← Product Code [KCT](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/KCT) · K061033

# MILTEX RIGID STERILIZATION CONTAINER SYSTEM, (47 COMPONENTS) (K061033)

_Miltex, Inc. · KCT · Dec 12, 2006 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/KCT/K061033

## Device Facts

- **Applicant:** Miltex, Inc.
- **Product Code:** [KCT](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/KCT.md)
- **Decision Date:** Dec 12, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6850
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Miltex Rigid Sterilization Container System is intended for use in hospitals and health care facilities to contain other medical devices that are to be sterilized and to allow sterilization of the enclosed medical devices, including surfaces and lumens, using ethylene oxide gas sterilizers. Metal lumens 3-mm in diameter and up to 200-mm in length were validated. The container is intended for use with non-porous materials such as stainless steel surgical instruments. Sterilized devices may be stored and transported within the container. The container is intended to maintain sterility of the medical devices within it until used. Containers should not be stacked when used within ethylene oxide sterilizers.

## Device Story

Miltex Rigid Sterilization Container System consists of reusable, sealed aluminum alloy, stainless steel, and silicone containers; includes perforated/non-perforated bottoms, lids, and baskets. Used in hospitals/healthcare facilities by sterile processing staff to house surgical instruments during ethylene oxide (EO) sterilization. System utilizes filter-based lid/bottom perforations to allow EO gas penetration while preventing microbial ingress. After sterilization, containers protect instruments during storage and transport until point of use. Benefits include effective sterilization packaging, sterility maintenance, and organized instrument handling. Does not support stacking during EO cycles.

## Clinical Evidence

Bench testing only. Validation included EO sterilization efficacy, aeration time, EO residuals, and sterility maintenance (30-day real-time with weekly handling). Tested with 3mm x 200mm metal lumens and inoculated stainless steel instruments. Load capacity validated up to 25 lbs.

## Technological Characteristics

Materials: Aluminum alloy, stainless steel, silicone. Design: Rigid, reusable, sealed container with filter-based lid/bottom perforations for EO gas exchange. Dimensions: Various sizes (1/2, 3/4, 1/1). Connectivity: None. Sterilization: Ethylene oxide gas. Standards: Conforms to AAMI ST 77 (Draft).

## Regulatory Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

## Predicate Devices

- SteriTite® Rigid Sterilization Container System with MediTray Products- Case Medical, Incorporated ([K023614](/device/K023614.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the word "Miltex" in a stylized, cursive font. A horizontal line extends from the top right of the "t" in "Miltex" over the "ex". The word appears to be a logo or brand name. The font is bold and black, contrasting with the white background.

K061033

DEC 1 2 2006

589 Davies Drive York, PA 17402 phone 717 840-9335 toll-free 800 221-1344 fax 717 840-9347

# 510(k) Summary [21 CFR §807.92]

Prepared: April 13, 2006

Device Trade Name: Miltex Rigid Sterilization Container System.

Device Common Name: Rigid Sterilization Container.

Classification Name: Sterilization wrap containers, trays, cassettes, and other accessories.

Class of Device: Class II device, product code KCT

Predicate Device: SteriTite® Rigid Sterilization Container System with MediTray Products- Case Medical, Incorporated- K023614

Official Contact: Charles Weaver, Regulatory Affairs Specialist

## Device Description:

The Miltex Rigid Sterilization Container System consists of a family of rigid, re-usable, sealed containers that provides an effective sterilization packaging method for medical devices. Container bottoms and lids, within a given size, are interchangeable. The system is composed of the following components:

- Container bottoms (both perforated and non-perforated versions) 트
- 제 Container baskets,
- Container lids (perforated only), and .
- u Container color-coding "labels."

The container system is designed for sterilant penetration through perforations in the lid and container bottom models that are perforated.

### Intended Use:

The Miltex Rigid Sterilization Container System is intended for use in hospitals and health care facilities to contain other medical devices that are to be sterilized and to allow sterilization of the enclosed medical devices, including surfaces and lumens, using ethylene oxide gas sterilizers. Metal lumens 3-mm in diameter and up to 200-mm in length were validated. The container is intended for use with non-porous materials such as stainless steel surgical instruments.

Sterilized devices may be stored and transported within the container. The container is intended to maintain sterility of the medical devices within it until used.

Containers should not be stacked when used within ethylene oxide sterilizers.

Page 8-1, rev 1

{1}------------------------------------------------

# Technological Characteristics:

A comparison of the technology characteristics of the Miltex Rigid Sterilization Containers to the predicate device's.

:

| Properties                                                                               | Miltex System                                  | SteriTite System                               |
|------------------------------------------------------------------------------------------|------------------------------------------------|------------------------------------------------|
| Intended use to contain instruments<br>being sterilized in ethylene oxide<br>sterilizers | Yes                                            | Yes                                            |
| Intended to be re-used                                                                   | Yes                                            | Yes                                            |
| Closed System                                                                            | Yes                                            | Yes                                            |
| Sealed                                                                                   | Yes                                            | Yes                                            |
| Design                                                                                   |                                                |                                                |
| Incorporates a filter system to permit<br>entry of sterilant agent                       | Yes                                            | Yes                                            |
| Incorporates a filter system to prevent<br>microbial migration during transport.         | Yes                                            | Yes                                            |
| Materials                                                                                |                                                |                                                |
| Container                                                                                | Aluminum alloy, Stainless<br>Steel, & Silicone | Aluminum alloy, Stainless<br>Steel, & Silicone |

{2}------------------------------------------------

# Performance Data:

A comparison of the non-clinical performance of the Miltex Rigid Sterilization Containers to the predicate device's.

| Properties                                    | Miltex System                                                                                                | SteriTite System                                                                                                                                                                                           |
|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Standards                         |                                                                                                              |                                                                                                                                                                                                            |
| Conformance to appropriate AAMI<br>standards  | Yes, conforms to AAMI ST<br>77 Draft- Containment<br>Devices for Reusable<br>Medical Device<br>Sterilization | Yes, conforms to AAMI ST 33-<br>Guidelines for the Selection and Use<br>of Reusable Rigid Sterilization<br>Container Systems for ETO<br>Sterilization and Steam Sterilization<br>in Health Care Facilities |
| Validation Testing                            |                                                                                                              |                                                                                                                                                                                                            |
| Ethylene Oxide Sterile Efficacy Testing       | Yes                                                                                                          | Yes                                                                                                                                                                                                        |
| Aeration Time and EO Residuals                | 8-hours/ below maximum<br>acceptable levels                                                                  | 12-hours/ below maximum<br>acceptable levels                                                                                                                                                               |
| Sterility Maintenance                         | 30-day real time w/ weekly<br>handling events.                                                               | 90-day real time w/ weekly handling<br>events. 30-day real time w/ daily<br>handling events.                                                                                                               |
| Load                                          | Up to 16-lbs. (small)<br>Up to 20-lbs. (med.)<br>Up to 25-lbs. (large)                                       | Up to 22-lbs.                                                                                                                                                                                              |
| Test Organisms/ Inoculated Product            |                                                                                                              |                                                                                                                                                                                                            |
| Inoculated Lumens                             | Yes--3-mm I.D. x 200-<br>mm, metal                                                                           | Yes-- 2.2-mm I.D. x 457-mm,<br>metal                                                                                                                                                                       |
| Inoculated Stainless Steel Medical<br>Devices | Yes (hinge area of medical<br>pliers and knurled<br>instruments)                                             | Yes (blades)                                                                                                                                                                                               |

# Conclusion:

The Miltex Rigid Sterilization Containers is substantially equivalent to the SteriTite Container (K023614).

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name encircling a stylized eagle emblem. The eagle is depicted with three overlapping profiles, creating a sense of depth and movement. The seal is simple, yet recognizable, representing the department's role in protecting the health of all Americans.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Charles Weaver Regulatory Affairs Specialist Miltex, Incorporated 589 Davies Drive York, Pennsylvania 17402

DEC 1 2 2006

Re: K061033

Trade/Device Name: Miltex Rigid Sterilization Container System Regulation Number: 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: November 20, 2006 Received: November 21, 2006

Dear Mr. Weaver:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Weaver

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html,

Sincerely yours.

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

#### Indications for Use

510(k) Number (if known): K061033

Device Name: Miltex Rigid Sterilization Container System

Indications for Use:

The Miltex Rigid Sterilization Container System is intended for use in hospitals and health care facilities to contain other medical devices that are to be sterilized and to allow sterilization of the enclosed medical devices, including surfaces and lumens, using ethylene oxide gas sterilizers. Metal lumens 3-mm in diameter and length up to 200-mm in length were validated. The container intended for use with non-porous materials such as stainless steel surgical instruments.

Sterilized devices may be stored and transported within the container. The container is intended to maintain sterility of the medical devices within it until used.

Containers should not be stacked when used within cthylene oxide sterilizers

The device models that are the subject of this pre-market notification on listed on page 2 of this Indications for Use statement.

Prescription Use Over-The-Counter Use XX AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

#### Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 2

Shula Murphy 15

K 061033

Page 4-1, rev 2

{6}------------------------------------------------

:

| Part Number | Description                                         | UM |
|-------------|-----------------------------------------------------|----|
| 3-5100-10   | STERI CONT 1/2 4 NON-PERFORATED BOTTOM, SILVER      | EA |
| 3-5100-13   | STERI CONT 1/2 5 NON-PERFORATED BOTTOM, SILVER      | EA |
| 3-5100-15   | STERI CONT 1/2 6 NON-PERFORATED BOTTOM, SILVER      | EA |
| 3-5110-10   | STERI CONTAINER 1/2 4 PERFORATED BOTTOM, SILVER     | EA |
| 3-5110-13   | STERI CONTAINER 1/2 5 PERFORATED BOTTOM, SILVER     | EA |
| 3-5110-15   | STERI CONTAINER 1/2 6 PERFORATED BOTTOM, SILVER     | EA |
| 3-5172-05   | STERI CONTAINER WIRE BASKET 1/2 2-INCHES            | EA |
| 3-5172-07   | STERI CONTAINER WIRE BASKET 1/2 3-INCHES            | EA |
| 3-5172-10   | STERI CONTAINER WIRE BASKET 1/2 4-INCHES            | EA |
| 3-5174-05   | STERI CONTAINER PERFORATED BASKET 1/2 2-INCHES      | EA |
| 3-5174-07   | STERI CONTAINER PERFORATED BASKET 1/2 3-INCHES      | EA |
| 3-5174-09   | STERI CONTAINER PERFORATED BASKET 1/2 4-INCHES      | EA |
| 3-5300-10   | STERI CONTAINER 3/4 4 NON-PERFORATED BOTTOM, SILVER | EA |
| 3-5300-13   | STERI CONTAINER 3/4 5 NON-PERFORATED BOTTOM, SILVER | EA |
| 3-5300-15   | STERI CONTAINER 3/4 6 NON-PERFORATED BOTTOM, SILVER | EA |
| 3-5310-10   | STERI CONTAINER 3/4 4 PERFORATED BOTTOM, SILVER     | EA |
| 3-5310-13   | STERI CONTAINER 3/4 5 PERFORATED BOTTOM, SILVER     | EA |
| 3-5310-15   | STERI CONTAINER 3/4 6 PERFORATED BOTTOM, SILVER     | EA |
| 3-5372-05   | STERI CONTAINER WIRE BASKET 3/4 2-INCHES            | EA |
| 3-5372-07   | STERI CONTAINER WIRE BASKET 3/4 3-INCHES            | EA |
| 3-5372-10   | STERI CONTAINER WIRE BASKET 3/4 4-INCHES            | EA |
| 3-5374-05   | STERI CONTAINER PERFORATED BASKET 3/4 2-INCHES      | EA |
| 3-5374-07   | STERI CONTAINER PERFORATED BASKET 3/4 3-INCHES      | EA |
| 3-5374-09   | STERI CONTAINER PERFORATED BASKET 3/4 4-INCHES      | EA |
| 3-5500-10   | STERI CONTAINER 1/1 4 NON-PERFORATED BOTTOM, SILVER | EA |
| 3-5500-13   | STERI CONTAINER 1/1 5 NON-PERFORATED BOTTOM, SILVER | EA |
| 3-5500-15   | STERI CONTAINER 1/1 6 NON-PERFORATED BOTTOM, SILVER | EA |
| 3-5510-10   | STERI CONTAINER 1/1 4 PERFORATED BOTTOM, SILVER     | EA |
| 3-5510-13   | STERI CONTAINER 1/1 5 PERFORATED BOTTOM, SILVER     | EA |
| 3-5510-15   | STERI CONTAINER 1/1 6 PERFORATED BOTTOM, SILVER     | EA |
| 3-5572-05   | STERI CONTAINER WIRE BASKET 1/1 2-INCHES            | EA |
| 3-5572-07   | STERI CONTAINER WIRE BASKET 1/1 3-INCHES            | EA |
| 3-5572-10   | STERI CONTAINER WIRE BASKET 1/1 4-INCHES            | EA |
| 3-5574-05   | STERI CONTAINER PERFORATED BASKET 1/1 2-INCHES      | EA |
| 3-5574-07   | STERI CONTAINER PERFORATED BASKET 1/1 3-INCHES      | EA |
| 3-5574-09   | STERI CONTAINER PERFORATED BASKET 1/1 4-INCHES      | EA |
| 3-5940-01   | STERI CONTAINER LABEL, SILVER                       | EA |
| 3-5940-02   | STERI CONTAINER LABEL, BLUE                         | EA |
| 3-5940-03   | STERI CONTAINER LABEL, RED                          | EA |
| 3-5940-04   | STERI CONTAINER LABEL, GOLD                         | EA |
| 3-5940-05   | STERI CONTAINER LABEL, GREEN                        | EA |
| 3-5940-06   | STERI CONTAINER LABEL, BLACK                        | EA |
| 3-6161-00   | STERI CONTAINER LID 1/2 PERFORATED, SILVER          | EA |
| 3-6361-00   | STERI CONTAINER LID 3/4 PERFORATED, SILVER          | EA |
|             |                                                     |    |

:

:

.

:

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/KCT/K061033](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/KCT/K061033)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).