K241278 · Pfm Medical, Inc. · PNG · Jan 14, 2025 · Gastroenterology, Urology
Device Facts
Record ID
K241278
Device Name
ASEPT® Glide Peritoneal Drainage System
Applicant
Pfm Medical, Inc.
Product Code
PNG · Gastroenterology, Urology
Decision Date
Jan 14, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of an indwelling peritoneal drainage catheter is for drainage of refractory ascites with long-term occurrence from the peritoneal cavity.
Device Story
The ASEPT Glide Peritoneal Drainage System facilitates drainage of accumulated abdominal fluid. The system consists of an indwelling peritoneal catheter and a drainage kit. The catheter is surgically placed into the peritoneal cavity; it features a surface modification from the distal tip to the polyester cuff. Patients or clinicians perform periodic drainage in home or hospital settings. The device provides a method for symptom palliation in patients with recurrent ascites. By enabling controlled fluid removal, the system helps manage symptoms related to refractory ascites, potentially improving patient quality of life.
Clinical Evidence
No clinical tests were required. Substantial equivalence was demonstrated through bench testing, including biocompatibility testing per ISO 10993 and sterilization validation per ISO 11135:2014.
Technological Characteristics
Indwelling peritoneal catheter with surface modification from distal tip to polyester cuff. Provided sterile. Biocompatibility per ISO 10993; sterilization per ISO 11135:2014. Mechanical drainage system.
Indications for Use
Indicated for intermittent, long-term drainage of symptomatic, recurrent, malignant and non-malignant ascites in adults unresponsive to medical management. Non-malignant use limited to patients intolerant/resistant to medical therapy, refractory to large volume paracentesis (LVP), and not candidates for trans-jugular intrahepatic portosystemic shunt or LVP.
Regulatory Classification
Identification
(1) A peritoneal dialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions, and that consists of a peritoneal access device, an administration set for peritoneal dialysis, a source of dialysate, and, in some cases, a water purification mechanism. After the dialysate is instilled into the patient's peritoneal cavity, it is allowed to dwell there so that undesirable substances from the patient's blood pass through the lining membrane of the peritoneal cavity into this dialysate. These substances are then removed when the dialysate is drained from the patient. The peritoneal dialysis system may regulate and monitor the dialysate temperature, volume, and delivery rate together with the time course of each cycle of filling, dwell time, and draining of the peritoneal cavity or manual controls may be used. This generic device includes the semiautomatic and the automatic peritoneal delivery system.(2) The peritoneal access device is a flexible tube that is implanted through the abdominal wall into the peritoneal cavity and that may have attached cuffs to provide anchoring and a skin seal. The device is either a single use peritioneal catheter, intended to remain in the peritoneal cavity for less than 30 days, or a long term peritoneal catheter. Accessories include stylets and trocars to aid in the insertion of the catheter and an obturator to maintain the patency of the surgical fistula in the abdominal wall between treatments. (3) The disposable administration set for peritoneal dialysis consists of tubing, an optional reservoir bag, and appropriate connectors. It may include a peritoneal dialysate filter to trap and remove contaminating particles. (4) The source of dialysate may be sterile prepackaged dialysate (for semiautomatic peritoneal dialysate delivery systems or “cycler systems”) or dialysate prepared from dialysate concentrate and sterile purified water (for automatic peritoneal dialysate delivery systems or “reverse osmosis” systems). Prepackaged dialysate intended for use with either of the peritoneal dialysate delivery systems is regulated by FDA as a drug.
Special Controls
*Classification.* Class II (special controls). The following accessories are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9: A catheter finger grip that is non-patient contacting and intended for single use with a peritoneal catheter; a continuous ambulatory peritoneal dialysis (CAPD) belt; and a catheter stand that does not include weigh scales.
K221779 — ASEPT Peritoneal Drainage System · Pfm Medical, Inc. · May 2, 2023
K110396 — ASPIRA PERITONEAL DRAINAGE SYSTEM · C.R. Bard, Inc. · May 6, 2011
K212675 — Aspira Peritoneal Drainage System · Merit Medical Systems, Inc. · Dec 14, 2021
K201155 — PleurX Peritoneal Catheter System · Care Fusion · Oct 21, 2020
K162457 — Rocket Indwelling Peritoneal Catheter (IPC) Insertion Kit · Rocket Medical Plc · Nov 14, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 14, 2025
PFM Medical, Inc. Madison Reid Regulatory Affairs Coordinator 1916 Palomar Oaks Way, Suite 150 Carlsbad, CA 92008
Re: K241278
Trade/Device Name: ASEPT® Glide Peritoneal Drainage System Regulation Number: 21 CFR§ 876.5630 Regulation Name: Peritoneal Dialysis System and Accessories Regulatory Class: II Product Code: PNG Dated: December 13, 2024 Received: December 13, 2024
Dear Madison Reid:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
l devices and radiation-emitting products, including
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Maura Rooney -S
Maura Rooney Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K241278
Device Name
ASEPT® Glide Peritoneal Drainage System
Indications for Use (Describe)
The ASEPT Glide Peritoneal Drainage System is indicated for intermittent, long-term drainage of symptomatic, recurrent, malignant and non-malignant ascites that does not respond to medical management of the underlying disease and for the palliation of symptoms related to recurrent ascites.
The use of the ASEPT Glide Peritoneal Catheter for non-malignant ascites is limited to patients who are intolerant or resistant to maximum medical therapy, refractory to large volume paracentesis (LVP) and are not candidates for trans-juqular intrahepatic portosystemic shunt or LVP. The ASEPT Glide Peritoneal Catheter is indicated for adults only.
Type of Use (Select one or both, as applicable)
< Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for PFM Medical. The letters 'pfm' are in a blue sans-serif font, with the 'p' and 'f' connected. The word 'medical' is in a black sans-serif font and is connected to the 'm' in 'pfm'. The logo is simple and modern.
This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92.
| I. | SUBMITTER | PFM Medical, Inc<br>1916 Palomar Oaks Way, Suite 150<br>Carlsbad, CA 92008, USA |
|----|------------------------|-----------------------------------------------------------------------------------------------|
| | Contact Person: | Madison Reid<br>Regulatory Affairs Coordinator<br>PFM Medical, Inc<br>mreid@pfmmedicalusa.com |
| | Date Summary Prepared: | January 9, 2024 |
### II. DEVICE
| Trade or Proprietary Name | ASEPT® Glide Peritoneal Drainage System |
|---------------------------|------------------------------------------------------------------------------------|
| Regulation Name | Peritoneal dialysis system and accessories |
| Device Class | Class II |
| Regulation Number | 21 CFR §876.5630 |
| FDA Product Code | PNG, Peritoneal, Drainage Catheter for Refractory<br>Ascites, Long-Term Indwelling |
### III. LEGALLY MARGETED PREDICATE DEVICES
| Predicate Device | | | |
|------------------|----------------------------------|----------------|--|
| 510(k) | Product Name | Clearance Date | |
| K221779 | ASEPT Peritoneal Drainage System | May 02, 2023 | |
{5}------------------------------------------------
#### IV. DEVICE DESCRIPTION
The ASEPT Glide Peritoneal Drainage System provides patients with a method to drain accumulated fluid from the abdomen. The primary components of the system are the indwelling ASEPT Glide Peritoneal Catheter and the ASEPT Drainage Kit. The catheter is placed in the patient's peritoneal cavity enabling the patient to perform periodic peritoneal drainage at home or in the hospital. The device is provided sterile.
The ASEPT Glide Peritoneal Drainage Catheter has a surface modification applied from the distal tip to the polyester cuff.
#### V. INTENDED USE
The intended use of an indwelling peritoneal drainage catheter is for drainage of refractory ascites with long-term occurrence from the peritoneal cavity.
#### VI. INDICATIONS FOR USE
The ASEPT Peritoneal Drainage System is indicated for intermittent, long-term drainage of symptomatic, recurrent, malignant, and non-malignant ascites that does not respond to medical management of the underlying disease and for the palliation of symptoms related to recurrent ascites.
The use of the ASEPT Glide Peritoneal Catheter for non-malignant ascites is limited to patients who are intolerant or resistant to maximum medical therapy, refractory to large volume paracentesis (LVP) and are not candidates for a trans-jugular intrahepatic portosystemic shunt or LVP. The ASEPT Glide Peritoneal Catheter is indicated for adults only.
{6}------------------------------------------------
#### VII. TECHNICAL COMPARISON TO PREDICATE
The technological design features of the subject device including intended use, design, materials, drainage function and method, and fundamental scientific technology were qualified and compared to the predicate device and it was demonstrated that they are substantially equivalent.
#### PERFORMANCE DATA VIII.
Bench testing was performed on the ASEPT Glide Peritoneal Catheter and replacement valve to demonstrate substantial equivalence. The performance testing requirements were determined by the predicate and reference devices and assessment of risk. Biocompatibility was tested in accordance with ISO 10993 standards. The device sterilization process was validated in accordance with ISO 11135:2014.
#### IX. SUMMARY OF CLINICAL TESTS REFERENCED
No clinical tests were required to confirm the safety and effectiveness of the subject device.
#### X. CONCLUSION
Based on the information provided in this 510(k) submission, it has been determined that the subject device is as safe, as effective and performs as well or better than the legally marketed predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.