CENTURION PROTECTIVE RESTRAINT, SCR SERIES

K080448 · Tri-State Hospital Supply Corp. · FMQ · Aug 28, 2008 · General Hospital

Device Facts

Record IDK080448
Device NameCENTURION PROTECTIVE RESTRAINT, SCR SERIES
ApplicantTri-State Hospital Supply Corp.
Product CodeFMQ · General Hospital
Decision DateAug 28, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6760
Device ClassClass 1
AttributesPediatric

Indications for Use

The Centurion® protective restraint is a device, including a wristlet, anklet, body/limb holder, or other type of strap, that is intended for medical purposes and that limits the patient's movements to the extent necessary for treatment, examination, or protection of the patient or others. This device is intended for use on neonatal/newborn patients by or on the order of a physician in a hospital or clinic setting.

Device Story

Centurion® Protective Restraint is a physical restraint device (wristlet, anklet, or body/limb holder) used in hospital or clinic settings. Operated by or under the order of a physician, it limits patient movement to facilitate medical treatment, examination, or safety. The device is constructed from hook and loop material, 100% nylon fabric with a polyester foam core, and nylon tricot backing. It functions as a mechanical barrier to restrict limb or body motion. Healthcare providers apply the restraint to neonatal patients to prevent interference with medical procedures or self-injury. The device provides a non-invasive method for patient management.

Clinical Evidence

Bench testing only. Comparative performance testing evaluated physical integrity and functional requirements against the predicate device. Results confirmed the device is as safe and effective as the predicate.

Technological Characteristics

Materials: hook and loop material, 100% nylon fabric, polyester foam core, nylon tricot backing. Form factor: wristlet, anklet, or body/limb holder straps. Mechanical restraint principle. No energy source, software, or connectivity.

Indications for Use

Indicated for neonatal/newborn patients in hospital or clinic settings to limit movement for treatment, examination, or protection. Contraindicated for patients with dislocations, fractures, open wounds on affected limbs, or compromised I.V. sites.

Regulatory Classification

Identification

A protective restraint is a device, including but not limited to a wristlet, anklet, vest, mitt, straight jacket, body/limb holder, or other type of strap, that is intended for medical purposes and that limits the patient's movements to the extent necessary for treatment, examination, or protection of the patient or others.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Koto441 Image /page/0/Picture/1 description: The image shows a logo for Tri-State Hospital Supply Corporation. The logo features a graphic of a person in the upper left corner. Below the company name is the text "MANUFACTURERS OF" and then the word "CENTURION" in large, bold letters. Below that is the text "HEALTHCARE PRODUCTS". P.O. Box 170 · Howell, MI 48844 (517) 546-5400 • (800) 248-4058 • Fax (517) 546-9388 www.tshsc.com • e-mail: info@tshsc.com | 510(k) Summary | AUG 2 8 2008 | |------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer: | Tri-State Hospital Supply Corporation<br>301 Catrell Drive<br>Howell, MI 48843 | | Contact: | Mr. Matthew K. Price<br>Director of Quality Assurance & Regulatory Affairs<br>Tri-State Hospital Supply Corporation<br>301 Catrell Drive<br>Howell, MI 48843<br>Phone: (517) 546-1135<br>Facsimile: (517) 546-3356 | | Date Summary Prepared: | February 15, 2008 | | Proprietary Name: | Centurion® Protective Restraint | | Common Name of Device: | Protective Restraint | | Classification Name: | Restraint, Protective | | Device Classification: | Class I | | Regulation: | 880.6760 | | Product Code: | FMQ | | Panel: | General Hospital | ## Predicate SE Device(s): This product is similar in design, composition, function, and method of use to the Posey Company Pediatric Limb Holder (K963439). ### Description: The Centurion® Protective Restraint is a device comprised of hook and loop material, 100% nylon fabric with polyester foam core, and nylon tricot backing. ### Intended Use: The Centurion® protective restraint is a device, including a wristlet, anklet, body/limb holder, or other type of strap, that is intended for medical purposes and that limits the patient's movements to the extent necessary for treatment, examination, or protection of the patient or others. This device is intended for use on neonatal/newborn patients by or on the order of a physician in a hospital or clinic setting. The device is contraindicated for use on patients with dislocations, fractures, open wounds on the affected limb, or if I.V. site can be compromised. ## Summary of Technological Characteristics between Subject and Predicate Device: The predicate device (K963439) provides the same functions, characteristics described herein for the device. Although there are some dimensional differences between the predicate device and the Centurion® Protective Restraint, the difference is minor and raises no new questions of safety or effectiveness. {1}------------------------------------------------ # Summary of Testing: Comparative bench performance testing was performed to evaluate the physical integrity and performance of the device and its predicate. The results demonstrate that the device satisfies all performance, physical, and functional requirements and the device is as safe, as effective, and performs as well as the predicate device. No new issues of safety or efficacy were found. . . . {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. SEP 9-2008 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Matthew K. Price Director of Quality Assurance & Regulatory Affairs Tri-State Hospital Supply Corporation 301 Catrell Drive Howell, Michigan 48843 Re: K080448 > Trade/Device Name: Centurion® Protective Restraint Regulation Number: 21 CFR 880.6760 Regulation Name: Protective Restraint Regulatory Class: I Product Code: FMO Dated: August 20, 2008 Received: August 21, 2008 Dear Mr. Price: This letter corrects our substantially equivalent letter of August 28, 2008. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Price Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, JHommels-Lendm-D foell Chu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ #### Indications for Use 510(k) Number (if known): K080448 Device Name: Centurion® Protective Restraint ### Indications for Use: The Centurion® Protective Restraint is a device, including a wristlet, anklet, body/limb holder, or other type of strap, that is intended for medical purposes and that limits the patient's movements to the extent necessary for treatment, examination, or protection of the patient or others. This device is intended for use on neonatal/newborn patients by or on the order of a physician in a hospital or clinic setting. ### Contraindications: The device is contraindicated for use on patients with dislocations, fractures, open wounds on the affected limb, or if I.V. site can be compromised. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K080448 Page 1 of Centurion® Protective Restraint 510(k) Submission 4-2 (REVISED) Indications for Use
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...