ENFit Oral / Enteral Syringe

K190502 · Jiangsu Caina Medical Co.,Ltd · PNR · Aug 6, 2019 · Gastroenterology, Urology

Device Facts

Record IDK190502
Device NameENFit Oral / Enteral Syringe
ApplicantJiangsu Caina Medical Co.,Ltd
Product CodePNR · Gastroenterology, Urology
Decision DateAug 6, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5980
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

ENFit Oral/Enteral Syringe is indicated for use as a dispenser, a measuring device, and an oral/enteral fluid transfer device. It is used to deliver fluids into the body orally or enterally. It is intended to be used in clinical or home care settings by users ranging from clinicians to laypersons (under the supervision of a clinician) in all age groups.

Device Story

Disposable enteral feeding syringe; sizes 0.5ml to 100ml; components include barrel, plunger, piston, and tip cap. Features female ENFit connector (standard or low-dose tip) for connection to enteral access devices. Operated manually by clinicians or laypersons (under supervision) in clinical or home settings. Used to measure and deliver fluids orally or enterally. Sterile (EO gas) or non-sterile options available. Transparent or amber barrels; purple plungers. Benefits include standardized connection to prevent misconnections and accurate fluid delivery.

Clinical Evidence

No clinical data. Bench testing only. Compliance verified via ISO 80369-3, ISO 80369-20, ISO 7886-1, ISO 10993-5/7/10, ISO 14971, ASTM F88/F88M, and ASTM F1929.

Technological Characteristics

Materials: Polypropylene (barrel/plunger), polyisoprene (piston). Manual operation. Sizes 0.5ml-100ml. Connectors: Female ENFit (standard or low-dose). Sterilization: Ethylene Oxide (SAL 10^-6). Standards: ISO 80369-3, ISO 80369-20, ISO 7886-1, ISO 10993, ASTM F88, ASTM F1929.

Indications for Use

Indicated for all age groups in clinical or home care settings for dispensing, measuring, and transferring oral/enteral fluids, under clinician supervision for laypersons.

Regulatory Classification

Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. August 6, 2019 Jiangsu Caina Medical Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai. 200120 CHINA Re: K190502 > Trade/Device Name: ENFit Oral/Enteral Syringe Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: Class II Product Code: PNR Dated: June 21, 2019 Received: July 2, 2019 Dear Diana Hong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for Martha W. Betz, Ph.D. Acting Assistant Division Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K190502 Device Name ENFit Oral/Enteral Syringe ### Indications for Use (Describe) ENFit Oral/Enteral Syringe is indicated for use as a dispenser, a measuring device, and an oral/enteral fluid transfer device. It is used to deliver fluids into the body or enterally. It is intended to be used in clinical or home care settings by users ranging from clinicians to laypersons (under the supervision of a clinician) in all age groups. | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------| | <div> <span> </span> <span>X Prescription Use (Part 21 CER 801 Subpart D)</span> </div> | <div> <span> </span> <span>Over-The-Counter Use (21 CER 801 Subpart C)</span> </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92. The assigned 510(k) Number: K190502 - 1. Date of Preparation: 08/05/2019 - 2. Sponsor Identification Jiangsu Caina Medical Co., Ltd.s No.23, Huanxi Rd. Zhutang Town, 214415 Jiangyin City, Jiangsu, China Establishment Registration Number: 3005670221 Contact Person: Xinyan Ruan Position: QA Engineer Tel: +86-0510-8686 6666-8027 Fax: +86-0510-8686 6666-8009 Email: sherry.ruan@cainamed.com - 3. Designated Submission Correspondent Ms. Diana Hong (Primary Contact Person) Mr. Chengyu Wang (Alternative Contact Person) Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China Tel: +86-21-22815850, Fax: +1-360-925-3199 Email: info@mid-link.net {4}------------------------------------------------ #### 4. Identification of Proposed Device Trade Name: ENFit Oral/Enteral Syringe Common Name: Enteral Feeding Syringe with ENFit Connector ### Regulatory Information Classification Name: Gastrointestinal tube and accessories Classification: II; Product Code: PNR Regulation Number: 21CFR 876.5980 Review Panel: Gastroenterology/Urology; Indications for Use Statement: ENFit Oral/Enteral Syringe is indicated for use as a dispenser, a measuring device, and an oral/enteral fluid transfer device. It is used to deliver fluids into the body orally or enterally. It is intended to be used in clinical or home care settings by users ranging from clinicians to laypersons (under the supervision of a clinician) in all age groups. Device Description: The proposed device is a disposable enteral feeding syringe provided in a variety of sizes from 0.5ml to 100ml. It consists of plunger, piston, barrel and tip cap, and used to deliver fluids into the body orally or connected to an enteral access device with male ENFit connector. The syringes (size 5ml, 6ml, 12ml, 20ml, 30ml, 35ml, 50ml, 60ml and 100ml) incorporate a female Standard ENFit connector; the syringes (size 0.5ml, 1ml, 3ml, 5ml and 6ml) incorporate a female low dosing ENFit connector. There are 2 types of the syringe: Side connector ENFit syringe and central connector ENFit syringe. The sizes of the Side connector ENFit syringe range from 10ml to 100ml; and central connector ENFit syringe range from 0.5ml to 100ml. The proposed syringe is sterile or non-sterile. Sterile device was sterilized by Ethylene Oxide Gas to achieve a SAL of 10° and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of five years. The sterile syringes are provided with transparent barrel, and the non-sterile syringes are provided with transparent/ amber color barrel. {5}------------------------------------------------ #### 5. Identification of Predicate Device 510(k) Number: K161039 Product Name: Oral/Enternal Feeding Syringes with ENFit Connector (12ml to 100ml) and Low Dose Tip Oral/Enternal Feeding Syringes with ENFit Connector (0.5ml to 6ml) - 6. Non-Clinical Test Conclusion Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: ISO 80369-3:2016 Small-Bore Connectors for Liquids and Gases in Healthcare Applications-Part 3: Connectors for Enteral Applications ISO 7886-1:2017 Sterile Hypodermic Syringes for Single Use-Part 1: Syringes for Manual Use; ISO 80369-20:2015 Small-Bore Connectors for Liquids And Gases in Healthcare Applications-Part 20: Common Test Methods; ISO 10993-5:2009 Biological Evaluation of Medical Devices-Part 5: Tests for in Vitro Cytotoxicity; ISO 10993-10:2010 Biological Evaluation of Medical Devices-Part 10: Tests for Irritation and Skin Sensitization: ISO 10993-7:2008, Biological evaluation of medical devices- Part 7: Ethylene oxide sterilization residuals ISO 14971:2007 Medical Devices-Application of Risk Management to Medical Devices; ASTM F88/F88M-15 Standard Test Method for Seal Strength of Flexible Barrier Materials; ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration; - 7. Clinical Test Conclusion No clinical study is included in this submission. {6}------------------------------------------------ #### 8. Substantially Equivalent (SE) Comparison ## Table 1 Comparison of Technology Characteristics | Item | Proposed Device | Predicate Device K161039 | |------------------------|-----------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------| | Product Code | PNR | PNR | | Regulation Number | 21CFR 876.5980 | 21CFR 876.5980 | | Indications for Use | ENFit Oral/Enteral Syringe is indicated | The device is indicated for use as a | | | for use as a dispenser, a measuring | dispenser, a measuring device, and a fluid | | | device, and an oral/enteral fluid | transfer device. It is used to deliver fluids | | | transfer device. It is used to deliver | into the body orally or enterally. It is | | | fluids into the body orally or enterally. | intended to be used in clinical or home care | | | It is intended to be used in clinical or | setting by users ranging from clinicians to | | | home care settings by users ranging | laypersons (under the supervision of a | | | from clinicians to laypersons (under the supervision of a clinician) in all age groups. | clinician) in all age groups. | | Configuration | Piston; | Piston; | | | Plunger; | Plunger; | | | Barrel with ENFit connector; | Barrel with ENFit connector; | | | Tip cap | Tip cap | | Material | Plunger: polypropylene (PP); | Unknown | | | Barrel: polypropylene (PP); | | | | Piston: polyisoprene | | | Color | Plunger: purple | Plunger: white | | | Barrel: transparent and amber color | Barrel: transparent and amber color | | | Low does tip ENFit Syringe: 0.5ml,<br>1ml, 3ml, 5ml and 6ml | Low dose tip ENFit syringe: 0.5ml~6ml; | | Product Size | Standard ENFit syringe: 5ml, 6ml,<br>10ml, 12ml, 20ml, 30ml, 35ml, 50ml,<br>60ml and 100ml | Standard ENFit syringe: 12ml~100ml; | | Product<br>Performance | Complied with:<br>ISO 80369-3;<br>ISO 80369-20;<br>ISO 7886-1; | Complied with:<br>ISO 80369-3;<br>ISO 80369-20;<br>ISO 7886-1; | | Operation Mode | For Manual Use Only | For Manual Use Only | | Sterile | Sterile or non-sterile | Sterile or non-sterile | | Sterile Method | EO Sterilized | EO Sterilized | | SAL | 10-6 | 10-6 | | Single Use | Single Use | Single Use | {7}------------------------------------------------ The patient contact material for the predicate devices is unknown, and the plunger color is different between the proposed device and predicate device. However, the Cytotoxicity, Irritation, and Sensitization Testing on proposed device were completed according to ISO 10993, the results did not show any adverse effect. Therefore, this difference is not considered to effect substantially equivalency. - 9. Substantially Equivalent (SE) Conclusion Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...