← Product Code [FYC](/submissions/SU/subpart-e%E2%80%94surgical-devices/FYC) · K212717

# Paneffort AAMI Level 3 Isolation Gown (K212717)

_Paneffort, LLC · FYC · Nov 30, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FYC/K212717

## Device Facts

- **Applicant:** Paneffort, LLC
- **Product Code:** [FYC](/submissions/SU/subpart-e%E2%80%94surgical-devices/FYC.md)
- **Decision Date:** Nov 30, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The Paneffort AAMI Level 3 Isolation Gowns are intended to be worn by healthcare personnel in isolation applications to provide moderate barrier protection for health care personnel and patients from the transfer of microorganisms, body fluids, and particulate material. Paneffort AAMI Level 3 Isolation Gowns meet the requirements of level 3 Liquid Barrier Performance as per AAMI PB70:2012 and are provided non-sterile and are single use only.

## Device Story

Single-use, non-sterile isolation gown; constructed of nonwoven Spunbond-Meltblown-Spunbond (SMS) polypropylene; features tie-neck, elastic cuffs, belt tie, and taped seams. Provides AAMI Level 3 liquid barrier protection. Used in healthcare facilities by personnel to prevent transfer of microorganisms, body fluids, and particulates. Protects both wearer and patient. No electronic or software components.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM and AAMI standards including hydrostatic pressure (AATCC 127), impact penetration (AATCC 42), bursting strength (ASTM D3786), tensile/breaking strength (ASTM D5034), tearing strength (ASTM D5587), seam strength (ASTM D1683), linting (ISO 9073-10), air permeability (ASTM D737), flammability (16 CFR 1610), and thermal/evaporative resistance (ASTM F1868). Biocompatibility confirmed per ISO 10993-5 and ISO 10993-10.

## Technological Characteristics

Material: SMS non-woven (polypropylene). Barrier: AAMI Level 3 per ANSI/AAMI PB70:2012. Form factor: Disposable gown with tie-neck, elastic cuffs, belt tie, and taped seams. Sterilization: Non-sterile. Standards: ASTM F2407, ASTM D3776, ASTM D5034, ASTM D5587, AATCC 127, AATCC 42, ISO 9073-10, ASTM D737, ASTM D1683, ASTM F1868, 16 CFR 1610, ISO 13938-2.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- PRIMAGARD Isolation Gown (AAMI PB70 Level 3) ([K160361](/device/K160361.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 30, 2022

Paneffort, LLC % Shilpa Gampa Delivery Manager and US Agent Freyr Solutions 150 College Rd W #102 Princeton, New Jersey 08540

Re: K212717

Trade/Device Name: Paneffort AAMI Level 3 Isolation Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FYC Dated: October 19, 2022 Received: October 20, 2022

Dear Shilpa Gampa:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Bifeng Qian -S

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212717

Device Name Paneffort AAMI Level 3 Isolation Gown

#### Indications for Use (Describe)

The Paneffort AAMI Level 3 Isolation Gowns are intended to be worn by healthcare personnel in isolations to provide moderate barrier protection for health care personnel and patients from the transfer of microorganisms, body fluids, and particulate material. Paneffort AAM Level 3 Isolation Gowns meet the requirements of level 3 Liquid Barrier Performance as per AAMI PB70:2012 and are provided non-sterile and are single use only.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

> Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary – K212717 Paneffort AAMI Level 3 Isolation Gown

# 1. Submitter Information:

| Application Correspondent:<br>Phone:<br>E-mail: | Shilpa Gampa<br>Delivery Manager and US Agent<br>Freyr Inc.<br>150 College Rd W #102, Princeton, NJ 08540<br>+1 908-483-7958 Ext 1780<br>+1 760-583-4401<br>usagent@freyrsolutions.com |  |  |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Manufacturer:                                   | Paneffort (Cambodia) Garment Co. Ltd.,<br>National Road No.2, Kleang Sambatt,<br>Pot Sar Ward, Bati District,<br>Takeo 21309, Cambodia                                                 |  |  |
| Specification Developer<br>& Product Owner      | Paneffort, LLC<br>700 Elmridge Center Drive<br>Rochester, NY 14626                                                                                                                     |  |  |
| Phone:                                          | +855 12284810                                                                                                                                                                          |  |  |
| Contact Person:                                 | Harry Xu                                                                                                                                                                               |  |  |
| E-mail:                                         | Phil@paneffort.com                                                                                                                                                                     |  |  |
| Date Prepared:                                  | 28-Nov-2022                                                                                                                                                                            |  |  |
| 2. Device Identification:                       |                                                                                                                                                                                        |  |  |
| Device Trade Name:                              | Paneffort AAMI Level 3 Isolation Gown                                                                                                                                                  |  |  |
| Device Common Name:                             | Gowns, Isolation, Surgical                                                                                                                                                             |  |  |
| Classification Name:                            | Surgical Isolation Gown/Surgical Apparels                                                                                                                                              |  |  |
| Device Class:                                   | Class II                                                                                                                                                                               |  |  |
| Regulation Number:                              | 21 CFR 878.4040                                                                                                                                                                        |  |  |
| Product Code:                                   | EVC                                                                                                                                                                                    |  |  |

{4}------------------------------------------------

#### Predicate Device:

| Device Name                                     | 510(k) Number |
|-------------------------------------------------|---------------|
| PRIMAGARD Isolation Gown<br>(AAMI PB70 Level 3) | K160361       |

#### Table 1 List of Predicate Devices

#### 3. Device Description

The Paneffort AAMI Level 3 Isolation Gown is a surgical isolation gown with moderate barrier protection identified by Regulation 21 CFR 878.4040 under FDA product code FYC. The Paneffort AAMI Level 3 Isolation Gown is a single use, disposable medical device provided non-sterile. Paneffort AAMI Level 3 Isolation Gown is offered in one color (blue) and six different sizes of Small, Medium, Regular/Large, X-Large, 2X-Large, and 3X-Large. Each model is constructed of a nonwoven Spunbond-Meltblown-Spunbond (SMS) material that is tested according to ANSI/AAMI PB70:2012 Liquid Barrier and Performance Classification of Protective Apparel and Surgical Drapes Intended for Use in Health Care Facilities and meets AAMI Level 3.

The Paneffort AAMI Level 3 Isolation Gown consists of one critical zone throughout the entire gown including seams, but excluding cuffs, hems, and bindings as per AAMI PB70:2012. It is a protective tape sealed gown along with other design features like tieneck, elastic cuffs, and belt tie, that meets AAMI PB70 Level 3 barrier performance for protective apparel.

#### 4. Intended Use/Indications for Use

The Paneffort AAMI Level 3 Isolation Gowns are intended to be worn by healthcare personnel in isolation applications to provide moderate barrier protection for health care personnel and patients from the transfer of microorganisms, body fluids, and particulate material. Paneffort AAMI Level 3 Isolation Gowns meet the requirements of level 3 Liquid Barrier Performance as per AAMI PB70:2012 and are provided non-sterile and are single use only.

#### 5. Device and Predicate Device Technical Characteristics

The fundamental scientific technology, materials of construction and mechanism of operation are identical between the subject Paneffort AAMI Level 3 Isolation Gown and the predicate device. Table 2 summarizes the comparison of technological characteristics between the subject and predicate devices.

{5}------------------------------------------------

| S.No. | Parameters                                          | PRIMAGARD Isolation Gown<br>(AAMI PB70 Level 3)<br>(Predicate Device)                                                                                                                                                                                                                                                                                                                                                                                                                              | Paneffort AAMI Level 3 Isolation<br>Gown<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                                                                      | Remarks   |
|-------|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| 1.    | Manufacturer                                        | Primed Medical Products Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Paneffort (Cambodia) Garment Co. Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                             | N/A       |
| 2.    | Product Class<br>and Code                           | Class II, FYC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Class II, FYC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same      |
| 3.    | 510(k) Number                                       | K160361                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | K212717                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | N/A       |
| 4.    | Regulation Number                                   | 21CFR 878.4040                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 21CFR 878.4040                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same      |
| 5.    | Intended use/<br>Indications for use                | PRIMAGARD Level 3 Isolation<br>Gowns are intended to be worn by<br>healthcare personnel in isolation<br>applications to provide moderate<br>barrier protection for health care<br>personnel and patients from the<br>transfer of microorganisms, body<br>fluids, and particulate material.<br>The PRIMAGARD Isolation<br>Gowns of this 510(k) meet the<br>requirements of level 3 Liquid<br>Barrier Performance as per AAMI<br>PB70:2012 and are provided non-<br>sterile and are single use only. | The Paneffort AAMI Level 3 Isolation<br>Gowns are intended to be worn by<br>healthcare personnel in isolation<br>applications to provide moderate barrier<br>protection for health care personnel and<br>patients from the transfer of<br>microorganisms, body fluids, and<br>particulate material. Paneffort AAMI<br>Level 3 Isolation Gowns meet the<br>requirements of level 3 Liquid Barrier<br>Performance as per AAMI PB70:2012 and<br>are provided non-sterile and are single use<br>only. | Same      |
| 6.    | Level of barrier<br>protection AAMI PB70            | Level 3                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Level 3                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same      |
| 7.    | Material composition                                | Non-woven Polypropylene<br>(SMS)                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | SMS non-woven (polypropylene)                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same      |
| 8.    | Design                                              | 5 Models:<br>Apron Neck Closure,<br>Tape Tab Neck Closure,<br>Hook and Loop Closure                                                                                                                                                                                                                                                                                                                                                                                                                | Tie neck,<br>Full back,<br>Elastic cuffs,<br>Blue belt tie                                                                                                                                                                                                                                                                                                                                                                                                                                        | Similar1  |
| S.No. | Parameters                                          | PRIMAGARD Isolation Gown<br>(AAMI PB70 Level 3)<br>(Predicate Device)                                                                                                                                                                                                                                                                                                                                                                                                                              | Paneffort AAMI Level 3 Isolation<br>Gown<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                                                                      | Remarks   |
|       |                                                     | Apron Neck Closure,<br>Tie Neck Closure                                                                                                                                                                                                                                                                                                                                                                                                                                                            | taped seams.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |           |
| 9.    | Color                                               | Blue, Yellow                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Blue                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Similar²  |
| 10.   | Sterility                                           | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same      |
| 11.   | Use                                                 | Single Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Single Use; Disposable                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same      |
| 12.   | Basis weight ASTM<br>D3776                          | $39.97 ± 1.61 g/m²$<br>$(1.17 oz/yd² ± 0.05)$                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Tested to ASTM D3766. Met the<br>acceptance criteria.                                                                                                                                                                                                                                                                                                                                                                                                                                             | Similar³  |
| 13.   | LBP: Hydrostatic<br>Pressure AATCC 127              | ≥ 50 cm H₂ O                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Pass, Met the standard's requirement<br>(≥ 50 cm H₂ O)                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same      |
| 14.   | LBP: Impact Penetration<br>AATCC 42                 | ≤ 1 g                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Pass, Met the standard's requirement<br>(≤ 1.0 g)                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same      |
| 15.   | Flammability: 16 CFR<br>Part 1610.7                 | Class I                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Class I                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same      |
| 16.   | Breaking strength (MD)<br>ASTM D5034                | Tested to ASTM D5034.<br>Met acceptance criteria                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Tested to ASTM D5034. Met<br>acceptance criteria                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Similar⁴  |
| 17.   | Breaking strength (CD)<br>ASTM D5034                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Similar⁵  |
| 18.   | Tearing strength (MD)<br>ASTM D5587-08              | Tested to ASTM D5587.<br>Met acceptance criteria                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Tested to ASTM D5587. Met acceptance<br>criteria                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Similar⁶  |
| 19.   | Tearing strength (CD)<br>ASTM D5587                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Similar⁷  |
| 20.   | Linting (ISO 9073-10)                               | Tested to NWSP 160.1 R0                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Met acceptance criteria.                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Similar⁸  |
| S.No. | Parameters                                          | PRIMAGARD Isolation Gown<br>(AAMI PB70 Level 3)<br>(Predicate Device)                                                                                                                                                                                                                                                                                                                                                                                                                              | Paneffort AAMI Level 3 Isolation<br>Gown<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                                                                      | Remarks   |
| 21.   | Seam Strength                                       | Tested to ASTM D5035.<br>Met acceptance criteria.                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Tested to ASTM D5035. Met<br>acceptance criteria.                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Similar9  |
| 22.   | Thermal and<br>Evaporative Resistance<br>ASTM F1868 | Tested to ASTM F1868<br>PART C.<br>Met acceptance criteria                                                                                                                                                                                                                                                                                                                                                                                                                                         | Tested to ASTM F1868 PART<br>C.<br>Met acceptance criteria                                                                                                                                                                                                                                                                                                                                                                                                                                        | Similar10 |
| 23.   | Air Permeability<br>ASTM D737                       | Tested to ASTM D737.<br>Met acceptance criteria.                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Tested to ASTM D737. Met acceptance<br>criteria.                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Similar11 |
| 24.   | Irritation                                          | Under the conditions of the study,<br>not an irritant.                                                                                                                                                                                                                                                                                                                                                                                                                                             | Under the conditions of the study, not an<br>irritant.                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same      |
| 25.   | Sensitization                                       | Under the conditions of the study,<br>not a sensitizer.                                                                                                                                                                                                                                                                                                                                                                                                                                            | Under the conditions of the study, not a<br>sensitizer.                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same      |
| 26.   | Cytotoxicity                                        | Under the conditions of the study<br>the device is non-cytotoxic.                                                                                                                                                                                                                                                                                                                                                                                                                                  | Under the conditions of the study the<br>device is non-cytotoxic.                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same      |

# Table 2 Technological Characteristics Comparison of Proposed device and Predicate device

{6}------------------------------------------------

{7}------------------------------------------------

#### Note:

N/A –refers that the section is Not applicable/ available.

Similar': The subject device design is sinilar to the additional features. The additional features do not raise related the safety and effectiveness of the device. Performance testing includion has been performed on the final finished device which includes all construction materials and color additives.

Similar': The subject device is offered only in single color while the predicate is offered in both Blue and Yellow colors. Peformance testing including biocompatibility evaluation has been the final finished device which includes all construction materials and color additives. Therefore, the difference does not raise any questions on the safety and effectiveness of the device.

Sinitar': The gowns were tested using the ASTM D3776 standard test method. As there linits for this test, we compared the test values obtained with a similar FDA approved (K203415) device. The test values for predicate and subject were less than the when evel

{8}------------------------------------------------

#### Traditional 510(k) Paneffort AAMI Level 3 Isolation Gown

device, so we considered the difference in basis as acceptable. Therefore, the difference does not raise any gressions related to the safety and effectiveness of the device.

Similar' and Similar': The subject devices were tested for Breaking Strength using the ASTM F2407 recommended ASTM DS134 test nethod. The breaking strength of subject device differed from the recommended range of the ASTM P2407 Standard Specification for Sugical Gowns Intended for Use in Healthcare Facilities. The difference does not raise any questions on the safety and effectiveness of the device.

Similar and Similar : The subject devices were tested for Breaking Strength using the ASTM F2407 recommended ASTM D5887 test nethod. The tearing strength of the subject than the predicates, but it is within the recommended range of the ASTM F2407 for Surgical Gowns Intended for Use in Healthere, the difference does not raise any questions related to the safety and efectiveres of the device. Similar: Both the subject device and Predicate device were to the Gown using different test nethods of ASTM D1683/D1683M-11 a and ASTM D5035, respectively. Here, the Subject device to test method jielded the values within the recommended range of the ASTM F240 Standard Specification for Surgical Gowns Intented for Use in Healther. While the Predicate device was tested according to ASTM D505, which specifically utilizes cut strip test proce and elong force and elongation of most textile fabrics within the recommended range of the ASTM F2407 Standard Specification. As both the performed to indicate the durability of the fabric within the acceptable ranges, the difference does not raise any questions on the safety and effectiveness of the device.

Similar": Both the subject device were tested for Lint generation using the ISO 9073-10 standard recommended test nethod. As per the annicable ISO testing standard, there is no acception to Linting, However, the standard states that devending on the choice of particle counter, the size ranges can fall within of 0.3 um or 0.5 um. As test values obtained were within this particle size limit, the difference does not raise any questions related to the safety and effectiveness of the device.

Similar": Both the subject device and Predicted for Thermal and Euporative Resistance using the ASTM F1868-17 test nethod. As per the ASTM F1868 standard, the acceptance criteria are was evaporative resistance neasured for the speciment (that includes the test specinen (fabriches in test plate covered with a ligated. There are no acceptance criteria for the Reserver, as bath the gowns vere tested as per the applicable standard mot raise any questions related to the safety and effectiveness of the device.

Similar": Both the subject device and Predicty were tested for Air Permeability using the ASTM D737 est method. There is no minimum performance expectation for the air permeability parameter and Ashation gowns. Therefore, the difference does not ruise any questions related to the safety and effectiveness of the device.

{9}------------------------------------------------

## 6. Non-Clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

- 1. ASTM F2407-20 Standard Specification for Surgical Gowns Intended for Use in Healthcare Facilities
- 2. AAMI ANSI PB70:2012 Liquid Barrier Performance Classification of Protective Apparel and Drapes Intended for Use in Health Care Facilities
- 3. ASTM D3776/D3776/D3776M-2020 Test Methods for Mass Per Unit Area (Weight) of Woven Fabric
- 4. ASTM D5034-09(2013)- Standard Test Method for Breaking Strength and Elongation of Textile Fabrics (Grab Test)
- ASTM D5587-08- Standard Test Method for Tearing Strength of Fabrics by ર. Trapezoid
- 6. AATCC 127-2017 Water Resistance: Hydrostatic Pressure Test
- 7. AATCC 42:2018 -Water Resistance: Impact Penetration Test
- ISO 9073 Part 10: 2003 (for non-woven fabric) Lint and other particles generation 8. in the dry state (Lint Generation test)
- 9. ASTM D737-18 Standard Test Method for Air Permeability of Textile Fabrics
- 10. ASTM D1683/D1683M-11a Standard Test Method for Failure in Sewn Seams of Woven Apparel Fabrics
- 11. ASTM F1868-17 Standard Test Method for Thermal and Evaporative Resistance of Clothing Materials Using a Sweating Hot Plate
- 12. 16 CFR Part 1610-2008 Standard for the Flammability of Clothing
- 13. ISO 13938-2:2019 Pneumatic method for determination of bursting strength and bursting distension

The test summary table mentioned below includes the brief information for all the ASTM F2407-20 - Standard Specification for Surgical Gowns Intended for Use in Healthcare Facilities recommended tests that demonstrates the performance of Level 3 Isolation gowns.

{10}------------------------------------------------

| S.<br>No | Test Performed                                  | Device Description/ Sample<br>Size                                                                                 | Test Methods/ Standards applicable                                                                                                                        | Acceptance<br>criteria                                                         | Unexpected<br>results/<br>Significant<br>deviations                                                                                                                                                | Conclusion                                   |
|----------|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| 1.       | Bursting<br>Strength                            | Paneffort AAMI Level 3<br>Isolation Gown,<br>32 samples each from 3<br>different lots                              | ASTM D3786 Standard Test Method for Bursting<br>Strength of Textile Fabrics                                                                               | ≥ 40 kPa (≥5.80 psi)                                                           | None                                                                                                                                                                                               | The test meets the<br>Acceptance<br>Criteria |
| 2.       | Basis Weight                                    | Paneffort AAMI Level 3<br>Isolation Gown<br>32 samples each from 3<br>different lots                               | ASTM D3776 / D3776M - 20 :<br>Standard Test Methods for Mass Per Unit Area<br>(Weight) of Fabric; Method C                                                | ≥35 g/m2                                                                       | None                                                                                                                                                                                               | The test meets the<br>Acceptance<br>Criteria |
| 3.       | Linting Test                                    | Paneffort AAMI Level 3<br>Isolation Gown<br>32 samples each from 3<br>different lots                               | ISO 9073-10:2003:<br>Test methods for nonwovens - Part 10: Lint and<br>other particles generation in the dry state                                        | Co-coefficient of<br>Linting ≤4.2                                              | None                                                                                                                                                                                               | The test meets the<br>Acceptance<br>Criteria |
| 4.       | Air<br>Permeability                             | Paneffort AAMI Level 3<br>Isolation Gown<br>32 samples each from 3<br>different lots                               | ASTM D737-18:<br>Standard Test Method for Air Permeability of<br>Textile Fabrics                                                                          | ≥15 ft3/min/ft2                                                                | None                                                                                                                                                                                               | The test meets the<br>Acceptance<br>Criteria |
| 5.       | Water<br>Resistance:<br>Hydrostatic<br>Pressure | Paneffort AAMI Level 3<br>Isolation Gown,<br>3 lots<br>Lot 1: 74 pieces;<br>Lot 2: 72 pieces;<br>Lot 3: 72 pieces. | AATCC 127-2018:<br>Test Method for Water Resistance: Hydrostatic<br>Pressure                                                                              | ≥ 50cm H2 O (Ас:<br>3, Re: 4);<br>AQL: 4% (Level 3<br>ANSI/<br>AAMI PB70:2012) | In Lot 1, three<br>samples (out<br>of 32) failed the<br>test.<br>In Lot 2, three<br>samples (out<br>of 32) failed the<br>test. In Lot 3, two<br>samples (out of 32<br>samples) failed<br>the test. | The test meets the<br>Acceptance<br>Criteria |
| 6.       | Water<br>Resistance:<br>Impact<br>Penetration   | Paneffort AAMI Level 3<br>Isolation Gown,<br>3 lots<br>Lot 1: 74 pieces;<br>Lot 2: 72 pieces:<br>Lot 3: 72 pieces  | AATCC 42-2017<br>Test Method for Water Resistance: Impact<br>Penetration                                                                                  | ≤1.0g (Ac: 3, Re:<br>4);<br>AQL: 4% (Level 3<br>ANSI/<br>AAMI PB70:2012)       | None                                                                                                                                                                                               | The test meets the<br>Acceptance<br>Criteria |
| 7.       | Tensile<br>Strength/<br>Breaking<br>Strength    | Paneffort AAMI Level 3<br>Isolation Gown,<br>32 samples each from 3<br>different lots                              | ASTM D5034-09 (2013):<br>Standard Test Method for Breaking Strength and<br>Elongation of Textile Fabrics (Grab Test)                                      | ≥30N (≥7 lbf).                                                                 | None                                                                                                                                                                                               | The test meets the<br>Acceptance<br>Criteria |
| 8.       | Tearing<br>Strength                             | Paneffort AAMI Level 3<br>Isolation Gown,<br>32 samples each from 3<br>different lots                              | ASTM D5587-08,<br>Option 1: Standard Test Method for Tearing<br>Strength of Fabrics by Trapezoid Procedure                                                | ≥10N (≥2.3 lbf).                                                               | None                                                                                                                                                                                               | The test meets the<br>Acceptance<br>Criteria |
| 9.       | Seam Strength                                   | Paneffort AAMI Level 3<br>Isolation Gown,<br>32 samples each from 3<br>different lots                              | ASTM D1683/ D1683M-11a:<br>Standard Test Method for Failure in Sewn Seams<br>of Woven Fabrics                                                             | ≥30N (≥7 lbf).                                                                 | None                                                                                                                                                                                               | The test meets the<br>Acceptance<br>Criteria |
| 10.      | Flammability                                    | Paneffort AAMI Level 3<br>Isolation Gown<br>32 samples each from 3<br>different lots                               | 16 CFR Part 1610<br>-2008: Standard for the Flammability of<br>Clothing Textiles                                                                          | Class I. Textile<br>exhibiting normal<br>flammability.                         | None                                                                                                                                                                                               | The test meets the<br>Acceptance<br>Criteria |
| 11.      | Water Vapour<br>resistance                      | Paneffort AAMI Level 3<br>Isolation Gown<br>32 samples each from 3<br>different lots                               | ASTM F1868-17,<br>Procedure Part B: Standard Test<br>Method for Thermal and Evaporative<br>Resistance of Clothing Materials Using a<br>Sweating Hot Plate | 0,0-1,0 kPa.m2 / W                                                             | None                                                                                                                                                                                               | The test meets the<br>Acceptance<br>Criteria |

{11}------------------------------------------------

{12}------------------------------------------------

Traditional 510(k) Paneffort AAMI Level 3 Isolation Gown

# 7. Biocompatibility Testing

Biocompatibility tests were performed as per the ISO 10993-1:2018 Biological Evaluation of Medical Devices- Part 1: Evaluation and testing within a risk management process. The In-Vitro cytotoxicity, Skin irritation and Skin sensitization parameters were tested as per the below standards:

- 1. ISO 10993-5:2009 Biological Evaluation of Medical Device, Part 5-Tests for In Vitro cytotoxicity
- 2. ISO 10993-10:2010 Biological Evaluation of Medical Device, Part 10-Test for irritation and skin sensitization

The Paneffort AAMI Level 3 Isolation gowns passed biocompatibility studies per ISO-10993

## 8. Clinical Testing

No clinical study is included in this submission.

#### 9. Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the device, the Paneffort AAMI Level 3 Isolation gown, is as safe, as effective, and performs as well as or better than the legally marketed predicate device (K160361).

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FYC/K212717](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FYC/K212717)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
