | X-Small | 245 mm | |
| | | | Small | 238 mm | |
| | | | Medium | 233 mm | |
| | | | Large | 240 mm | |
| Dimensions- Width | ASTMD3578-19 | Width Min 95+/- 10<br>mm (for medium<br>size) | Width Min 95+/-10<br>mm (for medium size) | Similar | |
| | | | Size | Average | |
| | | | X-Small | 77 mm | |
| | | | Small | 81 mm | |
| | | | Medium | 93 mm | |
| | | | Large | 101 mm | |
| Physical Properties-<br>Tensile Strength | ASTMD3578-19 | Before Ageing<br>Tensile Strength<br>> 18 Mpa After<br>Ageing Tensile<br>Strength<br>> 14 Mpa | Before Ageing<br>Tensile Strength > 18 Mpa | Similar | |
| | | | Size | Actual value | |
| | | | X-Small | 29.6 | |
| | | | Small | 25 | |
| | | | Medium | 26 | |
| | | | Large | 24.9 | |
| | | | After Ageing<br>Tensile Strength > 14 Mpa | | Similar |
| | | | Size | Actual<br>value | |
| | | | X-Small | 25.5 | |
| | | | Small | 24.1 | |
| | | | Medium | 22.2 | |
| | | | Large | 24.4 | |
{5}------------------------------------------------
| CHARACTERSTICS | STANDARDS | DEVICE PERFORMANCE | | Comparison | | | | | | | | | | | | | | | | | | | | |
|---------------------------------------------|---------------------------------------------------------|----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|---------|--|--|--|--|---------|
| | | PREDICATE | CURRENT | | | | | | | | | | | | | | | | | | | | | |
| 510(k) Number | -- | K202377 | K212438 | | | | | | | | | | | | | | | | | | | | | |
| Physical Properties-<br>Ultimate Elongation | ASTMD3578-19 | Before Ageing<br>Ultimate<br>Elongation<br>> 650%<br><br>After Ageing<br>Ultimate<br>Elongation<br>>500% | Before Ageing<br>Ultimate Elongation > 650%<br>Size Actual value X-Small 810 Small 760 Medium 680 Large 780<br>After Ageing<br>Ultimate Elongation > 500% Size Actual value X-Small 820 Small 760 Medium 650 Large 760 | | | | | | | | | | | | | | | | | | | | | Similar |
| Thickness | ASTMD3578-19 | Palm > 0.08 mm<br>Finger > 0.08 mm | Palm > 0.08 mm<br>Finger > 0.08 mm Size Palm (Actual value) Finger (Actual value) X-Small 0.102 mm. 0.114 mm. Small 0.087 mm. 0.109 mm. Medium 0.10 mm. 0.122 mm. Large 0.108 mm. 0.121 mm. | | | | | | | | | | | | | | | | Similar | | | | | |
| Powder Free Residue | ASTMD3578-19 | $\leq$ 2 mg/glove | 0.68 mg/glove | Similar | | | | | | | | | | | | | | | | | | | | |
| Biocompatibility | Primary Skin<br>Irritation-ISO<br>10993-10:2010(E) | Under the<br>condition of<br>study, not an<br>irritant | Under the condition of study<br>not an irritant | Same | | | | | | | | | | | | | | | | | | | | |
| | Dermal<br>Sensitization-ISO<br>10993-10:2010(E) | Under the<br>conditions of the<br>study, not a<br>sensitizer | Under the conditions of the<br>study, not a sensitizer | Same | | | | | | | | | | | | | | | | | | | | |
| | In vitro<br>cytotoxicity<br>ISO10993-5<br>:2009(E) | Under the<br>conditions<br>of the study, non-<br>cytotoxic | Under the conditions of the<br>study, non-cytotoxic | Same | | | | | | | | | | | | | | | | | | | | |
| | Acute Systemic<br>Toxicity Test ISO<br>10993-11:2017(E) | Under the<br>conditions of study<br>the device extracts<br>do not pose a<br>systemic toxicity<br>concern | Under the conditions<br>of study the device extracts<br>do not pose a systemic<br>toxicity concern | same | | | | | | | | | | | | | | | | | | | | |
{6}------------------------------------------------
| CHARACTERSTICS | STANDARDS | DEVICE PERFORMANCE | | COMPARISON<br>ON |
|-----------------------|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| | | PREDICATE | CURRENT | |
| 510(k) Number | | K202377 | K212438 | |
| Water Tight (1000 ml) | ASTM D5151-19<br>AQL-2.5 | Passes | Passes | Similar |
| Indication for Use | | Latex Examination<br>Powder Free Gloves are<br>disposable devices intended<br>for medical purpose that are<br>worn on the examiner's<br>hand to prevent<br>contamination between<br>patient and examiner. | ComfortPro latex<br>examination powder free<br>gloves is disposable<br>device intended for<br>medical purpose that is<br>worn on the examiner's<br>hand to prevent<br>contamination<br>between patient and<br>examiner. | Same |
| Material | | Natural Latex | Natural Latex | Identical |
| Color | | Natural (No color is added) | Natural color (No color is<br>added) | Similar |
| Size | ASTMD3578-19 | X Small, Small, Medium,<br>Large | X Small, Small,<br>Medium, Large | Similar |
| Single Use | Medical Glove<br>Guidance<br>Manual-<br>Labeling | Single Use | Single Use | Same |
| Sterile/nonsterile | | Nonsterile | Nonsterile | Same |
| Powder/Powder<br>free | | Powder free | Powder free | Same |
| Label and<br>Labeling | FDA Label<br>requirements | Meets FDA's label and<br>labeling requirements | Meets FDA's<br>label and labeling<br>requirements | Same |
| Manufacturer(s) | | Hi-Care Thai Gloves Co.<br>Ltd. | THAI RUBBER<br>INDUSTRY COMPANY<br>LIMITED | --- |
Both devices meet the ASTM standard D3578.
{7}------------------------------------------------
## G. COMPARISON BASED ON AN ASSESSMENT OF NON-CLINICAL PERFORMANCE DATA
### BENCH TEST DATA
| TEST METHOD | PURPOSE | ACCEPTANCE CRITERIA | RESULT | | | | | | | | | | | | | | | |
|--------------------------------------------------------------------------------|--------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| ASTM D3578-19<br>Standard<br>Specification for<br>Rubber Examination<br>Gloves | To determine the<br>length of the gloves | Min 230 mm for all<br>sizes | X-Small : 245 mm<br>Small : 238 mm<br>Medium : 233 mm<br>Large : 240 mm | | | | | | | | | | | | | | | |
| ASTM D3578-19<br>Standard<br>Specification for<br>Rubber Examination<br>Gloves | To determine the<br>width of the gloves | X-Small : 70+/-10 mm<br>Small : 80+/-10 mm<br>Medium : 95+/-10 mm<br>Large : 111+/-10 mm | X-Small : 77 mm<br>Small : 81 mm<br>Medium : 93 mm<br>Large : 101 mm | | | | | | | | | | | | | | | |
| ASTM D3578-19<br>Standard<br>Specification for<br>Rubber Examination<br>Gloves | To determine the<br>thickness of the<br>gloves | Palm<br>0.08 mm min for all<br>sizes<br>Finger<br>0.08 mm min for all<br>sizes | Size Palm Finger X-Small 0.102 mm 0.114 mm Small 0.087 mm 0.109 mm Medium 0.10 mm 0.122 mm Large 0.108 mm 0.121 mm | | | | | | | | | | | | | | | |
| ASTM D3578-19<br>Standard<br>Specification for<br>Rubber Examination<br>Gloves | To determine the<br>physical properties-<br>Tensile strength | <b>Before Ageing</b><br>Tensile Strength<br>18MPa Min for all<br>sizes<br><b>After Ageing</b><br>Tensile Strength<br>14MPa Min for all<br>sizes | Size Before<br>ageing After<br>ageing X-Small 29.6 MPa 25.5 MPa Small 25 MPa 24.1 MPa Medium 26 MPa 22.2 MPa Large 24.9 MPa 24.4 MPa | | | | | | | | | | | | | | | |
| | To determine the<br>physical properties-<br>Ultimate<br>Elongation | <b>Before Ageing</b><br>Ultimate Elongation<br>650% Min for all sizes<br><b>After Ageing</b><br>Ultimate Elongation<br>500% Min for all sizes | Size Before<br>ageing After<br>ageing X-Small 810% 820% Small 760% 760% Medium 680% 650% Large 780% 760% | | | | | | | | | | | | | | | |
{8}------------------------------------------------
| TEST METHOD | PURPOSE | ACCEPTANCE<br>CRITERIA | RESULT |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|------------------------|------------------------|
| ASTM D5151-19 Standard<br>Test Method for Detection<br>of Holes in Medical Gloves | To determine the holes<br>in the gloves | AQL 2.5 | Gloves Passes AQL 2.5 |
| ASTM D6124-06<br>(Reapproved 2017) Standard<br>Test Method for Residual<br>Powder on Medical Gloves | To determine the<br>residual powder in the<br>gloves | ≤2 mg/glove | Medium : 0.68 mg/glove |
| ASTM D5712-15<br>(Reapproved 2020) Standard<br>Test Method for Analysis of<br>Aqueous Extractable Protein<br>in Latex, Natural Rubber, and<br>Elastomeric Products Using<br>the Modified Lowry Method | To determine the<br>extractable protein in<br>the gloves. | 200 µg/ dm² Max | Medium : 187.1 µg/ dm² |
### BIOCOMPATIBILITY DATA
| TEST METHOD | PURPOSE | ACCEPTANCE<br>CRITERIA | RESULT |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 10993-10:2010(E)<br>Biological Evaluation Of<br>Medical Devices - Part 10,<br>Tests for Irritation and Skin<br>Sensitization.<br>Test done for irritation. | To evaluate the test<br>item, for skin irritation<br>test in New Zealand<br>White rabbits. | Under the condition<br>of study, not an<br>irritant | Under the condition of<br>study, not an irritant |
| ISO 10993 10:2010(E)<br>Biological Evaluation of<br>Medical Devices - Part 10,<br>Tests for Irritation and Skin<br>Sensitization.<br>Test done for skin sensitization | To evaluate the test<br>item, for the skin<br>sensitization in Guinea<br>pigs by maximization<br>test. | Under the conditions<br>of the study, not a<br>sensitizer | Under the conditions of<br>the study, not a<br>sensitizer |
| ISO10993-5:2009(E)<br>Biological Evaluation of<br>Medical Devices - Part 5,<br>Tests for In Vitro<br>Cytotoxicity. | To evaluate the test<br>item, for its ability to<br>induce cytotoxicity<br>using L-929 mouse<br>fibroblast cells by<br>Elution Method. | Under the conditions<br>of the study, non-<br>cytotoxic | Under the conditions of the<br>study cytotoxic for 100%<br>(neat) test item extract. As<br>a follow up, acute<br>systemic toxicity testing<br>was performed to<br>demonstrate the extract<br>did not present an acute<br>systemic toxicity concern. |
| ISO 10993-11:2017(E)<br>Biological Evaluation of<br>Medical Devices - Part 11,<br>Tests for Systemic Toxicity. | To evaluate the test<br>item, for acute systemic<br>toxicity in Swiss Albino<br>Mice. | Under the conditions<br>of study, the device<br>extracts do not pose a<br>systemic toxicity<br>concern | Under the conditions<br>of study the device<br>extracts do not pose a<br>systemic toxicity concern |
{9}------------------------------------------------
# 510(K) SUMMARY
#### AS REQUIRED BY: 21CFR§807.92(C)
The performance test data of the non-clinical tests meet following standards:
ASTM D3578-19 Standard Specification for Rubber Examination Gloves
ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves
ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medica1Gloves
ASTM D5712-15 (Reapproved 2020) Standard Test Method for Analysis of Aqueous Extractable Protein in Latex, Natural Rubber, and Elastomeric Products Using the Modified Lowry Method
ISO 10993-10:2010 (E)Biological Evaluation of Medical Devices - Part 10, Tests for Irritation and Skin Sensitization.
ISO 10993-5:2009(E) Biological Evaluation of Medical Devices - Part 5, Tests for In Vitro Cytotoxicity.
ISO 10993-11:2017 (E) Biological Evaluation of Medical Devices - Part 11, Tests for Systemic Toxicity.
#### H. COMPARISON BASED ON ASSESSMENT OF CLINICAL PERFORMANCE DATA
Not applicable - Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.
#### I. CONCLUSION
The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission ComfortPro Latex Examination Powder Free Gloves is as safe, as effective, and performs as well as or better than the legally marketed predicate device K202377.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.