MODIFICATION TO BAYER ADVIA 2120 HEMATOLOGY ANALYZER WITH AUTOSLIDE SYSTEM

K051693 · Bayer Healthcare, LLC · GKL · Dec 7, 2005 · Hematology

Device Facts

Record IDK051693
Device NameMODIFICATION TO BAYER ADVIA 2120 HEMATOLOGY ANALYZER WITH AUTOSLIDE SYSTEM
ApplicantBayer Healthcare, LLC
Product CodeGKL · Hematology
Decision DateDec 7, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5200
Device ClassClass 2
AttributesPediatric

Indications for Use

The ADVIA 2120 nRBC method is intended to provide an in vitro diagnostic, quantitative determination of nucleated red blood cells in peripheral whole blood.

Device Story

The ADVIA 2120 with nRBC Method is an automated hematology analyzer used in clinical laboratories. It processes peripheral whole blood samples to identify and quantify nucleated red blood cells (nRBCs). The device utilizes histogram analysis of the Peroxidase channel and an arithmetic algorithm combining counts from the Peroxidase and Basophil/Lobularity channels. nRBCs are identified by nuclear size in the peroxidase-negative region and by nuclear density/volume in the Basophil/Lobularity channel. The system outputs absolute nRBC counts and percentages, while automatically correcting the total White Blood Cell (WBC) count and WBC differential for the presence of nRBCs. This correction improves the accuracy of hematological reporting, preventing false elevations of WBC counts caused by nRBC interference. Healthcare providers use these corrected results to assess patient hematological status, particularly in cases of hemolytic crisis or neonatal care.

Clinical Evidence

No clinical trial data provided. Substantial equivalence is supported by performance data demonstrating the accuracy of the nRBC method compared to manual reference methods consistent with NCCLS H-20A standards.

Technological Characteristics

Automated hematology analyzer; uses Peroxidase and Basophil/Lobularity channel sensing. Analytical module performs aspiration, dilution, and analysis. Software-controlled; includes Autoslide module for automated smear preparation. Requires EDTA-anticoagulated whole blood. Connectivity via workstation interface. Calibration and QC performed using commercially available materials. Software version 5.2.x.

Indications for Use

Indicated for the quantitative determination of nucleated red blood cells in peripheral whole blood for patients requiring nRBC counts (specifically those with >= 2% nRBCs or >= 200 nRBC/μL with a WBC count >= 3,000/μL).

Regulatory Classification

Identification

An automated cell counter is a fully-automated or semi-automated device used to count red blood cells, white blood cells, or blood platelets using a sample of the patient's peripheral blood (blood circulating in one of the body's extremities, such as the arm). These devices may also measure hemoglobin or hematocrit and may also calculate or measure one or more of the red cell indices (the erythrocyte mean corpuscular volume, the mean corpuscular hemoglobin, or the mean corpuscular hemoglobin concentration). These devices may use either an electronic particle counting method or an optical counting method.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE A. 510(k) Number: K051693 B. Purpose for Submission: Addition of new parameter C. Manufacturer and Instrument Name: Bayer HealthCare LLC, Diagnostics Division, ADVIA 2120 with Autoslide System D. Type of Test or Tests Performed: Nucleated Red Blood Cells (nRBC) Method, quantitative E. System Descriptions: 1. Devise Description: The technological and methodological principles of the ADVIA 2120 with Autoslide used for the measurement of blood cells in whole blood specimens will be used without change except for the addition of a new methodology to quantitate nucleated red blood cells and correct white blood cells in peripheral whole blood. The ADVIA 2120 nRBC method reports nRBC counts for whole blood with either 200 or more nRBC/μL, or with at least 2% nRBCs with a WBC count of at least 3,000/ul. The method reports both an absolute nRBC count (10⁹/L) and a percentage count (#nRBC/100 WBC). The software corrects the WBC count for nRBC, recalculates the WBC differential, and recalculates %MN and %PMN. The uncorrected counts are also available and are designated by a lower case "u." 2. Principles of Operation: The ADVIA 2120 nRBC method uses histogram analysis routines to analyze the unstained region of the Peroxidase channel as well as arithmetic and algorithm that combines counts from the Peroxidase and Basophil/Lobularity channel to enumerate nRBCs. {1} The analysis corrects the White Blood Cell Count (WBC) for the presence of nRBCs, as well as the WBC Differential. It reports the corrected WBC count and the corrected differential. 3. Modes of Operation: Not applicable. 4. Specimen Identification: Not applicable. 5. Specimen Sampling and Handling: The ADVIA 2120 nRBC method requires whole blood collected in an evacuated tube containing EDTA as an anticoagulant. Samples requiring nRBC count should not be refrigerated and must be analyzed within eight hours of phlebotomy. 6. Calibration: Proper calibration of the ADVIA 2120 Baso WBC and Perox WBC counts are required for accurate nRBC counts. Calibration is as necessary using the instructions provided by the ADVIA 2120 Hematology System Operator’s Guide. 7. Quality Control: Quality control of the ADVIA 2120 nRBC method can be performed with commercially available control materials. In the event that control materials are not available, the method should be controlled following agency requirement for the user’s laboratory. 8. Software: The ADVIA 2120 software updates the Run Screen options and provides unit set options to support nRBC analysis. The customer is to refer to the ADVIA Hematology System Operator’s Guide for detailed instructions on configurations. FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types: Yes ☐ X ☐ or No ☐ {2} F. Regulatory Information: 1. Regulation section: 21 CFR 864.5200, Automated cell counter 21 CFR 864.5220, Automated differential cell counter 21 CFR 864.3800, Automated slide stainer 2. Classification: Class II 3. Product code: GKL, Counter, Cell, Automated (Particle Counter) GKZ, Counter, Differential Cell KPA, Slide Stainer, Automated 4. Panel: Hematology (81) G. Intended Use: 1. Indication(s) for Use: The ADVIA 2120 Nucleated Red Blood Cells (nRBC) Method nRBC method is intended to provide an in vitro diagnostic, quantitative determination of nucleated red blood cells in peripheral whole blood. 2. Special Conditions for Use Statement(s): Not applicable. H. Substantial Equivalence Information: 1. Predicate Device Name(s) and 510(k) numbers: ADVIA 2120 with Autoslide System, K042251 {3} 2. Comparison with Predicate Device: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | | ADVIA 2120 with Autoslide and nRBC Method | ADVIA 2120 with Autoslide | | Intended Use | Same/w Additional parameter for red blood cell analysis for percent nucleated Red Blood Cells (% nRBC) and absolute nRBC count (# nRBC per microliter) | Quantitative, automated hematology analyzer that provides a leukocyte differential count and reticulocyte analysis for in vitro diagnostic use in clinical laboratories. In addition, the system provides the added capability to automatically prepare and stain high quality blood smears on a microscope slide. | | Technology | Analytical Module that aspirates, dilutes and analyzes whole blood samples; an auto sampler that automatically mixes, identifies, and presents samples for processing, a computer workstation that provides the software to control the instrument provides user interface and supports the management of the data, and Autoslide module that provides the added capability to automatically prepare and stain high-quality blood smears on a microscope slide. | Same | | Reagents | Reagents to support Complete Blood Count, White Blood Cell Differential, Reticulocyte analysis and Staining reagents | Same | 4 {4} 5 I. Special Control/Guidance Document Referenced (if applicable): H20-A Reference Leukocyte Differential Count (Proportional) and Evaluation of Instrument Methods, Approved Standard, NCCLS J. Performance Characteristics: 1. Analytical Performance: a. Accuracy: Systematic bias of the ADVIA 2120 nRBC count was evaluated by comparing ADVIA 2120 nRBC counts to reference manual counts for 744 samples collected at clinical trial sites. Regression statistics are found in the table that follows: | Parameter | r | Slope | Intercept | Manual Mean | ADVIA 2120 Mean | Bias | | --- | --- | --- | --- | --- | --- | --- | | nRBC% | 0.970 | 0.98 | -0.6 | 11.9 | 11.1 | -0.8 | | #nRBC (x10^{9}/L) | 0.992 | 0.98 | -0.15 | 1.91 | 1.82 | -0.09 | | Corrected WBC | 0.997 | 1.02 | -0.14 | 11.54 | 11.64 | 0.10 | The ADVIA 2120 only provides an accurate nRBC count for samples with $\geq 2\%$ nRBC. b. Precision/Reproducibility: Within run precision was evaluated by performing duplicate assays of 45 samples with nRBC counts of $2\%$ or more. Precision results are as follows: | Parameter | Mean | SD | CV | | --- | --- | --- | --- | | nRBC% | 26.7 | 4.6 | 17.2 | | #nRBC (x10^{9}/L) | 4.0 | 0.4 | 10.0 | | Corrected WBC | 14.7 | 0.5 | 3.4 | c. Linearity: The analytical range or range of reportable values for nRBC% is 2 – 300%. #nRBC values are reportable if the nRBC% associated with the #nRBC value falls within those limits. In a study to evaluate the performance of the ADVIA 2120 nRBC% over the analytical range the nRBC method was found to have a specificity of 92% at a level of 2% nRBCs. {5} d. Carryover: The ADVIA 2120 nRBC method uses hydraulic cycles that are identical to those contained in software versions prior to version 5.2.x. Therefore, the expected carryover remains the same when using version 5.2.x (< 1% for all parameters). e. Interfering Substances: - Unlysed RBCs: The RBCs from peripheral blood samples from newborns and patients with certain disorders (e.g. thalassemia) may be resistant to lysis. The unlysed RBCs occupy the same region of the Peroxidase cytogram as the nRBCs and interfere with the count. - Large platelets: In certain disease states the peripheral blood may contain many large platelets that also occupy the same region of the Peroxidase channel as nRBCs. The large platelets will interfere with nRBC enumeration. - Lipids: Samples from patients undergoing total parenteral nutrition may contain lipids of high enough concentration to interfere with sample analysis on the ADVIA 2120. The effect is observed in both the ADVIA 2120 Peroxidase and Basophil/Lobularity channels and could compromise the nRBC count. The ADVIA 2120 software provides four sample/system flags to help identify samples that may have invalid nRBC counts due to interfering substances. 2. Other Supportive Instrument Performance Data Not Covered Above: K. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. L. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...