COLONGLIDE(R) LUBRICANT

K131617 · Pediatric Pharmaceuticals, Inc. · KMJ · Oct 3, 2013 · General Hospital

Device Facts

Record IDK131617
Device NameCOLONGLIDE(R) LUBRICANT
ApplicantPediatric Pharmaceuticals, Inc.
Product CodeKMJ · General Hospital
Decision DateOct 3, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6375
Device ClassClass 1

Indications for Use

Colonglide® Lubricant is intended for medical purposes for use to lubricate a body orifice of a patient in order to facilitate the entry of a diagnostic or therapeutic medical device (such as an endoscope). This device is intended for use on order of a physician.

Device Story

Colonglide® Lubricant is a clear, colorless, non-sterile, water-based patient lubricant packaged in a 250 mL plastic bottle. Used in clinical settings, typically during procedures like colonoscopies, to reduce friction and facilitate the insertion of medical devices (e.g., endoscopes). Applied by healthcare professionals to the orifice or device; acts as a physical lubricant to ease entry. Clinical benefits include increased patient safety, reduced need for external pressure maneuvers during procedures, and improved procedural efficiency. Does not involve electronic processing or automated decision-making.

Clinical Evidence

Open-label pilot study (n=30) evaluated Colonglide® during colonoscopy. Primary endpoints: successful cecal intubation (100%), median time to cecum (7 min), and reduction in external pressure. No adverse effects reported. Literature review of two prospective, randomized, controlled studies (Brocchi et al., 2005, 2008) involving 346 and 510 subjects respectively, compared standard lubrication with additional instillation methods (corn oil/water). Results showed improved cecal intubation rates and reduced procedure time/pain with additional lubrication, supporting the safety and utility of lubricants in colonoscopy.

Technological Characteristics

Water-based lubricant containing pH adjusters, thickening agents, and preservatives. Physical state: liquid/gel. pH: 5.88. Specific gravity: 1.042. Clear, odorless. Biocompatibility per ISO 10993-1:2009. Packaging: 250 mL plastic bottle meeting USP <661> requirements. Non-sterile.

Indications for Use

Indicated for patients requiring lubrication of a body orifice to facilitate the entry of diagnostic or therapeutic medical devices, such as endoscopes. Intended for use on the order of a physician.

Regulatory Classification

Identification

A patient lubricant is a device intended for medical purposes that is used to lubricate a body orifice to facilitate entry of a diagnostic or therapeutic device.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary # 1. Submitter/510(k) Holder Pediatric Pharmaceuticals, Inc. 120 Wood Avenue South Suite 200 lselin, NJ 08830 | Contact Person: | Robert C. Stites, President | |-----------------|-----------------------------| | Telephone: | (732) 603-7708 | | Fax: | (732) 603-7732 | Date Prepared: May 20, 2013 # OCT 0 3 2013 ### 2. Device Name | Trade Name: | Colonglide® Lubricant | |-----------------------|------------------------------------| | Common Name: | Patient lubricant | | Classification Name: | Lubricant, Patient, 21 CFR 880.637 | | Device Class: | Class I | | Classification Panel: | General Hospital Panel | | Product Code: | KMJ | #### 3. Predicate Devices PDI® Sterile Lubricating Jelly (Nice-Pack Products Inc., K974768) Aquagel® Lubricating Jelly (DePuy Healthcare, K951431) # 4. Device Description Colonglide® Lubricant is a clear, colorless, non-sterile, odorless, water-based patient lubricant. The subject device is packaged for individual use in a 250 mL round plastic bottle sealed with a screw cap with a foil induction innerseal liner. # 5. Indication for Use Colonglide® Lubricant is intended for medical purposes for use to lubricate a body orifice of a patient in order to facilitate the entry of a diagnostic or therapeutic medical device (such as an endoscope). This device is intended for use on order of a physician. # 6. Technological Characteristics A tabular summary comparing the technological characteristics of Colonglide® Lubricant to the predicate devices is presented in Table 5-1 below. {1}------------------------------------------------ #### 7. Non-Clinical Testing Performance testing conducted on Colonglide® Lubricant meets all the pre-determined acceptance criteria for appearance, pH, viscosity, specific gravity, preservatives, and microbial limits to support the safety and effectiveness of the finished device. A comparative pH test was performed on the device formulations for Colonglide® Lubricant and the predicate devices: the formulations were determined to be equivalent. The pH results (average) reported for Colonglide® Lubricant and the predicate devices are provided in Table 5-1. Physicochemical testing conducted on the device plastic bottle met all test method acceptance criteria in accordance with current USP requirements for Containers-Plastics <661>. The current shelf-life is 6-months with testing ongoing to support a 24-month shelf-life. Biocompatibility testing was conducted on the finished device according to ISO 10993-I :2009. The predicate device. PD19 Sterile Lubricating Jelly, was used as a control article in the biocompatibility safety tests evaluating the sensitization and irritation of Colonglide® Lubricant. The negative test results reported for both sensitization and irritation confirm that Colonglide® Lubricant is as safe as the predicate device, PD10 Sterile Lubricating Jelly. #### 8. Clinical Testing An open-label pilot study was conducted to evaluate the use of Colonglide® Lubricant in 30 patients undergoing a routine screening colonoscopy. The effectiveness of Colonglide® Lubricant to facilitate the performance of a colonoscopy was evaluated by three study investigators enrolling 10 patients each into the study. The effectiveness of Colonglide® Lubricant to facilitate the performance of a colonoscopy was reported for 57% of the total study population. Each study investigator reported that Colonglide® Lubricant facilitated the colonoscopy in at least 50% of the patients. Most patients (24/30) required either minimal or moderate amount of applied external pressure during their procedure. Successful intubation to the cecum was reported in 100% of the study patients. The median time needed to reach the cecum was 7 minutes. No adverse effects related to the use of Colonglide® Lubricant were reported by the study investigators. Also, there were no adverse effects or gastrointestinal problems reported by any of the study subjects contacted within 72 hours of the completed procedure. In a published study by Brocchi, et al. (2005), a prospective, randomized, controlled study was conducted to determine whether the performance of a colonoscopy could be facilitated by {2}------------------------------------------------ instillation of corn oil through the biopsy channel of the colonoscope during the procedure when compared to a standard lubricating method. A total of 346 subjects underwent a colonoscopy with a standard lubricating method (using water-soluble KY jelly; group A) or with the standard method plus seed oil (corn oil) instillations through the biopsy channel (group B). Successful intubation to the cecum was significantly more frequent (P < 0.005) in the corn oil lubrication group (group B, 159/168) than in the control group (group A, 145/170), and less time was needed for the corn oil group (P < 0.001). No significant differences were found with regard to time for examination at withdrawal and detection rates for colorectal diseases. Level of pain and degree of difficulty during colonoscopy were significantly lower in the com oil group (P < 0.001). In the crossover examinations done in subjects in whom total colonoscopy was not achieved, no statistical difference was found between the two groups. No adverse effects were reported for any of the study subjects in either the corn oil (group B) or the control group (group A). No damage to the instrument (i.e., colonoscope) was observed in the study. In another published study by Brocchi, et al. (2008), a prospective, randomized, controlled study was conducted in 510 eligible subjects who underwent a colonoscopy and were randomly assigned into three groups: a standard lubricating method (water-soluble KY jelly: group A, 170 subjects) was adopted in a control group, whereas the standard method plus instillation into the colon of corn seed oil (group B, 170 subjects) or warm water (group C, 170 subjects) was employed in the other groups. In group A, 6/170 subjects were excluded from the study (total colonoscopy judged "impossible") because of obstructive cancers (5 subjects) or impassable strictures (1 subject). Additionally; 4/170 subjects in group B and 7/170 in group C were excluded because of obstructive cancers (2 subjects in group B and 4 subjects in group C) and strictures (2 subjects in group B and 3 subjects in group C). Impassable strictures were always due to diverticular disease and adherences at the level of the sigmoid colon. Successful intubation to the cecum was reported in 138/164 subjects (84.1%) in group A, 155/166 subjects (93.4%) in group B, and 156/163 subjects (95.7%) in group C. The differences between groups A and B and between groups A and C were statistically significant (P < 0.01 and P < 0.001, respectively), whereas between groups B and C were not significant (P = 0.864). The median time needed to reach the cecum was significantly (P < 0.001) shorter in groups B (9.2 min) and C (9.1 min) when compared with group A (13.2 min). However, there was no significant difference in the time to reach the cecum between groups B and C. The time of withdrawal was similar in all three groups, with no significant difference. {3}------------------------------------------------ Pain and technical difficulties experienced during the procedure were significantly lower in groups B and C than in group A (P < 0.001); no difference was found between groups B and C. No adverse effects were noted in any of the subjects participating in the study. No damage to the instrument (i.e., colonoscope) was observed in the study. Thus, the use of patient lubricants in colonoscopy did not present any patient safety concerns in the study. #### 9. Conclusions Drawn from Non-Clinical and Clinical Tests Colonglide® Lubricant has the same intended use and similar technological characteristics as the predicate devices. Colonglide® Lubricant and the predicate devices are all water-soluble lubricants for medical and/or professional use. The formulation pHs for Colonglide® Lubricant and the predicates devices were evaluated and found to be equivalent. Biocompatibility testing of Colonglide® Lubricant showed it to be as safe as PDI Sterile Lubricating Jelly in both sensitization and irritation assays. The clinical investigation of Colonglide® Lubricant to facilitate the performance of a colonoscopy demonstrated: (1) increased patient safety; (2) decreased use of external pressure maneuvers; and (3) increased efficiency in performing the procedure. In clinical studies performed by Brocchi et al., the use of a standard lubricating method along with the instillation of either corn oil or warm water through the biopsy channel of a colonoscope demonstrated improved performance for facilitating the procedure when compared to the use of a standard lubricating method. No adverse effects were noted in any of the subjects participating in the studies. The similarity of the technological characteristics and performance data from comparative and biocompatibility testing confirm that Colonglide® Lubricant is substantially equivalent to the predicate devices. {4}------------------------------------------------ | | Colonglide® Lubricant | PDI® Sterile Lubricating Jelly<br>(Nice-Pack Products Inc.) | Aquagel® Lubricating Jelly<br>(DePuy Healthcare) | |-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------| | Regulatory Status | Proposed | K974768 | K951431 | | Product Code | KMJ | KMJ | KMJ | | Intended Use | Colonglide® Lubricant is<br>intended for medical purposes<br>for use to lubricate a body<br>orifice of a patient in order to<br>facilitate the entry of a<br>diagnostic or therapeutic<br>medical device (such as an<br>endoscope). This device is<br>intended for use on order of a<br>physician. | PDI® Sterile Lubricating Jelly is<br>intended for use to lubricate a<br>body orifice of a patient in order<br>to facilitate the entry of a<br>diagnostic or therapeutic<br>medical device (such as a<br>catheter, enema tip, or<br>endoscope). The device is<br>intended for use on order of a<br>physician. | Aquagel® is intended for use<br>in gynecological, digital and<br>instrument examinations and<br>general hospital procedures. | | Contains<br>Lubrication<br>Base/Solvent | Yes | Yes | Yes | | Contains pH<br>adjuster or buffer | Yes | Yes | Yes | | Contains<br>thickening agent | Yes | Yes | Yes | | Contains<br>preservatives | Yes | Yes | Yes | | Contains<br>sequestering agents | None | None | Yes | | Physical state | Liquid, gel | Liquid, paste/gel | Gel | | Clear Appearance | Yes | Yes | Yes | | Odorless | Yes | Yes | Yes | | pH* | 5.88 | 5.98 | 6.78 | | Water soluble | Yes | Yes | Yes | | Specific Gravity | 1.042 | 1.03 | 1.045 | | Biocompatible | Yes | Yes, sensitization and<br>irritation** | Yes | | Size and packaging | 250 mL plastic bottle | 2.7 g and 5.0 g packet | 142 g plastic tube, half gallon<br>(1.9 L) plastic jug with pump | | Shelf Life | 2 years | Manufacturer information is not<br>available | 2 years | | Sterile | No | Yes, gamma irradiation | No | # Table 5-1. Side-by-Side Comparison of Colonglide® Lubricant with Predicate Devices Comparative pH testing of Colonglide® Lubricant with predicate devices. * ** Comparative biocompatibility studies were conducted to evaluate the sensitivity and irritation of Colonglide® Lubricant and the predicate PDI® Sterile Lubricating Jelly. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002 #### October 3, 2013 Pediatric Pharmaceuticals, Incorporated C/O Mr. Robert M. Harris Vice President Harris FRC Corporation 2137 Route 35 HOLMDEL NJ 07733 Re: K131617 Trade/Device Name: Colonglide Lubricant Regulation Number: 21 CFR 880.6375 Regulation Name: Patient Lubricant Regulatory Class: I Product Code: KMJ Dated: August 15, 2013 Received: August 19, 2013 #### Dear Mr. Harris: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misseading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {6}------------------------------------------------ Page 2 - Mr. Harris Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (2 1 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mary S. Runner -S Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ # Indications for Use 510(k) Number (if known): Colonglide® Lubricant Device Name: Indications for Use: The Colonglide® Lubricant Is indicated for medical purposes for use to lubricate a body orifice of a patient in order to facilitate the entry of a diagnostic or therapeutic medical device (such as an endoscope). This device is intended for use on order of a physician. Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Image /page/7/Picture/10 description: The image shows the letters FDA in a stylized font. The letters are made up of small squares, giving them a pixelated appearance. The letters are arranged horizontally, with the F on the left, the D in the middle, and the A on the right. Richard C. Chapman 2013.10.03 09:56:43 -04'00' Colonglide® Lubricant, Traditional 510(k) Pediatric Pharmaceuticals, Inc. Page 4-1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...