← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K210267

# Disposable Surgical Mask (K210267)

_Xiamen Eagledon Pharmaceutical Co., Ltd. · FXX · Jul 8, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K210267

## Device Facts

- **Applicant:** Xiamen Eagledon Pharmaceutical Co., Ltd.
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** Jul 8, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The Disposable Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluid and particulate materials. This face mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device provided non-sterile.

## Device Story

Disposable surgical mask; single-use; non-sterile. Composed of three layers: outer/inner polypropylene non-woven cloth, middle melt-blown polypropylene filter. Features nose clip and polyester spandex ear loops. Worn by healthcare personnel and patients to cover nose and mouth. Acts as physical barrier against microorganisms, body fluids, and particulates. Used in clinical settings for infection control. Protects wearer and patient from cross-contamination. Bench-tested for fluid resistance, filtration efficiency, and flammability.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM F1862 (fluid resistance), ASTM F2101 (BFE), ASTM F2299 (particulate filtration), and EN 14683 (differential pressure). Biocompatibility confirmed via ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization).

## Technological Characteristics

3-layer construction: polypropylene non-woven outer/inner layers, melt-blown polypropylene middle layer. Nose clip: PE material with chrome-plated iron core. Ear loops: polyester spandex. Dimensions: 17.5cm x 9.5cm. Non-sterile. ASTM F2100 Level 2 compliant. Fluid resistance: 120 mmHg. BFE/PFE: ≥98%. Differential pressure: <6 mmH2O/cm2. Flammability: Class 1 (16 CFR Part 1610).

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Surgical Face Mask ([K153496](/device/K153496.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 8, 2021

Xiamen Eagledon Pharmaceutical Co., Ltd % Ruth Wu Consultant Kavalan Consulting Inc. 1100 First Ave. Ste 305 King of Prussia, Pennsylvania 19406

Re: K210267

Trade/Device Name: Disposable Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: January 29, 2021 Received: February 1, 2021

Dear Ruth Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ryan Ortega, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K210267

Device Name Disposal Surgical Mask

#### Indications for Use (Describe)

The Disposable Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluid and particulate materials. This face mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device provided non-sterile.

Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

The assigned 510(k) Number: K210267

Date of Preparation: June 30, 2021

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92. "

### 1. Submitter's Information:

Xiamen Eagledon Pharmaceutical Co., Ltd No.220-228, Meihe 3rd Road, Xiamen, Fujian 361100 China Tel: +86-592-7231999 Contact Person: Aggie Zhang Email: aggie@eaglehealthItd.com

### 2. Application Correspondent:

Kavalan Consulting Inc 1100 First Ave, Ste 305 King of Prussia, PA 19406 Tel: +1 610-310-2793 Contact Person: Ruth Wu Email: ruthwu@kavalangroup.com

### 3. Subject Device:

The 510(K) number: Traditional Common Names: Surgical Mask Disposable Surgical Mask Trade Name: Regulation Number: 21 CFR 878.4040 Review Panel: Surgical Apparel Classification Name: Mask, Surgical Regulatory Class: II Product Code: FXX

{4}------------------------------------------------

### 4. Predicate Device:

The 510(K) number: K153496 Manufacturer: Xiantao Rayxin Medical Products Co Ltd Common Names: Surgical Mask Surgical Face Mask Trade Name: Regulation Number: 21 CFR 878.4040 Review Panel: Surgical Apparel Classification Name: Mask, Surgical Regulatory Class: II Product Code: FXX

### 5. Device Description:

Disposable Surgical Mask is a single-use, disposable surgical mask to cover the nose and mouth to protect the wearer from microorganisms, body fluids and particulates.

Disposable Surgical Mask consists of the following materials: Non-woven cloth in the front and back layers, melt-blown cloth in the middle layer, a nose clip and two ear loops in Spandex. The Disposable Surgical Masks are provided non-sterile:

| Model | SKU#         | Product Type                                      | Package                        |
|-------|--------------|---------------------------------------------------|--------------------------------|
| YJ002 | YJ002-NS2-50 | Disposable Surgical Mask (Non-Sterile)<br>Level 2 | 50 pcs/box,<br>2000 pcs/carton |
| YJ002 | YJ002-NS2-10 | Disposable Surgical Mask (Non-Sterile)<br>Level 2 | 10 pcs/bag,<br>1500 pcs/carton |

### 6. Indication for Use:

The Disposable Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluid and particulate materials. This face mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device provided non-sterile.

{5}------------------------------------------------

| Device                               |                                                                                                              | Disposable Surgical Mask                                                                                                                                                                                                                                                                                                                    | Disposable Surgical Face Mask                                                                                                                                                                                                                                                                                                                                                                                     | Result     |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer                         |                                                                                                              | Xiamen Eagledon<br>Pharmaceutical Co., Ltd                                                                                                                                                                                                                                                                                                  | Xiantao Rayxin Medical<br>Products Co.,ltd.                                                                                                                                                                                                                                                                                                                                                                       | -          |
| 510K number                          |                                                                                                              | K210267                                                                                                                                                                                                                                                                                                                                     | K153496                                                                                                                                                                                                                                                                                                                                                                                                           | -          |
| Model Name                           |                                                                                                              | Disposable Surgical Mask                                                                                                                                                                                                                                                                                                                    | Surgical Face Mask                                                                                                                                                                                                                                                                                                                                                                                                | Same       |
| Classification                       |                                                                                                              | Class II Device, FXX (21<br>CFR878.4040)                                                                                                                                                                                                                                                                                                    | Class II Device, FXX (21<br>CFR878.4040)                                                                                                                                                                                                                                                                                                                                                                          | Same       |
| Intended use                         |                                                                                                              | The Disposable Surgical Masks<br>are intended to be worn to<br>protect both healthcare<br>personnel and patients from<br>transmission of microorganisms,<br>body fluid and particulates.<br>These surgical masks are<br>intended for use in infection<br>control practices to reduce the<br>potential exposure to blood and<br>body fluids. | The Surgical Face Masks are<br>intended to be worn to protect<br>both the patient and healthcare<br>personnel from transfer of<br>microorganisms, body fluids<br>and particulate material. These<br>face masks are intended for<br>use in infection control<br>practices to reduce the<br>potential exposure to blood and<br>body fluids. This is a single<br>use, disposable device(s),<br>provided non-sterile. | Same       |
| Model Type                           |                                                                                                              | Ear Loops, Flat Pleated, 3 layers                                                                                                                                                                                                                                                                                                           | Ear Loops, Tie-On, Flat Pleated,<br>3 layers                                                                                                                                                                                                                                                                                                                                                                      | Same       |
| Material                             | Outer layer                                                                                                  | Polypropylene non-woven cloth                                                                                                                                                                                                                                                                                                               | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                                           | Same       |
|                                      | Middle layer                                                                                                 | Polypropylene Melt blown cloth                                                                                                                                                                                                                                                                                                              | Melt blown polypropylene filter                                                                                                                                                                                                                                                                                                                                                                                   | Same       |
|                                      | Inner layer                                                                                                  | Polypropylene non-woven cloth                                                                                                                                                                                                                                                                                                               | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                                           | Same       |
|                                      | Nose piece                                                                                                   | PE material +1 chrome plated<br>iron core 2mm width                                                                                                                                                                                                                                                                                         | Malleable polyethylene wire                                                                                                                                                                                                                                                                                                                                                                                       | Different* |
|                                      | Ear loops                                                                                                    | Polyester Spandex                                                                                                                                                                                                                                                                                                                           | Polyester                                                                                                                                                                                                                                                                                                                                                                                                         | Different* |
| Color                                |                                                                                                              | Blue                                                                                                                                                                                                                                                                                                                                        | Blue                                                                                                                                                                                                                                                                                                                                                                                                              | Same       |
| Dimension (Width)                    | 17.5cm±1cm                                                                                                   | 17.5cm±1cm                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                                                                                              |            |
| Dimension (Length)                   | 9.5cm±1cm                                                                                                    | 9.5cm±1cm                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                                                                                                                                                              |            |
| OTC use                              | Yes                                                                                                          | Yes                                                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                                                                                                                                                                                                                                                                                              |            |
| Sterility                            | Non-Sterile                                                                                                  | Non-Sterile                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                                              |            |
| Use                                  | Single use, disposable                                                                                       | Single use, disposable                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                              |            |
| ASTM F2100 Level                     | Level 2                                                                                                      | Level 2                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                                                                                                              |            |
| Fluid Resistance<br>Performance      | 32/32 Passed at 120 mmHg<br>ASTM F1862                                                                       | 13/13 Passed at 120 mmHg<br>ASTM F1862                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                              |            |
| Particulate Filtration<br>Efficiency | 32/32 Passed at ≥98%<br>ASTM F2299-03                                                                        | 13/13 Passed at  ≥98%<br>ASTM F2299                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                                                                                                                                                                                                                                                                                              |            |
| Bacterial Filtration<br>Efficiency   | 32/32 Passed at ≥98%<br>ASTM F2101-14                                                                        | 13/13 Passed at  ≥98%<br>ASTM F2101                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                                                                                                                                                                                                                                                                                              |            |
| Differential Pressure                | 32/32 Passed at <6<br>mmH2O/cm2<br>EN 14683: 2019, Annex C                                                   | 13/13 Passed at <5<br>mmH2O/cm2 MIL-M36954C                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                                              |            |
| Flammability                         | 32/32 Passed ≥3 Seconds<br>burn Time-Class 1 16 CFR Part<br>1610                                             | 13/13 Passed ≥3 Seconds<br>burn Time-Class 1 16 CFR Part<br>1610                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                                                                                                                                                              |            |
| Biocompatibility                     |                                                                                                              |                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                   |            |
| Cytotoxicity                         | Under the conditions of the<br>study, the subject device extract<br>was determined to be non-cytotoxic       | Under the conditions of the<br>study, the subject device<br>extract was determined to be<br>non-cytotoxic                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                                                                                                                                                              |            |
| Irritation                           | Under the conditions of the<br>study, the subject device extract<br>was determined to be non-irritating      | Under the conditions of the<br>study, the subject device<br>extract was determined to be<br>non-irritating                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                                                                                              |            |
| Sensitization                        |                                                                                                              |                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                   |            |
|                                      | Under the conditions of the<br>study, the subject device extract<br>was determined to be non-<br>sensitizing | Under the conditions of the<br>study, the subject device<br>extract was determined to be<br>non-sensitizing                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                                              |            |

### 7. Comparison of Technological Characteristics:

{6}------------------------------------------------

{7}------------------------------------------------

* The Disposable Surgical Mask and the Surgical Face Mask compared in the above table are substantially similar in Performance and Indications for Use. The differences raised in this table does not raise the concern for safety or effectiveness of the Disposable Surgical Masks submitted by Xiamen Eagledon Pharmaceutical Co Ltd.

### 8. Summary of Non-Clinical Performance Testing:

Non-clinical tests were conducted to verify that the proposed device met all design specifications to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

Samples submitted to the testing lab and passed the following performance tests:

- . Color: Blue
- . Dimension: 17.5cm x 9.5cm (Ear loops: 17.5 cm)
- Lot numbers: 20210405, 20210504, 20210517 .

| Test item<br>(Performance<br>Level 2)        | Test Method                                                                                                                                                                                                                                | Pass Criteria                                | Results                                                   |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|-----------------------------------------------------------|
| Synthetic Blood<br>Penetration<br>Resistance | ASTM<br>F1862/F1862M-17<br>Standard Test Method<br>for Resistance of<br>Medical Face Masks<br>to Penetration by<br>Synthetic Blood<br>(Horizontal Projection<br>of Fixed Volume at a<br>Known Velocity)<br>according to ASTM<br>F2100:2019 | Level 2: fluid<br>resistant claim<br>120mmHg | 32/32 passed at 120<br>mm/Hg/Pass                         |
| Flammability of<br>Clothing Textiles         | 16 CFR Part 1610<br>Standard for the<br>Flammability of<br>Clothing according to<br>ASTM F2100:2019                                                                                                                                        | Class 1                                      | 32/32 Passed ≥3<br>Seconds burn<br>Time-Class 1 /<br>Pass |
| Differential<br>Pressure (Delta P)           | EN 14683: 2019,<br>Annex C Medical face<br>masks -<br>Requirements and<br>test methods<br>according to ASTM<br>F2100:2019                                                                                                                  | Level 2:<6.0                                 | 32/32 Passed at <6<br>mmH2O/cm2 / Pass                    |
| Bacterial Filtration<br>Efficiency (BFE)     | ASTM F2101-14<br>Standard Test Method<br>for Evaluating the<br>Bacterial Filtration<br>Efficiency (BFE) of<br>Medical Face Mask<br>Materials, Using a<br>Biological Aerosol of<br>Staphylococcus<br>aureus according to<br>ASTM F2100:2019 | Level 2: ≥ 98%                               | 32/32 Passed at ≥<br>98% / Pass                           |
| Latex Particle<br>Challenge                  | 16 CFR Part 1610<br>Standard for the<br>Flammability of<br>Clothing according to<br>ASTM F2100:2019                                                                                                                                        | Level 2: ≥ 98%                               | 32/32 Passed at ≥<br>98% / Pass                           |

{8}------------------------------------------------

### Biocompatibility testing

According to ISO 10993-1:2009, the nature of body contact for the subject device is Surface Device category, Skin Contact and duration of contact is A-Limited (≤24h). The following tests for the subject device were conducted to demonstrate that the subject device is biocompatible and safe for its intended use:

1) In vitro Cytotoxicity Test per ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicity,

{9}------------------------------------------------

2) Skin Sensitization Tests per ISO 10993-10:2010 Biological evaluation of medical devices— Part 10: Tests for irritation and skin sensitization,

3) Skin Irritation Tests per ISO 10993-10:2010 Biological evaluation of medical devices— Part 10: Tests for irritation and skin sensitization

### 9. Summary of Clinical Performance:

No clinical study is included in this submission.

### 10. Conclusions:

The Conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission K210267, Disposable Surgical Mask, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K153496.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K210267](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K210267)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
