K964782 · Medrad, Inc. · FOX · Feb 25, 1997 · General Hospital
Device Facts
Record ID
K964782
Device Name
MEDRAD CONTRAST HOLDER
Applicant
Medrad, Inc.
Product Code
FOX · General Hospital
Decision Date
Feb 25, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6990
Device Class
Class 1
Indications for Use
The Medrad Contrast Holders are an accessory device intended to be used in conjunction with Medrad injectors. The device is designed to aid in the transfer of common contrast media or flushing agents such as saline to the syringe. Users of this device are to be medical professionals trained in CT, MR, Angiographic Procedures, and are expected to have a good understanding of the events leading up to and including the use of the device. The Medrad Contrast Holder replaces the need for a separate contrast holder or Medrad Contrast Holder and is designed specifically for use with Medrad injectors.
Device Story
Accessory device for Medrad injector systems; mounts to injector pedestal or Counterpoise System (CS) via clamps. Features telescoping tube with height adjustment and hooks for hanging contrast/saline containers; optional accessory tray available. Used in clinical settings (CT, MR, Angiography) by trained medical professionals. Provides stable support for fluid containers during contrast media transfer to syringes. MR-compatible design allows use in magnetic resonance environments. Benefits include integration with existing injector hardware, eliminating need for separate free-standing IV stands.
Clinical Evidence
Bench testing only. Functional evaluation included adjustable height, magnetic interference, magnetic attraction, chemical resistance, rate of collapse, handling/tip testing, durability/life-cycle testing, support hook testing, installation testing, and shipping/handling testing. Testing confirmed device safety, compatibility with Medrad injector pedestals, weight-bearing capacity, and MR environment suitability.
Technological Characteristics
Materials: Aluminum pole, non-magnetic stainless steel hooks/hardware. Design: Two-section telescoping tube with hand-turned knob, clamp-based mounting system. MR-compatible. No electronic components or software.
Indications for Use
Indicated for use by medical professionals trained in CT, MR, and Angiographic procedures to hold contrast media or saline containers during injection procedures.
Regulatory Classification
Identification
The infusion stand is a stationary or movable stand intended to hold infusion liquids, infusion accessories, and other medical devices.
K031483 — MEDRAD MANUAL SYRINGE LOADER · Medrad, Inc. · Aug 1, 2003
K984418 — UNIVERSAL ADJUSTABLE T-RAIL TABLE BRACKET · Medrad, Inc. · Mar 11, 1999
K980336 — IV STANDS · Clinton Industries, Inc. · Apr 6, 1998
K042784 — MEDRAD SPECTRIS SOLARIS EP MR INJECTOR SYSTEM · Medrad, Inc. · Dec 10, 2004
K170431 — CT Syringe for Stellant Injectors (with & without Spikes, Prime Tubes, & Fill Tubes), CT Syringe Dual Pack for Stellant Injectors (with & without Spikes, Prime Tubes, & Fill Tubes), Single Fascia and Ram Tip, Dual Fascia and Ram Tip · Coeur, Inc. · Aug 18, 2017
Submission Summary (Full Text)
{0}
FEB 20 '97 03:38PM YEDRHO R&D CORP. 412 757 9999
P.8
MEDRAD
QA-96-1054-A
February 20, 1997
K964782
# SUMMARY OF SAFETY AND EFFECTIVENESS
Reference 807.92: Content and format of a 510(k) summary.
1) Ms. Sandra A. Pavlovic
Medrad, Inc.
One Medrad Drive
Indianola, PA 15051
(412) 767-2400
November 22, 1996
2) Device Name:
Trade/Proprietary Name: Medrad Contrast Holder
Common/Usual Name: Contrast Holder
Classification Name: Unknown
3) Predicate Device Information:
The Medrad Contrast Holder is compared to and is substantially equivalent to the EF Brewer IV Stand (K823270) as both devices are intended to hold IV/Contrast containers during examinations. Both the predicate device and the Medrad Contrast Holder have a hook assembly to hold the containers and are adjustable for height. The differences are that the Medrad Contrast holder is attached to the Injector mount and is made out of aluminum and stainless steel hardware that are compatible within an MR environment where as the EF Brewer IV Stand is free-standing and is made of chrome-plated steel.
4) Device Description:
The Medrad Contrast Holder will be used in conjunction with 510 (k) approved Medrad Injector systems. The Contrast Holder mounts to either a Medrad injector pedestal or the Counterpoise System (CS) by means of clamps that attach to the pedestal or CS.
The injector pedestal mounted Contrast Holder utilizes a two-section telescoping tube with a hand-turned knob to adjust the height. Two hooks are used to hang containers. One of the pedestal mounts uses two ring clamps to mount to the pedestal shaft. The other mounts to one of the pedestal legs and is supported by a clamp on the handle of the injector control unit.
The Contrast Holder for the CS uses a ring clamp to attach to the articulating counterbalance arm of the CS. A hook hangs vertically from the clamp.
{1}
FEB 20 '97 03:38PM MEDRAD R&D CORLYN 412 767 9999
P.9
MEDRAD
QA-96-1054-A
February 20, 1997
4) Device Description: (contd.)
An optional tray for holding accessories or bottles is available. The tray snaps into a clamp on the lower section of the telescoping tube and can be removed for cleaning.
5) Intended Use:
The Medrad Contrast Holders are an accessory device intended to be used in conjunction with Medrad injectors. The device is designed to aid in the transfer of common contrast media or flushing agents such as saline to the syringe. Users of this device are to be medical professionals trained in CT, MR, Angiographic Procedures, and are expected to have a good understanding of the events leading up to and including the use of the device. The Medrad Contrast Holder replaces the need for a separate contrast holder or Medrad Contrast Holder and is designed specifically for use with Medrad injectors.
6) Technological Characteristics:
There are no new technological characteristic differences between the Medrad Contrast Holder and the EF Brewer IV Stand. The most significant difference is that the Medrad Contrast Holder is made of an aluminum pole and non-magnetic stainless steel hooks and hardware allowing the device to be used in an MR environment. The EF Brewer stand is made of chrome-plated steel.
Performance Data
The Medrad Contrast Holder was tested by performing a visual inspection of the system as a well as a functional evaluation that included adjustable height, magnetic interference, magnetic attraction, chemical resistance, rate of collapse, handling/tip testing, durability and life-cycle testing, support hook testing, installation testing and shipping and handling testing. These tests were performed while the Medrad Contrast Holder was attached to the injector pedestal and with the Contrast Holder both in the retracted and extended mode. The purpose of these tests was to ensure that 1) the device is safe and effective when used as intended, 2) that the contrast holder is compatible with the Medrad injector pedestals and could maintain a known weight of contrast containers and, 3) can be safely used in an MR environment. The Medrad Contrast Holder meets all specifications and passed all tests performed indicating that the device is safe and effective for ins intended use.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.