RMI DUAL FLUID IRRIGATING SYRINGE

K964833 · Research Medical, Inc. · KYZ · Feb 19, 1997 · General Hospital

Device Facts

Record IDK964833
Device NameRMI DUAL FLUID IRRIGATING SYRINGE
ApplicantResearch Medical, Inc.
Product CodeKYZ · General Hospital
Decision DateFeb 19, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6960
Device ClassClass 1

Device Story

Device consists of two 10cc syringes pre-assembled in PVC holster with dual lumen applicator tip; uses two separate lines of Tecoflex tubing to deliver two solutions simultaneously to surgical site. Plunger ends coupled with LDPE yoke to ensure 1:1 ratio. Includes three-way stopcocks and fluid transfer sets for drawing solutions. Used in surgical settings by clinicians to deliver fluids. Benefits include simultaneous, equal-ratio delivery of two fluids, improving upon manual handling of separate syringes.

Clinical Evidence

Bench testing only. Biocompatibility testing included primary skin irritation, sensitization (weak sensitizer), cytotoxicity (MEM elution and agar overlay), acute systemic toxicity, hemocompatibility (hemolysis and thrombogenicity), pyrogenicity, muscle implantation, and AMES mutagenicity assay. Functional performance testing included leak tests at 2 PSI, flow tests, and pull tests (exceeding 5 lbs). Performance testing confirmed 1:1 ratio delivery compared to manual syringe methods.

Technological Characteristics

Materials: PVC holster, Tecoflex tubing, polycarbonate stopcocks, LDPE yoke, PVC/ABS fluid transfer sets. Dual lumen configuration. Sterilization: Ethylene oxide (EtO). Single-use, disposable. Manual operation via yoke-coupled plungers.

Regulatory Classification

Identification

An irrigating syringe is a device intended for medical purposes that consists of a bulb or a piston syringe with an integral or a detachable tube. The device is used to irrigate, withdraw fluid from, or instill fluid into, a body cavity or wound.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} k964833 FEB 19 1997 # Attachment 11.A. ## SMDA Summary of Safety and Effectiveness Information In compliance with requirements of the Safe Medical Device Act (SMDA) of 1990, the following information is submitted as a summary of safety and effectiveness information for this 510(k) premarket notification: ### 1. Product: RMI Dual Fluid Irrigating Syringe Manufactured by: Research Medical, Inc. 6864 South 300 West Midvale, UT 84047 Phone: (801) 562-0200 FAX: (801) 562-1122 ### 2. Description: The RMI Dual Fluid Irrigating Syringe consists of two Becton Dickinson 10cc syringes pre-assembled in a PVC holster. The dual lumen applicator tip of the Dual Fluid Irrigating Syringe consists of two separate lines of tecoflex tubing that form a ‘Y’ connection, then remain parallel to deliver two separate solutions or fluids simultaneously onto the surgical site. A three-way stopcock is connected to each syringe to accommodate attachable 36” fluid transfer sets used to draw prescribed solutions or fluids into each syringe. The plunger ends of the 10cc syringes are coupled together with an LDPE yoke. The RMI Dual Fluid Irrigating Syringe is ethylene oxide (EtO) sterilized and is sold as a disposable, single use, non-pyrogenic device. ### 3. Component Materials: The RMI Dual Fluid Irrigating Syringe is composed of the following fluid contacting components and materials: | Component | Materials | | --- | --- | | (2) 10cc SLP LL BD#301030 syringes | Standard Becton Dickinson | | (2) Three-way stopcocks | Polycarbonate | | (2) Parallel fluid pathway lines | Tecoflex tubing | | (2) Gish Fluid Transfer Set, CPVMF36S | PVC tubing, PVC female luer-lock, polycarbonate male luer-lock, and ABS end caps | ### 4. Predicate Device Identification: A claim of substantial equivalence of the RMI Dual Fluid Irrigating Syringe is made to the following predicate devices: - Micromedics, Inc. Surgical Sealant Dispenser (SSD); 510(k) Number K881020. [Also marketed as the Micromedics Fibrijet Delivery System] Marketed by: Micromedics, Inc. 1285 Corporate Center Drive #150 Eagan, Minnesota 55121 Phone: (612) 452-1977 Fax: (612) 452-1787 {1} Duoflo™ Dispenser; 510(k) Number K872565. Marketed by: Hemaedics, Inc. 2649 Seahorn Drive Malibu, CA 90265 Phone: (213) 454-4757 ## 5. Summary of Biocompatibility: - A Primary Skin Irritation Evaluation was conducted. Under conditions of the test, the Test Article extract did not produce a skin irritation and passed the CFR primary skin irritation. - The Guinea Pig Maximization (Magnusson/Kligman Method) was performed. Under conditions of the test, the Test Article, extracted in normal saline, was a class I (weak) sensitizer in guinea pigs using the Magnusson/Kligman Guinea Pig Maximization test. A response of 1 is not considered to reflect significant sensitization. - The Cytotoxicity Test (MEM Elution Test) was performed. Under conditions of the test, the Test Article is not cytotoxic. - The Cytotoxicity Test (Agar Overlay Test) was performed. Under conditions of the test, the Test Article is not cytotoxic. All components met the USP guidelines with a grade of 2 or lower. - The Acute Systemic Toxicity Test (USP Systemic Toxicity Evaluation) was conducted. Under conditions of the test, there was no reaction of the mice to the Test Article when observed at intervals for a period of 72 hours following treatment. - The Hemocompatibility (Hemolysis Test) for blood compatibility was conducted. Under conditions of the test, the Test Article had 0.20% Hemolysis. Experience with this test on a wide range of samples indicates that less than 5% hemolysis should not be considered significant. The 5% level is based on the reproducibility and ruggedness of the assay. - The Hemocompatibility (Thrombogenicity Test) was performed to quantitate the thrombogenic potential of the Test Article. Under conditions of the test, the average clotting time for the negative control sample was 137.3 seconds. The average clotting time for the Test Article was 129.5 seconds. Summarized, the test sample did demonstrate shortened clotting time (thrombogenic risk was detected.) - The Material Mediated Pyrogenicity Test (Limulus Test) was performed. Under conditions of the test, the Test Article is not pyrogenic. - USP muscle implantation tests (Seven Day Muscle Implant Evaluation) were performed on various components of the Test Article. Under conditions of the test, all tested components of the Test Article met the requirements of the 7 day USP muscle implant test. - The Salmonella Typhimurium Reverse Mutation Assay (AMES Test) was performed. Under conditions of the test, extract of the Test Article tested against five strains did not meet the criteria for a potential mutagen. {2} # 6. Functional/Performance and Comparative Information A. Summary of Functional Testing of the RMI Dual Fluid Irrigating Syringe: | Parameter | Results | | --- | --- | | DFIS Leak Test @ 2 PSI (13 units): | Passed | | DFIS Flow Test @ 1.0 CFH (13 units) | Passed | | DFIS Pull Test -- 5 lbs. minimum Left stopcock / tubing (13 units) Right stopcock / tubing (13 units) | All pulled to 17 lbs. or more. All pulled to 17 lbs. or more. | | Gish Fluid Transfer Set, CPVMF36S | Meets all IQA acceptance criteria | B. Summary of Performance Testing of the RMI Dual Fluid Irrigating Syringe: Performance [functionality] testing shows that the RMI Dual Fluid Irrigating Syringe (DFIS) is substantially equivalent in performance to: (1) two single 10cc syringes, one syringe held in each hand; and (2) two 10cc syringes taped together held in one hand. The testing showed that the DFIS not only delivers fluids simultaneously, but in an equal 1:1 ratio. {3} C. Summary of Comparison of RMI Dual Fluid Irrigating Syringe (DFIS) to Micromedics Fibrijet Delivery System: Table of Comparison | Characteristic | Research Medical, Inc. | Micromedics | | --- | --- | --- | | Model | DFIS | SA-4310 | | Syringe Volume | (2) 10cc | (2) 10cc | | Syringe Assembly | Pre-assembled in PVC holster | Syringes clipped into assembly | | Fluid Applicator Tip | Syringe pistons yoke-coupled | Syringe pistons yoke-coupled | | Sterilization | Pre-assembled on DFIS | SA-3615 selected representative | | | Characteristics | Characteristics | | | Dual lumen tip configuration | Dual lumen tip configuration | | | Parallel .039 ID Tecoflex tubing | 20ga. malleable cannula | | | ETO | Unknown |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...