← Product Code [QBW](/submissions/SU/subpart-e%E2%80%94surgical-devices/QBW) · K180217

# VESTEX Apparel (K180217)

_Vestagen Protective Technologies, Inc. · QBW · May 18, 2018 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/QBW/K180217

## Device Facts

- **Applicant:** Vestagen Protective Technologies, Inc.
- **Product Code:** [QBW](/submissions/SU/subpart-e%E2%80%94surgical-devices/QBW.md)
- **Decision Date:** May 18, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

VESTEX® Apparel ("VESTEX"), scrub tops, pants, skirts, and lab coats are apparel that provides fluid repellency and reduces the retention of Methicillin Resistant Staphylococcus aureus (MRSA) on the surface (outer layer) of the apparel. The ability of VESTEX to reduce the retention of MRSA on the fabric surface has not been shown to correlate with the reduction of infections. Clinical studies to evaluate reduction in infection have not been performed for this device. VESTEX Apparel is not intended to replace personal protective equipment.

## Device Story

VESTEX Apparel is nonsterile, reusable, fluid-repellent, antimicrobial clothing for healthcare workers; designed for continuous wear to protect against intermittent/unexpected exposure to blood, body fluids, and potentially infectious materials. Fabric features dual-action surface: fluid-repellent barrier to resist bacteria acquisition and antimicrobial agent to disrupt cell membranes/proteins/bio-energetics. Interior side is moisture-wicking. Device does not replace PPE for expected exposure episodes. Healthcare providers wear apparel during clinical duties; output is the physical barrier/antimicrobial surface reducing MRSA burden on clothing. Benefits include reduced pathogen retention on fabric surface, though clinical correlation to infection reduction is not established.

## Clinical Evidence

Clinical evidence includes a randomized, blinded, crossover trial (Bearman et al., 2012) in a medical ICU. Study compared VESTEX apparel to control clothing on healthcare workers. Results showed statistically significant reduction in MRSA accumulation on VESTEX apparel at both the beginning and end of 12-hour shifts. Bench testing included biocompatibility (ISO 10993-1:2009), fluid repellency (AATCC22), and splatter testing for MRSA retention.

## Technological Characteristics

Materials: 79% polyester, 18% rayon, 3% spandex. Dual-action surface: fluid-repellent barrier and antimicrobial agent covalently bonded to outer fabric. Interior: moisture-wicking. Standards: ISO 10993-1 (biocompatibility), ASTM D1424-09:2013 (tearing strength). Reusable; nonsterile.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Pro-Safe Professional Linens, Inc. Aprons, Scrub Slacks & Tops, Patient Drape ([K891212](/device/K891212.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 18, 2018

Vestagen Protective Technologies, Inc. Scott Pease Sr. Vice President of Regulatory and Quality 1301 W. Colonial Dr. Orlando, Florida 32804

Re: K180217

Trade/Device Name: VESTEX Apparel ("VESTEX") Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: QBW Dated: April 16, 2018 Received: April 17, 2018

Dear Scott Pease:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Geeta K. Pamidimukkala -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K180217

Device Name VESTEX® Apparel ("VESTEX")

Indications for Use (Describe)

VESTEX® Apparel ("VESTEX"), scrub tops, pants, skirts, and lab coats are apparel that provides fluid repellency and reduces the retention of Methicillin Resistant Staphylococcus aureus (MRSA) on the surface (outer layer) of the apparel.

The ability of VESTEX to reduce the retention of MRSA on the fabric surface has not been shown to correlate with the reduction of infections. Clinical studies to evaluate reduction in infection have not been performed for this device.

VESTEX Apparel is not intended to replace personal protective equipment.

| Type of Use (Select one or both, as applicable) |                                             |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D)    | Over-The-Counter Use (21 CFR 801 Subpart C) |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# VESTEX® Apparel "Vestex"

# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Vestagen Protective Technologies, Inc. 1301 W. Colonial Dr. Orlando. FL 32804

(o) 407.781.2572 (m) 678.689.8010

Contact Person: Scott Pease, Sr. Vice President of Regulatory and Quality (scott.pease@vestagen.com)

Date Prepared: May 3, 2018

| Name of Device                              | VESTEX® Apparel                                                                              |
|---------------------------------------------|----------------------------------------------------------------------------------------------|
| Common or Usual Name                        | Healthcare Practitioner Apparel (Scrub)                                                      |
| Classification Name                         | Suit, Surgical                                                                               |
| Product Code / Regulation<br>Classification | QBW / 21 CFR § 878.4040 Class II                                                             |
| Predicate Devices                           | K891212 (Aprons, Scrub Slacks & Tops, Patient Drape) -<br>Pro-Safe Professional Linens. Inc. |

#### Indications for Use

VESTEX® Apparel ("VESTEX"), scrub tops, pants, skirts, and lab coats are apparel that provides fluid repellency and reduces the retention of Methicillin Resistant Staphylococcus aureus (MRSA) on the surface (outer layer) of the apparel.

The ability of VESTEX to reduce the retention of MRSA on the fabric surface has not been shown to correlate with the reduction of infections. Clinical studies to evaluate reduction in infection have not been performed for this device.

VESTEX Apparel is not intended to replace personal protective equipment.

### Device Description / Technological Characteristics

VESTEX® Apparel is nonsterile, reusable, fluid repellant apparel containing an antimicrobial agent that is designed for continuous wear to provide protection to healthcare workers where intermittent or unexpected exposure to microorganisms from blood, body fluids and other potentially infectious material (OPIM) can occur.

VESTEX® does not replace PPE worn for specific use and during episodes of expected blood, body fluid and other potentially infectious material (OPIM) exposure.

VESTEX® has a dual mechanism of action that is fluid repellant and an antimicrobial agent. VESTEX® has a proprietary application method used to covalently bond a fluid repellent and an antimicrobial agent to the outer surface of the fabric.

{4}------------------------------------------------

The primary mechanism of action is a fluid repellant fabric surface which acts to resist bacteria acquisition on the fabric. As the fluids and pathogens are repelled, the numbers of microorganisms that remain on the fabric are significantly reduced.

The secondary mechanism of action is an antimicrobial agent. It has one of several possible modes of action, including disruption of the cell membrane, denaturation of cell surface and transmembrane proteins, and inactivation of bio-energetic systems to prevent the microorganisms from being retained on the fabric. This secondary mechanism of action reduces Methicillin Resistant Staphylococcus aureus (MRSA) on the surface (outer layer) of the apparel.

### Performance Data

Biocompatibility and functional bench testing performed by Vestagen Protective Technologies, Inc. demonstrates VESTEX® Apparel substantial equivalence, in terms of the safety and effectiveness, to the referenced predicate device. In vitro bench testing included an assessment of all design input requirements and confirmation that the output of the design process met all design input requirements was completed, including those relating to appropriate standards and guidance's, as follows:

#### Biocompatibility

- ISO 10993-1:2009 Biological Evaluation of Medical Devices- Part I: ● Evaluation and Testing
	- Skin Sensitization O
	- Irritation O
	- Cytotoxicity O

### Functional

- ASTM D1424-09:2013 Tearing Strength of Fabrics by Falling-Pendulum ● (Elemdorf-Type) Apparatus
- . Fabric Clinical Environment / Reuse
	- Pre-Conditioning o
	- O Fluid Repellency: Spray Test
	- Splatter Testing (Reduction of MRSA Retention) O
	- Cleaning Validation O

Additionally, in a randomized, blinded, crossover trial within a medical ICU, VESTEX® apparel demonstrated its effectiveness in reducing accumulation of Methicillin Resistant Staphylococcus aureus (MRSA) on health care workers (HCW) clothing 1 compared to control clothing. The MRSA reduction as statically significant both at the beginning and end of a 12 hour work shift.

- 1) Bearman, GML, et al. A Crossover Trial of Antimicrobial Scrubs to Reduce Methicillin-Resistant Staphylococcus aureus Burden on Healthcare Worker Apparel. Infection Control and Hospital Epidemiology. Vol. 33, No. 3 (March 2012), pp. 268-275.

{5}------------------------------------------------

| Feature                       | Subject Device (K180217)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate Device<br>(K891212)                                                                                        |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indications | VESTEX® Apparel (“VESTEX”),<br>scrub tops, pants, skirts, and lab coats<br>are apparel that provides fluid<br>repellency and reduces the retention of<br>Methicillin Resistant Staphylococcus<br>aureus (MRSA) on the surface (outer<br>layer) of the apparel.<br>The ability of VESTEX to reduce the<br>retention of MRSA on the fabric<br>surface has not been shown to<br>correlate with the reduction of<br>infections. Clinical studies to evaluate<br>reduction in infection have not been<br>performed for this device. | Protective apparel for<br>dentists to prevent the<br>migration of bacteria and<br>liquids.                           |
|                               | VESTEX Apparel is not intended to<br>replace personal protective equipment.                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                      |
| Materials of<br>Construction  | 79% Polyester, 18% Rayon, 3%<br>Spandex                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 65% polyester, 35% cotton<br>back with Barrier Supreme®<br>50% cotton 50% polyester<br>front.                        |
| Design Features               | One layer of fabric constructed as an<br>active barrier. Exterior has a dual<br>mechanism of action to include a fluid<br>barrier and antimicrobial agent.<br>Interior has moisture wicking on the<br>skin touching side of the fabric.                                                                                                                                                                                                                                                                                        | One layer of fabric Scrub<br>Top and Slacks constructed<br>with a fluid resistant fabric in<br>front of the garment. |
| Performance<br>Testing        | Biocompatibility:<br>Yes - ISO 10993-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Biocompatibility:<br>Not available                                                                                   |
|                               | Fluid Repellency (@ 0 & 50 washes):<br>Yes - Simulated Worst Case Study<br>(Variation of AATCC22)                                                                                                                                                                                                                                                                                                                                                                                                                              | Fluid Repellency (@ 0 & 50<br>washes):<br>Yes - AATCC22                                                              |
|                               | Durability & Tear Strength:<br>Yes - ASTM D1424                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Durability & Tear Strength:<br>Yes - ASTM D2261                                                                      |
|                               | Splatter Challenge (Reduction of<br>MRSA Retention @ 0 & 50 Washes):<br>Yes - Simulated Worst Case Study                                                                                                                                                                                                                                                                                                                                                                                                                       | Reduction of MSRA<br>Retention:<br>No                                                                                |
|                               | Clinical Trial<br>Yes - Bearman, GML, et al. A<br>Crossover Trial of Antimicrobial<br>Scrubs to Reduce Methicillin-<br>Resistant Staphylococcus aureus<br>Burden on Healthcare Worker Apparel                                                                                                                                                                                                                                                                                                                                  | Clinical Trial<br>No                                                                                                 |

# Summary of Substantial Equivalence

Based upon the device description, technical characteristics and test data provided within this submission, VESTEX® Apparel is substantially equivalent to the referenced predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/QBW/K180217](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/QBW/K180217)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
