K141109 · Leica Biosystems Imaging, Inc. · NOT · Jul 29, 2014 · Hematology
Device Facts
Record ID
K141109
Device Name
APERIO EPATHOLOGY EIHC IVD SYSTEM
Applicant
Leica Biosystems Imaging, Inc.
Product Code
NOT · Hematology
Decision Date
Jul 29, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.1860
Device Class
Class 2
Indications for Use
The Aperio ePathology eIHC IVD System is an automated digital slide creation, management, viewing and analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape. The IHC HER2 Image Analysis application is intended for use as an aid to the pathologist in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffinembedded neoplastic tissue. The IHC HER2 Image Analysis application is intended for use as an accessory to the Dako HercepTest™ to aid in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded neoplastic tissue. When used with the Dako HercepTest™, it is indicated for use as an aid in the assessment of breast cancer patients for whom HERCEPTIN® (Trastuzumab) treatment is being considered. The IHC ER Image Analysis application is intended for use as an aid to the pathologist in the detection and quantitative measurement of ER (Estrogen Receptor) in formalin-fixed paraffin-embedded neoplastic tissue. The IHC PR Image Analysis application is intended for use as an aid to the pathologist in the detection and quantitation measurement of PR (Progesterone Receptor) in formalin-fixed, paraffin-embedded neoplastic tissue. It is indicated for use as an aid in the management, prognosis, and prediction of therapy outcomes of breast cancer.
Device Story
System automates digital slide creation, management, viewing, and analysis for pathology. Components include digital microscope slide scanner (AT Turbo or CS2 models), computer, monitor, and eSlideManager software. Operator loads glass slides into carrier; scanner uses linear-array technique with 20x objective to generate high-resolution, seamless digital images. Software stores images in Digital Slide Repository (DSR) with SQL database metadata. Pathologists view images on monitors and use integrated image analysis applications to detect and quantify HER2, ER, and PR protein expression in breast tissue. System acts as adjunctive tool to assist pathologist reproducibility; results aid in breast cancer management and therapy prediction. Used in clinical laboratory settings by technicians and pathologists.
Clinical Evidence
No clinical tests submitted. Evidence consists of bench-level system verification, validation, and performance testing. Performance testing evaluated concordance of HER2, ER, and PR scores between updated ScanScope models and the predicate device using formalin-fixed, paraffin-embedded breast tissue slides. HER2 results analyzed via trichotomous scoring (0/1+, 2+, 3+); ER/PR results analyzed via dichotomous (positive/negative) analysis at 1% and 10% thresholds. Acceptance criteria based on percent agreement with exact 95% lower confidence intervals were met.
Technological Characteristics
System includes automated digital microscope slide scanner, computer, color monitor, and keyboard. Employs linear-array scanning with 20x objective. Energy source: 100vAC–240vAC, 50Hz/60Hz. Connectivity: Networked via eSlideManager server (SQL database, RAID storage). Software performs colorimetric and morphometric pattern recognition for cell detection. Sterilization: Not applicable (device is diagnostic imaging system).
Indications for Use
Indicated for use as an aid to pathologists in the display, detection, counting, and classification of tissues and cells in formalin-fixed, paraffin-embedded neoplastic breast tissue. Specifically used for semi-quantitative measurement of HER2/neu and quantitative measurement of Estrogen Receptor (ER) and Progesterone Receptor (PR) to aid in the management, prognosis, and prediction of therapy outcomes for breast cancer patients.
Regulatory Classification
Identification
Immunohistochemistry test systems (IHC's) are in vitro diagnostic devices consisting of polyclonal or monoclonal antibodies labeled with directions for use and performance claims, which may be packaged with ancillary reagents in kits. Their intended use is to identify, by immunological techniques, antigens in tissues or cytologic specimens. Similar devices intended for use with flow cytometry devices are not considered IHC's.
Special Controls
(2) Class II (special control, guidance document: “FDA Guidance for Submission of Immunohistochemistry Applications to the FDA,” Center for Devices and Radiologic Health, 1998). These IHC's are intended for the detection and/or measurement of certain target analytes in order to provide prognostic or predictive data that are not directly confirmed by routine histopathologic internal and external control specimens. These IHC's provide the pathologist with information that is ordinarily reported as independent diagnostic information to the ordering clinician, and the claims associated with these data are widely accepted and supported by valid scientific evidence. Examples of class II IHC's are those intended for semiquantitative measurement of an analyte, such as hormone receptors in breast cancer.
Predicate Devices
ScanScope® System XT (k071128, k073677)
Related Devices
K073677 — SCANSCOPE XT SYSTEM · Aperio Technologies · Aug 1, 2008
K071671 — SCANSCOPE XT SYSTEM · Aperio Technologies · Dec 28, 2007
K080254 — SCANSCOPE XT SYSTEM · Aperio Technologies · Oct 31, 2008
K071128 — SCANSCOPE XT SYSTEM · Aperio Technologies · Oct 10, 2007
K142965 — Virtuoso System for IHC PR (1E2) using iScan HT · Ventana Medical Systems, Inc. · Jul 16, 2015
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## A. 510(k) Number:
k141109
## B. Purpose for Submission:
Addition of the AT Turbo and CS2 scanner models to the cleared Aperio ScanScope® XT System for scoring ER, PR and HER2 immunohistochemically stained slides by the image analysis algorithm method.
## C. Manufacturer and Instrument Name:
Leica Biosystems Imaging, Inc., Aperio ePathology eIHC IVD System, AT Turbo and CS2 models
## D. Type of Test or Tests Performed:
Computer-assisted image analysis algorithm for scoring immunohistochemically stained HER2/neu slides, estrogen receptor (ER)/progesterone receptor (PR) slides.. The image analysis algorithm scoring is confirmed by the pathologist before reporting the scores.
## E. System Descriptions:
### 1. Device Description:
The Aperio ePathology eIHC IVD System is a closed system comprised of a ScanScope® digital slide scanner instrument (either the ScanScope AT Turbo or the ScanScope CS2) and eSlide Manager software, ImageScope software and Image Analysis Algorithms, as applicable. The Aperio ePathology eIHC IVD System is an automated digital slide creation, management, viewing and analysis system. The system is comprised of a slide scanner instrument and a computer executing eSlideManager, ImageScope and Image Analysis Algorithm software. The system capabilities include digitizing immunohistochemically stained slides, storing and managing the resulting digital slide images, retrieving and displaying digital slides, providing tools for annotating digital slides and entering and editing data associated with digital slides, and tools for image analysis of the regions that pathologists select on the digital slide images.
### 2. Principles of Operation:
Glass slides are digitized by an image acquisition system that provides digital slide images of the immunohistochemically stained slides; the image acquisition system provides support for patient / slide identification, slide loading, calibration, tissue identification, focusing and digital image storage and management. Pathologists select
{1}
regions on the digital slide images that will be automatically analyzed by an image analysis algorithm for HER2 or ER/PR (i.e., HER2 or ER/PR Image Analysis). The regions on the digital slide images selected by pathologists for image analysis should be representative for the entire tumor.
The HER2 Image Analysis algorithm detects the membrane staining for the individual tumor cells in the selected regions and quantifies the intensity and completeness of the membrane staining. Tumor cells are individually classified as 0, 1+, 2+, and 3+ based on their membrane staining intensity and completeness. The HER2 score is then calculated based on the percentages of 0, 1+, 2+, and 3+ cells according to the HER2 scoring scheme. A markup image highlights the detected cell features (black = nuclei and membrane) and the membrane staining which is color-coded according to the cell classification (blue = 0, yellow = 1+, orange = 2+, red = 3+). The pathologist is then provided with HER2 score and the percentages of 0, 1+, 2+, and 3+ cells. The pathologist makes the final interpretation based on both the qualitative and quantitative information and should follow all appropriate instructions in the Dako Hercep Test™ product insert
The ER/PR image analysis algorithms use color, intensity, size, pattern and shape information to detect, count, and classify cells of interest and determine a score that correlates with the score the pathologists provide. The image analysis algorithm reports the percentage of positive nuclei and average intensity score of 0, 1+, 2+ or 3+ for each slide. The pathologist makes the final interpretation based on the percent positivity of tumor nuclei and/or staining intensity according to the laboratory's established interpretation criteria.
3. Modes of Operation:
Semi-automated computer-assisted scoring of IHC HER2, ER or PR stained slide images performed by Image Analysis Applications with manual verification by the pathologist.
4. Specimen Identification:
Glass slides are identified by barcode label
5. Specimen Sampling and Handling:
The Autoloader component of the ScanScope® AT Turbo instrument is responsible for storing slides to be scanned, and contains a mechanism for mounting the current slide to be processed on the scanning stage. The slide autoloader is an assembly within the ScanScope AT Turbo that can hold up to 400 slides.
The ScanScope® CS2 has no Autoloader. The CS2 has a five slide capacity tray that can be inserted by the operator into the system.
6. Calibration:
{2}
The scanner is calibrated and verified at the factory before shipment. No further manual calibration or verification is required by the customer. Annual preventive maintenance, performed by Leica Biosystems Imaging Technical Services, is recommended to make sure the scanner is operating correctly. The scanner provides automatic calibration and verification every time a slide is scanned by performing prescan calibration during the scan.
7. Quality Control:
The pathologist must verify that the eSlides are of sufficient quality to perform his or her task. The pathologist should review the quality of the scanning by evaluating that the entire tissue sample is scanned and in focus. Positive and negative control slides are also scanned along with the patient slides which are reviewed by the pathologist before scoring.
8. Software:
FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types:
Yes ☐ X or No ☐
F. Regulatory Information:
1. Regulation section:
21 CFR §864.1860 Immunohistochemistry reagents and kits
2. Classification:
II
3. Product code:
NOT, NQN
4. Panel:
Pathology (88)
G. Intended Use:
1. Indication(s) for Use:
- HER2 Image Analysis
{3}
The Aperio ePathology eIHC IVD System is an automated digital slide creation, management, viewing and analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape.
The IHC HER2 Image Analysis application is intended for use as an aid to the pathologist in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded neoplastic tissue.
The IHC HER2 Image Analysis application is intended for use as an accessory to the Dako HercepTest™ to aid in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded neoplastic tissue. When used with the Dako HercepTest™, it is indicated for use as an aid in the assessment of breast cancer patients for whom HERCEPTIN® (Trastuzumab) treatment is being considered.
Note: The IHC HER2 Image Analysis application is an adjunctive computer-assisted methodology to assist the reproducibility of a qualified pathologist in the acquisition and measurement of images from microscope slides of breast cancer specimens stained for the presence of HER-2 receptor protein. The IHC HER2 Image Analysis application is intended to be used on images viewed on a computer monitor. The accuracy of the test result depends upon the quality of the immunohistochemical staining. It is the responsibility of a qualified pathologist to employ appropriate morphological studies and controls as specified in the instructions for the Dako HercepTest™ to assure the validity of the IHC HER2 Image Analysis application assisted HER-2/neu score. The actual correlation of the Dako HercepTest™ to Herceptin® clinical outcome has not been established.
- **ER/PR Image Analysis**
The Aperio ePathology eIHC IVD System is an automated digital slide creation, management, viewing and analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape.
The IHC ER Image Analysis application is intended for use as an aid to the pathologist in the detection and quantitative measurement of ER (Estrogen Receptor) in formalin-fixed paraffin-embedded neoplastic tissue.
The IHC PR Image Analysis application is intended for use as an aid to the pathologist in the detection and quantitation measurement of PR (Progesterone Receptor) in formalin-fixed, paraffin-embedded neoplastic tissue.
It is indicated for use as an aid in the management, prognosis, and prediction of
{4}
therapy outcomes of breast cancer.
Note: The IHC ER and PR Image Analysis applications are an adjunctive computer-assisted methodology to assist the reproducibility of a qualified pathologist in the acquisition and measurement of images from microscope slides of breast cancer specimens stained for the presence of estrogen and progesterone receptor proteins. The IHC ER and PR Image Analysis applications are intended to be used on images viewed on a computer monitor. The accuracy of the test result depends upon the quality of the immunohistochemical staining. It is the responsibility of a qualified pathologist to employ appropriate morphological studies and controls as specified in the instructions for the ER and PR reagent/kit used to assure the validity of the IHC ER and PR Image Analysis application assisted scores.
2. Special Conditions for Use Statement(s):
For prescription use only
Indicated for use with slides stained on the Dako Autostainer??
H. Substantial Equivalence Information:
1. Predicate Device Name(s) and 510(k) numbers: ScanScope® System XT; k071128, k073677
2. Comparison with Predicate Device:
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Aperio ScanScope® XT System (Predicate) | Aperio ePathology eIHC IVD ScanScope AT Turbo and CS2 System |
| Intended Use/ Indications for Use | The ScanScope System is an automated digital slide creation, management, viewing and analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest and shape.
The IHC HER2 Image Analysis application (k071128) is intended for use as an aid to the pathologist in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in | Same |
{5}
| | formalin-fixed, paraffin-embedded neoplastic tissue. The IHC HER2 Image Analysis application is intended for use as an accessory to the Dako HercepTest™ to aid in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded neoplastic tissue. When used with the Dako HercepTest™, it is indicated for use as an aid in the assessment of breast cancer patients for whom HERCEPTIN® (Trastuzumab) treatment is being considered.The IHC ER Image Analysis application (k073677) is intended for use as an aid to the pathologist in the detection and quantitative measurement of ER (Estrogen Receptor) in formalin-fixed paraffin-embedded neoplastic tissue. The IHC PR Image Analysis application (k073677) is intended for use as an aid to the pathologist in the detection and quantitation measurement of PR (Progesterone Receptor) in formalin-fixed, paraffin-embedded neoplastic tissue. It is indicated for use as an aid in the management, prognosis, and prediction of therapy outcomes of breast cancer. | |
| --- | --- | --- |
| Specimen Type | Formalin-fixed, paraffin-embedded breast tissue specimens stained by immunohistochemistry | Same |
| Assay Used | HER2: Dako HercepTest®ER: Dako Mouse Anti- Human Estrogen Receptor α (Clone 1D5)PR: Dako Monoclonal Mouse Anti-Human Progesterone Receptor (Clone PgR 636) | Same |
| Method of Interpretation | Pathologists select regions on the digital slide images that will be automatically analyzed by an image analysis algorithm for HER2 or ER/PR (i.e., HER2 or | Same |
| | the image. The image will be analyzed by an image analysis algorithm for HER2 or ER/PR (i.e., HER2 or ER/PR). | |
| Method of Evaluation | Pathologists select regions on the digital slide images that will be automatically evaluated by an image analysis algorithm for HER2 or ER/PR (i.e., HER2 or ER/PR). | Same |
{6}
| | ER/PR Image Analysis) | |
| --- | --- | --- |
| DeviceComponents | System components consist of anautomated digital microscope slide scanner, computer, color monitor, keyboard, image analysis software and digital pathology information management software | Same as predicate device with the exception of the following component updates: • Slide Autoloader • Stages • Macro and Area Camera • Motion Control • Controller Software • Console Software • eSlide Manager Software |
| Image Acquisition | Slide scanner based on line scanning technology | Same |
# I. Special Control/Guidance Document Referenced (if applicable):
None
# J. Performance Characteristics:
# 1. Analytical Performance:
The performance of the ScanScope AT Turbo and CS2 models were evaluated for 1) system accuracy (method comparison) 2) inter-system precision and 3) intra-system precision. The performance studies were conducted using three (3) ScanScope AT Turbo systems, three (3) ScanScope CS2 systems (Candidate Devices) and one (1) of the original FDA cleared ScanScope XT systems (Predicate Device). Two board certified pathologists participated in the accuracy studies, and provided the annotations that were used in the precision studies. All study scanning was completed by a qualified scanning professional in a simulated laboratory environment.
The sample sizes for the Method Comparison and Precision Studies were based on the number of samples required to obtain the appropriate confidence in the data.
# a. Accuracy:
The performance of the ScanScope AT Turbo and CS2 models were evaluated in comparison with the predicate device ScanScope XT systems using malignant breast cancer samples. The slides were scanned and analyzed by the predicate or ScanScope AT Turbo or CS2 models. Board certified pathologists provided the annotations. The data was recorded into contingency tables, and further analyzed for statistical significance based on the pre-established acceptance criteria. For the HER2 Method Comparison Study, 120 slides were obtained for the study, however 1 of the slide was removed for not having a malignant tumor yielding a total sample size of $n = 119$ . The results are shown in Table 1.
{7}
Table 1: Method comparison contingency table for HER2 distribution between the ScanScope XT (Predicate Device), the ScanScope AT Turbo, and the Scanscope CS2 (Candidate Devices).
| Scanscope XT | Scanscope AT Turbo | | | Scanscope CS2 | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | 0/1+ | 2+ | 3+ | 0/1+ | 2+ |
| | 0/1+ | 31 | 0 | 0 | 31 | 0 |
| | 2+ | 3 | 29 | 5 | 1 | 30 |
| | 3+ | 0 | 3 | 48 | 0 | 2 |
| | Total | 34 | 32 | 53 | 32 | 32 |
| | PA 95% CI | 90.70% (84.1, 95.3) | | | 92.40% (86.1, 96.5) | |
| PPA 95% CI | 96.60% (90.4, 99.3) | | | 98.90% (93.8, 99.9) | | |
| NPA 95% CI | 100% (88.8, 100) | | | 100% (88.8, 100) | | |
For the ER Method Comparison Study, 120 slides were obtained for the study, however 2 of the slides were removed for not having a malignant tumor yielding a total sample size of $n = 118$ . The results are shown in Table 2.
Table 2: Method comparison contingency table for ER distribution at two different percent positive thresholds between the ScanScope XT (Predicate Device), the ScanScope AT Turbo, and the Scanscope CS2 (Candidate Devices).
| Scanscope XT | Scanscope AT Turbo | | | | Scanscope CS2 | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | 1% cutoff | | 10% cutoff | | 1% cutoff | | 10% cutoff |
| | | <1% | 1-100% | 0-10% | 11-100% | <1% | 1-100% | 0-10% |
| | <1% | 45 | 2 | | 42 | 5 | | |
| | 1-100% | 3 | 68 | | | 5 | | 66 |
| | 0-10% | | 59 | 2 | | 59 | | |
| | 11-100% | | | 1 | | | 56 | 1 |
| | PA 95% CI | 95.80% (90.4, 98.6) | | 97.50% (92.8, 99.5) | | 91.50% (85.0, 95.9) | | 97.50% (92.8, 99.5) |
| PPA 95% CI | 95.80% (88.1, 99.1) | | 98.20% (90.6, 99.9) | | 93% (84.3, 97.7) | | 98.20% (90.6, 99.9) | |
| NPA 95% CI | 95.70% (85.5, 99.5) | | 96.70% (88.7, 99.6) | | 89.40% (76.9, 96.5) | | 96.70% (88.7, 99.6) | |
For the PR Method Comparison Study, 120 cases were obtained and included in the study $(n = 120)$ . The results are shown in Table 3.
{8}
Table 3: Method comparison contingency table for PR distribution at two different percent positive thresholds between the ScanScope XT (Predicate Device), the ScanScope AT Turbo, and the Scanscope CS2 (Candidate Devices).
| Scanscope AT Turbo | | | | | Scanscope CS2 | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | 1% cutoff | | 10% cutoff | | 1% cutoff | | 10% cutoff | |
| | <1% | 1-100% | 0-10% | 11-100% | <1% | 1-100% | 0-10% | 11-100% |
| <1% | 44 | 0 | | 44 | 0 | | | |
| 1-100% | 1 | 75 | | | 1 | | | 75 |
| 0-10% | | 53 | 0 | | 51 | 2 | | |
| 11-100% | | | 1 | | | 66 | 1 | 66 |
| PA95% CI | 99.20% (95.4, 99.9) | | 99.20% (95.4, 99.98) | | 99.20% (95.4, 99.9) | | 97.50% (92.9, 99.5) | |
| PPA95% CI | 98.70% (92.9, 99.9) | | 98.50% (92.0, 99.9) | | 99% (92.9, 99.9) | | 98.50% (92.0, 99.6) | |
| NPA95% CI | 100% (92.0, 100) | | 100% (93.3, 100) | | 100% (92.0, 100) | | 96.20% (87.0, 99.5) | |
# b. Precision/Reproducibility:
For the ScanScope AT Turbo HER2 Precision Studies, 40 slides were obtained, however 1 of the slides was removed for not having visible tumor on the slide yielding a sample size of $n = 39$ . The results are shown in Table 4 and Table 5.
Table 4: Inter-system precision contingency tables for HER2 distribution in three ScanScope AT Turbo devices
| | Scanscope AT Turbo | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | System 2 | | | System 3 | | | System 3 | | |
| | | 0/1+ | 2+ | 3+ | 0/1+ | 2+ | 3+ | 0/1+ | 2+ | 3+ |
| System 1 | 0/1+ | 10 | 0 | 0 | 10 | 0 | 0 | | | |
| | 2+ | 0 | 7 | 0 | 0 | 7 | 0 | | | |
| | 3+ | 0 | 0 | 22 | 0 | 0 | 22 | | | |
| System 2 | 0/1+ | 100% (91.0, 100) | 100% (91.0, 100) | 10 | 0 | 0 | | | | |
| | 2+ | | | | | 0 | 7 | 0 | | |
| | 3+ | | | | | 0 | 0 | 22 | | |
| PA 95% CI | | | | | | 100% (91.0, 100) | | | | |
{9}
Table 5: Intra-system precision contingency tables for HER2 distribution in three scans from the same ScanScope AT Turbo device
| | Scanscope AT Turbo | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Run 2 | | | Run 3 | | | Run 3 | | |
| | | | 0/1+ | 2+ | 3+ | 0/1+ | 2+ | 3+ | 0/1+ | 2+ | 3+ |
| Scanscope AT Turbo | Run 1 | 0/1+ | 10 | 0 | 0 | 10 | 0 | 0 | | | |
| | | 2+ | 0 | 7 | 0 | 0 | 7 | 0 | | | |
| | | 3+ | 0 | 0 | 22 | 0 | 0 | 22 | | | |
| | Run 2 | 0/1+ | 100% (91.0, 100) | 100% (91.0, 100) | 10 | 0 | 0 | | | | |
| | | 2+ | | | | | 0 | 7 | 0 | | |
| | | 3+ | | | | | 0 | 0 | 22 | | |
| | PA 95% CI | | 100% (91.0, 100) | | | 100% (91.0, 100) | | | 100% (91.0, 100) | | |
The ScanScope CS2 HER2 Precision studies included 40 slides $(n = 40)$ . The results are shown in Table 6 and Table 7.
Table 6: Inter-system precision contingency tables for HER2 distribution in three ScanScope CS2 devices
| | Scanscope CS2 | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | System 2 | | | System 3 | | | System 3 | | |
| | | | 0/1+ | 2+ | 3+ | 0/1+ | 2+ | 3+ | 0/1+ | 2+ | 3+ |
| Scanscope CS2 | System 1 | 0/1+ | 9 | 1 | 0 | 10 | 0 | 0 | | | |
| | | 2+ | 0 | 14 | 0 | 0 | 14 | 1 | | | |
| | | 3+ | 0 | 0 | 16 | 0 | 0 | 15 | | | |
| | System 2 | 0/1+ | 97.5% (86.8, 99.9) | 97.5% (86.8, 99.9) | 9 | 1 | 0 | | | | |
| | | 2+ | | | | | 0 | 14 | 1 | | |
| | | 3+ | | | | | 0 | 0 | 15 | | |
| | PA 95% CI | | 97.5% (86.8, 99.9) | | | 97.5% (86.8, 99.9) | | | 95% (83.1, 99.4) | | |
Table 7: Intra-system precision contingency tables for HER2 distribution in three scans from the same ScanScope CS2 device
| | Scanscope CS2 | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Run 2 | | | Run 3 | | | Run 3 | | |
| | | | 0/1+ | 2+ | 3+ | 0/1+ | 2+ | 3+ | 0/1+ | 2+ | 3+ |
| Scanscope CS2 | Run1 | 0/1+ | 10 | 0 | 0 | 10 | 0 | 0 | | | |
| | | 2+ | 0 | 14 | 0 | 0 | 14 | 0 | | | |
| | | 3+ | 0 | 1 | 15 | 0 | 0 | 16 | | | |
| | Run2 | 0/1+ | 97.5% (86.8, 99.9) | 100% (91.2, 100) | 10 | 0 | 0 | | | | |
| | | 2+ | | | | | 0 | 14 | 0 | | |
| | | 3+ | | | | | 0 | 1 | 15 | | |
| | PA 95% CI | | 97.5% (86.8, 99.9) | | | 100% (91.2, 100) | | | 97.5% (86.8, 99.9) | | |
For the ScanScope AT Turbo ER Precision Study, 80 slides were obtained and included
{10}
$(n = 80)$ . The results are shown in Table 8 and Table 9.
Table 8: Inter-system precision contingency tables for ER distribution at two different percent positive thresholds in three ScanScope AT Turbo devices
| Scanscope AT Turbo | 1% cutoff | Scanscope AT Turbo | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | System 2 | | System 3 | | System 3 | |
| | | | <1% | 1-100% | <1% | 1-100% | <1% | 1-100% |
| System 1 | <1% | | 28 | 1 | 28 | 1 | | |
| | 1-100% | | 4 | 47 | 4 | 47 | | |
| System 2 | <1% | | 93.75% (86.0, 97.9) | 93.75% (86.0, 97.9) | 31 | 1 | | |
| | 1-100% | | | | | 1 | 47 | |
| PA 95% CI | | | 93.75% (86.0, 97.9) | | 97.5% (91.3, 99.7) | | | |
| Scanscope AT Turbo | 10% cutoff | Scanscope AT Turbo | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | System 2 | | System 3 | | System 3 | |
| | | | 0-10% | 11-100% | 0-10% | 11-100% | 0-10% | 11-100% |
| System 1 | 0-10% | 44 | 0 | 44 | 0 | | | |
| | 11-100% | 1 | 35 | 1 | 35 | | | |
| System 2 | 0-10% | 98.75% (93.2, 99.9) | 98.75% (93.2, 99.97) | 45 | 1 | | | |
| | 11-100% | | | | 1 | 33 | | |
| PA 95% CI | | | 98.75% (93.2, 99.9) | | 97.5% (91.3, 99.7) | | | |
Table 9: Intra-system precision contingency tables for ER distribution at two different percent positive thresholds for three scans from the same ScanScope AT Turbo device
| Scanscope AT Turbo | 1% cutoff | Scanscope AT Turbo | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Run 2 | | Run 3 | | Run 3 | |
| | | | <1% | 1-100% | <1% | 1-100% | <1% | 1-100% |
| Run 1 | <1% | 29 | 0 | 27 | 2 | | | |
| | 1-100% | 1 | 50 | 2 | 49 | | | |
| Run 2 | <1% | 98.75% (93.2, 99.7) | 95% (87.7, 98.6) | 27 | 2 | | | |
| | 1-100% | | | | 3 | 48 | | |
| PA 95% CI | | | 98.75% (93.2, 99.7) | | 93.75% (86.0, 97.9) | | | |
| Scanscope AT Turbo | 10% cutoff | Scanscope AT Turbo | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Run 2 | | Run 3 | | Run 3 | |
| | | | 0-10% | 11-100% | 0-10% | 11-100% | 0-10% | 11-100% |
| Run 1 | 0-10% | 45 | 0 | 45 | 0 | | | |
| | 11-100% | 0 | 35 | 0 | 35 | | | |
| Run 2 | 0-10% | 100% (95.5, 100) | 100% (95.5, 100) | 45 | 0 | | | |
| | 11-100% | | | | 0 | 35 | | |
| PA 95% CI | | | 100% (95.5, 100) | | 100% (95.5, 100) | | | |
For the Scanscope CS2 ER Precision Study, 40 slides were obtained, however 2 slides were removed for not having a malignant tumor yielding a sample size of $n = 38$ . The
{11}
results are shown in Table 10 and Table 11.
Table 10: Inter-system precision contingency tables for ER distribution at two different percent positive thresholds in three ScanScope CS2 devices
| 1% cutoff | | | Scanscope CS2 | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | System 2 | | System 3 | | System 3 | | | |
| | | | <1% | 1-100% | <1% | 1-100% | <1% | 1-100% |
| System 1 | <1% | 12 | 1 | 13 | 0 | | | |
| | 1-100% | 0 | 25 | 0 | 25 | | | |
| System 2 | <1% | 97.50% (86.2, 99.9) | 100% (90.7, 100) | 12 | 1 | | | |
| | 1-100% | | | | 0 | 25 | | |
| PA 95% CI | | | | | 97.50% (86.2, 99.9) | | | |
| 10% cutoff | | | Scanscope CS2 | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | System 2 | | System 3 | | System 3 | | | |
| | | | 0-10% | 11-100% | 0-10% | 11-100% | 0-10% | 11-100% |
| System 1 | 0-10% | 16 | 1 | 16 | 1 | | | |
| | 11-100% | 0 | 21 | 0 | 21 | | | |
| System 2 | 0-10% | 97.37% (86.2, 99.9) | 97.37% (90.7, 100) | 15 | 0 | | | |
| | 11-100% | | | | 0 | 23 | | |
| PA 95% CI | | | | | 100% (86.2, 99.9) | | | |
Table 11: Intra-system precision contingency tables for ER distribution at two different percent positive thresholds for three scans from the same ScanScope CS2 device
| 1% cutoff | | | Scanscope CS2 | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Run 2 | | Run 3 | | Run 3 | | | |
| | | | <1% | 1-100% | <1% | 1-100% | <1% | 1-100% |
| Run 1 | <1% | 13 | 0 | 13 | 0 | | | |
| | 1-100% | 0 | 25 | 0 | 25 | | | |
| Run 2 | <1% | 100% (90.7, 100) | 100% (90.7, 100) | 13 | 0 | | | |
| | 1-100% | | | | 0 | 25 | | |
| PA 95% CI | | | | | 100% (90.7, 100) | | | |
| 10% cutoff | | | Scanscope CS2 | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Run 2 | | Run 3 | | Run 3 | | | |
| | | | 0-10% | 11-100% | 0-10% | 11-100% | 0-10% | 11-100% |
| Run 1 | 0-10% | 16 | 0 | 16 | 0 | | | |
| | 11-100% | 0 | 22 | 0 | 22 | | | |
| Run 2 | 0-10% | 100% (90.7, 100) | 100% (90.7, 100) | 16 | 0 | | | |
| | 11-100% | | | | 0 | 22 | | |
| PA 95% CI | | | | | 100% (90.7, 100) | | | |
{12}
For the ScanScope AT Turbo PR Precision Study, 40 slides were obtained and included $(n = 40)$ . The results are shown in Table 12 and Table 13.
Table 12: Inter-system precision contingency tables for PR distribution at two different percent positive thresholds in three ScanScope AT Turbo devices
| 1% cutoff | Scanscope AT Turbo | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | System 2 | | System 3 | | System 3 | |
| | | <1% | 1-100% | <1% | 1-100% | <1% | 1-100% |
| System 1 | <1% | 12 | 0 | 12 | 0 | | |
| | 1-100% | 0 | 28 | 1 | 27 | | |
| System 2 | <1% | 100% (91.2, 100) | 97.5% (86.8, 99.9) | 12 | 1 | | |
| | 1-100% | | | | 0 | 27 | |
| PA 95% CI | | | | | | 97.5% (86.8, 99.9) | |
| 10% cutoff | Scanscope AT Turbo | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | System 2 | | System 3 | | System 3 | |
| | | 0-10% | 11-100% | 0-10% | 11-100% | 0-10% | 11-100% |
| System 1 | 0-10% | 16 | 0 | 16 | 0 | | |
| | 11-100% | 0 | 24 | 0 | 24 | | |
| System 2 | 0-10% | 100% (91.2, 100) | 100% (91.2, 100) | 16 | 0 | | |
| | 11-100% | | | | 0 | 24 | |
| PA 95% CI | | | | | | 100% (91.2, 100) | |
Table 13: Intra-system precision contingency tables for PR distribution at two different percent positive thresholds for three scans from the same ScanScope AT Turbo device
| 1% cutoff | Scanscope AT Turbo | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Run 2 | | Run 3 | | Run 3 | |
| | | <1% | 1-100% | <1% | 1-100% | <1% | 1-100% |
| Run 1 | <1% | 12 | 0 | 12 | 0 | | |
| | 1-100% | 0 | 28 | 0 | 28 | | |
| Run 2 | <1% | 100% (91.2, 100) | 100% (91.2, 100) | 12 | 0 | | |
| | 1-100% | | | | 0 | 27 | |
| PA 95% CI | | | | | | 100% (91.2, 100) | |
| 10% cutoff | Scanscope AT Turbo | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Run 2 | | Run 3 | | Run 3 | |
| | | 0-10% | 11-100% | 0-10% | 11-100% | 0-10% | 11-100% |
| Run 1 | 0-10% | 16 | 0 | 16 | 0 | | |
| | 11-100% | 0 | 24 | 0 | 24 | | |
| Run 2 | 0-10% | 100% (91.2, 100) | 100% (91.2, 100) | 16 | 0 | | |
| | 11-100% | | | | 0 | 24 | |
| PA 95% CI | | | | | | 100% (91.2, 100) | |
{13}
For the Scanscope CS2 PR Precision Study, 40 slides were obtained and included (n =40). The results are shown in Table 14 and Table 15.
Table 14: Inter-system precision contingency tables for PR distribution at two different percent positive thresholds in three ScanScope CS2 devices
| 1% cutoff | Scanscope CS2 | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | System 2 | | System 3 | | System 3 | |
| | | | <1% | 1-100% | <1% | 1-100% | <1% | 1-100% |
| System 1 | <1% | 16 | 0 | 16 | 0 | | | |
| | 1-100% | 0 | 24 | 0 | 24 | | | |
| System 2 | <1% | 100% (91.2, 100) | 100% (91.2, 100) | 16 | 0 | | | |
| | 1-100% | | | | 0 | 24 | | |
| PA 95% CI | | | | | 100% (91.2, 100) | | | |
| 10% cutoff | Scanscope CS2 | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | System 2 | | System 3 | | System 3 | |
| | | | 0-10% | 11-100% | 0-10% | 11-100% | 0-10% | 11-100% |
| System 1 | 0-10% | 18 | 0 | 18 | 0 | | | |
| | 11-100% | 0 | 22 | 0 | 22 | | | |
| System 2 | 0-10% | 100% (91.2, 100) | 100% (91.2, 100) | 18 | 0 | | | |
| | 11-100% | | | | 0 | 22 | | |
| PA 95% CI | | | | | 100% (91.2, 100) | | | |
Table 15: Intra-system precision contingency tables for PR distribution at two different percent positive thresholds for three scans from the same ScanScope CS2 device
| 1% cutoff | Scanscope CS2 | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Run 2 | | Run 3 | | Run 3 | |
| | | | <1% | 1-100% | <1% | 1-100% | <1% | 1-100% |
| Run 1 | <1% | 16 | 0 | 16 | 0 | | | |
| | 1-100% | 0 | 24 | 0 | 24 | | | |
| Run 2 | <1% | 100% (91.2, 100) | 100% (91.2, 100) | 16 | 0 | | | |
| | 1-100% | | | | 0 | 24 | | |
| PA 95% CI | | | | | 100% (91.2, 100) | | | |
| 10% cutoff | Scanscope CS2 | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Run 2 | | Run 3 | | Run 3 | |
| | | | 0-10% | 11-100% | 0-10% | 11-100% | 0-10% | 11-100% |
| Run 1 | 0-10% | 18 | 0 | 18 | 0 | | | |
| | 11-100% | 0 | 22 | 0 | 22 | | | |
| Run 2 | 0-10% | 100% (91.2, 100) | 100% (91.2, 100) | 18 | 0 | | | |
| | 11-100% | | | | 0 | 22 | | |
| PA 95% CI | | | | | 100% (91.2, 100) | | | |
{14}
c. Linearity:
Not applicable
d. Carryover:
Not applicable
e. Interfering Substances:
Not applicable
2. Other Supportive Instrument Performance Data Not Covered Above:
None
K. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
L. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
15
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.