Omnipod® Insulin Management System, Omnipod DASH® Insulin Management System
K211575 · Insulet Corporation · LZG · Aug 13, 2021 · General Hospital
Device Facts
Record ID
K211575
Device Name
Omnipod® Insulin Management System, Omnipod DASH® Insulin Management System
Applicant
Insulet Corporation
Product Code
LZG · General Hospital
Decision Date
Aug 13, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Omnipod® Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin and for the quantitative measurement of glucose in fresh whole capillary blood (in vitro) from the finger. The glucose measurements should not be used for the diagnosis or screening for diabetes. The PDM glucose meter is intended for single patient use and should not be shared. Abbott FreeStyle® test strips are used with the built-in FreeStyle meter for the quantitative measurement of blood glucose in fresh whole capillary blood from the finger, upper arm, and palm. Abbott Freestyle Control Solutions are used to verify that the meter and test strips are working together properly and that the test is performed correctly. The Omnipod DASH™ Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin. Additionally, the Omnipod DASH System is interoperable with a compatible blood glucose meter to receive and display glucose measurements.
Device Story
System manages insulin therapy via disposable body-worn infusion pump (Pod) and wireless remote controller (Personal Diabetes Manager, PDM). Pod delivers subcutaneous insulin at set/variable rates; PDM controls basal/bolus delivery. PDM includes bolus calculator using carbohydrates, blood glucose, correction factor, insulin-to-carbohydrate ratio, target glucose, and Insulin on Board (IoB). Omnipod PDM features integrated glucose meter; DASH PDM uses Bluetooth Low Energy for interoperability with compatible meters. Users wear Pod on arm, back, abdomen, or thigh for up to 3 days. Healthcare providers use system to manage diabetes; patients operate PDM to adjust therapy. Output includes insulin delivery, glucose readings, and therapy history logs. Device benefits patients by automating insulin delivery and simplifying dosage calculations based on clinical parameters.
Clinical Evidence
Bench testing only. In-use stability and leachables testing conducted to validate compatibility of Lyumjev U100 insulin with the infusion systems. Safety assurance cases provided per FDA guidance for infusion pumps.
Technological Characteristics
Pod housing: Polycarbonate; Adhesive: Medical Grade Acrylic with polyester backing; Fluid path: FEP, silicone, stainless steel, polypropylene, PVC, cyclic olefin copolymer, EPDM, polyisoprene. Pumping: Step drive mechanism via microprocessor-controlled leadscrew. Communication: 433.92MHz RF (Omnipod) or Bluetooth Low Energy (DASH). Sterilization: Ethylene Oxide (SAL 10^-6).
Indications for Use
Indicated for subcutaneous insulin delivery at set/variable rates for management of diabetes mellitus in insulin-requiring patients. Omnipod system includes integrated glucose monitoring for capillary blood. Omnipod DASH system is interoperable with compatible glucose meters.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Omnipod Insulin Management System and Omnipod DASH Insulin Management System (K192659)
Related Devices
K192659 — Omnipod Insulin Management System, Omnipod DASH Insulin Management System · Insulet Corporation · Oct 21, 2019
K182630 — Omnipod Insulin Management System, Omnipod DASH Insulin Management System · Insulet Corporation · Jan 8, 2019
K180045 — Omnipod DASH Insulin Management System · Insulet Corporation · Jun 1, 2018
{0}
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information
A 510(k) Number
K211575
B Applicant
Insulet Corporation
C Proprietary and Established Names
Omnipod Insulin Management System and Omnipod DASH Insulin Management System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LZG | Class II | 21 CFR 880.5725 - Infusion Pump | CH - Clinical Chemistry |
| NBW | Class II | 21 CFR 862.1345 – Glucose Test System | CH - Clinical Chemistry |
| NDC | Class II | 21 CFR 868.1890 – Predictive pulmonary-function value calculator | CH - Clinical Chemistry |
## E Purpose for Submission
This submission seeks to expand the device labeling of the Insulet Omnipod and Omnipod DASH Insulin Management Systems (K192659) to include an additional insulin, Lyumjev U-100 (Eli Lilly).
## II Intended Use/Indications for Use
A Intended Use(s)
K211575 - Page 1 of 5
{1}
See Indications for Use below.
## B Indication(s) for Use
### Omnipod Insulin Management System
The Omnipod® Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin and for the quantitative measurement of glucose in fresh whole capillary blood (in vitro) from the finger.
The glucose measurements should not be used for the diagnosis or screening for diabetes. The PDM glucose meter is intended for single patient use and should not be shared.
Abbott FreeStyle® test strips are used with the built-in FreeStyle meter for the quantitative measurement of blood glucose in fresh whole capillary blood from the finger, upper arm, and palm. Abbott Freestyle Control Solutions are used to verify that the meter and test strips are working together properly and that the test is performed correctly.
### Omnipod DASH Insulin Management System
The Omnipod DASH™ Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin.
Additionally, the Omnipod DASH System is interoperable with a compatible blood glucose meter to receive and display glucose measurements.
## C Special Conditions for Use Statements
Both devices are for prescription use only.
The Omnipod® Systems are designed to use rapid-acting U-100 insulin. The following U-100 rapid-acting insulin analogs have been tested and found to be safe for use in the Pod: NovoLog® (insulin aspart), Fiasp® (insulin aspart), Humalog® (insulin lispro), Lyumjev™ (insulin lispro-aabc), Apidra® (insulin glulisine) or Admelog® (insulin lispro). Novolog, Fiasp, Humalog, Lyumjev, and Admelog are compatible with the Omnipod Systems for use up to 72 hours (3 days). Apidra is compatible with the Omnipod Systems for use up to 48 hours (2 days). If you have questions about using other insulins, contact your healthcare provider. Fiasp and Lyumjev have a faster initial absorption than other rapid-acting U-100 insulins; always consult with your healthcare provider and refer to the insulin labeling prior to use.
### Omnipod Insulin Management System
Insulin pump therapy is not recommended for people who are:
K211575 - Page 2 of 5
{2}
- Unable to perform at least four (4) blood glucose tests per day
- Unable to maintain contact with their healthcare provider
- Unable to use the System according to instructions
Do not use the built-in blood glucose meter for:
- Testing on neonates
- Testing arterial blood
- Diagnosing of or screening for diabetes mellitus
- Patients who are critically ill, dehydrated, or in diabetic ketoacidosis (DKA)
## Omnipod DASH® Insulin Management System
Insulin pump therapy is not recommended for people who are:
- Unable to monitor blood glucose levels as recommended by their healthcare provider
- Unable to maintain contact with their healthcare provider
- Unable to use the Omnipod DASH® System according to the instructions
## III Device Description
The subject devices provide for the management of insulin therapy and blood glucose monitoring by patients with diabetes mellitus. They are each comprised of two primary components: the disposable insulin infusion pump (Pod) and an associated wireless remote controller referred to as the Personal Diabetes Manager (PDM). The PDMs incorporate a suggested bolus calculator which aids the user in determining the insulin bolus dosage needed based on carbohydrates ingested, most recent blood glucose reading, programmable correction factor, insulin to carbohydrate ratio, target blood glucose value, and Insulin on Board (IoB).
The Pod is a body-wearable insulin pump that affixes to the user on the back of the arm, the lower back or abdomen, the thigh area, or any site that has a layer of fatty tissue available. It is held in place by an adhesive pad and provides up to three days of insulin before it is removed and replaced with a new Pod. The PDM is a handheld device that controls the Pod. The user interfaces with the device system through the PDM, where they control basal and bolus delivery and various insulin program settings and calculations. The PDM also has a food library to assist with carbohydrate calculations, and it maintains several variables in a history log for the viewer to track their diabetes therapy. The Omnipod Insulin Management System PDM has an integrated blood glucose meter and communicates with the Pod wirelessly using secure, low power, bi-directional radio frequency (RF) communications. The Omnipod DASH Insulin Management System PDM does not have an integrated blood glucose meter but is interoperable with a compatible blood glucose meter to receive and display glucose measurements. The Omnipod DASH PDM communicates to the Pod and a compatible blood glucose meter using Bluetooth Low Energy.
K211575 - Page 3 of 5
{3}
K211575 - Page 4 of 5
# IV Substantial Equivalence Information
## A Predicate Device Name(s)
Omnipod Insulin Management System and Omnipod DASH Insulin Management System
## B Predicate 510(k) Number(s)
K192659
## C Comparison with Predicate(s)
| Device & Predicate Device(s): | K211575 | K192659 |
| --- | --- | --- |
| Device trade name | Omnipod Insulin Management System and Omnipod DASH Insulin Management System | Omnipod Insulin Management System and Omnipod DASH Insulin Management System |
| General Device Characteristic Similarities | | |
| Intended Use | Intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin. | Same |
| General Device Characteristic Differences | | |
| Specific Drug/Biologic Use | U-100 Insulin.
System has been tested with NovoLog, Humalog, Apidra, Admelog, Fiasp, and Lyumjev. | U-100 Insulin.
System has been tested with NovoLog, Humalog, Apidra, Admelog, and Fiasp. |
# V Standards/Guidance Documents Referenced
- ISO 14971 Second Edition: 2007, Medical Devices – Application of Risk Management to Medical Devices
- ISO 10993-1 Fifth Edition: 2018, Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process
{4}
VI Performance Characteristics
In vitro stability and leachables testing were conducted with Lyumjev U-100 insulin to verify that the pumps and the insulin are compatible. The stability of the insulin product was evaluated under stressed, worst-case conditions. The studies observed acceptable results of degradation products and leachables, and support the compatibility of this insulin product with these pumps.
A comprehensive hazard analysis for this device was prepared in accordance with ISO 14971 Second Edition: 2007, Medical Devices – Application of Risk Management to Medical Devices. The risk management process included elements of risk analysis, risk evaluation, risk control, and production and post-production information. A safety assurance case was also prepared in accordance with FDA Guidance for Industry and FDA Staff: Infusion Pumps Total Product Lifecycle, December 2, 2014.
VII Proposed Labeling
The labeling supports the finding of substantial equivalence for this device.
VIII Conclusion
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K211575 - Page 5 of 5
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.