MOTEX POWDER-FREE SURGICAL GLOVES AND LIGHTLY POWDERED SURGICAL GLOVES
K063757 · Shanghai Motex Healthcare Co., Ltd. · KGO · Jul 27, 2007 · General, Plastic Surgery
Device Facts
Record ID
K063757
Device Name
MOTEX POWDER-FREE SURGICAL GLOVES AND LIGHTLY POWDERED SURGICAL GLOVES
Applicant
Shanghai Motex Healthcare Co., Ltd.
Product Code
KGO · General, Plastic Surgery
Decision Date
Jul 27, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4460
Device Class
Class 1
Intended Use
Motex Powder-Free Surgical Gloves are sterile disposable devices made of natural rubber latex (that may bear a trace amount of glove powder) and is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants. Motex Powdered Latex Surgical Gloves are sterile disposable devices made of natural rubber latex that bears powder to facilitate donning, and it is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants.
Device Story
Sterile, disposable natural rubber latex surgical gloves; available in powder-free and powdered versions. Designed for use by clinicians in surgical settings to provide a protective barrier against infectious materials and contaminants. Powdered version includes powder to facilitate donning. Device functions as a physical barrier; no electronic or mechanical components.
Clinical Evidence
No clinical data.
Technological Characteristics
Natural rubber latex surgical gloves. Meets ASTM D 3577-06 for dimensions, physical properties, and freedom from holes. Biocompatibility per ISO 10993-5/-10. Sterilization validation per ISO 11137. Powder-free variant: <= 2mg powder, <= 50ug/dm2 extractable protein. Powdered variant: <= 120mg powder, <= 120ug/dm2 extractable protein.
Indications for Use
Indicated for use by healthcare personnel in surgical settings as a protective barrier against infectious materials and contaminants. Suitable for use on hands of surgical staff.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
MEDISPO powdered surgeon's gloves and MEDISPO-PF powder-free surgeon's gloves (K050071)
Related Devices
K191898 — Sterile Latex Powder-Free Surgical Gloves · Modern Healthcare Corp. · Dec 18, 2019
K992020 — BRIGHTWAY BRAND LATEX SURGICAL GLOVES - STERILE (POWDERED) · Brightsway Holdings Sdn Bhd · Jul 26, 1999
K980564 — MEDI-GRIP (BEIGE 43)SURGICAL GLOVES · Ansell Perry · Mar 25, 1998
K032467 — MAXTER STERILE POWDERED LATEX SURGICAL GLOVES · Maxter Glove Manufacturing Sdn Bhd · Oct 10, 2003
K012135 — POWDER FREE BROWN LATEX SURGICAL GLOVES WITH PROTEIN CONTENT LABELING CLAIM (50 MICROGRAMS OR LESS) · Wrp Asia Pacific Sdn. Bhd. · Aug 1, 2001
Submission Summary (Full Text)
{0}------------------------------------------------
JUL 27 2007
## SHANGHAI MOTEX HEATHCARE CO., LTD.
No.369, Jiasong Zhong Road, Huaxin, Qingpu, Shanghai, 201708, China Telephone: 86-21-5979 9888 Fax: 86-21-5979 9728 E-mail: motex@public 1.sta.net.cn
#### દર્ 510(k) SUMMARY "
The assigned 510(k) number is : __ K063757
Submitter's Name: SHANGHAI MOTEX HEALTHCARE CO., LTD. No.369, Jiasong Zhong Road, Huaxin, Qingpu, Shanghai, 201708, China
### Date summary prepared: Device Name:
November 26, 2006
- Classification name: Surgeon's Gloves
- . Classification number: KGO, Class I
- Regulation Number: 878.4460
- . Proprietary name: Motex Powder-Free Surgical Gloves & Powdered Latex Surgical Gloves
- . Predicate Device: K050071. MEDISPO powdered surgeon's gloves and MEDISPO-PF powder-free surgeon's gloves
. Official Correspondent: Dr. Jen. Ke-Min
> E-mail: ceirs.jen@msa.hint.net (Tel) 886-3-5208829; (Fax) 886-3-5209783 Address: No.58, Fu Chiun Street, Hsin Chu City, 30067, Taiwan, ROC
#### ● Intended Use:
Motex Powder-Free Surgical Gloves are sterile disposable devices made of natural rubber latex (that may bear a trace amount of glove powder) and is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants.
{1}------------------------------------------------
# SHANGHAI MOTEX HEATHCARE CO., LTD.
No.369, Jiasong Zhong Road, Huaxin, Qingpu, Shanghai, 201708, China Fax: 86-21-5979 9728 Telephone: 86-21-5979 9888 E-mail: motex@public1.sta.net.cn
Motex Powdered Latex Surgical Gloves are sterile disposable devices made of natural rubber latex that bears powder to facilitate donning, and it is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants..
#### Technological Characteristics: ●
The Motex Powder-Free Surgical Gloves & Powdered Latex Surgical Gloves characteristics are summarized below compared to ASTM and ISO standards to the predicate device:
| Characteristic | Standard |
|--------------------------|-----------------------|
| Dimensions | meets ASTM D 3577-06, |
| Physical Properties | meets ASTM D 3577-06, |
| Freedom from Holes | meets ASTM D 3577-06, |
| Biocompatibility | meets ISO10993-5/-10 |
| Sterilization Validation | meets ISO11137 |
- . Clinical Data:
Not applicable.
#### ● Conclusions:
The Motex Powder-Free Surgical Gloves & Powdered Latex Surgical Gloves and predicate devices also meet the technological characteristics of ASTM D 3577-06, ISO10993-5/-10, and ISO11137 standards. Besides, for Powder-Free SURGICAL GLOVES contain no more than 2mg powder and no more than 50ug/dm2 extractable protein, and for Powdered I.atex SURGICAL GLOVES contuin no more than 120mg powder and no more than 120ug/dm2 extractable protein claim.
Thus the new devices are substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The eagle is black, and the text is in a smaller, sans-serif font.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### JUL 2 7 2007
Shanghai Motex Healthcare Company Limited C/O Dr. Jen Ke-Min Official Correspondent ROC Chinese-European Industrial No. 58, Fu Chiun Street Hsin Chu City, Taiwan 30067 CHINA
Re: K063757
Trade/Device Name: Motex Powder-Free Surgical Gloves and Powdered Latex Surgical Gloves Regulation Number: 878.4460 Regulation Name: Surgeon's Glove Regulatory Class: I Product Code: KGO Dated: June 15, 2007
Received: June 29, 2007
Dear Dr. Jen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent ffor the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
### Page 2 - Dr. Jen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sutterf. Mckano MD
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510 (K) Number ( If Known ):
Device Name: Motex Powder-Free Surgical Gloves and Powdered Latex Surgical Gloves
Indications for Use:
Motex Powder-Free Surgical Gloves are sterile disposable devices made of natural rubber latex (that may bear a trace amount of glove powder) and is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants.
Motex Powdered Latex Surgical Gloves are sterile disposable devices made of natural rubber latex that bears powder to facilitate donning, and it is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants.
Prescription Use
AND/OR
Over-The-Counter Use \$\sqrt{}\$
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
P.4
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Mision Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
Page 1 of
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.