INNOSENSE PELVIC FLOOR STIMULATION AND ELECTROMYOGRAPHY SYSTEM

K971527 · Empi · KPI · Jun 30, 1997 · Gastroenterology, Urology

Device Facts

Record IDK971527
Device NameINNOSENSE PELVIC FLOOR STIMULATION AND ELECTROMYOGRAPHY SYSTEM
ApplicantEmpi
Product CodeKPI · Gastroenterology, Urology
Decision DateJun 30, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5320
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The InnoSense System is indicated for acute and ongoing treatment of stress, urge or mixed urinary incontinence and where the following results may improve urinary control: Improvement of urethral sphincter closure, Strengthening of pelvic floor muscles, Inhibition of the detruser muscle through reflexive mechanisms. It is also indicated during incontinence treatment for assessing EMG activity of the pelvic floor and accessory muscles such as the abdominal or gluteal muscles.

Device Story

InnoSense Pelvic Floor Stimulation and Electromyography (EMG) System provides electrical stimulation and biofeedback for urinary incontinence. Device inputs include electrical signals from pelvic floor muscles via vaginal/rectal electrodes and accessory muscle activity via surface EMG electrodes. System features one-channel stimulation (0-100mA peak pulse intensity) and two-channel EMG sensing. Battery-powered (two 1.5V AA) unit includes LCD panel with low battery indicator. Used in clinical or home settings under prescription; operated by patient or clinician. Output consists of stimulation pulses and visual/data feedback of muscle activity. Clinicians use EMG data to assess muscle performance and guide treatment; stimulation aims to strengthen muscles and inhibit detrusor activity, potentially improving urinary control.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Battery-powered (2x 1.5V AA) external stimulator/EMG reader. One-channel stimulation (0-100mA peak pulse intensity); two-channel EMG sensing. Electrodes: vaginal, rectal, and surface EMG. LCD interface. Non-implanted electrical continence device.

Indications for Use

Indicated for patients with stress, urge, or mixed urinary incontinence requiring pelvic floor muscle strengthening, urethral sphincter closure improvement, or detrusor muscle inhibition. Also indicated for EMG assessment of pelvic floor and accessory (abdominal/gluteal) muscles during incontinence treatment.

Regulatory Classification

Identification

A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUN 30 1997 Image /page/0/Picture/1 description: The image shows the word "Empi" in a bold, stylized font. Above the word "Empi" is the number "K971527". To the right of the word "Empi" is the number "142" in a handwritten font. The image appears to be a product label or identifier. # Cost Effective Health Care Solutions # SUMMARY OF SAFETY AND EFFECTIVENESS InnoSense Pelvic Floor Stimulation and Electromyography System Date of Summary June 17, 1997 Page 1 of 2 Empi, Inc. 599 Cardigan Road St. Paul. Minnesota 55126-3965 USA 612-415-9000 FAX 612-415-7305 #### General Provisions A. | Submitter's Name: | Empi, Inc. | |----------------------|-------------------------------------------------------------------| | Submitter's Address: | 599 Cardigan Road<br>St. Paul, Minnesota 55126-3965 | | Contact Person: | Carolyn M. Steele Husten<br>Regulatory Affairs Manager | | Proprietary Name: | InnoSense Pelvic Floor Stimulation and<br>Electromyography System | | Common Name: | Pelvic Floor Stimulation and Biofeedback Device | | Classification Name: | Non-Implanted Electrical Continence Device<br>Biofeedback Device | #### Name of Predicate Devices B. - Verimed Myoexerciser III, K892649 ◆ - Thought Technology InControl, K935213 � - Verimed Veristim, K893220 � - Innova Pelvic Floor Stimulation System, K941911 � - Minnova Pelvic Floor Stimulation System, K970307 � #### C. Device Description ### List of Components The InnoSense Pelvic Floor Stimulation and Electromyography (EMG) System is comprised of a external stimulator/electromyography reader, leadwires (for surface EMG only) and electrodes. The device has a one channel stimulation function and a two channel EMG function. It is battery powered by two 1.5 volt (AA) batteries, with a low battery indication shown on the LCD Panel. The electrodes used to stimulate the pelvic floor muscles consist of either a vaginal or rectal electrode. The EMG function uses two electrodes consisting at the most one Vaginal or Rectal electrode and/or one or two surface EMG electrodes. One electrode is used to sense pelvic floor muscle activity and one is to sense accessory muscle activity such as the abdominal or gluteal muscles. # Electrical Characterization ## External Stimulator The external stimulator function provides stimulus output to the electrode and can be adjusted to deliver stimulation at an peak pulse intensity range of 0 to 100mA. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human figures, represented by flowing lines, suggesting a sense of community and support. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 30 1997 Re: K971527 InnoSense Pelvic Floor Stimulation and Electromyography System -- Dated: April 25, 1997 ------Received: April 28, 1997 Regulatory class: II 21 CFR §876.5320/Product code: 78 KPI 21 CFR §884.1425/Product code: 85 HIR Ms. Carolyn M. Steele Husten Regulatory Affairs Manager Empi, Inc." 599 Cardigan Road St. Paul, Minnesota 55126-3965 Dear Ms. Husten: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. . If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address http://www.fda.gov/cdrl/dsmamain.html". Sincerely yours, h.J.Liau Yir Lillian Yin, Ph.D. Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health {2}------------------------------------------------ 510(k) Number: (if known): K971527 Device Name: InnoSense Pelvic Floor Stimulation and Electromyography System Indications for Use: The InnoSense System is indicated for acute and ongoing treatment of stress, urge or mixed urinary incontinence and where the following results may improve urinary control: Improvement of urethral sphincter closure, Strengthening of pelvic floor muscles, Inhibition of the detruser muscle through reflexive mechanisms. It is also indicated during incontinence treatment for assessing EMG activity of the pelvic floor and accessory muscles such as the abdominal or gluteal muscles. PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Robert D. Salling (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number K911527 Prescription Use (Per 21 CFR 801.109) OR Over-The Counter Use _________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...