QAJ · Cutaneous Electrode Stimulator For Urinary Incontinence

Gastroenterology, Urology · 21 CFR 876.5330 · Class 2

Overview

Product CodeQAJ
Device NameCutaneous Electrode Stimulator For Urinary Incontinence
Regulation21 CFR 876.5330
Device ClassClass 2
Review PanelGastroenterology, Urology

Identification

A transcutaneous electrical continence device consists of cutaneous electrodes that are used to apply external stimulation to reduce urinary incontinence.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

In combination with the general controls of the FD&C Act, the transcutaneous electrical continence device is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must characterize the electrical stimulation, including the following: Waveforms, output modes, maximum output voltage, maximum output current, pulse duration, frequency, net charge per pulse, maximum phase charge at 500 ohms, maximum current density, maximum average current, and maximum average power density. (2) The patient-contacting materials must be demonstrated to be biocompatible. (3) Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, thermal safety, and mechanical safety of the device. (4) Software verification, validation, and hazard analysis must be performed. (5) Labeling must include the following: (i) Instructions for use, including specific instructions regarding the proper placement of electrodes; (ii) A summary of electrical stimulation parameters; and (iii) Cleaning instructions and reuse information.

Recent Cleared Devices (5 of 5)

RecordDevice NameApplicantDecision DateDecision
K253285Elitone for MenElidah, Inc.Jan 23, 2026SESE
K223884ELITONE Urge Urinary Incontinence DeviceElidah, Inc.Feb 24, 2023SESE
K192357INNOVOAtlantic Therapeutics, Ltd.Jan 16, 2020SESE
K183585Elitone DeviceElidah, Inc.Feb 11, 2019SESE
DEN170049InnovoBio-Medical Research, Ltd.Nov 6, 2018DENG

Top Applicants

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...