Hach CM130 Chlorine Monitoring System

K162471 · Hach Company · PSX · Mar 30, 2017 · Gastroenterology, Urology

Device Facts

Record IDK162471
Device NameHach CM130 Chlorine Monitoring System
ApplicantHach Company
Product CodePSX · Gastroenterology, Urology
Decision DateMar 30, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5665
Device ClassClass 2

Indications for Use

The Hach CM130 is an instrument that is intended for use by hemodialysis professionals to automatically monitor low levels of total chlorine (i.e. total chlorine) in feed water used to prepare dialysate in hemodialysis systems. The CM130 is a component of the complete water treatment system for hemodialysis and does not treat or alter the water used in dialysate. The CM130 instrument's automated monitoring records total chlorine values in feed water at intervals between 5 and 20 minutes.

Device Story

The CM130 is a microprocessor-controlled analyzer for monitoring total chlorine in hemodialysis feed water. It operates by introducing N,N-Diethyl-p-phenylenediamine (DPD) indicator, potassium iodide, and a buffer to water samples; the resulting red dye intensity is measured photometrically. The device automatically conditions samples, delivers reagents, and performs measurements at 5-20 minute intervals. It is installed in water rooms, with a Remote Indicator (RI) placed in patient treatment areas to provide audible and visual alerts/alarms. Alerts trigger at >0.07 mg/L; alarms trigger at >0.1 mg/L. The system provides continuous monitoring, allowing staff to address water system maintenance before chlorine levels reach unsafe concentrations, thereby protecting patients from chlorine exposure during dialysis.

Clinical Evidence

Bench testing only. Performance verification included bias, repeatability, and reproducibility testing per CLSI-EP standards. Interference testing evaluated effects of iron, metal cations, and inorganics. Alarm performance was tested per CLSI EP12-A2. Human factors usability testing was conducted per ANSI/AAMI HE75:2009 using simulated and actual use tasks. Software verification and validation were performed per FDA guidance. Electrical safety and EMC were evaluated per IEC61326-1 and IEC61010-1.

Technological Characteristics

Microprocessor-controlled photometer. Uses DPD colorimetric method. Measures total chlorine (0.03-0.20 mg/L). Features automated sample conditioning, reagent delivery, and remote indicator. Electrical safety per IEC61010-1; EMC per IEC61326-1. Enclosed in protective housing.

Indications for Use

Indicated for hemodialysis professionals to automatically monitor low levels of total chlorine (free chlorine and combined chloramines) in feed water used to prepare dialysate in hemodialysis systems.

Regulatory Classification

Identification

A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.

Special Controls

*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, with flowing lines beneath them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 30, 2017 HACH Company Jeff Ryberg Vice President RA /OA 100 Dayton Avenue Ames, IA 50010 Re: K162471 > Trade/Device Name: CCM130 Chlorine Monitor Regulation Number: 21 CFR§ 876.5665 Regulation Name: Water Purification System For Hemodialysis Regulatory Class: II Product Code: PSX, FIP Dated: February 24, 2017 Received: February 27, 2017 Dear Jeff Ryberg: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. Joyce M. Whang -S for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K162471 Device Name CM130 Chlorine Monitor #### Indications for Use (Describe) The Hach CM130 is an instrument that is intended for use by hemodialysis professionals to automatically monitor low levels of total chlorine (i.e. total chlorine) in feed water used to prepare dialysate in hemodialysis systems. The CM130 is a component of the complete water treatment system for hemodialysis and does not treat or alter the water used in dialysate. The CM130 instrument's automated monitoring records total chlorine values in feed water at intervals between 5 and 20 minutes. Type of Use (Select one or both, as applicable) | <label> <input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </label> | |-------------------------------------------------------------------------------------------------------------| | <label> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </label> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Section 6: # 510 (k) SUMMARY | 1. Submitted by: | Hach Company<br>100 Dayton Avenue<br>Ames<br>Iowa, 50010 USA | Phone: | (714) 516 - 7659 | |-----------------------------------------|---------------------------------------------------------------------------------------|--------|-------------------------| | | | Fax: | (714) 516 - 7474 | | | Contact: Jeff Ryberg, VP RA/QA | | jeff.ryberg@danaher.com | | 2. Date<br>Prepared: | March 30, 2017 | | | | 3. Device Name: | Water Purification System for Hemodialysis<br>Auxiliary Component | | | | Classification<br>of Device: | Class II<br>21 CFR 876.5665<br>Primary Product Code PSX<br>Secondary Product Code FIP | | | | 4. Trade Name of<br>Proposed<br>Device: | Hach CM130 Chlorine Monitor | | | #### 6. Predicate Device: | Sub-System | Predicate Device Name | Description | 510(k)<br>Number | |------------|--------------------------------------|--------------------------|------------------| | Chemistry | SteriChek®<br>Total Chlorine DPD Kit | Chlorine<br>Measurements | K983997 | ### 7. Reference Devices: | Sub-System | Reference Device<br>Name | Description | 510(k)<br>Number | |----------------------------------------------|---------------------------------------------------------------|--------------------------------------------|------------------| | Water Sample<br>Conditioning and<br>Delivery | Gambro Central<br>Water Treatment<br>System, CWP 100<br>WRO H | Water Flow Control | K974899 | | Reagent Delivery | AP 34 Multi-Therapy<br>Infusion Pump | Fluid movement using a<br>Peristaltic pump | K082182 | {4}------------------------------------------------ | Sub-System | Reference Device<br>Name | Description | 510(k)<br>Number | |------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|------------------| | Measurement<br>Chamber | DensiChek Plus Fully<br>Automated Short<br>Term Incubation<br>Cycle Antimicrobial<br>Susceptibility Device | Provides measurements<br>based on the optical<br>density of a<br>microorganism<br>suspension | K093244 | | Electronic Hardware<br>Software Controls<br>with Enclosure | Dialysate Meter,<br>Model D-6 | Microprocessor<br>controlled automation<br>in a water room | K100237 | | Remote Indicator | AmeriWater Alarm | Remotely mounted<br>device “state” indicator | K121022 | # 8. Proposed Device Description: The Hach CM130 Chlorine Monitor is a microprocessor-controlled analyzer used to monitor the chlorine content of water which is used to prepare dialysate for hemodialysis. The monitor is mounted in the water room in a location that allows sampling of feed water between the primary (scrubbing) and secondary (polishing) carbon tanks. Water that flows into the CM130 exits the device to a drain. The water samples, once tested, are also discarded to a drain. The monitor measures and displays total chlorine (free chlorine and combined chloramines) as Cl2, ranging from 0.03 to 0.20 mg/L. Image /page/4/Picture/6 description: The image shows a HACH CM130 instrument. The instrument has a digital display that reads "0.05 mg/L". The instrument is enclosed in a gray box with a blue frame and has two white bottles inside. Section 6 – Figure 1: CM130 Chlorine Monitor {5}------------------------------------------------ ## CM130 Chlorine Monitor 510 (k) March 30, 2017 The CM130 Chlorine Monitor employs a DPD Colorimetric Method for the detection of total chlorine. The analyzer introduces N,N-Diethyl-pphenylenediamine (DPD) indicator with potassium iodide and a buffer to the water sample under test. The indicator and buffer are allowed to react with the chlorine and chloramines present in the water. The reaction results in a dye which forms a red color in the water sample. The intensity of the red color is proportional to the total chlorine concentration. It is measured photometrically and the results are automatically recorded and displayed. The Remote Indicator (RI) is a subsystem that is mounted in the patient treatment area, typically near the ceiling, so that it can be easily seen. Its primary function is to communicate the status of the chlorine concentration of the feed water and the status of the analyzer to the staff while they are working in the patient treatment room. Image /page/5/Figure/4 description: The image shows a device with visual and audible signals. There are two blue water drop icons at the top, indicating the state of the device. Below the water drop icons are two yellow squares with a curved line inside, which are also visual signals. At the bottom of the device, there is an arrow pointing to the audible signal. Section 6: Figure 2: CM130's Remote Indicator (RI) The CM130 Chlorine Monitor and the Remote Indicator will each provide both an audible and a visual signal. {6}------------------------------------------------ #### Alert An alert is a signal to inform staff that total chlorine measurements indicate concentrations have increased (medium – high), but are still acceptable. This predefined "alert" level is greater than .07 mg/L but less than 0.1 mg/L Cl2. Notification at this point with an amber illumination and alert tone, allows staff to address maintenance of the water system at a time that is relatively convenient. #### Alarm The analyzer and the Remote Indicator each provide both an audible and a visual "Alarm" signal when the total Chlorine exceeds the standard allowable level of 0.1 mg/L Cl2. #### 9. Indications for Use: The Hach CM130 is an instrument that is intended for use by hemodialysis professionals to automatically monitor low levels of total chlorine (i.e. total chloramines plus free chlorine) in feed water used to prepare dialysate in hemodialysis systems. The CM130 is a component of the complete water treatment system for hemodialysis and does not treat or alter the water used in dialysate. The CM130 instrument's automated monitoring records total chlorine values in feed water at intervals between 5 and 20 minutes. # 10. Summary of Technological Characteristics of New Device Compared to Predicate and Reference Devices The proposed CM130 Chlorine Monitor uses the same chlorine detection technology (DPD color reaction) as the predicate. Both products are used to monitor feed water used to prepare dialysate. However, providing automatic measurements, the CM130 Chlorine Monitor adds technological features not found in the predicate device. These technological features include: - 1. Automated water sample conditioning and delivery - 2. Automated reagent delivery - 3. Automated measurements - 4. Electronic hardware and software used to control device operation - 5. A Remote Indicator (RI) to allow individuals in the patient treatment area, to monitor the status of equipment in the water room - 6. An Enclosure to protect electronics from incidental splashing or spills {7}------------------------------------------------ # 11. Summary of Performance Testing Supporting Substantial Equivalence System Performance Testing Non-clinical testing was conducted on the CM130 Chlorine Monitor. Performance verification and validation testing demonstrate that the CM-130 Chlorine Monitor system and subsystems are safe and effective for their intended use. | Description | Performance Test Comment | |-----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | CM130 Bias Verification | Demonstrates the ability of the CM130 to match a<br>known reference reading (Accuracy) and<br>compares performance to the Predicate Devices<br>ability to match a known reading. | | CM130 Repeatability -<br>Reproducibility<br>Verification | Demonstrates within instrument and between<br>instrument variations. | | CM130 Reproducibility<br>of the Predicate | Comparative reproducibility demonstrates there is<br>less variation with the CM130 | | Effects of Iron Cations<br>on Total Chlorine<br>Measurements<br>Note: Minerals are<br>grouped see intro to<br>each protocol | Interference Testing<br>Interferences are tested to limits allowed in water<br>by standards. The testing demonstrates the device<br>performs acceptably even when potentially<br>"interfering" chemicals are present.<br>The results are published in the instructions for<br>use | | Effects of Metal Cations<br>on Total Chlorine<br>Measurements | | | Effects of Additional<br>Inorganics on Total<br>Chlorine Measurements | | Performance testing was conducted on the CM130 Chlorine Monitor and on the predicate device using the same solutions and test setups with modifications as appropriate for the differences between manual and automated testing. The CM130 Chlorine Monitor is designed to test automatically, generating a measurement of a new sample of water every 5 minutes during use. The predicate is a manual process that takes an unspecified amount of time to collect and process each water sample. The devices have similar measurement capability in that the predicate scale is manually read to the nearest .02 mg/L and is interpolated to .01 mg/L while the CM130 Chlorine Monitor measures to .01mg/L Cl2. {8}------------------------------------------------ ### CM130 Chlorine Monitor 510 (k) March 30, 2017 Accuracy (Bias) and Precision (Repeatability) verifications were conducted under CLSI-EP Standards. These protocols address equivalency testing with the CM130 Chlorine Monitor results compared to the predicate. Bias testing covered the entire measurement range of the CM130 Chlorine Monitor. Test measurements for each device are compared to a reference method used to independently verify the "true" measurement. #### Electrical Safety Testing The CM-130 Chlorine Monitor has been evaluated against the requirements for electromagnetic compatibility and electrical safety in accordance with IEC61326-1 and IEC61010-1. #### Software Verification and Validation Testing The CM-130 Chlorine Monitor software was developed, verified and validated in accordance with the applicable FDA guidance documents. The results of the verification and validation activities demonstrate the software performs as intended. #### Alarm Testing The CM130 system was tested for alarm performance following CLSI EP12-A2. The positive and negative alarm performance demonstrate the device is safe and effective for intended use. ### Reagent Shelf Life and Use Life Testing The CM130 reagents were tested to confirm that they met the specified shelf life and use life. The reagents meet all specifications when stored upright in the marketing container at the specified temperatures. The results of the shelf life and use life testing demonstrate the device is safe and effective for intended use. #### Human Factors Testing The CM130 was tested for human factors usability per ANSI/AAMI HE75:2009. The CM130 has been found to be safe and effective for the intended users, uses and use environments as demonstrated in summative simulated and actual use testing of the user interfaces with representative users. The Human Factors Engineering and Usability Engineering process involved several methodologies, including observational user research, formative user testing and summative user testing. Formative and summative user testing included simulated use and actual use tasks for both user groups. Based on subjective and objective data it can be concluded that all users can safely and effectively use the device. #### 12. Conclusion CM-130 Chlorine Monitor non-clinical performance, electrical safety, software verification and validation, alarm testing, reagent and use life testing and {9}------------------------------------------------ # CM130 Chlorine Monitor 510 (k) March 30, 2017 human factors testing demonstrate the device is safe and effective for its intended use. The proposed device uses the same active chemistry, has the same intended use, and has labeling similar to the predicate device. The CM130 Chlorine Monitor is also similar to the reference devices in specific technological features. The predicate and subject devices were tested to plan with passing results, therefore demonstrating substantial equivalence.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...