FECAL OCCULT BLOOD CARD TEST, MODEL F735-A, FECAL OCCULT BLOOD CARD KIT, MODEL F735-25
Applicant
Teco Diagnostics
Product Code
KHE · Hematology
Decision Date
Jul 14, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.6550
Device Class
Class 2
Indications for Use
Teco Rapid Fecal Occult Blood (FOB) Card Test is an immunochromatographic assay intended for the determination of human hemoglobin in feces by professional laboratories or physician’s offices. It is useful to determine gastrointestinal bleeding founding a number of gastrointestinal disorders such as colorectal carcinoma, colon polyps, diverticulitis and ulcerative colitis.
Device Story
Teco Rapid Fecal Occult Blood (FOB) Card Test is an immunochromatographic assay for qualitative detection of human hemoglobin in fecal samples. Device used in professional laboratories or physician offices by healthcare personnel. Sample applied to test card; hemoglobin presence indicated by visual color change on test strip. Results assist clinicians in identifying gastrointestinal bleeding associated with conditions like colorectal cancer, polyps, diverticulitis, or ulcerative colitis. Provides rapid diagnostic information to guide further clinical investigation or patient monitoring.
Clinical Evidence
No clinical studies were performed. Evidence is based on analytical performance testing. Precision/reproducibility studies (n=120) showed 98.3% to 100% agreement across physician office laboratories. Method comparison against the predicate device using 120 spiked specimens demonstrated 98.3% total agreement, 96.3% positive percent agreement, and 100% negative percent agreement. Analytical specificity testing confirmed no interference from various animal hemoglobins or horseradish peroxidase at specified concentrations. Limit of detection is 50 ng/ml in buffer or 5 µg/g in feces.
Technological Characteristics
Lateral flow immunochromatographic sandwich assay. Employs monoclonal and polyclonal antibodies for human hemoglobin detection. Sample matrix: feces in PBS extraction buffer. Sensitivity: 50 ng/ml. Visual readout via rose-pink dye-conjugate bands. Procedural control included. Standalone, non-instrumented test device.
Indications for Use
Indicated for routine physical examination, hospital monitoring for bleeding, and screening for colorectal cancer or gastrointestinal bleeding from any source in patients requiring fecal occult blood testing.
Regulatory Classification
Identification
An occult blood test is a device used to detect occult blood in urine or feces. (Occult blood is blood present in such small quantities that it can be detected only by chemical tests of suspected material, or by microscopic or spectroscopic examination.)
Special Controls
*Classification.* Class II (special controls). A control intended for use with an occult blood test is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
Related Devices
K113506 — IND ONE STEP OCCULT BLOOD (FOB) TEST · Ind Diagnostics, Inc. · Dec 19, 2012
K051806 — CARE FECAL OCCULT BLOOD TEST, MODEL KT313 · Epitope Diagnostics, Inc. · Oct 3, 2005
K100031 — IND ONE STEP FECAL OCCULT BLOOD TEST MODEL 440-10 · Ind Diagnostic, Inc. · Jul 19, 2010
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
K061065
B. Purpose for Submission:
Marketing in the U.S.
C. Measurand:
Human Hemoglobin
D. Type of Test:
Immunological test for the qualitative detection of monoclonal antibodies for human hemoglobin
E. Applicant:
TECO Diagnostics
F. Proprietary and Established Names:
Teco Rapid Fecal Occult Blood (FOB) Card Test
G. Regulatory Information:
1. Regulation section:
21 CFR § 864.6550 Occult Blood Test
2. Classification:
Class II
3. Product code:
KHE
4. Panel:
Hematology 81
{1}
H. Intended Use:
1. Intended use(s):
Teco Rapid Fecal Occult Blood (FOB) Card Test is an immunochromatographic assay intended for the determination of human hemoglobin in feces by professional laboratories or physician’s offices. It is useful to determine gastrointestinal bleeding founding a number of gastrointestinal disorders such as colorectal carcinoma, colon polyps, diverticulitis and ulcerative colitis.
2. Indication(s) for use:
Teco Rapid Fecal Occult Blood (FOB) Card Test is recommended for use in 1) routine physical examination, 2) hospital monitoring for bleeding in patients, and 3 screening for colorectal cancer or gastrointestinal bleeding from any source.
3. Special conditions for use statement(s):
N/A
4. Special instrument requirements:
N/A
I. Device Description:
The Teco Rapid Fecal Occult Blood (FOB) Card Test is a sandwich immunoassay; it employs a combination of monoclonal and polyclonal antibodies to selectively identify human hemoglobin in test samples with a high degree of sensitivity. It consists of the test device and buffer in a sample collection tube.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Alfa Scientific, Inc. Instant-View™ Fecal Occult Blood Test
2. Predicate 510(k) number(s):
K021423
{2}
3
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Sample | Feces in extraction buffer | Same |
| Intended Use | Qualitative detection of fecal occult blood in feces by professional laboratories and physician’s offices | Same |
| Extraction Buffer | PBS | Same |
| Sensitivity | 50 ng hHb/ml buffer | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Test Principle | Immunoassay utilizing monoclonal and polyclonal antibodies for the detection of human hemoglobin | Immunoassay utilizing monoclonal for the detection of human hemoglobin |
K. Standard/Guidance Document Referenced (if applicable):
NCCLS EP9-A2 Method Comparison and Bias Estimation using Patient Samples; Approved Guideline
NCCLS EP2-P User Protocol for Evaluation of Qualitative Test performance; Approved Guideline
L. Test Principle:
The Teco Rapid Fecal Occult Blood Card Test employs polyclonal and monoclonal antibodies to detect human hemoglobin in feces. A fecal sample is collected and placed into the fecal collection tube containing a buffer solution. The resulting fluid with extracted hemoglobin (if any) is then added to the test device. The sample fluid mixes with anti-hemoglobin dye-conjugate in the test membrane forming an antigen-dye complex which migrates through the test device. The complex is captured in the test zone by immobilized anti-hemoglobin antibodies. The captured dye-complex becomes visible as a rose-pink band within the test zone, which indicates the test has detected human hemoglobin. In the absence of hemoglobin no line will form in the test zone.
A procedural control is built into the test device to indicate proper test performance. It appears as a rose-pink sandwich dye conjugate reaction and should appear regardless of the test result.
{3}
A rose-pink band in the test zone and in the control zone at or before five minutes is considered a positive result by the criteria of the test. A band only in the control zone at five minutes is a negative result.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Fresh human hemoglobin free stool extraction specimens were spiked with human hemoglobin at three different concentrations: 0, 100, and 1000 ng/ml. Each specimen concentration was tested in each day for 20 days. There were 20 total replicates for each concentration specimen and the agreement was 100% at each concentration.
Reproducibility studies were also conducted at three physician office laboratories with 120 human hemoglobin free stool extracts separated into 4 groups. Each group of specimens was spiked with 0, 50, 60, and 1000 ng/ml of human hemoglobin. Tests were conducted by laboratory technicians with various degrees of educational background and experience and compared to the results of a professional user. There was 99.2% agreement between the results from site 1 and the professional user. There was 98.3% agreement between the results from site 2 and the professional user. There was 100% agreement between the results from site 3 and the professional user.
#### b. Linearity/assay reportable range:
N/A
#### c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Internal Control: A procedural control is built into the test device to indicate proper test performance. It appears as a rose-pink sandwich dye conjugate reaction and should appear regardless of the test result.
External Control: Positive and negative controls should be run to insure that the captured and conjugated antibodies are present and reactive.
#### d. Detection limit:
The minimal detectable concentration of hHb is 50 ng/ml in buffer or 5 µg/g in feces.
#### e. Analytical specificity:
This FOB test is specific to human hemoglobin. The following substances when spiked in both positive and negative specimens did not interfere with the test results.
4
{4}
| Substance | Concentration (ug/ml) |
| --- | --- |
| Chicken Hb | 500 ug/ml |
| Pork Hb | 500 ug/ml |
| Horse Hb | 500 ug/ml |
| Beef Hb | 500 ug/ml |
| Rabbit Hb | 500 ug/ml |
| Goat Hb | 500 ug/ml |
| Horseradish Peroxidase | 2000 ug/ml |
f. Assay cut-off:
N/A
2. Comparison studies:
a. Method comparison with predicate device:
The Teco Rapid FOB Card Test was compared to a predicate device with 120 human hemoglobin free feces extraction specimens in house. The specimens were divided into 6 groups containing 20 specimens. The six groups were spiked with human hemoglobin in the following concentrations: 0, 20, 40, 50, 100, and 2000 ng/ml. The specimens in each group were further divided into two groups: Group A was tested immediately while Group B was stored for 10 days at 2-8°C. The total agreement was 98.3%. The positive percent agreement was 96.3% and the negative percent agreement was 100%.
b. Matrix comparison:
N/A
3. Clinical studies:
a. Clinical Sensitivity:
N/A
b. Clinical specificity:
N/A
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
{5}
5. Expected values/Reference range:
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
1. The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
6
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.