MONOCLONAL RABBIT ANTI HUMAN ESTROGEN RECEPTOR ALPHA ANTIBODY CLONE SP1, MODEL M3634
K081286 · Dako North America, Inc. · MYA · May 8, 2009 · Hematology
Device Facts
Record ID
K081286
Device Name
MONOCLONAL RABBIT ANTI HUMAN ESTROGEN RECEPTOR ALPHA ANTIBODY CLONE SP1, MODEL M3634
Applicant
Dako North America, Inc.
Product Code
MYA · Hematology
Decision Date
May 8, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.1860
Device Class
Class 2
Indications for Use
Monoclonal Rabbit Anti-Human Estrogen Receptor α (ER α) antibody, Clone SP1, may be used in the semi-quantitative detection of human estrogen receptor in formalin-fixed, paraffin-embedded tissue sections of human breast cancer by immunohistochemistry. The information gained by this assay can aid in assessing the likelihood of response to therapy as well as in the prognosis and management of breast cancer patients. Clinical interpretation of any positive staining or its absence should be complemented by morphological and histological studies with proper controls. Evaluations should be within the context of the patient’s clinical history and other diagnostic tests by a qualified pathologist.
Device Story
Monoclonal Rabbit Anti-Human Estrogen Receptor a (Clone SP1) is an immunohistochemical (IHC) assay; identifies ER expression in formalin-fixed, paraffin-embedded breast cancer tissue. Operates via antibody-antigen binding to nuclear estrogen receptors; requires detection chemistry for visualization. Used in clinical pathology laboratories; interpreted by qualified pathologists. Output provides semi-quantitative staining results; assists clinicians in determining breast cancer prognosis and potential response to endocrine therapy.
Clinical Evidence
Clinical performance supported by a published study (Cheang et al., 2006) evaluating 4,150 breast cancer samples. Study demonstrated SP1 positivity rate of 69.5% vs 63.1% for 1D5. Ten-year Breast Cancer Specific Survival (BCSS) for SP1 ≥ 1% nuclear staining was 76.5% (95% CI: 73.9-79.0%) vs 65% (95% CI: 62.6-68%) for < 1% staining. A bridging study (n=27) confirmed 100% agreement between the publication's protocol and the Dako SP1 protocol. Analytical performance included precision/reproducibility studies showing 100% agreement across sites.
Technological Characteristics
Semi-quantitative IHC assay; utilizes monoclonal rabbit anti-human ERα antibody (Clone SP1) in concentrated liquid form with 0.015mol/L sodium azide. Detection system uses dextran-polymer backbone conjugated with secondary goat anti-rabbit antibody and horseradish peroxidase. Requires Dako Autostainer for processing. Results interpreted via light microscopy.
Indications for Use
Indicated for the semi-quantitative detection of estrogen receptor in formalin-fixed, paraffin-embedded breast cancer tissue sections to aid in therapy response assessment, prognosis, and management of breast cancer patients.
Regulatory Classification
Identification
Immunohistochemistry test systems (IHC's) are in vitro diagnostic devices consisting of polyclonal or monoclonal antibodies labeled with directions for use and performance claims, which may be packaged with ancillary reagents in kits. Their intended use is to identify, by immunological techniques, antigens in tissues or cytologic specimens. Similar devices intended for use with flow cytometry devices are not considered IHC's.
Special Controls
(2) Class II (special control, guidance document: “FDA Guidance for Submission of Immunohistochemistry Applications to the FDA,” Center for Devices and Radiologic Health, 1998). These IHC's are intended for the detection and/or measurement of certain target analytes in order to provide prognostic or predictive data that are not directly confirmed by routine histopathologic internal and external control specimens. These IHC's provide the pathologist with information that is ordinarily reported as independent diagnostic information to the ordering clinician, and the claims associated with these data are widely accepted and supported by valid scientific evidence. Examples of class II IHC's are those intended for semiquantitative measurement of an analyte, such as hormone receptors in breast cancer.
Predicate Devices
Monoclonal Mouse ER 1D5/ER-2-123 in the Dako ER/PR pharmDx™ Kit (K042884)
K120663 — MONOCLONAL RABBIT ANTI-HUMAN ESTROGEN RECEPTOR (ER) A, CLONE EPI · Dako North America, Inc. · Feb 12, 2013
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k081286
B. Purpose for Submission:
New device
C. Measurand:
Estrogen Receptor on formalin-fixed paraffin-embedded breast cancer specimens
D. Type of Test:
Semi-quantitative, immunohistochemical
E. Applicant:
Dako North America, Inc
F. Proprietary and Established Names:
Dako Monoclonal Rabbit Anti-Human Estrogen Receptor, Clone SP1
G. Regulatory Information:
1. Regulation section:
21 CFR 864.1860, Immunohistochemistry reagents and kits
2. Classification:
Class II
3. Product code:
MYA, Immunohistochemistry Antibody Assay, Estrogen Receptor
4. Panel:
Pathology 88
H. Intended Use:
1. Intended use(s):
Monoclonal Rabbit Anti-Human Estrogen Receptor α (ER α) antibody, Clone SP1, may be used in the semi-quantitative detection of human estrogen receptor in formalin-fixed, paraffin-embedded tissue sections of human breast cancer by immunohistochemistry. The information gained by this assay can aid in assessing the likelihood of response to therapy as well as in the prognosis and management of breast cancer patients.
Clinical interpretation of any positive staining or its absence should be complemented by morphological and histological studies with proper controls.
Evaluations should be within the context of the patient’s clinical history and other diagnostic tests by a qualified pathologist.
2. Indication(s) for use:
Not applicable.
3. Special conditions for use statement(s):
For prescription use only.
4. Special instrument requirements:
Dako Autostainer
I. Device Description:
Dako Monoclonal Rabbit Anti-Human Estrogen Receptor α antibody, Clone SP1, is a semi-quantitative immunohistochemical (IHC) assay kit to identify estrogen receptor (ER) expression in normal and neoplastic tissues routinely processed and paraffin-
{1}
embedded. The monoclonal rabbit antibody is provided in concentrated liquid form as cell culture supernatant and contains 0.015mol/L sodium azide.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
Dako ER/PR pharmDx™ Kit-Monoclonal Mouse ER 1D5/ER-2-123 reagent only
2. Predicate $510(\mathbf{k})$ number(s):
k042884
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Antibody type | Monoclonal | Same |
| Intended Use | Semi-quantitative detection of ER | Same |
| Technology | Immunohistochemistry | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Clone (ER) | 1D5/ER-2-123 | SP1 |
# K. Standard/Guidance Document Referenced (if applicable):
Guidance for Submission of Immunohistochemistry Applications to the FDA; Final
# L. Test Principle:
Monoclonal Rabbit Anti-Human Estrogen Receptor $\alpha$ (ER $\alpha$ ) antibody, Clone SP1, may be used in the semi-quantitative detection of human estrogen receptor in tissue sections of human breast cancer by immunohistochemistry. The information gained by this assay can aid in assessing the likelihood of response to therapy as well as in the prognosis and management of breast cancer patients. Steroid receptors exhibit a high affinity and specificity for their ligands. The human estrogen receptor (ER) is a dimeric protein of $65\mathrm{kDa}$ located primarily in the nucleus and belongs to a class of trans-acting proteins which stimulate transcription by binding to specific DNA elements, also known as hormone response elements. Through binding estrogen, the ER is induced to stimulate gene transcription, hence is also known as an inducible enhancer factor.
Immunohistochemical staining is performed on routinely processed, paraffin-embedded specimens. Immunohistochemistry is a well established, widely accepted laboratory methodology. The assay detects the presence of Estrogen receptor (ER) through first, the binding of an antibody to the ER antigen and second, visualization of the bound primary antibody through a reagent based on dextran technology. This reagent consists of both secondary goat anti-rabbit antibody molecules and horseradish peroxidase molecules linked to a common dextran polymer backbone, thus eliminating the need for sequential application of link antibody and peroxidase conjugate. The enzymatic conversion of the subsequently added chromogen results in formation of a visible reaction product at the antigen site. The specimen is then counterstained and coverslipped. Results are interpreted using a light microscope.
{2}
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision:
Serial sections from each of three different formalin-fixed paraffin embedded blocks of breast carcinoma were collected for testing from the Dako tissue bank. Test slides from the same specimen must not display staining intensity deviation of ± 0.5 grade for a total of 1 grade variation on a 0-4 scale.
Intra-run Precision: Three slides from each tissue block were stained with Monoclonal Rabbit Anti-Human ERα Clone SP1 by three analysts. Concurrently one slide from each block was stained with a negative control reagent.
Inter-run Precision: One slide from each tissue block was stained on two additional days by one analyst. Concurrently, one slide from each tissue block was stained with a negative control reagent.
The precision experiment results met the acceptance criteria described above. Additionally, the estimated percent of tumor cells stained was evaluated and all results were in agreement.
Reproducibility:
Fifteen breast carcinoma specimens (5 Negative, 1 Low Positive, and 9 Positive) were stained with Dako Monoclonal Rabbit Anti-Human ERα, Clone SP1 and scored at 3 sites. Between each site comparison pair, 100% (95% C.I. 78.2-100%) overall agreement was obtained.
b. Linearity/assay reportable range:
Not applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Positive and negative controls should be performed with each staining run. The pathologist is responsible for assuring that the assay is performing properly.
d. Detection limit:
Not applicable.
e. Analytical specificity:
A total of 87 formalin-fixed and paraffin-embedded tissues covering a wide range of normal human tissue types were tested with the ER antibody. The antibody demonstrated negative immunoreactivity with most tissues. Positive immunoreactivity was noted with some normal tissues which are typically positive, like uterus, ovary and ductal epithelial cells of the breast.
f. Assay cut-off:
A positive staining result is defined as more than 1% of tumor cells with stained nuclei of any intensity.
2. Comparison studies:
{3}
a. Method comparison with predicate device:
Estrogen receptor status was evaluated at two sites on a total of 228 cases using formalin-fixed paraffin-embedded cases of breast carcinoma. ERα, SP 1 testing was performed with EnVision™ FLEX and scored with positive staining result defined as more than 1% of tumor cells with stained nuclei of any intensity. ER 1D5/2-123 was performed using the ER/PR pharmDx™ Kit and scored using the Allred scoring guideline. Method comparison data are presented below:
Study Site 1:
| | ER Component of ER/PR pharmDx | | |
| --- | --- | --- | --- |
| | | Positive | Negative |
| ER Clone SP1 | Positive | 69 | 6 |
| | Negative | 0 | 34 |
| | Total | 69 | 40 |
Positive Percent Agreement = 69/69 = 100% (95% CI: 94.8-100%)
Negative Percent Agreement = 34/40 = 85% (95% CI: 70.2-94.3%)
Total Percent Agreement = 103/109 = 94.5% (95% CI: 88.4-97.95%)
Study Site 2:
| | ER Component of ER/PR pharmDx | | |
| --- | --- | --- | --- |
| | | Positive | Negative |
| ER Clone SP1 | Positive | 27 | 1 |
| | Negative | 7 | 84 |
| | Total | 34 | 85 |
Positive Percent Agreement = 27/34 = 79.4% (95% CI: 62.1-91.3%)
Negative Percent Agreement = 84/85 = 98.8% (95% CI: 93.6-99.9%)
Total Percent Agreement = 111/119 = 93.3% (95% CI: 87.2-97.1%)
Clone SP1 has been noted in the literature to have a higher sensitivity than Clone 1D5.
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
A study published in the Journal of Clinical Oncology was used to establish clinical performance of Dako Monoclonal Rabbit Anti-Human Estrogen Receptor α (ER α) antibody, Clone SP1. Estrogen receptor status was evaluated in 4,150 breast cancer patient samples on formalin-fixed, paraffin-embedded tissue samples with anti-ERα, Clone SP1 and anti-ER α, Clone
{4}
1D5. The rate of positivity for SP1 was 69.5% compared to 63.1% for 1D5. For SP1 the weighted estimate for ten-year Breast Cancer Specific Survival (BCSS) for ≥ 1% nuclear staining in tumor cells is 76.5% (95% C.I.: 73.9-79.0%) and for < 1% nuclear staining in tumor cells is 65% (95% C.I.: 62.68%).
A bridging study was performed to establish that there was no difference between the immunostaining protocol cited in the publication and the protocol used to stain Dako anti-ERα, Clone SP1. Twenty-seven formalin-fixed paraffin embedded breast carcinoma tissue specimens with percent tumor cells stained ranging from 1% to 100% were tested with both protocols. Results showed 100% agreement between the two protocols and the published study can therefore be used to support the clinical performance of the Dako anti-ERα, Clone SP1 (see below):
| | Protocol 2 | | |
| --- | --- | --- | --- |
| | | Positive | Negative |
| Protocol 1 | Positive | 23 | 0 |
| | Negative | 0 | 4 |
| | Total | 23 | 4 |
Total Percent Agreement: 27/27 = 100% (95% CI = 87.2-100%)
¹Cheang MCU, Treaba DO, Spears CH, Olivotto IA, Bajdik CD, Chia SK, Goldstein LC, Gelmon KA, Huntsmann D, Gilks CB, Nielson TO, Gown AM. Immunohistochemistry detection using the new rabbit monoclonal antibody SP1 of estrogen receptor in breast cancer is superior to mouse monoclonal antibody 1D5 in predicting survival. J Clin Oncol 2006;24:5637-5644.
4. Clinical cut-off:
A positive result is defined as nuclear staining in ≥ 1% of tumor cells.
5. Expected values/Reference range:
Not applicable.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.