K160305 · Natus Medical Incorporated · LBI · May 27, 2016 · General Hospital
Device Facts
Record ID
K160305
Device Name
neoBLUE® LED Phototherapy System
Applicant
Natus Medical Incorporated
Product Code
LBI · General Hospital
Decision Date
May 27, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5700
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The neoBLUE® LED Phototherapy System is indicated for the treatment of hyperbilirubinemia for neonates and infants in a hospital environment, and administered by trained, professional medical staff, on the order of a licensed medical practitioner. The light can be used with a bassinet, incubator, open bed, or radiant warmer.
Device Story
Mobile, floor-standing phototherapy system; delivers high-intensity blue light via LEDs to treat neonatal hyperbilirubinemia. Device consists of light source and optional roll stand; can be placed on incubators. Blue LEDs (400-550 nm, peak 450-475 nm) target bilirubin absorption spectrum; yellow LEDs added to improve visual color rendering. Features high/low intensity settings and cumulative hour timer. Operated by hospital staff; light output facilitates bilirubin degradation. Benefits include effective jaundice treatment with minimal UV/IR exposure, reducing skin damage and overheating risks.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including compliance with IEC 60601-2-50, IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-6, alongside design verification and validation testing.
Technological Characteristics
Floor-standing phototherapy unit; blue and yellow LED light source. Intensity settings: 35 µW/cm²/nm (high) and 15 µW/cm²/nm (low) at 12 inches. Spectral range 400-550 nm. Complies with IEC 60601-2-50, IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-6. Includes cumulative hour timer.
Indications for Use
Indicated for treatment of hyperbilirubinemia in neonates and infants in hospital settings, administered by trained medical staff per physician order.
Regulatory Classification
Identification
A neonatal phototherapy unit is a device used to treat or prevent hyperbilirubinemia (elevated serum bilirubin level). The device consists of one or more lamps that emit a specific spectral band of light, under which an infant is placed for therapy. This generic type of device may include supports for the patient and equipment and component parts.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle or bird symbol, with three overlapping profiles suggesting a sense of community or collaboration. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 27, 2016
Natus Medical Incorporated Ms. Judy Buckham Regulatory Affairs Specialist 5900 First Avenue South Seattle, Washington 98108
Re: K160305
Trade/Device Name: neoBLUE® LED Phototherapy System Regulation Number: 21 CFR 880.5700 Regulation Name: Neonatal Phototherapy Unit Regulatory Class: Class II Product Code: LBI. Dated: April 14, 2016 Received: April 18, 2016
Dear Ms. Buckham:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang
-s
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K160305
Device Name
neoBLUE® LED Phototherapy System
#### Indications for Use (Describe)
Indications for Use: The neoBLUE® LED Phototherapy System is indicated for the treatment of hyperbilirubinemia for neonates and infants in a hospital environment, and administered by trained, professional medical staff, on the order of a licensed medical practitioner. The light can be used with a bassinet, incubator, open bed, or radiant warmer.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# K160305 510(K) SUMMARY
| Manufacturer's<br>Name: | Natus Medical Incorporated<br>5900 First Avenue South<br>Seattle, WA 98108 |
|------------------------------------|----------------------------------------------------------------------------|
| Corresponding<br>Official: | Judy Buckham<br>Regulatory Affairs Specialist |
| Telephone | 206 268 5187 |
| Number:<br>Fax Number: | 206 268 5104 |
| Preparation Date: | April 14, 2016 |
| Trade Name: | neoBLUE® LED Phototherapy System |
| Common or Usual<br>Name: | Neonatal Phototherapy Unit |
| Classification<br>Name and Number: | Unit, Neonatal Phototherapy<br>21 CFR 880.5700<br>Product Code: LBI |
| Predicate Device: | K022196 Natus Blue Light Phototherapy Unit |
## Device Description
The neoBLUE® Phototherapy System consists of two products – the neoBLUE LED Phototherapy light source (light) and the neoBLUE LED Phototherapy roll stand.
The neoBLUE LED Phototherapy System is a floor-standing, mobile phototherapy light that delivers a narrow band of high-intensity blue light emitting diodes (LEDs) to provide treatment for neonatal hyperbilirubinemia. The light can be used independently of the roll stand, and can be placed directly on an incubator with a flattopped surface.
### Intended Use
The neoBLUE LED Phototherapy System is indicated for the treatment of hyperbilirubinemia for neonates and infants in a hospital environment, and administered by trained, professional medical staff, on the order of a licensed medical practitioner. The light can be used with a bassinet, incubator, open bed, or radiant warmer.
{4}------------------------------------------------
The indications for use of the predicate (K022196) reference the phototherapy system as emitting "blue light." The indications for use for the subject device (neoBLUE LED Phototherapy System) do not include this statement because the system emits blue and yellow light. Removal of this sentence does not raise new questions of safety or effectiveness.
# Technological Characteristics
The neoBLUE light has intensity settings, high and low, which are factory callbrated to provide intensity of 35 µW/cm²/nm at the high setting and 15 µW/cm²/nm at the low setting at a distance of 12 inches (30.5 cm) from the light enclosure to the baby. The light output can also be adjusted to accommodate different distances.
Blue LEDs emit light in the range of 400 - 550 nm (peak wavelength 450-475 nm). This range corresponds to the spectral absorption of light by bilirubin, and is thus considered to be the most effective for the degradation of bilirubin. Blue LEDs do not emit significant energy in the ultraviolet (UV) region of the spectrum, reducing the potential risk of skin damage. In addition. blue LEDs do not emit significant energy in the infrared (IR) region of the spectrum, minimizing concern about excessive warming of the infant.
The blue LEDs in the device produce the same therapeutic output as in the predicate. Yellow LEDs were intermixed with the blue LEDs in the new device so that the light output visually appears less blue in color, the light source was separated from the roll stand for independent use, and a timer was added to record the cumulative number of working hours for the LED lights.
# Clinical Tests
N/A
# Nonclinical Tests
The neoBLUE LED Phototherapy System passed all testing and complies with the following referenced standards:
- IEC 60601-2-50:2009-03 Edition 2; Medical Electrical Equipment Part 2-50: Particular . Requirements for the Basic Safety and Essential Performance of Infant Phototherapy Equipment
- IEC 60601-1:2005/(R)2012 and C1:2009/(R)2012; Medical Electrical Equipment -Part 1: ● General Requirements for Basic Safety
- IEC 60601-1-2: 2007/(R)2012; Medical Electrical Equipment Part 1-2: General ● Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
- IEC 60601-1-6:2013-10 Edition 3.1: Medical Electrical Equipment Part 1-6: General . Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability
In addition to compliance with the relevant standards, design verification and validation testing was performed to ensure that the neoBLUE LED Phototherapy System meets its performance specifications and demonstrates equivalence to the predicate device (K022196).
{5}------------------------------------------------
# Conclusions
The performance verification/validation testing data and risk analysis documentation provided in this 510(k) support the conclusion that the neoBLUE LED Phototherapy system performs as intended and is substantially equivalent to the predicate device.
The neoBLUE LED Phototherapy System is substantially equivalent to the Natus Blue Light Phototherapy Unit cleared under K022196.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.