OnGuard2 Chemfort Closed Administration (CADM)

K210707 · Simplivia Healthcare , Ltd. · ONB · Sep 23, 2021 · General Hospital

Device Facts

Record IDK210707
Device NameOnGuard2 Chemfort Closed Administration (CADM)
ApplicantSimplivia Healthcare , Ltd.
Product CodeONB · General Hospital
Decision DateSep 23, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5440
Device ClassClass 2

Indications for Use

The OnGuard®2 Chemfort™ Closed Administration is a single use, sterile Closed System Transfer Device (CSTD) that mechanically prohibits the release of drugs, including antineoplastic and hazardous drugs, in vapor, aerosol or liquid form during administration, thus minimizing exposure of individuals, healthcare personnel, and the environment to hazardous drugs. OnGuard®2 Chemfort™ Closed Administration prevents the introduction of microbial and airborne contaminants into the drug or fluid path for up to 7 days.

Device Story

OnGuard®2 Chemfort™ Closed Administration (CADM) is a sterile, single-use Closed System Transfer Device (CSTD) used in hospitals and clinics by healthcare professionals. It functions as a multi-component system (Bag Adaptor, Y-Set, Secondary Set, Spike Port) that connects infusion containers to primary sets. The device utilizes elastomeric seals and a resealing septum mechanism; when components are joined, septums are pressed and pierced by a needle from the Chemfort™ system, creating a secure, closed fluid path. This design prevents the release of hazardous drug vapors, aerosols, or liquids and blocks microbial/airborne contaminants for up to 7 days. Unlike the predicate, the CADM allows for safe disconnection of the patient's IV set after administration and reconnection to a new bag via a new BACP device. The system minimizes hazardous drug exposure to personnel and patients during administration.

Clinical Evidence

Bench testing only. Compliance demonstrated with ISO 8536-4 (infusion sets), ISO 80369-7 (small-bore connectors), and USP <788> (particulate matter). Biocompatibility testing (ISO 10993-1) included cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, and hemolysis. Sterility (EtO, SAL 10^-6), shelf-life (3 years, ASTM F1980-16), and package integrity (ASTM F88, ASTM F1929, ISTA 3A) were validated.

Technological Characteristics

Multi-component system with resealing septums. Materials comply with ISO 10993-1. Connectivity via mechanical snap connections and small-bore connectors (ISO 80369-7). Sterilization via Ethylene Oxide (EtO) to SAL 10^-6. Shelf life 3 years. No software or electronic components.

Indications for Use

Indicated for use by pharmacists or other healthcare professionals in hospitals, compounding centers, and clinics to prevent exposure to hazardous drugs (including antineoplastic) and microbial/airborne contamination during drug administration.

Regulatory Classification

Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Simplivia Healthcare LTD. Shay Shaham VP QA / RA North Industrial Zone Kiryat Shmona, 1101801 Israel Re: K210707 Trade/Device Name: OnGuard®2 Chemfort™ Closed Administration (CADM) Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: ONB Dated: August 19, 2021 Received: August 24, 2021 ### Dear Shay Shaham: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K210707 Device Name OnGuard®2 Chemfort™ Closed Administration (CADM) Indications for Use (Describe) The OnGuard®2 Chemfort™ Closed Administration is a single use, sterile Closed System Transfer Device (CSTD) that mechanically prohibits the release of drugs, including antineoplastic and hazardous drugs, in vapor, aerosol or liquid form during administration, thus minimizing exposure of individuals, healthcare personnel, and the environment to hazardous drugs. OnGuard®2 Chemfort™ Closed Administration prevents the introduction of microbial and airborne contaminants into the drug or fluid path for up to 7 days. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------|------------------------------------------------------------------------------| | <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### K210707 510(k) SUMMARY | Preparation Date: | September 23, 2021 | |----------------------------------------------|----------------------------------------------------------------------------------------| | Submitter Name | Simplivia Healthcare LTD.<br>North Industrial Zone<br>Kiryat Shmona, 1101801<br>Israel | | Contact Person: | Shay Shaham<br>VP QA / RA | | Telephone Number:<br>E-mail Address:<br>Fax: | (972) 4 6908826<br>Shay.Shaham@simplivia.com<br>(972) 74 7652161 | | Trade Name: | OnGuard®2 Chemfort™ Closed Administration (CADM) | | Common Name: | Closed Antineoplastic and Hazardous Drug Reconstitution and<br>Transfer System | | Regulation Name: | Intravascular Administration Set | | Regulation Number: | 21CFR 880.5440 | | Product Code: | ONB | | Device Class: | Class II | | Predicate Device: | K192866, Chemfort™ Closed System Transfer Device (CSTD) | ## Device Description: The OnGuard®2 Chemfort™ Closed Administration (CADM) devices allow drug transfer to the IV bag and drug administration to the patient. The use of elastomeric seals in CADM prevents hazardous drugs contamination of healthcare professionals, the patient and the environment. The OnGuard®2 Chemfort™ Closed Administration contains four devices that connect between infusion containers and primary sets: - Bag Adaptor Chemfort™ Port (BACP) - - -Closed Y Inline Set (Y-Set) - Closed IV Secondary Set (Secondary) - - Closed Adaptor Spike Port (CASP) - CADM devices are an addition to the cleared Chemfort™ system (K192866). CADM provides closed system protection during the following procedures: - 1) Drug transfer to a container (e.g. IV bag) through the Chemfort™ Syringe Adaptor (K192866) and CADM Bag Adaptor Chemfort™ Port (BACP). - 2) Drug administration, with one of the CADM sets after it is attached to the BACP and creates a closed system. The main differences and unique features between the subject device and the predicate: {4}------------------------------------------------ - In the Chemfort™ system (K192866) the infusion set's spike is connected to an IV bag via the ● Chemfort™ Bag Adaptor SP's (BASP) tail. This is a one-time connection which remains sealed during the entire administration procedure (and after). The CADM set's Syringe Adaptor component allows the healthcare professional to have the option for safe disconnection of the patient's IV set after the drug was administrated, and then re-connecting it to a new IV bag containing saline or a new drug for administration via a new CADM BACP device, while keeping the closed system. - The CADM BACP (K210707) device consists of a distal Chemfort™ port which serves as both the ● drug transfer to the IV bag path and the infusion outflow path (drug administration), while the predicate device Chemfort™ BASP (K192866) has a Chemfort™ port for drug transfer and a separate path for drug administration. Each of the OnGuard®2 Chemfort™ Closed Administration devices are available separately. | Characteristics | Subject Device- K210707<br>OnGuard®2 Chemfort™ Closed<br>Administration (CADM) | Predicate Device- K192866<br>Chemfort™ Closed System Transfer<br>Device (CSTD) | | |------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indication for<br>Use | The OnGuard®2 Chemfort™ Closed<br>Administration is a single use, sterile<br>Closed System Transfer Device (CSTD)<br>that mechanically prohibits the release of<br>drugs, including antineoplastic and<br>hazardous drugs, in vapor, aerosol or liquid<br>form during administration, thus<br>minimizing exposure of individuals,<br>healthcare personnel, and the environment<br>to hazardous drugs.<br>OnGuard®2 Chemfort™ Closed<br>Administration prevents the introduction of<br>microbial and airborne contaminants into<br>the drug or fluid path for up to 7 days. | Chemfort™ is a Closed System Transfer<br>Device (CSTD) that mechanically prohibits<br>the release of drugs, including<br>antineoplastic and hazardous drugs, in<br>vapor, aerosol or liquid form during<br>preparation, reconstitution, compounding<br>and administration, thus minimizing<br>exposure of individuals, healthcare<br>personnel, and the environment to<br>hazardous drugs.<br>Chemfort™ prevents the introduction of<br>microbial and airborne contaminants into<br>the drug or fluid path for up to 7 days. | | | Prescription<br>Only or Over<br>the Counter | Prescription Only | Prescription Only | | | | Proposed Device- K210707<br>OnGuard®2 Chemfort™<br>Closed Administration | Predicate Device- K192866<br>Chemfort™ Closed System<br>Transfer Device (CSTD) | Equivalence to<br>predicate | | Indications for<br>use | The OnGuard®2 Chemfort™<br>Closed Administration is a single<br>use, sterile Closed System<br>Transfer Device (CSTD) that<br>mechanically prohibits the release<br>of drugs, including antineoplastic<br>and hazardous drugs, in vapor,<br>aerosol or liquid form during<br>administration, thus minimizing<br>exposure of individuals,<br>healthcare personnel, and the<br>environment to hazardous drugs.<br>OnGuard®2 Chemfort™ Closed<br>Administration prevents the<br>introduction of microbial and<br>airborne contaminants into the<br>drug or fluid path for up to 7 days. | Chemfort™ is a Closed System<br>Transfer Device (CSTD) that<br>mechanically prohibits the<br>release of drugs, including<br>antineoplastic and hazardous<br>drugs, in vapor, aerosol or liquid<br>form during preparation,<br>reconstitution, compounding and<br>administration, thus minimizing<br>exposure of individuals,<br>healthcare personnel, and the<br>environment to hazardous drugs.<br>Chemfort™ prevents the<br>introduction of microbial and<br>airborne contaminants into the<br>drug or fluid path for up to 7<br>days. | Different - the<br>difference is in<br>the product<br>name<br>Chemfort™<br>System versus<br>OnGuard®2<br>Chemfort™<br>Closed<br>Administration<br>and an addition<br>of "is a single<br>use, sterile" | | Target users | Pharmacists or other healthcare<br>professionals | Pharmacists or other healthcare<br>professionals | Same | | Environment of<br>use | Hospitals, compounding centres<br>and clinics | Hospitals, compounding centres<br>and clinics | Same | | Components | Bag Adaptor Chemfort™ Port<br>(BACP)<br>Closed Y Inline Set (Y Set)<br>Closed Adaptor Spike Port<br>(CASP)<br>Closed Secondary IV Set<br>(Secondary) | Vial Adaptor 20 mm with 13<br>mm Vial Converter<br>Vial Adaptor 28 mm<br>Vial Adaptor 32 mm<br>Syringe Adaptor<br>Syringe Adaptor Lock<br>Luer Lock Adaptor<br>Bag Adaptor SP | Different device<br>(CADM will be<br>added to the<br>Chemfort™<br>System) | | Spike & body<br>design | The BACP spike itself and the spike body (finger placement) design<br>are based on the Bag Adaptor SP (BASP) design cleared under<br>K192866 | | Same | | Chemfort™ Port<br>location | BACP- the Chemfort™ port is<br>located in the distal part of the<br>device | BASP- the Chemfort™ port is<br>located in the middle of the<br>device | Different - see<br>comment 1 | | Drug delivery to<br>bag | Through BACP Chemfort™ port | Through BASP Chemfort™ port | Same | | Drug<br>administration<br>to patient | Through BACP Chemfort™ port<br>and the connected CADM IV set | Through BASP tail, located in<br>the distal part of the device and<br>the connected IV set | Different - See<br>comment 2 | | Residual volume | BACP residual volume is <<br>0.01% of 1 L saline bag | BACP residual volume is <<br>0.01% of 1 L saline bag | Same | | Interaction with<br>other devices | For most uses, the OnGuard®2<br>Chemfort™ Closed Administration<br>devices will connect to an IV<br>solution container and a primary | For most uses, the Vial Adaptor<br>will connect to a vial, the<br>Syringe Adaptor will connect to<br>a syringe, the Luer Lock<br>Adaptor will connect to a | Similar<br>connections to<br>Chemfort™<br>System's<br>devices | | | Proposed Device- K210707<br>OnGuard®2 Chemfort™<br>Closed Administration | Predicate Device- K192866<br>Chemfort™ Closed System<br>Transfer Device (CSTD) | Equivalence to<br>predicate | | | administration set. A connection<br>also can be made with:<br>- Syringe Adaptor (SA)<br>(K192866)<br>- Syringe Adaptor Lock (SAL)<br>(K192866) | needless injection site, the Bag<br>Adaptor SP will connect to an<br>IV solution container and a<br>primary administration set.<br>A connection also can be made<br>with<br>- Syringe Adaptor (K192866)<br>- Syringe Adaptor Lock<br>(K192866) | | | Re-use capability | All devices can be used up to 10<br>times, but once connected to a<br>non-Chemfort™ device (such as<br>IV bag) they can't be<br>disconnected | All devices can be used up to 10<br>times, but once connected to a<br>non-Chemfort™ device (such as<br>drug vial) they can't be<br>disconnected | Same | | Principles of<br>Operation | Multi-component system, devices<br>are intended to be used as a<br>system, manually manipulated | Multi-component system,<br>devices are intended to be used<br>as a system, manually<br>manipulated | Same | | Interaction with<br>patient | No direct interaction - interaction<br>with the patient is achieved by the<br>passage of IV fluids through the<br>central tubing of the<br>administration set | No direct interaction -<br>interaction with the patient is<br>achieved by the passage of IV<br>fluids through the central tubing<br>of the administration set | Same | | Interconnecting<br>features | Mechanical snap connections | Mechanical snap connections | Same | | Technology | All of the devices are sealed with<br>resealing septum. When devices<br>are joined together the two<br>septums are pressed and then<br>pierced by needle (from the<br>Chemfort™ system's SA or SAL<br>device), thus creating a secured<br>fluid path. | All of the devices are sealed<br>with resealing septum. When<br>devices are joined together the<br>two septums are pressed and<br>then pierced by needle (from the<br>Chemfort™ system's SA or<br>SAL device), thus creating a<br>secured fluid path. | Same | | Safety features | Needle tip protector<br>Septum to septum contact<br>Spike cap | 0.2 micron venting membrane<br>Charcoal cloth<br>Needle tip protector<br>Septum to septum contact<br>Spike cap | Similar-<br>Needle tip<br>protector<br>Septum to<br>septum contact<br>Spike cap | | Sterilization<br>method | Ethylene Oxide validated cycle<br>SAL 10-6 | Ethylene Oxide validated cycle<br>SAL 10-6 | Same | | Biocompatibility | All parts that are in contact with<br>patient comply with the<br>requirements of ISO 10993-1 | All parts that are in contact with<br>patient comply with the<br>requirements of ISO 10993-1 | Same | | Shelf life | 3 years | 3 years | Same | | Prescription use | Rx only | Rx only | Same | | | Proposed Device- K210707<br>OnGuard®2 Chemfort™<br>Closed Administration | Predicate Device- K192866<br>Chemfort™ Closed System<br>Transfer Device (CSTD) | Equivalence to<br>predicate | | Meets the<br>NIOSH<br>and ISOPP<br>definition of a<br>CSTD | Yes | Yes | Same | #### Intended Use / Indications for Use Discussions of differences in Indications for Use statement There are only editorial differences to the indications for use statement between the predicate and the subject device which do not change the indications. ## Summary of Technological Characteristics The table below includes a comparison of the technological characteristics between the new device and those of the predicate device: {5}------------------------------------------------ {6}------------------------------------------------ {7}------------------------------------------------ Discussions of differences in technological characteristics - Comment 1- The physical location of the predicate device's Chemfort™ port is different than that of the proposed device. The port of the Bag Adaptor SP (K192866) is located in the middle of the device body, in the distal part there is a tail to connect to an IV set, whereas in the proposed device, Bag Adaptor CP (K210707), the port is located in the distal part of the device. Tests conducted to evaluate the difference: air & fluid tightness, bidirectional flow. - Comment 2- In the predicate device, Bag Adaptor SP (K192866), drug administration is performed by connecting an IV set to the tail, located in the distal part of the device. In the proposed device, Bag Adaptor CP (K210707), the Chemfort™ port also serves for drug administration by connecting one of CADM sets to the port located in the distal part BACP. Tests conducted to evaluate the difference: tests according to ISO 8536-4, specifically; leakage, tensile strength, and flow rate. ## Performance Data Simplivia Healthcare conducted several performance tests to demonstrate that the OnGuard(®2 Chemfort" Closed Administration devices comply with the following standards and that they function as intended. - ISO 8536-4:2010, Infusion equipment for medical use -Part 4: Infusion sets for single use, . gravity feed. - o Tests conducted to ensure compliance with the standard: leakage, tensile strength, closure piercing device, flow rate, tubing, drip chamber & drip tube, flow regulator & protective сар. - ISO 80369-7:2016, Small-bore connectors for liquids and gases in healthcare applications -Part 7: Connectors for intravascular or hypodermic applications. - Tests conducted to ensure compliance with the standard: positive pressure liquid leakage, O sub-atmospheric pressure air leakage, stress cracking, resistance to separation from axial load, resistance to separation from unscrewing & resistance to overriding. - USP <788> Particulate Matter in Injections ● - Particulate matter testing was conducted in accordance USP <788> and met the USP o acceptance criteria. - ISO 14971:2007, Medical devices- Application of risk management to medical devices ● {8}------------------------------------------------ ## Biocompatibility In accordance with ISO 10993-1, the CADM devices are classified as: Blood path, indirect, Contact Duration: Prolonged (24hrs to 30days). The following testing was conducted: - . Cytotoxicity - Sensitization . - Irritation or Intracutaneous reactivity ● - Acute systemic toxicity . - . Material mediated pyrogenicity - Subacute/subchronic toxicity - Hemolvsis ● ## Sterility, Shipping and Shelf-Life - The OnGuard®2 Chemfort™ Closed Administration devices are supplied sterile for single-use. . The devices are sterilized by Ethylene Oxide (EtO) gas to achieve a sterility assurance level (SAL) of at least 10-6. The process underwent a full sterilization validation according to the 'overkill' (half cycle) approach. - Residuals of Ethylene Oxide (EtO) and Ethylene Chlorhydrine (ECH) were tested after aeration . and were found to comply with the requirements of ISO 10993-7:2008 for prolonged exposure devices (Category B). - The bacterial endotoxins test (LAL) was performed using the kinetic turbidimetric methods for 10 samples (in pool) and was found to be less than 20 EU per device. - A shelf life of three vears has been established using the FDA recognized standard ASTM ● F1980-16 "Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices", by exposing sterilized samples of CADM devices to accelerated aging equivalent to 3 years. Following the accelerated aging performance, functional and packaging integrity tests were performed. All tests passed according to the predetermined acceptance criteria. - . Package integrity testing, after environmental conditioning and simulated transportation in accordance with ISTA 3A, was conducted on the final, packaged, and sterile devices. All packaging deemed acceptable for protection of product and sterility maintenance. Sterile Barrier Packaging Testing was performed on the proposed device and were found to be in compliance according to the following standards: - 0 Seal strength ASTM F88/F88-15 - Dye penetration ASTM F1929-15 o ## Conclusions Simplivia Healthcare's OnGuard®2 Chemfort™ Closed Administration has the same intended use, indications for use, similar technological characteristics and principles of operation as its predicate device, K192866. Performance data demonstrated that the OnGuard®2 Chemfort™ Closed {9}------------------------------------------------ Administration is as safe and effective as its predicate and does not raise any new safety and effectiveness issues. Thus, Simplivia Healthcare's OnGuard®2 Chemfort™ Closed Administration is substantially equivalent to its predicate device, K192866.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...