ivWatch Model 400

K142374 · Ivwatch, LLC · MRZ · Feb 13, 2015 · General Hospital

Device Facts

Record IDK142374
Device NameivWatch Model 400
ApplicantIvwatch, LLC
Product CodeMRZ · General Hospital
Decision DateFeb 13, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2

Indications for Use

The ivWatch Model 400 is indicated for the detection of subcutaneous infiltrations and extravasations of 10 cc or less of optically clear, uncolored infusates, as an adjunctive device to the clinical evaluation in the hospital setting of patients 18 years old or greater with peripherally-inserted IVs (PIVs) on the forearm or dorsal aspect of the hand. The device is indicated to assess patients for subcutaneous infiltrations and extravasations but should not serve as a substitute for regular clinician assessment of the PIV site. The ivWatch Model 400 is intended for use by physicians, or under the direction of a physician, who have been trained in the use of the ivWatch Model 400.

Device Story

ivWatch Model 400 provides continuous, noninvasive monitoring of tissue near PIV insertion sites to detect infiltration/extravasation. System includes pole-mounted Patient Monitor (IPM), reusable fiberoptic sensor cable, and single-use sterile sensor receptacle. IPM emits visible and NIR light via LEDs through fiberoptics to skin; measures reflected light to detect optical property changes indicative of infusate pooling. Proprietary signal processing algorithms analyze reflected light; if thresholds are exceeded, IPM triggers audible/visual alerts. Used in hospital settings by trained clinicians. Alerts prompt manual site inspection. Benefits include early detection of infiltration events, potentially reducing patient injury from extravasation.

Clinical Evidence

Six IRB-approved clinical studies conducted. Verification studies evaluated infiltration characteristics and patient attributes; validation studies assessed performance metrics including sensitivity and false notification rate. Results demonstrated device utility in identifying early-stage PIV infiltration. No adverse events reported during clinical studies.

Technological Characteristics

System: IPM, fiberoptic cable, single-use receptacle. Materials: Biocompatible plastics/adhesives (ISO 10993 compliant). Sensing: Visible/NIR light reflectance. Energy: Low-power LEDs. Connectivity: Standalone (pole-mounted). Standards: IEC 60601-1, IEC 60601-1-2, ANSI Z136.1 (optical safety). Software: Proprietary signal processing algorithm.

Indications for Use

Indicated for detection of subcutaneous infiltrations/extravasations (≤10 cc, clear/uncolored infusates) in hospital patients ≥18 years old with PIVs on forearm or dorsal hand. Adjunctive to clinical assessment; not a substitute for regular site inspection.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three faces in profile, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 13, 2015 ivWatch, LLC Javier Garriz Vice President of Product Management 469 McLaws Circle Williamsburg, VA 23185 Re: K142374 Trade/Device Name: ivWatch Model 400 Regulation Number: 21 CFR 880.5725 Regulation Name: Accessories, Pump, Infusion Regulatory Class: Class II Product Code: MRZ Dated: January 16, 2015 Received: January 20, 2015 Dear Mr. Garriz: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Susan Runno DDS, MA Erin Keith Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K142374 Device Name ivWatch Model 400 #### Indications for Use (Describe) The iv Watch Model 400 is indicated for the detection of subcutaneous infiltrations and extravasations of 10 cc or less of optically clear, uncolored infusates, as an adjunctive device to the clinical evaluation in the hospital setting of patients 18 years old or greater with peripherally-inserted IVs (PIVs) on the forearm or dorsal aspect of the hand. The device is indicated to assess patients for subcutaneous infiltrations and extravasations but should not serve as a substitute for regular clinician assessment of the PIV site. The ivWatch Model 400 is intended for use by physicians, or under the direction of a physician, who have been trained in the use of the iv Watch Model 400. Type of Use (Select one or both, as applicable) | <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) | |---------------|----------------------------------------------| | <span></span> | Over-The-Counter Use (21 CFR 801 Subpart | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." C) {3}------------------------------------------------ ivWatch, LLC 469 McLaws Circle Williamsburg, VA 23185 1.855.IVWATCH (489.2824) Fax: 757.645.4760 www.ivwatch.com # 5 510(K) SUMMARY ## 5.1 Administrative | Submitter Name | ivWatch, LLC | |---------------------------|--------------------------------------------------------| | Applicant Address | 469 McLaws Circle<br>Williamsburg, VA 23185 | | Phone | 855-489-2824 | | Fax | 757-645-4760 | | Primary Contact | Javier Garriz, Vice President of<br>Product Management | | Primary Contact<br>e-mail | javier.garriz@ivwatch.com | | Primary Contact Phone | 855-489-2824 x7016 | | Date Prepared | August 21, 2014 | ## 5.2 Device | Trade Name | ivWatch Model 400 | |---------------------|-----------------------------| | Manufacturer | ivWatch, LLC | | 510(k) Number | K142374 | | Device Class | II | | Classification Name | Accessories, Pump, Infusion | | Regulation Number | 21 CFR 880.5725 | | Product Code | MRZ | #### 5.3 Predicate Device | Trade Name | Baxter SIGMA Spectrum<br>Infusion Pump with Master<br>Drug Library Model 35700 | |---------------------|--------------------------------------------------------------------------------| | Manufacturer | Baxter Healthcare Corporation | | 510(k) Number | K133801 | | Device Class | II | | Classification Name | Infusion Pump | | Regulation Number | 21 CFR 880.5725 | | Product Code | FRN | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for IVWatch. The logo consists of a stylized teardrop shape in black, with the letters "iv" in white inside the teardrop. The "i" in "iv" is represented by a grid of small squares. To the right of the teardrop is the word "Watch" in a simple, sans-serif font, also in black. A small circle is to the right of the word "Watch". ivWatch. IIC 469 McLaws Circle Williamsburg, VA 23185 1.855.JVWATCH (489.2824) Fax: 757.645.4760 www.ivwatch.com # 5.4 Device Description The ivWatch Model 400 is a medical device that provides continuous, noninvasive monitoring of human tissue adjacent to peripheral intravenous (PIV) insertion sites on the forearm and dorsal aspect of the hand to aid in the early detection of infiltration and extravasation events. The ivWatch Model 400 ("Device") consists of the ivWatch Patient Monitor (IPM), a reusable optical sensor cable, and a single-use sensor receptacle. The device uses visible and near-infrared light to measure changes in the optical properties of the tissue near a PIV insertion site; the measured changes are processed by proprietary ivWatch signal processing algorithms to determine if an infiltration event may have occurred. The device is indicated for use by medical professionals who are experienced with administering or monitoring peripheral IV therapy. The IPM mounts to an IV pole, typically above the infusion pump. The sensor is secured in the receptacle and subsequently placed on the patient's skin near the peripheral IV insertion site for the duration of the IV monitoring. The fiberoptic sensor cable follows the IV line back to the pole-mounted IPM. One end of the sensor cable attaches to the IPM; the other end snaps into the ivWatch sensor receptacle, which secures the sensor to the patient's skin. The sensor cable is a multiple-use disposable with a total useful life of approximately 240 hours of IV monitoring; it is intended to be cleaned between uses. The ivWatch sensor receptacle is molded out of biocompatible plastics and includes biocompatible adhesives for attaching the receptacle to the patient's skin. The sensor receptacle is a single-use disposable and supplied sterile. The IPM contains an optical system that generates light signals that are sent to the patient's skin (via the sensor cable) and measures the light returning from the patient's skin (also via the sensor cable). The Device uses low power LEDs as sources of visible and near-infrared (NIR) light. Optical fibers in the sensor cable deliver the light from the IPM to the sensor and transmit reflected light back to the IPM. The IPM also contains the hardware for executing the proprietary ivWatch signal processing algorithm. If changes in the optical properties of the tissue near the peripheral IV insertion site are consistent with an infusate pooling in the subcutaneous tissue, the IPM emits audible and visual notifications intended to prompt the clinician to inspect the peripheral IV site for a possible infiltration event. # 5.5 Indications For Use The ivWatch Model 400 is indicated for the detection of subcutaneous infiltrations and extravasations of 10 cc or less of optically clear, uncolored infusates, as an adjunctive device to the clinical evaluation in the hospital setting of patients 18 years old or greater with peripherally-inserted IVs (PIVs) on the {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for IVWatch. The logo consists of a stylized water droplet shape on the left, with the letters "iv" inside the droplet. To the right of the droplet is the word "Watch" in a simple, sans-serif font. The logo is clean and modern, with a focus on the company name. 169 McLaws Circle Williamsburg, VA 23185 1.855.IVWATCH (489.2824) Fax: 757.645.4760 www.ivwatch.com forearm or dorsal aspect of the hand. The device is indicated to assess patients for subcutaneous infiltrations and extravasations but should not serve as a substitute for regular clinician assessment of the PIV site. The ivWatch Model 400 is intended for use by physicians, or under the direction of a physician, who have been trained in the use of the ivWatch Model 400. ## 5.6 Comparison of Technological Characteristics with the Predicate Device At a high level, the subject device (ivWatch Model 400) and the predicate device (Baxter SIGMA Spectrum with Master Drug Library Model 35700) have the same intended use – ensuring the delivery of IV fluids to the patient's vein. The predicate device includes technology to detect fault conditions (air-in-line, upstream and downstream occlusions) which may prevent the IV fluid from being delivered to the intended administration route(s). The subject device is intended for use in detecting one particular fault condition (IV infiltration) that prevents the IV fluid from being delivered to the intended intravenous administration route. In this sense, the subject device augments the type of faults detected by the predicate device (and indeed, any infusion pump) by detecting IV infiltration, consistent with the subject device's classification as an accessory to an infusion pump. Both the subject and predicate devices monitor deviations in particular characteristics of the IV fluid which are correlative to possible faults in the delivery of the IV fluid to the intended intravenous administration route. The subject device monitors changes in the optical properties of the tissue surrounding the IV insertion site for signs of infusate accumulation in the subcutaneous space (i.e., IV infiltration); the predicate device uses pressure and ultrasonic waves to detect occlusion/blockage and air-in-line, respectively. However, this does not alter the intended use of the (accessory) subject and predicate devices-ensuring the delivery of IV fluids to the intended administration route (the patient's vein). Both the subject and predicate devices issue visual and audible notifications when the measured values of the monitored characteristics exceed preset threshold values for those monitored characteristics (diffuse reflectance in the case of the subject device, pressure in the case of the predicate device). {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for IVWatch. The logo consists of a stylized teardrop shape in black, with the letters "iv" in white inside the teardrop. To the right of the teardrop is the word "Watch" in a simple, sans-serif font. The logo is clean and modern, and the use of negative space makes it visually appealing. ivWatch. IIC 169 McI aws Circle Williamsburg, VA 23185 1.855.JVWATCH (489.2824) Fax: 757.645.4760 www.ivwatch.com ### 5.6.1 Reference Device To specifically address any safety concerns arising from the use of a red and infrared light source, a reference device was selected (Nellcor OxiMAX N-560 Pulse Oximeter, K021090). The safety of using infrared technology (within specified operating conditions) to determine human tissue properties has been well established by a variety of medical devices, most notably oximeters. The Nellcor OxiMAX N-560 Pulse Oximeter uses near-infrared (NIR) light within similar wavelengths as the ivWatch Model 400 to provide continuous monitoring of oxygenation levels and has been demonstrated as a safe and effective medical device. | Trade name | OxiMAX N-560 Pulse<br>Oximeter | |---------------------|--------------------------------| | Manufacturer | Nellcor | | 510(k) Number | K021090 | | Device class | II | | Classification name | Oximeter | | Regulation Number | 21 CFR 870.2700 | | Product Code | DQA | ## 5.7 Performance Data The following performance data were provided in support of substantial equivalence determination and to demonstrate safety and effectiveness. #### 5.7.1 Biocompatibility The biocompatibility evaluation of the patient-contacting components of the ivWatch Model 400 device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,''' May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The following tests were performed: - Sensitization - Irritation - Cytotoxicity There are two skin-contacting components of the ivWatch Model 400: the ivWatch sensor receptacle and the ivWatch sensor cable. The ivWatch sensor receptacle complies with the biocompatibility requirements for prolonged duration contact (greater than 24 hours but less than 30 days) on intact skin. The {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for IVWatch. The logo consists of a black teardrop shape with white squares inside, followed by the text "ivWatch" in a stylized font. The "i" and "v" are connected with a wavy line, and there is a small circle after the word "Watch". 169 McI aws Circle Williamsburg, VA 23185 1.855.JVWATCH (489.2824) Fax: 757.645.4760 www.ivwatch.com ivWatch sensor cable complies with the biocompatibility requirements for 24hour contact on intact skin. 5.7.2 Electrical Safety and Electromagnetic Compatibility (EMC) Electrical safety and EMC testing were conducted on the ivWatch Model 400, consisting of the ivWatch Patient Monitor and sensor cable. The system conforms and meets the requirements of IEC 60601-1-2: 3rd Ed. 2007. ivWatch Model 400 was also tested and determined to be compliant to ANSI/AAMI ES60601-1:2005, 3rd ed. (Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance); IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007). ### 5.7.3 Software Verification and Validation Testing Software verification and validation testing for the ivWatch Model 400 were conducted and documented in accordance with the FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Software validation is compliant to IEC 60601-1 3rd edition - Amendment 1, and therefore IEC 62304. The software code on the IPM was segregated into two types: - 1. Software whose function is consistent with the "moderate level of concern" classification, as stated in the FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (May, 2005). As stated in the quidance document, failures or latent design flaws in software having the "moderate" level of concern may result in minor to moderate injury to the patient, or it may indirectly affect the patient in that incorrect or delayed information could result in injury to the patient. - 2. Software which presents a "minor level of concern." The device is intended as a supplement (aid) to, and not a replacement for regular clinician assessment of the PIV site (i.e., regular clinician assessment is still in effect). Accordingly, only the modules that involved setting the level and duration of the emitted light were treated as having "moderate" level of concern, since the only software-controlled output that contacts the patient in any way is the light emitted by the LEDs. #### 5.7.4 Bench Testing Computer simulations were used to identify the optimal design for the ivWatch sensor. Simulation results were validated in bench tests using an ex vivo porcine foot model. Optical safety testing was performed and indicated that the optical radiation emitted by the ivWatch device is significantly less than the limits set forth by industry standards (ANSI Z136.1 - 2007). {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for IVWatch. The logo consists of a stylized black teardrop shape with the letters "iv" in white inside. To the right of the teardrop is the word "Watch" in a simple, sans-serif font. The logo is clean and modern, suggesting a focus on technology and healthcare. ivWatch. IIC 469 McLaws Circle Williamsburg, VA 23185 1.855.JVWATCH (489.2824) Fax: 757.645.4760 www.ivwatch.com ## 5.7.5 Clinical Studies A series of six IRB-approved verification and validation clinical studies were performed for the development and validation of the ivWatch Model 400. Verification studies investigated the specific effect of infiltration characteristics or patient attributes on the performance of the ivWatch Model 400 device. The validation studies assessed the performance (e.g., sensitivity, false notification rate) of the ivWatch Model 400 device. The results demonstrated that the ivWatch Model 400 could assist clinicians in identifying the early stages of a PIV infiltration. There were no adverse events during the clinical studies. ### 5.7.6 Summary The results of the performance testing indicate that the ivWatch Model 400 is substantially equivalent to the Baxter SIGMA Spectrum Infusion Pump with MDL and raises no concerns about safety or efficacy of the device. ## 5.8 Conclusions The ivWatch Model 400 is found to be substantially equivalent to the predicate device, as both have the same intended use and mechanism of action. Any safety concerns arising from the use of a red and infrared light source are addressed using a reference device. Any differences between the ivWatch Model 400 and the predicate were addressed through performance testing and such differences were determined to have no effect on the safety or efficacy of the device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...