CONFIRM ANTI-PROGESTORONE RECEPTOR (1E2) RABBIT MONOCLONAL PRIMARY ANTIBODY

K103818 · Ventana Medical Sytems, Inc. · MXZ · Nov 16, 2011 · Hematology

Device Facts

Record IDK103818
Device NameCONFIRM ANTI-PROGESTORONE RECEPTOR (1E2) RABBIT MONOCLONAL PRIMARY ANTIBODY
ApplicantVentana Medical Sytems, Inc.
Product CodeMXZ · Hematology
Decision DateNov 16, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.1860
Device ClassClass 2

Indications for Use

This antibody is intended for in vitro diagnostic (IVD) use. CONFIRM anti-Progesterone Receptor (PR) (1E2) Rabbit Monoclonal (IgG) Primary Antibody is intended for laboratory use for the qualitative detection of progesterone receptor (PR) antigen in sections of formalin-fixed, paraffin-embedded tissue on a VENTANA automated slide stainer with VENTANA detection kits and ancillary reagents. CONFIRM anti-PR (1E2) is directed against an epitope present on human progesterone receptor protein located in the nucleus of PR positive normal and neoplastic cells. CONFIRM anti-PR (1E2) is indicated as an aid in the management, prognosis, and prediction of hormone therapy for breast carcinoma. This product should be interpreted by a qualified pathologist in conjunction with histological examination, relevant clinical information, and proper controls. Prescription use only.

Device Story

Qualitative immunohistochemical (IHC) assay; detects progesterone receptor (PR) antigen in formalin-fixed, paraffin-embedded (FFPE) breast tissue sections. Uses rabbit monoclonal antibody (clone 1E2); applied via Ventana BenchMark XT or BenchMark ULTRA automated slide stainers. Process: primary antibody binds nuclear PR; biotin-conjugated secondary antibody binds primary; streptavidin-enzyme conjugate binds biotin; precipitating enzyme product visualized via light microscopy. Used in clinical pathology laboratories; interpreted by pathologists. Output: visual staining pattern in tissue; aids in breast cancer hormone therapy decision-making.

Clinical Evidence

Multi-site, multi-reader study comparing CONFIRM anti-PR (1E2) on BenchMark ULTRA and XT instruments against Dako FLEX anti-PgR (636) on Dako Autostainer Plus. Sample size: ~120 negative and 216 positive breast cancer cases. Primary endpoint: staining performance agreement. Results: positive, negative, and overall agreement rates were >85% for all instrument comparisons.

Technological Characteristics

Rabbit monoclonal antibody (clone 1E2) in 0.05M Tris-HCl, 2% carrier protein, 0.1% ProClin 300. Indirect IHC detection system using biotin-conjugated secondary antibody and streptavidin-enzyme conjugate. Automated staining on Ventana BenchMark XT/ULTRA platforms. Qualitative visual readout via light microscopy. Shelf life 18 months.

Indications for Use

Indicated for use as an aid in the management, prognosis, and prediction of hormone therapy for breast carcinoma patients.

Regulatory Classification

Identification

Immunohistochemistry test systems (IHC's) are in vitro diagnostic devices consisting of polyclonal or monoclonal antibodies labeled with directions for use and performance claims, which may be packaged with ancillary reagents in kits. Their intended use is to identify, by immunological techniques, antigens in tissues or cytologic specimens. Similar devices intended for use with flow cytometry devices are not considered IHC's.

Special Controls

(2) Class II (special control, guidance document: “FDA Guidance for Submission of Immunohistochemistry Applications to the FDA,” Center for Devices and Radiologic Health, 1998). These IHC's are intended for the detection and/or measurement of certain target analytes in order to provide prognostic or predictive data that are not directly confirmed by routine histopathologic internal and external control specimens. These IHC's provide the pathologist with information that is ordinarily reported as independent diagnostic information to the ordering clinician, and the claims associated with these data are widely accepted and supported by valid scientific evidence. Examples of class II IHC's are those intended for semiquantitative measurement of an analyte, such as hormone receptors in breast cancer.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k103818 B. Purpose for Submission: New device C. Measurand: Human progesterone receptor protein D. Type of Test: Qualitative E. Applicant: Ventana Medical Systems, Inc. F. Proprietary and Established Names: CONFIRM anti-Progesterone Receptor (PR) (1E2) Rabbit Monoclonal Primary Antibody G. Regulatory Information: 1. Regulation section: 21 CFR § 864.1860, Immunohistochemistry reagents and kits 2. Classification: Class II 3. Product code: MXZ, Immunohistochemistry Assay, Antibody, Progesterone Receptor 4. Panel: Pathology 88 H. Intended Use: 1. Intended use(s): CONFIRM anti-Progesterone Receptor (PR) (1E2) Rabbit Monoclonal Primary Antibody is intended for laboratory use for the qualitative detection of progesterone receptor (PR) antigen in sections of formalin-fixed, paraffin-embedded tissue on a Ventana automated slide stainer with Ventana detection kits and ancillary reagents. CONFIRM anti-PR (1E2) is directed against an epitope present on human PR protein located in the nucleus of PR positive normal and neoplastic cells. This product should be interpreted by a qualified pathologist in conjunction with histological examination, relevant clinical information, and proper controls. 2. Indication(s) for use: CONFIRM anti-PR (1E2) is indicated as an aid in the management, prognosis, and prediction of hormone therapy for breast carcinoma. 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: Ventana BenchMark XT or BenchMark ULTRA automated slide stainer I. Device Description: The Ventana CONFIRM anti-Progesterone Receptor (PR) (1E2) Rabbit Monoclonal Primary Antibody is a qualitative immunohistochemical assay used to identify the {1} progesterone receptor (PR) antigen in normal and neoplastic tissues sections that have been formalin-fixed and paraffin-embedded. These antibodies are available as ready-to-use reagents and can only be used on the Ventana Benchmark immunostainer platforms. The antibody is diluted in 0.05M Tris-HCl with 2% carrier protein, and 0.1% ProClin 300, a preservative. There is trace (~0.2%) fetal calf serum of U.S. origin from the stock solution. Total protein concentration of the reagent is approximately 10 mg/mL. Specific antibody concentration is approximately 1 µg/mL. It also contains a biotin conjugated secondary antibody formulation which recognizes rabbit or mouse immunoglobulins and a streptavidin-enzyme conjugate which binds to the biotin on the secondary antibody. CONFIRM anti-PR (1E2) is a rabbit monoclonal antibody produced as a cell culture supernatant. There is no known non-specific antibody reactivity observed in this product. ## J. Substantial Equivalence Information: 1. Predicate device name(s): Dako FLEX Monoclonal Mouse Anti-Human Progesterone Receptor Clone PGR636 2. Predicate 510(k) number(s): k020023 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Target | Human PR protein | Human PR protein | | Assay method | IHC with secondary detection | IHC with secondary detection | | Detection system | Indirect | Indirect | | Sample | Formalin-fixed paraffin-embedded breast cancer specimens | Formalin-fixed paraffin-embedded breast cancer specimens | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Antibody Type & clone | Monoclonal (Rabbit) 1E2 | Monoclonal (Mouse) 636 | | Assay format | Ventana BenchMark XT and BenchMark Ultra autostainers | Dako Autostainer | ## K. Standard/Guidance Document Referenced (if applicable): Guidance for Industry - "Guidance for Submission of Immunohistochemistry Applications to the FDA" ## L. Test Principle: CONFIRM anti-PR (1E2) assay is intended to be used on breast carcinoma specimens. The CONFIRM anti-PR (1E2) primary antibody binds to human progesterone receptor located in the nuclear region of normal and neoplastic cells. Automated immunohistochemistry staining is performed on formalin-fixed, paraffin- {2} embedded tissue. The antigen in tissue sections is demonstrated through several steps. The specific antibody binds to the antigen present in tissue sections. The bound primary antibody is located by a biotin conjugated secondary antibody formulation which recognizes rabbit or mouse immunoglobulins. This step is followed by the addition of a streptavidin-enzyme conjugate which binds to the biotin on the secondary antibody. The primary antibody-secondary antibody-avidin enzyme complex is visualized by using a precipitating enzyme generated product. This is readily detected by light microscopy. ## M. Performance Characteristics: ### 1. Analytical performance: #### a. Precision/Reproducibility: Studies to assess the intra-run (within-run, intra-day) and inter-run (day-to-day) analytical precision of CONFIRM anti-PR (1E2) antibody was conducted using two Ventana IHC stainers: BenchMark XT and BenchMark Ultra instruments and the iView DAB Detection Kit with biotin blocker reagents (A/B Blocker). In each study, nine slides were stained with the CONFIRM anti-PR (1E2) antibody and one slide was stained with Negative Control Rabbit Ig antibody from six individual breast carcinoma tissue blocks. Of the six tissues, two were PR negative, two were PR low expression and two were PR high expression based on a cutoff of <1% tumor cells staining for negative, 1-10% for low and >10% for high expression. A total of 30 slides were stained on each of the two IHC stainers - BenchMark XT and BenchMark Ultra instruments. An acceptance criteria of at least 90% overall agreement rates for positive and negative staining status was set by the sponsor. #### Intra-run Precision: Intra-run precision of CONFIRM anti-PR (1E2) antibody on BenchMark XT was 100% concordant for all six cases. The background was reported as 0 on 100% of the tissues stained. Intra-run precision of CONFIRM anti-PR (1E2) antibody on BenchMark ULTRA was 100% concordant for all six cases. Background was below 0.25 on 100% of the tissues stained. The acceptance criteria set by the sponsor were met in these studies. #### Inter-run Precision: Inter-run precision was assessed in five separate nonconsecutive runs conducted over a 20 day period on the same BenchMark XT. There were a total of 30 slides per run. The same testing configuration was also performed on a BenchMark ULTRA instrument. Inter-run precision of CONFIRM anti-PR (1E2) antibody on BenchMark XT was 100% concordant for all six cases. The background was reported as below 0.5 on 100% of the tissues stained. Inter-run precision of CONFIRM anti-PR (1E2) antibody on BenchMark ULTRA was 100% concordant for all six cases. Background was below 0.25 on 100% of the tissues stained. The acceptance criteria set by the sponsor were met in these studies. ### Intra-platform Reproducibility Four slides from six cases were stained with CONFIRM anti-PR (1E2) {3} antibody and one slide from each case was stained with Negative Control Rabbit Ig antibody on three separate BenchMark XT instruments. The same testing configuration was used for the BenchMark ULTRA instrument. The acceptance criteria set by the sponsor was as follows: The percentage of cells staining shall not vary in binning categories (as defined in protocol) on a minimum of 80% (positive) tissues tested. Background shall be ≤0.5 on a minimum of 90% of samples stained. This study met the acceptance criteria set by the sponsor and showed acceptable reproducibility with no variation in staining quality and patterns. b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): Controls: Positive and negative control slides should be stained with each staining run. The pathologist is responsible for assuring the proper performance of this test. Stability: Real-time stability tests and ship-stress tests were conducted to determine the shelf life of the reagent. Three lots of the reagent were tested. The acceptance criteria set by the sponsor were as follows: Positive cases – nuclear staining of 51% - 100% of the tumor cells and no more than 0.5 difference in staining intensity when compared to the Day 0 slide stain and less than or equal to 0.5+ non-specific staining. Background staining (0-4+) must not be greater than 0.5+. The acceptance criteria were met in these studies. Based on these studies, the expiration date of this reagent is set at 18 months. d. Detection limit: Not applicable e. Analytical specificity: A total of 90 formalin-fixed and paraffin-embedded tissues covering a wide range of normal human tissues types were tested with the CONFIRM anti-PR (1E2) antibody. The antibody demonstrated negative immunoreactivity with most tissues. Positive immunoreactivity was noted in breast, uterus, and cervix (uterine) tissues as expected. f. Assay cut-off: A negative staining result is defined as <1% tumor cells staining and a positive staining result is defined as ≥ 1% tumor cells staining of any intensity. 2. Comparison studies: a. Method comparison with predicate device: The Ventana CONFIRM anti-PR (1E2) antibody was compared to the predicate device Dako FLEX Monoclonal Mouse Anti-Human Progesterone Receptor Clone PGR636. A total of 336 breast carcinoma slides were tested in 3 different sites on the two Ventana platforms and the Dako Autostainer Plus. The acceptance criteria set by the sponsor were as follows: greater than 85% positive and negative agreement rates across all sites for CONFIRM anti-PR (1E2) staining 4 {4} on both the BenchMark ULTRA and XT compared to the Dako Autostainer Plus, with the lower bound 95% CI of at least 77% for all comparisons. CONFIRM anti-PR (1E2) on BenchMark ULTRA and FLEX anti-PR Clone PgR 636 on Dako Autostainer Plus | FLEX anti-PR Clone PgR 636 | | | | | --- | --- | --- | --- | | CONFIRM anti-PR (1E2) | Positive | Negative | Total | | Positive | 200 | 7 | 207 | | Negative | 9 | 104 | 113 | | Total | 209 | 111 | 320 | | Positive Percent Agreement | 200/209 | 95.7% (92.0-97.7) | (95% CI) | | Negative Percent Agreement | 104/111 | 93.7% (87.6-96.9) | (95% CI) | CONFIRM anti-PR (1E2) on BenchMark XT and FLEX anti-PR Clone PgR 636 on Dako Autostainer Plus | FLEX anti-PR Clone PgR 636 | | | | | --- | --- | --- | --- | | CONFIRM anti-PR (1E2) | Positive | Negative | Total | | Positive | 186 | 9 | 195 | | Negative | 18 | 100 | 118 | | Total | 204 | 109 | 313 | | Positive Percent Agreement | 186/204 | 91.2% (86.5-94.3) | (95% CI) | | Negative Percent Agreement | 100/109 | 91.7% (85.0-95.6) | (95% CI) | An additional assessment of the staining of the ductal carcinoma in-situ (DCIS) component was also performed. The primary objective of this study was to determine if the performance of the staining in areas of human breast carcinoma specimens using the CONFIRM anti-PR (1E2) with BenchMark ULTRA was equivalent to that using the FLEX anti-PR (636) with the Dako Autostainer Plus. All slides from the method comparison study that contained invasive tumor as well as DCIS component were assessed. There were no pre-set acceptance criteria. Of the PR positive cases, >99% of cases had positive staining in the DCIS component as well as the invasive component. CONFIRM anti-PR (1E2) on BenchMark ULTRA and FLEX anti-PR Clone PgR 636 on Dako Autostainer Plus | FLEX anti-PR Clone PgR 636 (Dako) | | | | | --- | --- | --- | --- | | CONFIRM anti-PR (1E2) | Positive | Negative | Total | | Positive | 97 | 3 | 100 | | Negative | 2 | 17 | 19 | | Total | 99 | 20 | 119 | | Positive Percent Agreement | 97/99 | 98.0% | | | Negative Percent Agreement | 17/20 | 85.0% | | {5} 6 CONFIRM anti-PR (1E2) on BenchMark XT and FLEX anti-PR Clone PgR 636 on Dako Autostainer Plus | FLEX anti-PR Clone PgR 636 (Dako) | | | | | --- | --- | --- | --- | | CONFIRM anti-PR (1E2) | Positive | Negative | Total | | Positive | 94 | 1 | 95 | | Negative | 5 | 19 | 24 | | Total | 99 | 20 | 119 | | Positive Percent Agreement | 94/99 | 94.9% | | | Negative Percent Agreement | 19/20 | 95.0% | | b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Not applicable N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...