FORSURE ONE STEP FECAL OCCULT BLOOD (FOB) SCREEN CARD TEST
K063693 · New Bay Bioresearch Co. , Ltd. · KHE · May 11, 2007 · Hematology
Device Facts
Record ID
K063693
Device Name
FORSURE ONE STEP FECAL OCCULT BLOOD (FOB) SCREEN CARD TEST
Applicant
New Bay Bioresearch Co. , Ltd.
Product Code
KHE · Hematology
Decision Date
May 11, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.6550
Device Class
Class 2
Indications for Use
Forsure One Step Fecal Occult Blood (FOB) Screen Card Test is a rapid, immunochromatographic assay for the qualitative detection of intact hHb (human hemoglobin) in fecal specimens. It is a convenient and hygienic method for detecting human fecal occult blood, which may be in the dication of gastrointestinal disease associated with bleeding such as colorectal carcinoma, Crohn's disease, ulcerative colitis, and colon polyps in humans.
Device Story
Lateral flow immunochromatographic assay; detects human hemoglobin in fecal samples. Employs sandwich colloidal gold conjugate immunoassay with monoclonal antibodies. Sample flows through absorbent device; gold-labeled antibody-conjugate binds hemoglobin; complex binds to anti-hemoglobin antibody in reaction zone. Produces pink-rose color band at hemoglobin concentrations ≥50 ng/ml. Control zone confirms reagent function. Visually read by healthcare provider or user. Results aid in clinical assessment of gastrointestinal bleeding.
Clinical Evidence
No clinical data provided; substantial equivalence established via bench testing and comparison of analytical properties to predicate device.
Indicated for qualitative detection of human hemoglobin in fecal specimens to aid in identifying gastrointestinal bleeding associated with conditions like colorectal carcinoma, Crohn's disease, ulcerative colitis, and colon polyps.
Regulatory Classification
Identification
An occult blood test is a device used to detect occult blood in urine or feces. (Occult blood is blood present in such small quantities that it can be detected only by chemical tests of suspected material, or by microscopic or spectroscopic examination.)
Special Controls
*Classification.* Class II (special controls). A control intended for use with an occult blood test is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
Predicate Devices
WHPM Hemosure One Step Fecal Occult Blood (FOB) Test (K041202)
Related Devices
K110309 — FOB ONE STEP RAPID TEST · Orient Gene Biotech · Sep 14, 2011
K173212 — Instant-view-PLUS immunochemical Fecal Occult Blood Test · Alfa Scientific Designs, Inc. · Feb 15, 2018
Submission Summary (Full Text)
{0}
K063693 MAY 11 2007
510(k) Submission For
Forsure One Step Fecal Occult Blood (FOB) Screen Card Test
New Bay Bioresearch Co., Ltd.
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
**510(k) Number:** Designed (k) Number to be Determined
## Submitter:
Tianjin New Bay Bioresearch Co., Ltd.
#3 Jian She Rd, Ba Li Tai Industry Area Jin Nan District,
Tianjin, China
Telephone: 86-22-28751515
Facsimile: 86-222-875-1516
## Contact Person:
Rodrigo Berlie
New Product Development Director
Telephone: (760) 828-0990
Facsimile: (760) 602-2999
## Preparation Date:
December 08, 2006
## Device Information:
Trade or Proprietary Name:
Forsure One Step Fecal Occult Blood (FOB) Screen Card Test
Common/Usual Name:
Lateral flow immunochromatographic assay for detection of human hemoglobin or fecal occult blood in feces
Device Classification Name:
Immunoassay of human hemoglobin or fecal occult blood
Regulatory Name:
Human Hemoglobin or Fecal Occult Blood (FOB) Test System
Regulation Section: 21 CFR § 866.6550
Regulatory Class: Class II
Product Code: KHE
CONFIDENTIAL
{1}
510(k) Submission For
Forsure One Step Fecal Occult Blood (FOB) Screen Card Test
New Bay Bioresearch Co., Ltd.
Panel: Hematology (81)
## Predicate Devices:
Forsure One Step Fecal Occult Blood (FOB) Screen Card Test is substantially equivalent to WHPM Hemosure One Step Fecal Occult Blood (FOB) Test cleared by FDA (K041202) for its stated intended use.
## Device Description:
Forsure One Step Fecal Occult Blood (FOB) Screen Card Test is a qualitative, sandwich colloidal gold conjugate immunoassay for the determination of human hemoglobin in feces. The method employs a unique combination of monoclonal antibodies to selectively identify hemoglobin in test sample with a high degree of sensitivity. In less than 5 minutes, elevated levels of human hemoglobin as low as 50 ng/ml can be detected, and positive results for high levels of hemoglobin can be seen in the test as early as two to three minutes. As the test sample flows through the absorbent device, the Colloidal Gold labeled antibody-conjugate binds to the hemoglobin in the specimen forming an antibody-antigen complex. This complex binds to another anti-hemoglobin antibody in the positive reaction zone and produces a pink-ross color band when hemoglobin concentration is greater than 50 ng/ml. In the absence of hemoglobin, there is no line in the positive reaction zone. The reaction mixture continues flowing through the absorbent device past the positive reaction zone and negative control zone. Unbind conjugate binds to the reagents in the negative control zone, producing a pink rose color band, demonstrating that the reagents and device are functioning correctly.
A NEGATIVE specimen produces one distinct color bands in control area. A POSITIVE specimen produces two color bands in the control and test area. There is no meaning attributed to color or its intensity for either line.
## Intended Use:
Forsure One Step Fecal Occult Blood (FOB) Screen Card Test is a rapid, immunochromatographic assay for the qualitative detection of intact hHb (human hemoglobin) in fecal specimens. It is a convenient and hygienic method for detecting human fecal occult blood, which may be in the dication of gastrointestinal disease associated with bleeding such as colorectal carcinoma, Crohn's disease, ulcerative colitis, and colon polyps in humans.
## Comparison to Predicate Device(s):
Both devices (Forsure and WHPM) are for the qualitative determination of the Fecal Occult Blood. The specific antibodies against human hemoglobin were used in both devices. All test devices are visually-read single use device. The cutoff of each analyte is same (50 ng/ml or 50 µg/g of feces).
CONFIDENTIAL
{2}
510(k) Submission For
Forsure One Step Fecal Occult Blood (FOB) Screen Card Test
New Bay Bioresearch Co., Ltd.
## Summary:
The information provided in this pre-market notification demonstrates that Forsure One Step Fecal Occult Blood (FOB) Screen Card Test is substantially equivalent to WHPM Hemosure One Step Fecal Occult Blood (FOB) Test. Substantial equivalence was demonstrated through comparison of intended use and physical properties to the commercially available and analytical predicate devices. The information supplied in this pre-market notification provides reasonable assurance that the Forsure One Step Fecal Occult Blood (FOB) Screen Card Test is safe and effective for its stated intended use.
CONFIDENTIAL
11
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.