← Product Code [OXZ](/submissions/SU/subpart-e%E2%80%94surgical-devices/OXZ) · K210321

# Dukal Corporation Level 1 Pediatric Face Mask (K210321)

_Dukal Corporation · OXZ · Mar 29, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/OXZ/K210321

## Device Facts

- **Applicant:** Dukal Corporation
- **Product Code:** [OXZ](/submissions/SU/subpart-e%E2%80%94surgical-devices/OXZ.md)
- **Decision Date:** Mar 29, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Pediatric

## Indications for Use

The Dukal Level 1 Pediatric Face Masks are intended to be worn by the patient/child to cover the nose and mouth to provide a barrier for the respiratory tract from microorganisms, body fluids, and particulate material. Recommended ages are 4-12. The face mask is specifically for use with patients whose age or illness may prevent them from taking necessary precautions in situations where transfer of microorganisms, body fluids, and particulates can occur. The face masks are recommended for use in a healthcare setting with appropriate adult supervision. The Dukal Level 1 Pediatric Face Masks are single use, disposable devices that are provided non-sterile.

## Device Story

Pediatric face mask; three-layer pleated design; inner/outer layers polypropylene (spunbond); middle layer polypropylene (meltblown) filter; knitted polyester/spandex ear loops; malleable aluminum nosepiece. Used in healthcare settings; worn by children ages 4-12; requires adult supervision. Provides physical barrier against microorganisms, body fluids, and particulates. Single-use; disposable; non-sterile. Decorative patterns printed with colored inks. Benefits patient by reducing exposure to contaminants in clinical environments.

## Clinical Evidence

No clinical data. Bench testing only. Conformance demonstrated via ASTM F2101 (BFE >99%), ASTM F2299 (PFE >99%), EN 14683 (Delta-P ≤3.1 mmH2O/cm²), ASTM F1862 (80 mmHg), 16 CFR 1610 (Class I), and ISO 10993 biocompatibility standards. Anthropometric analysis confirmed coverage for children ages 4-12.

## Technological Characteristics

Three-layer pleated construction: polypropylene spunbond (inner/outer), polypropylene meltblown (middle). Polyester/spandex ear loops; aluminum nosepiece. Ultrasonic hot sealing. Non-sterile. Biocompatibility per ISO 10993-5/10. Tested per ASTM F2100-19 Level 1 requirements. No software or electronic components.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Prestige Ameritech Pediatric/Child Face Mask (level 1) ([K160100](/device/K160100.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 29, 2022

Dukal. LLC Megan Quevedo Quality and Regulatory Affairs Engineer 2 Fleetwood Court Ronkonkoma, New York 11779

Re: K210321

Trade/Device Name: Dukal Level 1 Pediatric Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: OXZ Dated: February 22, 2022 Received: February 28, 2022

Dear Megan Quevedo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian, M.D., Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K210321

Device Name Dukal Level 1 Pediatric Face Mask

#### Indications for Use (Describe)

The Dukal Level 1 Pediatric Face Masks are intended to be worn by the patient/child to cover the nose and mouth to provide a barrier for the respiratory tract from microorganisms, body fluids, and particulate material. Recommended ages are 4-12. The face mask is specifically for use with patients whose age or illness may prevent them from taking necessary precautions in situations where transfer of microorganisms, body fluids, and particulates can occur. The face masks are recommended for use in a healthcare setting with appropriate adult supervision. The Dukal Level 1 Pediatric Face Masks are single use, disposable devices that are provided non-sterile.

| Type of Use (Select one or both, as applicable)                       |                                                                                 |
|-----------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY K210321 510(k) Premarket Notification for Dukal Level 1 Pediatric Face Mask

| 1. Submitter:                  | Dukal, LLC<br>2 Fleetwood Court<br>Ronkonkoma NY 11779<br>Phone: 631-656-3800<br>Fax: 631-656-3810                                                                                                                                                                                                    |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. FDA Registration Number:    | 2435946                                                                                                                                                                                                                                                                                               |
| 3. Regulatory Affairs Contact: | Megan Quevedo<br>Quality and Regulatory Affairs Supervisor<br>2 Fleetwood Court<br>Ronkonkoma NY 11779<br>Telephone Number: 631-656-3800 ext. 133<br>Fax Number: 631-656-3810                                                                                                                         |
| 4. Date Summary Prepared:      | March 29, 2022                                                                                                                                                                                                                                                                                        |
| 5. Name of Device:             | Dukal Level 1 Pediatric Face Mask                                                                                                                                                                                                                                                                     |
| 6. Trade Name:                 | Dukal Level 1 Pediatric Face Mask                                                                                                                                                                                                                                                                     |
| 7. Common/Classification Name: | Pediatric/Child Facemask                                                                                                                                                                                                                                                                              |
| 8. Regulation Number:          | 21 CFR §878.4040                                                                                                                                                                                                                                                                                      |
| 9. Device Class:               | Class II                                                                                                                                                                                                                                                                                              |
| 10. Regulation Name:           | Surgical Apparel                                                                                                                                                                                                                                                                                      |
| 11. Product Code:              | OXZ                                                                                                                                                                                                                                                                                                   |
| 12. Predicate Device:          | Prestige Ameritech Pediatric/Child Face Mask (level 1)<br>• 510k #K160100, cleared on 10/7/2016                                                                                                                                                                                                       |
| 13. Device Description:        | Pediatric face masks are intended to be worn by the<br>patient/child to cover the nose and mouth to provide a<br>barrier for the respiratory tract from microorganisms,<br>body fluids, and particulate material. The mask is<br>available in the following product sizes:<br>Regular Size (14.5x9cm) |

Extra Small Size (12.5x8cm)

{4}------------------------------------------------

|                          | The Dukal Level 1 Pediatric Face Mask (both sizes) are composed of three layers of materials and pleated to form the mask. The inner layer is composed of Polypropylene (Spunbond), the middle layer is Polypropylene (Meltblown) filter material, and the outer layer is polypropylene (Spunbond).                                                                                                                                                                                                                                                                                                                                                                                    |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                          | The masks (both sizes) are held in place on wearer with knitted polyester/spandex elastic ear loop and contain a malleable aluminum nosepiece strip.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                          | The masks (both sizes) have decorative patterns, printed with colored inks.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| 14. Packaging:           | 50 masks/box                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| 15. Indications for Use: | The Dukal Level 1 Pediatric Face Masks are intended to be worn by the patient/child to cover the nose and mouth to provide a barrier for the respiratory tract from microorganisms, body fluids, and particulate material. Recommended ages are 4-12. The face mask is specifically for use with patients whose age or illness may prevent them from taking necessary precautions in situations where transfer of microorganisms, body fluids, and particulates can occur. The face masks are recommended for use in a healthcare setting with appropriate adult supervision. The Dukal Level 1 Pediatric Face Masks are single use, disposable devices that are provided non-sterile. |

# 16. Comparison of Technological Characteristics with the Predicate Device:

| Element of<br>Comparison                       | Predicate Device Ameritech<br>Pediatric/ Child Face's Mask<br>(K160100)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Subject Device Dukal Level 1<br>Pediatric Face Masks                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Comparison |
|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indications for<br>Use                         | The Prestige Ameritech<br>Pediatric/Child Face Mask is<br>intended to be worn by the<br>patient/child. The<br>Pediatric/Child Facemask is a<br>single use, disposable device,<br>provided non-sterile. The<br>Pediatric/Child Facemask is                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | The Dukal Level 1 Pediatric Face<br>Masks are intended to be worn by<br>the patient/child to cover the nose<br>and mouth to provide a barrier for<br>the respiratory tract from<br>microorganisms, body fluids, and<br>particulate material. Recommended<br>ages are 4-12. The face mask is                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Similar    |
|                                                | intended to be worn by the patient (ages 4-12) to cover the nose and mouth to provide a barrier for the respiratory tract for microorganisms and particulate materials. The mask is specifically for use with patients whose age or illness may prevent them from taking necessary precautions in situations where transfer of microorganisms, body fluids, and particulates can occur.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | specifically for use with patients whose age or illness may prevent them from taking necessary precautions in situations where transfer of microorganisms, body fluids, and particulates can occur. The face masks are recommended for use in a healthcare setting with appropriate adult supervision. The Dukal Level 1 Pediatric Face Masks are single use, disposable devices that are provided non-sterile.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |            |
| Intended Use<br>Sites                          | Healthcare setting- This Face<br>Mask is recommended for use in a healthcare setting with appropriate adult supervision.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Healthcare setting- The Dukal Level 1 Pediatric Face Mask is recommended for use in a healthcare setting with appropriate adult supervision.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same       |
| Sterilization and<br>Shelf Life                | The Prestige Ameritech<br>Pediatric/Child Face Mask<br>is not provided sterile, has no<br>proposed shelf life/expiration<br>date, and is not a reprocessed<br>single use device. The<br>device is manufactured from<br>non-woven materials that are<br>not impacted by storage<br>conditions or aging and thus<br>does not have a<br>shelf life. Storage conditions<br>will not affect device safety of<br>effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | The Dukal Level 1 Pediatric Face<br>Mask is not provided sterile, has no<br>proposed shelf life/expiration date,<br>and is not a reprocessed single use<br>device.<br>The device is manufactured from<br>non-woven materials that are not<br>impacted by storage conditions or<br>aging and thus the device does not<br>have a shelf life Storage conditions<br>                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same       |
| Anthropometry                                  | Based on an anthropometric<br>study sample that is<br>representative of<br>the current US population in<br>both gender and racial<br>distribution,<br>the Prestige Ameritech<br>Child's/Pediatric face mask will<br>provide adequate coverage to<br>children between the ages of 4-<br>12 years old, of weight<br>between 24-153 pounds, and<br>of height between 3'3" and<br>5'4".                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Based on an anthropometric study<br>sample that is representative of the<br>US population in both gender and<br>racial distribution, the Dukal Level 1<br>Pediatric Face Mask will provide<br>adequate coverage to children<br>between the ages of 4-12 years old of<br>weight between 25-176 pounds, and<br>of height between 3'00" – 5'08".                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Similar    |
| Material<br>Composition                        | The Prestige Ameritech<br>Pediatric/Childs Face Mask                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | The Dukal Level 1 Pediatric Face Mask                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Similar    |
|                                                | The Prestige Ameritech<br>Pediatric/Childs Face Mask is<br>manufactured using ultrasonic<br>bonding, composed of three<br>layers of materials and pleated<br>to form the mask. The inner<br>layer is composed of<br>nonwoven, the middle layer is<br>meltblown polypropylene filter<br>material, and the outer<br>layer is cellulose.<br>Decorative patterns are<br>printed with colored inks.<br>Masks are held in place on<br>wearer with knitted<br>polyester/spandex elastic<br>earloop and contain a<br>malleable aluminum nosepiece<br>strip. The<br>Pediatric/Child's Face<br>Mask is appropriately sized to<br>the smaller faces of children<br>across a diverse population.<br>All of the materials used in this<br>device are typical materials<br>commonly used in the<br>construction of Surgical Masks<br>and are being used in current<br>legally marketed devices. This<br>product is not made with<br>natural rubber<br>latex. | The Dukal Level 1 Pediatric Face<br>Mask is manufactured using<br>ultrasonic hot sealing, composed of<br>three layers of materials and pleated<br>to form the mask. The inner layer is<br>composed of Polypropylene<br>(Spunbond), the middle layer is<br>Polypropylene (Meltblown) filter<br>material, and the outer<br>layer is polypropylene (Spunbond).<br>Decorative patterns are<br>printed with colored inks. Masks are<br>held in place on wearer with<br>polyester/spandex elastic earloop<br>and contain a malleable aluminum<br>wire nose piece. The Pediatric/Child's<br>Face<br>Mask is appropriately sized to the<br>smaller faces of children across a<br>diverse population.<br>All of the materials used in this<br>device are typical materials<br>commonly used in the construction<br>of Surgical Masks and are being used<br>in current legally marketed devices.<br>This product is not made with natural<br>rubber latex. | Similar    |
| Regulation,<br>Classification,<br>Product Code | Regulation Number: 21 CFR<br>§878.4040<br>Device Class: Class II<br>Regulation Name: Surgical<br>Apparel<br>Product Code: OXZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Regulation Number: 21 CFR<br>§878.4040<br>Device Class: Class II<br>Regulation Name: Surgical Apparel<br>Product Code: OXZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same       |
| Product Color                                  | Decorative patterns are<br>printed with colored inks                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Decorative patterns are<br>printed with colored inks                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Similar    |
| Mask Style                                     | Pleated                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Pleated                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same       |
| Design Features                                | Ear Loop                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Ear Loop                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same       |
| Sterility                                      | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same       |
| Use                                            | Single Use; Disposable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Single Use; Disposable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same       |
| Biocompatibility                               | Under the conditions of each<br>study, the Prestige Ameritech                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Under the conditions of each study,<br>the Dukal Level 1 Pediatric Face Mask                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same       |

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

| Pediatric/Child Face Mask is<br>non-cytotoxic (ISO 10993-5), is<br>non-irritating (ISO 10993-10),<br>and is non-sensitizing (ISO<br>10993-10) and has met the<br>requirements per ISO- 10993-1. | is non-cytotoxic (ISO 10993-5), is<br>non-irritating (ISO 10993-10), is non-<br>sensitizing (ISO 10993-10), and has<br>met the requirements per ISO-<br>10993-1. |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|

## Comparison Summary of Non-Clinical Testing Results

| Test Standards                                                                      | Predicate Device Ameritech<br>Pediatric/ Child Face's Mask<br>(K160100)                                                                                                                                                                                                                                                                             | Subject Device Dukal Level 1<br>Pediatric Face Masks                                                                                                                                                                                                                                                                                                | Remark  |
|-------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| ASTM F2101- Bacterial<br>Filtration Efficiency (BFE)                                | 96.32%                                                                                                                                                                                                                                                                                                                                              | >99%                                                                                                                                                                                                                                                                                                                                                | Similar |
| ASTM F2299- Particulate<br>Filtration Efficiency                                    | 98%                                                                                                                                                                                                                                                                                                                                                 | >99%                                                                                                                                                                                                                                                                                                                                                | Similar |
| EN14683- Differential<br>Pressure                                                   | 1.74 mmH2O/cm² (using<br>Differential Pressure, Mil<br>M36954C)                                                                                                                                                                                                                                                                                     | ≤3.1mmH2O/cm²                                                                                                                                                                                                                                                                                                                                       | Similar |
| ASTM F1862- Fluid<br>Resistance                                                     | n/a                                                                                                                                                                                                                                                                                                                                                 | Passed at 80 mmHg                                                                                                                                                                                                                                                                                                                                   | Similar |
| 16 CFR 1610-Flammability                                                            | Class I                                                                                                                                                                                                                                                                                                                                             | Class I                                                                                                                                                                                                                                                                                                                                             | Similar |
| Biocompatibility-10993-5-<br>Cytotoxicity & 10993-10-<br>Irritation & Sensitization | Tested under ISO 10993 Standard:<br>Under the conditions of the study, the device did not show cytotoxicity potential. Under the conditions of the study, the irritation response category of the device was categorized as negligible. Under the conditions of the study, the device showed no significant evidence of causing skin sensitization. | Tested under ISO 10993 Standard:<br>Under the conditions of the study, the device did not show cytotoxicity potential. Under the conditions of the study, the irritation response category of the device was categorized as negligible. Under the conditions of the study, the device showed no significant evidence of causing skin sensitization. | Similar |
| Total Lead Content<br>Analysis                                                      | Tested under CPSC-CH-E1002-08                                                                                                                                                                                                                                                                                                                       | Tested under CPSC-CH-E1002-08.3 & CPSC-CH-                                                                                                                                                                                                                                                                                                          | Similar |
|                                                                                     |                                                                                                                                                                                                                                                                                                                                                     | E1001-08.3                                                                                                                                                                                                                                                                                                                                          |         |
| Phthalate Analysis                                                                  | Tested under CPSC-CH-C1001-<br>09.3                                                                                                                                                                                                                                                                                                                 | Tested under CPSC-CH-<br>C1001-09.4                                                                                                                                                                                                                                                                                                                 | Similar |
| Safety of Toys – Migration<br>of certain elements                                   | Tested under EN 71-3                                                                                                                                                                                                                                                                                                                                | Tested under EN 71-3                                                                                                                                                                                                                                                                                                                                | Similar |
| Mechanical Hazards (Small<br>parts, sharp edges, and<br>sharp points)               | Tested under ASTM F963-17 (sec.<br>4.6, 4.7, and 4.9), 16 CFR Part<br>1500, 16 CFR Part 1501                                                                                                                                                                                                                                                        | Tested under ASTM F963-17<br>(sec. 4.6, 4.7, and 4.9), 16 CFR<br>Part 1500, 16 CFR Part 1501                                                                                                                                                                                                                                                        | Similar |

{8}------------------------------------------------

### Non-Clinical Test Results:

The subject pediatric face masks were tested and found conformance with following standards:

- ASTM F1862/1862M-17, Standard Test Method for Resistance of Medical Face Masks to ● Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)
- ASTM F2101-19, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus
- ISO 22609:2004, Clothing for protection against infectious agents Medical face masks -● Test method for resistance against penetration by synthetic blood (fixed volume, horizontally projected)
- EN 14683: 2019, Medical Face Masks-Requirements and Test Methods
- AS4381:2015, Single-Use Face Masks for Use in Health Care ●
- 16 CFR 1610, Standard for the Flammability of Clothing Textiles ●
- ASTM F2299/F2299M-03, Standard Test Method for Determining the Initial Efficiency of Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres
- ASTM F2100-19, Standard Specification for Performance of Materials Used in Medical Face Masks
- ISO 10993-5: 2009, Biological evaluation of medical devices Part 5: Tests for in vitro ● cvtotoxicitv
- ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
- CPSC-CH-E1002-08.3, Standard Operating Procedure for Determining Total Lead (Pb) in Nonmetal Children's Products
- CPSC-CH-C1001-09.4, Standard Operating Procedure for the Determination of Phthalates
- CPSC-CH-E1001-08.3, Standard Operating Procedure for Determining Total Lead (Pb) in ● Children's Metal Products (Including Children's Metal Jewelry)
- EN 71-3:2019, Safety of Toys-Migration of Certain Elements ●
- ASTM F963-17, Standard Consumer Safety Specification For Toy Safety
- 16 CFR Part 1500, Hazardous Substances and Articles: Administration and Enforcement Regulations
- . 16 CFR Part 1501, Method For Identifying Toys And Other Articles Intended For Use By Children Under 3 Years Of Age Which Present Choking, Aspiration, Or Ingestion Hazards Because Of Small Parts

{9}------------------------------------------------

# Summary for Non-Clinical Testing

| Test Item                                                           | Test Standard<br>Methods                                                      | Test Requirements                                                                                                                                                                                                                                                                                                                                             | Subject Device Dukal Level 1<br>Pediatric Face Masks (meets<br>ASTM F2100-19 Level 1<br>requirements)                                                                                                                                                                                                                                           | Remark               |
|---------------------------------------------------------------------|-------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Bacterial<br>Filtration<br>Efficiency<br>Performance<br>(BFE) (%)   | ASTM F2101                                                                    | ≥95%                                                                                                                                                                                                                                                                                                                                                          | >99%                                                                                                                                                                                                                                                                                                                                            | Meets<br>requirement |
| Particulate<br>Filtration<br>Efficiency<br>Performance<br>(PFE) (%) | ASTM F2299                                                                    | ≥95%                                                                                                                                                                                                                                                                                                                                                          | >99%                                                                                                                                                                                                                                                                                                                                            | Meets<br>requirement |
| Differential<br>Pressure (Delta-<br>P) (mm<br>H2O/cm²)              | EN 14683                                                                      | <5.0                                                                                                                                                                                                                                                                                                                                                          | ≤3.1mmH2O/cm²                                                                                                                                                                                                                                                                                                                                   | Meets<br>requirement |
| Fluid<br>Resistance<br>Performance<br>(mmHg)                        | ASTM F1862                                                                    | 80 mmHg                                                                                                                                                                                                                                                                                                                                                       | Passed at 80 mmHg                                                                                                                                                                                                                                                                                                                               | Meets<br>requirement |
| Flammability                                                        | 16 CFR part 1610                                                              | Class I<br>No Flame Spread                                                                                                                                                                                                                                                                                                                                    | Class I<br>No Flame Spread                                                                                                                                                                                                                                                                                                                      | Meets<br>requirement |
| Biocompatibility                                                    | ISO 10993-5<br>ISO 10993-10                                                   | -Under the conditions<br>of the study, the<br>device does not show<br>cytotoxicity potential.<br>-Under the conditions<br>of the study, the<br>irritation response<br>category of the device<br>is classified as<br>Negligible.<br>-Under the<br>conditions of the<br>study, the device<br>shows no significant<br>evidence of causing<br>skin sensitization. | -Under the conditions of the<br>study, the device did not show<br>cytotoxicity potential.<br>-Under the conditions of the<br>study, the irritation response<br>category of the device was<br>classified as Negligible.<br>-Under the conditions of the<br>study, the device showed no<br>significant evidence of causing<br>skin sensitization. | Meets<br>requirement |
| Mechanical                                                          | Tested under ASTM                                                             | Pass under                                                                                                                                                                                                                                                                                                                                                    | Pass                                                                                                                                                                                                                                                                                                                                            | Meets                |
| Hazards (Small<br>parts, sharp<br>edges, and<br>sharp points)       | F963-17 (sec. 4.6,<br>4.7, and 4.9), 16 CFR<br>Part 1500, 16 CFR<br>Part 1501 | requirements for<br>CPSIA Section 106<br>ASTM F963-17<br>Mechanical Hazards:<br>under As received,<br>Normal Use, Impact,<br>Torque, and Tension<br>conditions                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                 | requirement          |
| Total Lead<br>Content<br>Analysis                                   | Tested under CPSC-<br>CH-E1002-08.3 &<br>CPSC-CH-E1001-08.3                   | Each accessible<br>component shall<br>not contain more<br>than 100 ppm of<br>lead in children's<br>products.                                                                                                                                                                                                                                                  | Pass                                                                                                                                                                                                                                                                                                                                            | Meets<br>requirement |
| Phthalate<br>Analysis                                               | Tested under CPSC-<br>CH-C1001-09.4                                           | Children's toy or<br>child care article<br>does not contain<br>phthalate<br>concentrations of<br>more than 0.1<br>percent (1000<br>ppm) in<br>plasticized<br>components.                                                                                                                                                                                      | Pass                                                                                                                                                                                                                                                                                                                                            | Meets<br>requirement |
| Safety of Toys –<br>Migration of<br>certain<br>elements             | Tested under EN 71-<br>3                                                      | The migration of<br>elements from<br>toy materials<br>shall not exceed<br>the limits for<br>restricted<br>elements.                                                                                                                                                                                                                                           | Pass                                                                                                                                                                                                                                                                                                                                            | Meets<br>requirement |

{10}------------------------------------------------

Summary for Clinical Testing: Not Applicable. See Anthropometry information in section 16 of this 510k summary. Relevant anthropometric data was searched, appraised, analyzed and summarized to demonstrate that the Dukal Level 1 Pediatric Face Mask will provide adequate coverage to children between the ages of 4-12 years old of weight between 25-176 pounds, and of height between 3'00" – 5′08″.

Conclusions: The conclusion drawn from the nonclinical tests demonstrates that the subject device in this 510(k) submission, Dukal Level 1 Pediatric Face Masks, are as safe, as effective, and perform as well as or better than the legally marketed predicate device cleared under K160100.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/OXZ/K210321](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/OXZ/K210321)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
