WellDoc BlueStar (WellDoc DiabetesManager System and DiabetesManager-Rx System
K162225 · Welldoc, Inc. · MRZ · Nov 22, 2016 · General Hospital
Device Facts
Record ID
K162225
Device Name
WellDoc BlueStar (WellDoc DiabetesManager System and DiabetesManager-Rx System
Applicant
Welldoc, Inc.
Product Code
MRZ · General Hospital
Decision Date
Nov 22, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
DiabetesManager® (OTC Use): The WellDoc DiabetesManager® System is indicated for use by healthcare providers (HCPs) and their adult patients - aged 21 years and older -who have type 2 diabetesManager® System is intended to provide secure capture, storage, and transmission of blood glucose data as well as information to aid in diabetes self-management. The DiabetesManager® System analyzes and reports blood glucose test results and supports medication adherence. It includes software intended for use on mobile phones or personal computers in the home or in professional healthcare settings. The software also allows for entry of other diabetes-related healthcare information and provides educational information. The DiabetesManager® System is not intended to replace the care provided by a licensed healthcare professional, including prescriptions, diagnosis, or treatment. DiabetesManager®-Rx (Prescription Use): The WellDoc DiabetesManager®-Rx System is indicated for use by healthcare providers (HCPs) and their adult patients - aged 21 years and older - who have type 2 diabetes. The DiabetesManager®-Rx System is intended to provide secure capture, storage, and transmission of blood glucose data as well as information to aid in diabetes self-management. The DiabetesManager®-Rx System analyzes and reports blood glucose test results and supports medication adherence. In addition, the DiabetesManager®-Rx System provides coaching messages (motivational, behavioral, and educational) based on real-time blood glucose values and trends. It includes software intended for use on mobile phones or personal computers in the home or in professional healthcare settings. The software also allows for entry of other diabetes-related healthcare information and provides educational information. The DiabetesManager®-Rx System is not intended to replace the care provided by a licensed healthcare professional, including prescriptions, diagnosis, or treatment.
Device Story
WellDoc BlueStar is a software system (mobile app/web portal) for type 2 diabetes management. Inputs include manual blood glucose entry or automated data via Bluetooth from OneTouch Verio Flex meter. System stores data, provides educational resources, and offers coaching messages (motivational/behavioral/educational) based on real-time glucose trends. Rx version includes insulin dose calculator. Used by patients at home and HCPs in clinical settings. Output includes glucose reports, medication adherence tracking, and coaching feedback to assist patient self-management and HCP decision-making. Data can be transmitted to OneTouch Reveal Server.
Clinical Evidence
Bench testing only. Verification, validation, and human factors testing confirmed successful data transmission from OneTouch Verio Flex meter, compatibility with mobile operating systems, and overall system functionality.
Technological Characteristics
Software-based system for mobile (iOS/Android) and web platforms. Features Bluetooth connectivity for meter integration, cloud-based data transmission, and insulin dose calculation algorithms. Operates as standalone software.
Indications for Use
Indicated for adult patients aged 21+ with type 2 diabetes and their healthcare providers to aid in diabetes self-management via blood glucose data capture, storage, transmission, analysis, and medication adherence support.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
WellDoc DiabetesManager® System and Diabetes Manager®-Rx System (K100066)
Reference Devices
One Touch Verio Flex Blood Glucose Meter (K150214)
Related Devices
K162532 — WellDoc BlueStar, WellDoc BlueStar Rx · Welldoc, Inc. · Jan 12, 2017
K190013 — WellDoc BlueStar · Welldoc, Inc. · Nov 4, 2019
K141273 — WELLDOC DIABETESMANAGER SYSTEM AND DIABETESMANAGER- RX SYSTE · Welldoc, Inc. · Jul 8, 2014
K120314 — WELLDOC DIABETES MANAGER SYSTEM AND DIABETES MANAGER RX SYSTEM · Welldoc, Inc. · Feb 24, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. The logos are simple and professional in appearance.
May 17, 2024
WellDoc, Incorporated Danielle Dorfman Manager, Regulatory Affairs and Quality System 10221 Wincopin Circle Suite 150 Columbia, Maryland 21044
Re: K162225
Trade/Device Name: WellDoc BlueStar (WellDoc DiabetesManager® System and DiabetesManager®-Rx System) Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion pump Regulatory Class: Class II Product Code: MRZ, NDC
Dear Danielle Dorfman:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated November 22, 2016. Specifically, FDA is updating this SE letter as an administrative correction. A secondary product code, LNX, was inadvertently included.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Juliane Lessard, OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices, 240-402-6126, Juliane.Lessard(@fda.hhs.gov.
Sincerelv.
Juliane C. Lessard -S
Juliane C. Lessard, Ph.D. Director DHT3C: Division of Drug Delivery, General Hospital and Human Factors Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a symbol. The symbol consists of three stylized human profiles facing to the right, stacked on top of each other, with the top profile being the largest and the bottom profile being the the smallest.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 22, 2016
WellDoc, Incorporated Danielle Dorfman Manager, Regulatory Affairs and Quality System 10221 Wincopin Circle, Suite 150 Columbia, Maryland 21044
Re: K162225
Trade/Device Name: WellDoc Bluestar (WellDoc DiabetesManager® System and DiabetesManager®-Rx System) Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: MRZ, LNX, NDC Dated: October 26, 2016 Received: October 27, 2016
Dear Danielle Dorfman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely, Tina Kiang -5
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162225
#### Device Name
WellDoc Bluestar (WellDoc DiabetesManager® System and DiabetesManager®-Rx System)
#### Indications for Use (Describe)
DiabetesManager® (OTC Use): The WellDoc DiabetesManager® System is indicated for use by healthcare providers (HCPs) and their adult patients - aged 21 years and older -who have type 2 diabetesManager® System is intended to provide secure capture, storage, and transmission of blood glucose data as well as information to aid in diabetes self-management. The DiabetesManager® System analyzes and reports blood glucose test results and supports medication adherence. It includes software intended for use on mobile phones or personal computers in the home or in professional healthcare settings. The software also allows for entry of other diabetes-related healthcare information and provides educational information.
The DiabetesManager® System is not intended to replace the care provided by a licensed healthcare professional, including prescriptions, diagnosis, or treatment.
DiabetesManager®-Rx (Prescription Use): The WellDoc DiabetesManager®-Rx System is indicated for use by healthcare providers (HCPs) and their adult patients - aged 21 years and older - who have type 2 diabetes. The DiabetesManager®-Rx System is intended to provide secure capture, storage, and transmission of blood glucose data as well as information to aid in diabetes self-management. The DiabetesManager®-Rx System analyzes and reports blood glucose test results and supports medication adherence. In addition, the DiabetesManager®-Rx System provides coaching messages (motivational, behavioral, and educational) based on real-time blood glucose values and trends. It includes software intended for use on mobile phones or personal computers in the home or in professional healthcare settings. The software also allows for entry of other diabetes-related healthcare information and provides educational information.
The DiabetesManager®-Rx System is not intended to replace the care provided by a licensed healthcare professional, including prescriptions, diagnosis, or treatment.
| Type of Use (Select one or both, as applicable) | | | | |
|-------------------------------------------------|-----------------------------------------------|--|--|--|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | | | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED | | | | |
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for WellDoc. The logo consists of a blue and teal intertwined symbol on the left, resembling two hands coming together. To the right of the symbol is the word "WellDoc" in a dark blue, sans-serif font. A small registration mark is present in the upper right corner of the word "WellDoc".
# 510(k) Summary
## General Information
Date of Summary Preparation: November 18, 2016
| Name of Manufacturer: | WellDoc, Inc. |
|----------------------------|-----------------------------------------------------------------------------------------------|
| Address: | 10221 Wincopin Circle Suite 150<br>Columbia, MD 21044 |
| Contact Person: | Kevin McRaith<br>Chief Executive Officer |
| Phone: | (443) 692-3100 |
| Fax: | (443) 692-3099 |
| Trade or Proprietary Name: | WellDoc BlueStar (DiabetesManager® System and Diabetes<br>Manager®-Rx System) |
| Common or Usual Name: | Medical computers and software<br>Infusion pump accessories |
| Product Codes: | LNX, MRZ, NDC |
| Classification Name: | LNX is unclassified and therefore has no regulation number<br>21 CFR 880.5725 (Infusion Pump) |
| Regulatory Class: | II |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for WellDoc. The logo consists of a blue and teal intertwined hands graphic on the left, followed by the word "WellDoc" in blue, with a registered trademark symbol to the right of the word.
General Hospital Classification Panel: Predicate Device: K 100066 (WellDoc DiabetesManager® System and Diabetes Manager®-Rx System)
### Device Description
The WellDoc BlueStar DiabetesManager® System is a stand-alone, over-the-counter (OTC) software system that has the capability of providing additional functions, if and when, the functions are prescribed by the prescribing healthcare provider. Once the additional prescription (Rx) functions are activated, the entire software system is called WellDoc BlueStar DiabetesManager-Rx® System.
Both BlueStar DiabetesManager System and DiabetesManager-Rx System are implemented through an enterprise such as a health plan or large physician group in tandem with healthcare providers (HCPs) and are comprised of the following applications:
- Enterprise Director Portal .
- . HCP Service
- Patient Mobile Application .
- . Patient Web Portal
The Enterprise Director application is used for administrative purposes. The HCP Service houses the Medication Reconciliation feature (changed under K141273), which allows for the process of identifying the most accurate list of all medications that the patient is taking, including name, dosage and frequency, by comparing the medical record to an external list of medications obtained from a patient, hospital, or other provider.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the WellDoc logo. The logo features a stylized symbol on the left, resembling two hands intertwined in a circular shape, rendered in shades of blue. To the right of the symbol is the word "WellDoc" in a bold, blue font. A small trademark symbol appears next to the word "Doc".
The Patient Web Portal and the Patient Mobile application have a similar feature set. Data entered into these applications is stored in the database and can be retrieved for display in either application. Both applications require the initial web- or mobile-based registration before the patient can access them. Patientified by healthcare providers and an access code is provided to allow access to both the web and mobile applications. On the patient applications (Mobile and Web), the basic DiabetesManager System functions as an information repository [logbook and Personal Health Record (PHR)], diabetes education resource (learning library and health tips), educational messaging/coaching, and secure communication system (Message Center). If and when a prescription is obtained, additional functions become available to the patient as DiabetesManager-Rx System. Prescription functions include additional medication information (dose and schedule), enhanced coaching [(Blood Glucose (BG) real-time coaching feedback), messaging (Message Center Content), and workflow and decision support for healthcare providers. The patient web portal and mobile application also provide an insulin dosing calculator to allow patients to use their prescribed regimen to calculate a dose of insulin for a given amount of carbohydrates and/or blood glucose value.
With this 510(k), BlueStar will have FDA clearance to connect to the One Touch Verio Flex Blood Glucose Meter (K150214) via Bluetooth. This will allow users to send data from their meter to the BlueStar app, which will provide coaching messages (motivational, behavioral, and educational) based on the real-time blood glucose values and trends. The BlueStar Server will also have the ability to transmit data to the OneTouch Reveal Server. These modifications do not change the fundamental scientific technology or the indications for use of the device.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the WellDoc logo. The logo consists of a blue and teal intertwined symbol on the left, followed by the word "WellDoc" in blue. The symbol appears to be two hands intertwined. The logo is simple and modern.
### Indications for Use
DiabetesManager® (OTC Use): The WellDoc DiabetesManager® System is indicated for use by healthcare providers (HCPs) and their adult patients - aged 21 years and older -who have type 2 diabetes. The DiabetesManager® System is intended to provide secure capture, storage, and transmission of blood glucose data as well as information to aid in diabetes self-management. The DiabetesManager® System analyzes and reports blood glucose test results and supports medication adherence. It includes software intended for use on mobile phones or personal computers in the home or in professional healthcare settings. The software also allows for entry of other diabetes-related healthcare information and provides educational information.
The DiabetesManager® System is not intended to replace the care provided by a licensed healthcare professional, including prescriptions, diagnosis, or treatment.
DiabetesManager®-Rx (Prescription Use): The WellDoc DiabetesManager®-Rx System is indicated for use by healthcare providers (HCPs) and their adult patients – aged 21 years and older - who have type 2 diabetes. The DiabetesManager®-Rx System is intended to provide secure capture, storage, and transmission of blood glucose data as well as information to aid in diabetes self-management. The DiabetesManager®-Rx System analyzes and reports blood glucose test results and supports medication adherence. In addition, the DiabetesManager®-Rx System provides coaching messages (motivational, behavioral, and educational) based on realtime blood glucose values and trends. It includes software intended for use on mobile phones or personal computers in the home or in professional healthcare settings. The software also allows for entry of other diabetes-related healthcare information and provides educational information.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for WellDoc. The logo consists of a blue and light blue intertwined symbol on the left, followed by the word "WellDoc" in blue font. The logo is simple and professional, and it is likely used to represent the company's brand.
The DiabetesManager®-Rx System is not intended to replace the care provided by a licensed healthcare professional, including prescriptions, diagnosis, or treatment.
#### Summary of Technological Characteristics (compared to the predicate)
Intended use, design, materials, and performance are substantially equivalent to the predicate devices referenced. The only major differences between the subject and predicate devices are the ability to connect to the OneTouch Verio Flex Blood Glucose Meter via Bluetooth and the ability to transmit data from the BlueStar Server to the OneTouch Reveal Server. This modification does not change the fundamental scientific technology of the BlueStar application, as outlined in the Indications for Use:
- . Provides secure capture, storage, and transmission of blood glucose data as well as information to aid in diabetes self-management
- Analyzes and reports blood glucose test results and supports medication . adherence
- . Provides coaching messages (motivational, behavioral, and educational) based on real-time blood glucose values and trends
The subject change simply provides an alternative means of blood glucose inputs but processes the inputs in the same way as the predicate device. It should be noted that the user can continue to enter blood glucose values manually (as cleared under K100066); the subject modification allows for another means of BG inputs and is not necessary to use the app and receive feedback.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the WellDoc logo. The logo features a stylized symbol of two hands intertwined in a circular shape on the left. To the right of the symbol is the word "WellDoc" in a bold, blue font. The logo appears to be for a healthcare or wellness-related company.
Please refer to the table below for a comparison of the subject and predicate devices. Note that while both the predicate and subject device contain an insulin dose calculator with substantially equivalent functionality, the predicate device was not identified with the NDC product code at the time of clearance.
| Feature | Subject Device | Predicate Device,<br>K100066 |
|---------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| Indications for Use | Diabetes Manager (OTC Use):<br>The WellDoc DiabetesManager® System is<br>indicated for use by healthcare providers<br>(HCPs) and their adult patients - aged 21<br>years and older - who have type 2 diabetes.<br>The Diabetes Manager System is intended to<br>provide secure capture, storage, and<br>transmission of blood glucose data as well as<br>information to aid in diabetes self-<br>management. The DiabetesManager System<br>analyzes and reports blood glucose test<br>results and supports medication adherence. It<br>includes software intended for use on mobile<br>phones or personal computers in the home or<br>in professional healthcare settings. The<br>software also allows for entry of other<br>diabetes-related healthcare information and<br>provides educational information.<br>The DiabetesManager System is not intended<br>to replace the care provided by a licensed<br>healthcare professional, including<br>prescriptions, diagnosis, or treatment.<br><br>Diabetes Manager-Rx (Prescription Use):<br>The WellDoc DiabetesManager -Rx System<br>is indicated for use by healthcare providers<br>(HCPs) and their adult patients - aged 21<br>years and older - who have type 2 diabetes.<br>The DiabetesManager-Rx System is intended<br>to provide secure capture, storage, and<br>transmission of blood glucose data as well as<br>information to aid in diabetes self-<br>management. The DiabetesManager -Rx<br>System analyzes and reports blood glucose<br>test results and supports medication<br>adherence. In addition, the DiabetesManager<br>-Rx System provides coaching messages<br>(motivational, behavioral, and educational) | Identical |
| | | |
| OTC or Rx | Both | Both |
| Product Code | LNX, MRZ, NDC | LNX, MRZ |
| Classification | 21 CFR 880.5725 | Same |
| Class | II | II |
| Connection to BT meter | Yes | No |
| Data transmission to<br>OneTouch Reveal Server | Yes | No |
| Manual Data Entry | Yes | Yes |
| Logbook | Yes | Yes |
| Coaching Messages<br>(Motivational, Behavioral, and Educational) based on<br>real-time blood glucose<br>values and trends | Yes | Yes |
| Supported Mobile<br>Platforms | iPhone, Android | Same |
| Mobile Form Factors | Candy Bar, Touch Screen | Same |
| Mobile Environmental<br>Operating Systems | iOS 9.X and above<br>Android 4.3 and above | iOS 7.X and 8.X<br>Android 4.X |
| Insulin dose calculator | Yes | Yes |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the logo for WellDoc. The logo consists of a blue and teal graphic on the left, resembling two hands intertwined to form a circle. To the right of the graphic is the text "WellDoc" in blue, with a trademark symbol next to the "c".
### Non-Clinical Performance Data
Documentation according to FDA's Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices and Guidance for Industry and FDA Staff: Content of Premarket Submissions for the Management of Cybersecurity in Medical Devices was provided. Verification, validation and human factors testing were performed to ensure that the device meets its requirements and intended use. Testing showed that data was transmitted successfully from the OneTouch Verio Flex BG meter and to
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the WellDoc logo. The logo features a stylized image of two hands cradling each other in a circular shape on the left side. To the right of the hands is the text "WellDoc" in a blue, sans-serif font. The logo is clean and professional, suggesting a focus on care and support.
the One Touch Reveal Server, that the app is compatible with the mobile environmental operating systems, and that the system functions as intended.
## Conclusions Drawn from Non-Clinical Tests
The non-clinical testing demonstrated that the product is substantially equivalent to the predicate and continues to meet the cleared indications for use.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.