← Product Code [FNM](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FNM) · K973169

# PRESSURE GUARD SITE SELECT (A) (K973169)

_Span-America Medical Systems, Inc. · FNM · Feb 20, 1998 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FNM/K973169

## Device Facts

- **Applicant:** Span-America Medical Systems, Inc.
- **Product Code:** [FNM](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FNM.md)
- **Decision Date:** Feb 20, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5550
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

For the prevention and treatment of pressure ulcers.

## Device Story

PressureGuard Site Select is an air flotation, alternating pressure mattress system. It consists of an air inflation system housed within a foam shell, featuring a foam topper and bolsters. Unlike the predicate, which uses lengthwise cylinders, this device utilizes side-to-side air cylinders organized into four zones: head, upper torso, lower torso, and heel. The system provides multiple operational modes, including Normal (alternating on/off), Heel Relief (alternating on/off), Contour (customized firmness per zone), and CPR (firm surface). The device is intended for use in clinical or home settings to manage patient pressure distribution. Healthcare providers or caregivers adjust air pressure set points to meet individual patient needs. By providing alternating pressure and customized zone firmness, the device aims to reduce interface pressure, thereby preventing or treating pressure ulcers.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Air flotation alternating pressure mattress. Components: foam shell with topper and bolsters, side-to-side air cylinders. Four-zone configuration (head, upper torso, lower torso, heel). Operational modes: Normal, Heel Relief, Contour, CPR. Connectivity: None specified. Power: Air inflation system.

## Regulatory Identification

An alternating pressure air flotation mattress is a device intended for medical purposes that consists of a mattress with multiple air cells that can be filled and emptied in an alternating pattern by an associated control unit to provide regular, frequent, and automatic changes in the distribution of body pressure. The device is used to prevent and treat decubitus ulcers (bed sores).

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 880.9.

## Predicate Devices

- PressureGuard IV ([K953503](/device/K953503.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

SPAN-AMERICA | K973169 MEFICAL SYSTEMS; TNC.

博

K973169

FEB 20 1998

## PressureGuard Site Select 510K Summary

| 1. | Submitter's Name:        | Span-America Medical Systems, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|----|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|    | Address:                 | 70 Commerce Center<br>Greenville, SC 29615                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|    | Telephone Number:        | (864) 288-8877                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|    | Contact Person:          | James D. Ferguson, President & CEO<br>Wanda Totton, Director of Quality                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|    | Date Prepared:           | October 21, 1997                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| 2. | Trade Name:              | PressureGuard Site Select, Model A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|    | Common Name:             | Alternating Pressure Mattress                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|    | Classification Name:     | Alternating Pressure Air Flotation Mattress,<br>CFR 880.5550, Classification No. 80 FMN.                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| 3. | Predicate Device:        | PressureGuard IV, 510K953503                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| 4. | Description:             | The PressureGuard Site Select is an air flotation,<br>alternating pressure support system which provides for<br>customized selection of surface firmness for a head, upper<br>torso, lower torso, and heel zones.                                                                                                                                                                                                                                                                                                                         |
| 5. | Indications for Use:     | For the prevention and treatment of pressure ulcers.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| 6. | Substantial Equivalence: | The product is similar in function and intended use to the<br>PressureGuard IV, 510K953503. Both systems consists of<br>an air inflation system within a foam shell. The foam shell<br>includes a foam topper with bolsters that frame the<br>mattress. For the PressureGuard IV, four air cylinders<br>run lengthwise within the mattress and operate as two-<br>pairs. For the PressureGuard Site Select, the air cylinders<br>run side-to-side and operate as four zones, defined as<br>head, upper torso, lower torso, and heel zones |

{1}------------------------------------------------

Span-America Medical Systems, Inc. PressureGuard Site Select 510K Summary Page 2 of 2

> For both systems, air in the air cylinders is adjusted to defined pressure set points for four defined modes of operation. For the PressureGuard IV, a Center Mode (supine surface), CPR Mode (firm surface for performing CPR with a crash board), Fixed Roll ( 30 degree left and right rolls with two-hour dwell times), and Custom Roll (provides for options for roll pattern, roll angles, and dwell times) are provided.

> For the PressureGuard Site Select, Normal Mode, Alternating Off; Normal Mode, Alternating On; Heel Relief Mode, Alternating Off; Heel Relief Mode, Alternating On; Contour, and CPR modes are provided. The Normal Mode, Alternating Off Mode is comparable to the Center Mode for PressureGuard IV. The Heel Relief Mode, Alternating Off is comparable to the Center Mode for PressureGuard IV with lowered internal air pressures for the heel zone. The CPR Mode is comparable to the CPR Mode for PressureGuard IV. In addition, alternating pressure modes are options for both the Normal Mode and Heel Relief Modes. The Contour Mode provides for customized adjustment of firmness settings for each of the four zones to meet individual patient needs.

> Both products are indicated for the prevention and treatment of pressure ulcers.

lho

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a stylized eagle. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The eagle is depicted with its wings spread, symbolizing protection and service.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 20 1998

Ms. Wanda J. Totton Director of Quality Span-America Medical Systems, Inc. P.O. Box 5231 29606 Greenville, South Carolina

Re: K973169 PressureGuard Site Select Trade Name: Requlatory Class: II Product Code: _ ENM _ _ _ _ _ Dated: January 26, 1998 Received: January 27, 1998

Dear Ms. Totton:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ದ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Ms. Wanda J. Totton

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other general premarket notification" (21 CFR 807.97). information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Cella M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page ol

510(k) Number (if known):

Device Name: Pressure Guard Site Alert

Indications For Use:

For the prevention and treatment of
Pressure ulcers.

L.C. Patterson
8-7-97

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

tion. (GLE)

(Division Sign-Off)
Division of General Restorative Devices K973169:
510(k) Number.

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use X

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FNM/K973169](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FNM/K973169)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
