← Product Code [FMZ](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMZ) · K072512

# MODIFICATION TO: OHMEDA MEDICAL GIRAFFE INCUBATOR (GIRAFFE AND PANDA UNINTERRUPTIBLE POWER SUPPLY) (K072512)

_Datex-Ohmeda, Inc. · FMZ · Oct 3, 2007 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMZ/K072512

## Device Facts

- **Applicant:** Datex-Ohmeda, Inc.
- **Product Code:** [FMZ](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMZ.md)
- **Decision Date:** Oct 3, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 880.5400
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Giraffe Incubator is an Infant Incubator. Incubators provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. They achieve this by providing an enclosed temperature controlled environment to the infant. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide a stable oxygen concentration within the infant compartment at the value set by the operator (21-65%)

## Device Story

Infant incubator providing thermal support for neonates unable to thermo-regulate; maintains temperature via circulating heated air in enclosed compartment. Optional Servo Controlled Oxygen Delivery System regulates oxygen concentration (21-65%) using fuel cell sensors; microprocessor compares sensor output to operator-set value and adjusts valves accordingly; compensates for temperature/humidity fluctuations. Includes Giraffe and Panda Uninterruptible Power Supply (UPS) for intra-hospital mobility; UPS provides short-term battery power. Used in clinical settings; operated by healthcare professionals. Benefits include stable thermal and oxygen environment for vulnerable infants.

## Clinical Evidence

No clinical or animal testing performed. Safety and effectiveness demonstrated via extensive bench testing and software validation in accordance with 21 CFR 820, Subpart C-Design Controls.

## Technological Characteristics

Neonatal incubator; air/skin temperature control modes; fuel cell oxygen sensors; microprocessor-based control system. Includes UPS with medical-grade battery. Bench tested against consensus standards. Software validated per design controls.

## Regulatory Identification

A neonatal incubator is a device consisting of a rigid boxlike enclosure in which an infant may be kept in a controlled environment for medical care. The device may include an AC-powered heater, a fan to circulate the warmed air, a container for water to add humidity, a control valve through which oxygen may be added, and access ports for nursing care.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K0725/2 p. lot=2

## 510(k) Summary

### Submitter Information and Date Prepared

Agata Smieja GE Healthcare 8880 Gorman Road Laurel, MD 20723 USA

Phone: 410 888 5218 Mobile: 410 456 0329

Prepared: September 4, 2007

Device Identification

Proprietary Name: Common Name:

Ohmeda Medical Giraffe Incubator Giraffe Incubator

Predicate Device Information

| <br>- Children Children Children<br><br>:<br>ี<br>Status Career Station Company Comers<br><br>Dron<br>:<br><br><br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | <br><br>.<br><br><br>14 13 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11<br>.<br>and antinent and answer and American and American and an-<br><br>SANT SHOW SHOW<br>A<br><br><br><br> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| A LEAR BELL AND I A<br>GITATTA<br>ANNUAL AND A LA LA LA LA LEA LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE L                                                                                  |                                                                                                                                                                                                                                                                                                                 |

Intended Use Statement

The Giraffe Incubator is an Infant Incubator. Incubators provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. They achieve this by providing an enclosed temperature controlled environment to the infant. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide a stable oxygen concentration within the infant compartment at the value set by the operator (21-65%)

**B** 001

{1}------------------------------------------------

K622572 p. dof 3

### Product Description

The Ohmeda Medical Giraffe Incubator is an infant bed which provides thermal support for infants who are unable to provide for their own heat requirements. The device manintains the infant's temperature by circulating heated air within the enclosed bed compartment. The operator may select either the air or skin temperature control method. Depending on the control method selected, heat is regulated based on either the air temperature or the infant's skin temperature compared to the operator selected control temperature. Physical access to the patient is obtained through the side portholes or by opening one of the side doors. The optional Giraffe Servo Control Oxygen Delivery System is a fully integrated option available on the Giraffe Incubator. The Giraffe Servo Control Oxygen Delivery System is capable of oxygenating the entire infant compartment at oxygen concentrations of 21%-65% by volume. The device uses fuel cell type sensors that generate specific voltages depending on the oxygen concentrations they contact. The microprocessor stores the sensor output and compares it with the value corresponding to the concentration set by the operator. The valves that supply oxygen to the infant compartment are opened and closed as necessary to maintain the oxygen concentration at the set value. Fluctuations in fuel cell performance due to temperature and humidity are are compensated for by the microprocessor.

The Giraffe and Panda Uninterruptible Power Supply (UPS) provides a short term source of electrical power Giraffe Incubator, thus aiding its intra hospital mobility. The Giraffe UPS does not change the indications for use, control mechanisms, operating principles, performance specifications, or other features of the Giraffe Incubator.

The UPS serves as an extension to the Giraffe Incubator by providing uninterrupted electric power to the device. The UPS comprises a medical grade battery and a shelf.

# Performance Data

Since care of newborns in incubators, with or without supplemental oxygen, is a well established clinical practice, Ohmeda submits that clinical or animal testing to demonstrate safety and effectiveness is not necessary. The product was subject to extensive bench testing, the software was validated and, to the best of Ohmeda Medical's knowledge, the requirements of 21 CFR 820, Subpart C-Design Controls were satisfied.

{2}------------------------------------------------

KC7-2572 p. 3 of 3

The UPS battery has a life of 15 minutes in nominal operating conditions (37C, 70% RH), 10 minutes at the worst case operation condition (39C, 95% RH) and 7 minutes at the worst case operating condition with the addition of the phototherapy accessory and with all available lights on (39C, 95% RH) . The recharge time is 6 hours.

Performance of the Giraffe Incubator with the addition of UPS has been established by bench testing against product specifications and recognized consensus standards.

Prepared by: AGATA SMIEYA

Date 09/04/07

.3

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is facing to the right. The logo is simple and recognizable.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Agata Smieja Global Compliance Leader Datex-Ohmeda, Incorporated A Division of Datex Ohmeda, A GE Healthcare Company 8880 Gorman Road Laurel, Maryland 20723

2007

Re: K072512

Trade/Device Name: Ohmeda Medical Giraffe Incubator Regulation Number: 21 CFR 880.5400 Regulation Name: Neonatal Incubator Regulatory Class: II Product Code: FMZ Dated: September 4, 2007 Received: September 6, 2007

Dear Ms. Smieja:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Smieja

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known):_K 072512

Device Name: Ohmeda Medical Giraffe Incubator

Indications For Use:

The Giraffe Incubator is an Infant Incubator. Incubators provide heat in a rric Girane incubates who are unable to thermo-regulate based on their controlled manner to neonates his by providing an enclosed temperature own physiology. They as a the infant. This device may incorporate a Servo controlled Oxygen Delivery System. This is indicated to provide a stable oxygen ്റ്സീറ്റില്‍ ഗ്രാമ്യൂട്ടി Dellvery Bysts.ompartment at the value set by the operator (21-65%)

X Prescription Use

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

d d

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Chuton vem

(Division Sign-Off) Division of Anesthesiology, General Hospital Page 1 of 1 Infection Control, Dental Devices

2072512 510(k) Number:

A 001

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMZ/K072512](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMZ/K072512)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
