SPERMCHECK VASECTOMY

K073039 · Princeton BioMeditech Corp. · POV · Jan 24, 2008 · Hematology

Device Facts

Record IDK073039
Device NameSPERMCHECK VASECTOMY
ApplicantPrinceton BioMeditech Corp.
Product CodePOV · Hematology
Decision DateJan 24, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5220
Device ClassClass 2

Indications for Use

SpermCheck® Vasectomy is a rapid qualitative test that detects low concentrations of sperm at or above 250,000 sperm/mL in human semen as an aid for vasectomized men. For in vitro, over the counter home use.

Device Story

SpermCheck® Vasectomy is an immunochromatographic test for rapid, qualitative detection of sperm concentration in human semen. Designed for OTC home use by men who have undergone a vasectomy. The device detects sperm concentrations at or above 250,000 sperm/mL. The user performs the test at home; results are interpreted qualitatively by the presence or absence of a color indicator. The test serves as an aid to confirm the success of a vasectomy procedure by identifying the presence of residual sperm. It provides a convenient, non-clinical method for monitoring post-vasectomy status.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing and comparison of technological principles to predicate devices.

Technological Characteristics

Solid-phase chromatographic immunoassay. Components include nitrocellulose membrane, lyophilized detector antibody conjugated to gold dye, and immobilized capture antibodies. Detects SP10 protein. Ambient temperature operation. Visual readout via colored bands. No electronic components or software.

Indications for Use

Indicated for vasectomized men to qualitatively detect sperm concentration at or above 250,000 sperm/mL in human semen. For OTC home use.

Regulatory Classification

Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K073039 B. Purpose for Submission: Traditional, original 510(k) submission C. Measurand: Sperm concentration D. Type of Test: This is a one step, immuno-chromatographic assay. E. Applicant: Princeton BioMeditech Corporation F. Proprietary and Established Names: SpermCheck® Vasectomy G. Regulatory Information: 1. Regulation section: No regulation 2. Classification: Class II 3. Product code: MNA 4. Panel: Obstetrics and Gynecology (85) {1} H. Intended Use: 1. Intended use(s): SpermCheck Vasectomy is a rapid qualitative test that detects low concentrations of sperm at or above 250,000 sperm/mL, in human semen as an aid for vasectomized men. For in vitro, over the counter home use. 2. Indication(s) for use: N/A 3. Special conditions for use statement(s): N/A 4. Special instrument requirements: N/A I. Device Description: The SpermCheck Vasectomy test is designed to detect the presence or absence of sperm in semen following a vasectomy. Since sperm persist in the male reproductive tract for some time after a vasectomy it is necessary to continue other methods of contraception until those sperm have been voided. With this simple Sperm Check Vasectomy test the user can quickly know when the amount of sperm in his ejaculate has dropped to very low levels and it is safe to discontinue other forms of contraception. The SpermCheck Vasectomy test device is designed to give a positive result at or above 250,000 sperm/mL and negative result below 250,000 sperm/mL. J. Substantial Equivalence Information: 1. Predicate device name(s): a. Hemacytometer b. Fertell Male Fertility Test c. FertilMARQ™ Male Fertility Test {2} 2. Predicate K number(s): a. N/A b. K041039 c. K011679 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Test Type | Qualitative Test | Fertell: Same FertilMARQ: Same | | Sample Type | Human semen | Hemacytometer: Same Fertell: Same FertilMARQ: Same | | Assay Design | Chromatographic immunoassay | Fertell: Same | | Assay Principle | Colored label bound to sperm | Fertell: Same FertilMARQ: Same | | Test Temperature | Ambient temperature | Hemacytometer: Same FertilMARQ: Same | | Result Reading | Qualitative by visual line | Fertell: Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Test Type | Qualitative test | Hemacytometer: Quantitative test | | Assay Design | Chromatographic Immunoassay | Hemacytometer: Manual count FertilMARQ: Chemical reaction | | Assay Principle | Color label bound to sperm | Manual count | | Test Temperature | Ambient temperature | 37°C | {3} | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Result Reading | Qualitative by visual line | Hemacytometer: Manual count FertilMARQ: Qualitative by visual color | | Positive Detection Level | ≥250,000 sperm/mL | Hemacytometer: Any number Fertell: ≥10 M sperm/mL FertilMARQ: ≥20 M sperm/mL | | Performance Comparison | 96% compared to Hemacytometer | Fertell: Accuracy reported to its predicate: 95.7% FertilMARQ: Reported: Sensitivity: 94% Specificity: 61% Accuracy: 78% | ## K. Standard/Guidance Document Referenced (if applicable): WHO Laboratory Manual for the Examination of Human Semen and Sperm-Cervical Mucus Interaction, fourth ed., Cambridge University Press, Cambridge, U.K. ## L. Test Principle: SpermCheck Vasectomy test is a rapid qualitative test detecting the presence of sperm in human semen at 250,000 sperm/mL and above. The test employs solid-phase chromatographic immunoassay technology and sperm are detected using antibodies against sperm acrosome-specific protein analyte SP10. This protein is a post meiotic gene product expressed selectively in spermatids. In the test procedure, a semen sample mixed with a test solution is dispensed into the sample well, the sample migrates into the pads containing lyophilized detector antibody that is conjugated to gold dye. If the sample contains sperm, upon hydration by the applied sample, complexes are formed consisting of antibody-dye + antigen (SP10). The complexes and excess hydrated antibodies migrate forward via capillary action on the nitrocellulose membrane. The membrane serves as a solid support upon which capture antibody (test band) and anti-species IgG antibodies (control band) are immobilized. The immobilization of these reagents on the nitrocellulose membrane, in defined areas, allows for the formation of distinct colored bands that can be read visually. The complex of antibody-dye + antigen is captured by antibody on the membrane and a red line appears at the test band on the membrane. {4} M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: The precision study was performed by carrying out the test for two days by three different operators with two lots of devices. The samples were prepared by diluting the pooled normal semen to the appropriate sperm concentrations. The samples were randomly distributed. The tested concentrations were 0. 125,000, 251,000, 353,000, and 706,000 sperm/mL. The results agreed 100% with the expected results. No difference in the performance was found between lots, days, and operators. b. Linearity/assay reportable range: N/A c. Traceability, Stability, Expected values (controls, calibrators, or methods): Real time stability testing has been performed. At each test time point, negative, very low positive (approximately the level of sensitivity), and medium positive (2 times the level of sensitivity) controls were tested. At each time point, each sample level was tested in triplicate. Two lots of test devices have been tested and the test results are acceptable up to 16 months at ambient temperature. d. Detection limit: The SpermCheck Vasectomy test was designed to show positive results from samples containing sperm at concentrations at or above the sensitivity limit of 250,000 sperm/mL. Semen samples were collected from normal men and pooled. The initial counting of pooled semen by Hemacytometer was 93.6 million sperm/mL. The semen was diluted with semen from a pool of vasectomized male donors to a sperm concentration of 406,000 sperm/mL. All other concentrations were prepared from this semen sample by serially diluting with pooled semen from vasectomized men. The concentrations prepared were 406,000, 258,000, 129,000, 64,500, and 21,500 sperm/mL. At each concentration the SpermCheck Vasectomy test was repeated 10 times with each lot of devices. Two lots of SpermCheck Vasectomy devices were used for the experiment. Therefore, 20 replicates at each concentration and a total of 120 tests were tested. 5 {5} Table 1. Sensitivity Study | Sperm Concentration (Sperm/mL) | Percent Positive (N) | Percent Negative (N) | | --- | --- | --- | | 0 | 0 (20) | 100 (20) | | 21,500 | 0 (20) | 100 (20) | | 64,500 | 0 (20) | 100 (20) | | 129,000 | 0 (20) | 100 (20) | | 258,000 | 100 (200) | 0 (20) | | 406,000 | 100 (20) | 0 (20) | # e. Analytical specificity: The cross reactivity and potential interference of various substances on the SpermCheck Vasectomy test were tested by spiking each substance into the 250,000 sperm/mL semen sample and into the semen from a vasectomized male (no sperm in the semen) to the desired concentration. Each concentration was tested in triplicate. All substances showed no cross-reactivity or interference at the level tested. Table 2. Substances Tested for Cross Reactivity | Substances | Source and Cat. # | Test Concentration | | --- | --- | --- | | Escherichia coli | ATCC35218 | 1 x 10^6/mL | | Corynebacterium diphtheria | ATCC13812 | 1 x 10^6/mL | | Neisseria gonorrhea | ATCC35541 | 1 x 10^6/mL | | Chlamydia trachomatis | ATCC VR-880 | 3.1 x 10^4/mL | | 19-norethindrone acetate | Sigma N-6127 | 20 μg/mL | | Testosterone | Cerilliant T-037 | 20 μg/mL | | β-Estradiol | Sigma E-1132 | 20 μg/mL | | D(-) Norgestrel | Sigma N-2260 | 20 μg/mL | | White Blood Cells | - | 1.9 x 10^9/mL | | Whole Blood | - | 10% | | Urine | - | 10% | | Saliva | - | 10% | {6} f. Assay cut-off: N/A ## 2. Comparison studies: a. Method comparison with predicate device: A study was designed to test the performance of the SpermCheck Vasectomy test compared to the WHO standard procedure of determining sperm concentration with a Hemacytometer (the predicate device). SpermCheck Vasectomy is a semi-quantitative test that reports samples as containing sperm concentrations above or below 250,000 sperm/mL. The study was performed at two clinical sites by professional laboratory personnel. A total of 145 semen samples were tested, however one test result was excluded from the analysis since the result was not read at the correct reading time given in the test protocol. Since the detection level of SpermCheck Vasectomy is 250,000 sperm/mL, the data comparison was made with this concentration as a decision level between positive and negative. Table 3. SpermCheck Vasectomy vs. Hemacytometer | | SpermCheck Vasectomy | | Total | | | --- | --- | --- | --- | --- | | | | Positive | | Negative | | Hemacytometer | Positive | 37 | 3 | 40 | | | Negative | 3 | 101 | 104 | | Total | | 40 | 104 | 144 | Negative Predictive Value: 0.971 Positive Predictive Value: 0.925 Sensitivity: 93% Specificity: 97% Accuracy: 96% b. Matrix comparison: N/A ## 3. Clinical studies: a. Clinical Sensitivity: N/A {7} b. Clinical specificity: N/A c. Other clinical supportive data (when a. and b. are not applicable): I. A consumer use study was conducted to determine the test performance when used by unassisted lay users following the instructions in the package insert. The study was conducted in two ways: 1) Performance of the consumer at home with his own semen sample; 2) Performance of the consumer with provided semen samples. The minimum target number was 50 people for each study. In the study of the consumer at home with his own semen sample 51 subjects were enrolled with 31 who had been vasectomized and 20 who had not. Laboratory personnel performed a hemacytometer count on unused sample returned to the lab by the consumer. There was 100% agreement between the SpermCheck Vasectomy results obtained by the home users and those obtained by laboratory personnel retesting the same semen samples. For the method comparison, there was 98% agreement between the SpermCheck Vasectomy results and the hemacytometer counts with one disagreement among the 50 tests. In the study of consumers provided with semen samples 59 people (49 male and 10 female) participated. All results showed 100% agreement. Table 4. Consumer Performance with Provided Semen Samples | Sample ID | Concentration Sperm/mL | Negative | Positive | Invalid | Total | % Correct | | --- | --- | --- | --- | --- | --- | --- | | A | 350,000 | 0 | 59 | 0 | 59 | 100 | | B | 17,000 | 59 | 0 | 0 | 59 | 100 | | C | 1,000,000 | 0 | 59 | 0 | 59 | 100 | | Total | | 59 | 118 | 0 | 177 | 100 | II. A High Dose Hook Effect study was conducted to assess whether the Sperm Check Vasectomy test would give false negative results with very high concentrations of sperm due to the hook effect, and semen samples of various sperm concentrations of up to 253 million sperm/mL were tested. The results were read by an optical density reader. The signal intensity increased as the sperm concentration increased and no signal reduction was observed. In a separate study, a clinical sample with a sperm concentration of 387 million sperm/mL was tested and the test result showed a very strong positive signal. A false negative result due to the hook effect was not observed with this very high sperm concentration. {8} 4. Clinical cut-off: N/A 5. Expected values/Reference range: N/A N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. P. Other Supportive Device and Instrument Information: N/A 9
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...