Revi System

K252391 · Bluewind Medical , Ltd. · QXM · Dec 12, 2025 · Gastroenterology, Urology

Device Facts

Record IDK252391
Device NameRevi System
ApplicantBluewind Medical , Ltd.
Product CodeQXM · Gastroenterology, Urology
Decision DateDec 12, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5305
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Revi System is indicated for the treatment of patients with symptoms of urgency incontinence alone or in combination with urinary urgency.

Device Story

Revi System is an implantable tibial electrical urinary continence device. Implanted near the tibial neurovascular bundle, it receives wireless power from an external, wearable unit. The wearable unit, powered by a replaceable internal battery, transmits energy to the implant, which delivers electrical pulses to the tibial nerve. These pulses travel to the sacral plexus and spinal cord to modulate nerve function and relieve symptoms. The system includes a clinician programmer and a HealthGo Micro Hub for data communication to the cloud. The device is used by adult patients under prescription. Modifications include an updated wearable unit with a new UI (three enlarged buttons, four linear LED indicators), internal circuit/firmware improvements for power consumption, and redesigned enclosure using plastic injection molding. The implantable component remains unchanged.

Clinical Evidence

Bench testing only. Testing included electrical/battery performance, mechanical testing, system performance, transportation/storage (ASTM D4169), product stability, software verification, cybersecurity, electrical safety (IEC 60601-1 Ed. 3.2), EMC (IEC 60601-1-2:2020, IEC TS 60601-4-2), and wireless testing. A human factors/usability evaluation with a comparative use-related risk analysis (URRA) confirmed no impact on critical tasks.

Technological Characteristics

Implant: Zirconia ceramic, titanium, gold capsule, perylene coating, platinum-iridium electrodes, silicone suture wings. Wearable Unit: Polycarbonate, silicone, fabric cover. Energy: Wireless power transfer from wearable unit to implant; wearable unit uses replaceable rechargeable battery. Connectivity: Wireless communication to clinician programmer and HealthGo Micro Hub. Standards: ASTM D4169, IEC 60601-1, IEC 60601-1-2, IEC TS 60601-4-2.

Indications for Use

Indicated for adult patients with symptoms of urgency incontinence alone or in combination with urinary urgency.

Regulatory Classification

Identification

An implanted tibial electrical urinary continence device is an implanted prescription device that receives power from a non-implanted external power source to provide electrical stimulation of the tibial nerve in proximity to the ankle. The device is intended for the treatment of overactive bladder related symptoms of urge urinary incontinence, urinary urgency, urinary frequency and nocturia.

Special Controls

In combination with the general controls of the FD&C Act, an implanted tibial electrical urinary continence device is subject to the following special controls:

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION December 12, 2025 BlueWind Medical Ltd. % Elissa Burg CEO / Regulatory Consultant BioVision Ltd. Perach Halilach 183 Had - Nes, 1295000 ISRAEL Re: K252391 Trade/Device Name: Revi System Regulation Number: 21 CFR 876.5305 Regulation Name: Implanted Tibial Electrical Urinary Continence Device Regulatory Class: II Product Code: QXM Dated: July 31, 2025 Received: November 12, 2025 Dear Elissa Burg: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252391 - Elissa Burg Page 2 Additional information about changes that may require a new premarket notification are provided in the Food and Drug Administration (FDA) guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the {2} K252391 - Elissa Burg Page 3 Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, JESSICA K. NGUYEN -S Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. # Indications for Use Submission Number (if known) K252391 Device Name Revi™ System Indications for Use (Describe) The Revi System is indicated for the treatment of patients with symptoms of urgency incontinence alone or in combination with urinary urgency. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov “An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.” {4} K252391 Page 1 of 5 510(k) Summary BlueWind Medical Ltd.'s Revi System I. Company: BlueWind Medical Ltd. 6 Maskit Street Herzliya, 4614002 Israel Contact: Elissa Burg Regulatory Consultant BioVision Ltd. Had Nes 183 Israel 1295000 Tel. (972) 526633572 Email:elissa@biovision.co.il Date Prepared: December 11, 2025 II. Proprietary Trade Name: Revi System III. Common or Usual Name: Implantable tibial electrical urinary continence device IV. Classification Name: Implantable tibial electrical urinary continence device (21 CFR 876.5305) V. Classification: Class II VI. Product Code: QXM VII. Predicate Device: Revi System (K240037) Note: The predicate device has not been subject to a design-related recall VIII. Indications for Use The Revi System is indicated for the treatment of patients with symptoms of urgency incontinence alone or in combination with urinary urgency. Note: The indications for use statement is identical to the indications for use statement of the predicate device. Traditional 510(k) Submission – Revi System – BlueWind Medical Ltd. {5} K252391 Page 2 of 5 # IX. Product Description: The Revi System is an implanted tibial electrical urinary continence device that wirelessly receives power from a non-implanted external wearable unit to provide electrical stimulation of the tibial nerve in proximity to the ankle. The device is intended for the treatment of urgency incontinence, alone or in combination with urinary urgency. The implantable device is implanted in the vicinity of the tibial neurovascular bundle. The treatment effect of the system is achieved by the implantable wireless neurostimulation component, which sends pulses to the tibial nerve when energized by the wearable unit transmitted power. The electrical pulses stimulate the nerve along the leg, reaching the sacral plexus and entering the spinal cord. This stimulation is theorized to have the power to modulate nerve function, relieving symptoms. The modifications reported in this 510(k) submission include introducing an updated Wearable Unit, an associated software update to support the updated Wearable Unit and related labeling. These modifications do not affect the implantable component of the Revi System or the indications for use. # X. Comparison of Technological Characteristics with Predicate | Description | Subject Device Revi System | Predicate Device Revi System (K240037) | | --- | --- | --- | | Indications for Use | The Revi System is indicated for the treatment of patients with symptoms of urgency incontinence alone or in combination with urinary urgency. | The Revi System is indicated for the treatment of patients with symptoms of urgency incontinence alone or in combination with urinary urgency. | | System Components | • Implant – neurostimulator • Rechargeable Wearable Unit - non-implanted, rechargeable, powers the Implant to provide electrical stimulation • Clinician Programmer (CP) - proprietary application that interfaces with the Wearable Unit and Cloud | • Implant – neurostimulator • Rechargeable Wearable Unit - non-implanted, rechargeable, powers the Implant to provide electrical stimulation • Clinician Programmer (CP) - proprietary application that interfaces with the Wearable Unit and Cloud | Traditional 510(k) Submission – Revi System – BlueWind Medical Ltd. {6} K252391 Page 3 of 5 | Description | Subject Device Revi System | Predicate Device Revi System (K240037) | | --- | --- | --- | | | • HealthGo Micro Hub - communicates with the Wearable Unit to acquire data and transmit to the Cloud | • HealthGo Micro Hub – communicates with the Wearable Unit to acquire data and transmit to the Cloud | | Principles of Operation | Implanted device received power from non- implanted external wearable device providing electrical stimulation to tibial nerve in proximity to ankle. | Implanted device received power from non- implanted external wearable device providing electrical stimulation to tibial nerve in proximity to ankle. | | Energy Source | Wearable Unit powered by internal rechargeable battery that is replaceable. | Wearable Unit powered by internal rechargeable battery. | | Software Version | v5.0.0.3 | v4.3 | | Intended population | Adult Users | Adult Users | | Use Environment | Rx only - Prescription Use | Rx only - Prescription Use | | Materials | Implant Zirconia ceramic, titanium, and gold capsule, coated with perylene; platinum- iridium electrodes, and silicone suture wings. Wearable Unit Polycarbonate and silicone covered with fabric | Implant Zirconia ceramic, titanium, and gold capsule, coated with perylene; platinum- iridium electrodes, and silicone suture wings. Wearable Unit Polycarbonate and silicone covered with fabric | | Battery component | Replaceable battery | Non-replaceable battery | | User Interface (UI) | Three enlarged buttons with four separate, shape-specific LED indicators in a linear arrangement | Three buttons with a single muti-color LED (red/yellow/green) in a triangular layout | | Device Power Consumption | Enhanced power consumption via internal circuit and firmware improvements | Same device power consumption as DEN220073 | | Design for Manufacturability and Scalability | • Replacement of electronic components with newer equivalents • Integration of printed circuit boards • Redesigned enclosure using plastic injection molding | Same device design and manufacturing process as DEN220073 | Traditional 510(k) Submission – Revi System – BlueWind Medical Ltd. {7} K252391 Page 4 of 5 # XI. Performance Testing Non-clinical benchtop testing was conducted to evaluate the performance characteristics of the modified Revi System. The following tests were performed: - Electrical bench testing including battery testing to support the performance and safety of the battery throughout its service life and stimulation (RF) characteristics to ensure accuracy with device specification. - Mechanical bench testing including dimensional and tolerance analysis, and leg band performance. - System performance bench testing to ensure system performance is in compliance with the technical specification. - Transportation and storage per ASTM D4169 - Standard Practice for Performance Testing of Shipping Containers and Systems. - Product stability to support the device durability and battery lifetime. - Software verification to verify software requirements, including cybersecurity testing according to FDA Guidance documents titled "Content of Premarket Submissions for Device Software Functions" and "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" - Electrical safety per IEC 60601-1 Ed. 3.2 en: 2020. - Electromagnetic Compatibility (EMC) testing per IEC 60601-1-2:2020 and IEC TS 60601-4-2 Edition 1.0 2024-03. - Wireless Testing according to FDA Guidance document "Radio Frequency Wireless Technology in Medical Devices - Guidance for Industry and FDA Staff" The test results demonstrate that the modified Revi System complies with its specifications and requirements and substantiate its equivalence to the cleared Revi System. A human factors/usability evaluation of the Revi System with a comparative use-related risk analysis (URRA) found that no critical tasks were impacted or created by the subject device. Traditional 510(k) Submission – Revi System – BlueWind Medical Ltd. {8} K252391 Page 5 of 5 # XII. Conclusion The Revi System and its predicate device have the same intended use/indications for use and similar technological characteristics. The modifications introduced to the subject Revi System do not affect its principles of operation, underlying technology, essential mode of operation and principal functionality. The differences between the Revi System and its predicate do not present different questions of safety or effectiveness and the performance tests conducted on the subject device support the device is as safe and effective as the predicate. Therefore, the Revi System is substantially equivalent to the predicate device. Traditional 510(k) Submission – Revi System – BlueWind Medical Ltd.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...