15.5F X 19 cm Symetrex Long Term Hemodialysis Catheter with Sideholes, 15.5F X 23 cm Symetrex Long Term Hemodialysis Catheter with Sideholes, 15.5F X 28 cm Symetrex Long Term Hemodialysis Catheter with Sideholes, 15.5F X 33 cm Symetrex Long Term Hemodialysis Catheter with Sideholes, 15.5F X 37 cm Symetrex Long Term Hemodialysis Catheter with Sideholes

K173667 · Medical Components, Inc. (dba MedComp) · MSD · May 17, 2018 · Gastroenterology, Urology

Device Facts

Record IDK173667
Device Name15.5F X 19 cm Symetrex Long Term Hemodialysis Catheter with Sideholes, 15.5F X 23 cm Symetrex Long Term Hemodialysis Catheter with Sideholes, 15.5F X 28 cm Symetrex Long Term Hemodialysis Catheter with Sideholes, 15.5F X 33 cm Symetrex Long Term Hemodialysis Catheter with Sideholes, 15.5F X 37 cm Symetrex Long Term Hemodialysis Catheter with Sideholes
ApplicantMedical Components, Inc. (dba MedComp)
Product CodeMSD · Gastroenterology, Urology
Decision DateMay 17, 2018
DecisionSESK
Submission TypeTraditional
Regulation21 CFR 876.5540
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Long term, greater than 30 days, vascular access for Hemodialysis and Apheresis treatments.

Device Story

Symetrex™ Long Term Hemodialysis Catheter with Sideholes is a 15.5 French, dual-lumen, radiopaque polyurethane catheter; features polyester retention cuff for subcutaneous anchoring; includes two female luer adapters. Device provides vascular access for chronic hemodialysis and apheresis; inserted percutaneously or via cut down by clinicians. Symmetrical side channels and distal tip configuration separate intake and output flow. Sideholes at distal tip enhance flow dynamics. Device is used in clinical settings for long-term (>30 days) blood access. Healthcare providers use the catheter to facilitate extracorporeal blood treatment; output is the physical conduit for blood flow between patient and dialysis/apheresis machine. Benefits include reliable, long-term vascular access for patients with end-stage renal disease or those requiring apheresis.

Clinical Evidence

Bench testing only. No clinical data. Performance testing included in vitro recirculation studies (demonstrating <1% recirculation in forward/reverse flow), air/liquid leak testing, extension/hub strength, and biocompatibility per ISO 10993-1 (hemocompatibility, genotoxicity, cytotoxicity, irritation, systemic toxicity, implantation).

Technological Characteristics

15.5 French dual-lumen catheter; materials: Carbothane (hub/lumen) with 20% BaSO4, polyester felt (cuff), Pellethane (extension), Isoplast (luer), ABS (ID rings), polypropylene (clamp). EO sterilization. ISO 10555-1, ISO 11607-1/2, ISO 594-1/2 compliance. Radiopaque.

Indications for Use

Indicated for chronic hemodialysis and apheresis in patients requiring long-term vascular access. May be inserted percutaneously or by cut down. Catheters >37cm are indicated for femoral placement.

Regulatory Classification

Identification

A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).

Special Controls

*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided. (ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated. (B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided. (C) Priming volumes must be established. (D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use. (E) Air leakage testing and liquid leakage testing must be conducted. (F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage. (G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern. (H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established. (iii) Performance data must demonstrate the sterility of the device. (iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing. (v) Labeling of implanted blood access devices for hemodialysis must include the following: (A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated. (B) Labeling must specify the forward and reverse recirculation rates. (C) Labeling must provide the arterial and venous priming volumes. (D) Labeling must specify an expiration date. (E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device. (F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient. (G) Labeling must include a patient implant card. (H) The labeling must contain comprehensive instructions for the following: ( *1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;( *2* ) Proper care and maintenance of the device and device exit site;( *3* ) Removal of the device;( *4* ) Anticoagulation;( *5* ) Management of obstruction and thrombus formation; and( *6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following: (A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port. (B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device. (C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use. (vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following: (A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied. (B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties. (C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens. (D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness. (viii) The following must be included for A-V shunt cannulae (with vessel tips): (A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply. (B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation. (C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use. (2) Class II (performance standards) for the nonimplanted blood access device. (3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section. (4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym is in a blue box, and the full name of the agency, "U.S. Food & Drug Administration," is in blue text. May 17, 2018 Medical Components, Inc. Dba Medcomp® Courtnev Nix Regulatory Affairs Manager, North America and Europe 1499 Delp Drive Harleysville, Pennsylvania 19438 Re: K173667 Trade/Device Name: Symetrex™ Long Term Hemodialysis Catheter with Sideholes Regulation Number: 21 CFR 876.5540 Regulation Name: Blood Access Device and Accessories Regulatory Class: Class II Product Code: MSD Dated: April 12, 2018 Received: April 17, 2018 Dear Courtney Nix: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Joyce M. Whang -S for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K173667 Device Name SymetrexTM Long Term Hemodialysis Catheter with Sideholes Indications for Use (Describe) The Symetrex™ Long Term Hemodialysis Catheter with Sideholes is a symmetric tip dual lumen catheter designed for chronic hemodialysis and apheresis. It may be inserted percutaneously or by cut down. Catheters with greater than 37cm implant length are indicated for femoral placement. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image contains a logo for medCOMP. The logo features a red geometric shape that resembles a stylized letter "C" or an abstract building design. Below the geometric shape, the text "medCOMP" is written, with "med" in red and "COMP" in black. The overall design is clean and modern. 1499 Delp Drive Harleysville, PA 19438 Tel: 215-256-4201 Fax. 215-256-1787 www.medcompnet.com Medcomp®: Symetrex™ Long Term Hemodialysis Catheter with Sideholes | Section 6 | 510(k) SUMMARY K173667 Traditional 510K | |------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------| | 1. Submitter Information: | | | Submitter: | Medical Components Inc. dba Medcomp® 1499 Delp Drive Harleysville, PA 19438 Tel: (215) 256-4201, x 2285 Fax: (215) 256-9191 | | Registration Number: | 2518902 | | Contact: | Courtney Nix Cnix@Medcompnet.com Regulatory Affairs Manager: North America and EU | | Date Prepared: | November 27, 2017 | | 2. Proposed or Subject Device Information: | | | Trade Name: | Symetrex™ Long Term Hemodialysis Catheter with Sideholes | | Device: | Catheter, Hemodialysis, Implanted | | Product Code: | MSD | | Regulation Description: | Blood access device and accessories | | C.F.R. Section: | 21 CFR 876.5540 | | Class: | II | | Regulation Medical Specialty and Review Panel: | Gastroenterology/Urology | | 3. Predicate Device Information: | | | 510(k) Number: | K171618 | | 510(k) Holder: | Medcomp® | | | 6-1 | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains a logo for medCOMP. The logo features a red geometric shape that resembles a stylized letter "G" or an abstract building. Below the shape, the word "med" is written in red, followed by "COMP" in black. The overall design is simple and modern. Harlevsville, PA 19438 Tel: 215-256-4201 Fax. 215-256-1787 www.medcompnet.com Medcomp®: Symetrex™ Long Term Hemodialysis Catheter with Sideholes | Trade Name: | Symetrex™ Long Term Hemodialysis Catheter | |------------------------------------------------------|-------------------------------------------| | Device: | Catheter, Hemodialysis, Implanted | | Product Code: | MSD | | Regulation Description: | Blood access device and accessories | | C.F.R Section: | 21 CFR 876.5540 | | Class: | II (Special Controls) | | Regulation Medical<br>Specialty and Review<br>Panel: | Gastroenterology/Urology | #### 4. Device Description: The Symetrex™ Long Term Hemodialysis Catheter with Sideholes is a chronic, 15.5 French, dual lumen, radiopaque catheter made of polyurethane. It has a polyester retention cuff and two female luer adapters. The retention cuff promotes tissue ingrowth to anchor the catheter in the subcutaneous tunnel. The luer adapters are identical in color to indicate the reversibility of this catheter. This catheter features symmetrical side channels with a distal tip configuration designed to separate the intake flow from the output flow in both directions. #### 5. Indications for Use: The Symetrex™ Long Term Hemodialysis Catheter with Sideholes is a symmetric tip dual lumen catheter designed for chronic hemodialysis and apheresis. It may be inserted percutaneously or by cut down. Catheters with greater than 37cm implant length are indicated for femoral placement. #### 6. Intended Use: Long term, greater than 30 days, vascular access for Hemodialysis and Apheresis treatments. #### 7. Comparison to Predicate Devices: Symetrex™ Long Term Hemodialysis Catheter with Sideholes is substantially equivalent to the predicate device, Symetrex™ Long Term Hemodialysis Catheter (K171618), in terms of indications for use, labeling, intended use. anatomical location, performance, components, and method of sterilization. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a red geometric shape that resembles a stylized letter "G" or a house with a missing side. Below the geometric shape, the word "medCOMP" is written in a combination of red and black letters. The "med" portion is in red, while the "COMP" portion is in black. - 1499 Delp Drive Harleysville, PA 19438 Tel: 215-256-4201 Fax. 215-256-1787 www.medcompnet.com Medcomp®: Symetrex™ Long Term Hemodialysis Catheter with Sideholes The difference or changes between the Symetrex™ Long Term Hemodialysis Catheter Sideholes and the predicate, Symetrex™ Long Term Hemodialysis Catheter (K171618) are the sideholes at the distal tip on both the arterial and venous lumen. Comparison table on the following page. ## Table 6.1: 510(K) Summary: Design Comparison Matrix | Device | Proposed Device:<br>Symetrex™ Long Term<br>Hemodialysis Catheter<br>with Sideholes | Predicate Device:<br>Symetrex™ Long Term<br>Hemodialysis Catheter<br>(K171618) | |-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Design | Dual Kidney Shaped<br>Lumen Catheter | Dual Kidney Shape<br>Lumen Catheter | | Tip Design | Symmetric Tip, Sideholes<br>at Distal Tip | Symmetric Tip, No Side<br>Holes | | Lengths | 19cm-42cm | 19cm-42cm | | Indications<br>for Use | The Symetrex™ Long<br>Term Hemodialysis<br>Catheter with Sideholes<br>is a symmetric tip dual<br>lumen catheter designed<br>for chronic hemodialysis<br>and aphersis. It may be<br>inserted percutaneously<br>of by cut down. Catheters<br>with greater than 37cm<br>implant length are<br>indicated for femoral<br>placement. | The Symetrex™ Long<br>Term Hemodialysis<br>Catheter is a symmetric<br>tip dual lumen catheter<br>designed for chronic<br>hemodialysis and<br>aphersis. It may be<br>inserted percutaneously<br>of by cut down. Catheters<br>with greater than 37cm<br>implant length are<br>indicated for femoral<br>placement. | | Catheter<br>O.D. | 15.5FR | 15.5FR | | Sterilization<br>Method | EO Sterilization | EO Sterilization | | Materials | Hub (Green):<br>Carbothane w/ 20%<br>BaSO4<br>Lumen (White):<br>Carbothane w/ 20%<br>BaSO4<br>Cuff:<br>Polyester Felt | Hub (Green):<br>Carbothane w/ 20%<br>BaSO4<br>Lumen (White):<br>Carbothane w/ 20%<br>BaSO4<br>Cuff:<br>Polyester Felt | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for medCOMP. The logo is red and white. The logo has a stylized letter "C" above the word "medCOMP". Harleysville, PA 19438 Tel: 215-256-4201 Fax. 215-256-1787 www.medcompnet.com Medcomp®: Symetrex™ Long Term Hemodialysis Catheter with Sideholes | Extension (Clear):<br>Pellethane<br>Barbed Luer (Green) :<br>Isoplast<br>ID Ring (White with red ink):<br>ABS<br>ID Ring (White with blue ink):<br>ABS<br>Clamp (White):<br>Polypropylene | Extension (Clear):<br>Pellethane<br>Barbed Luer (Green) :<br>Isoplast<br>ID Ring (White with red ink):<br>ABS<br>ID Ring (White with blue ink):<br>ABS<br>Clamp (White):<br>Polypropylene | | | | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|-----------------| | Components | Dilator, Stylet, Tunneler,<br>Valved Introducer, and<br>Guidewire | Dilator, Stylet, Tunneler,<br>Valved Introducer, and<br>Guidewire | | | | Packaging<br>Systems | Clamshell (Catheters<br>under 33cm):<br>One unit gets a snap lid<br>one unit is sealed into a<br>header bag<br>5 Header bags are<br>placed in a shipping<br>carton<br><br>Translumbar tray<br>(Catheters greater than<br>33cm):<br>One unit gets a snap lid<br>One unit is sealed into a<br>header bag.<br>5 units are placed into a<br>shipping carton. | Clamshell (Catheters<br>under 33cm):<br>One unit gets a snap lid<br>one unit is sealed into a<br>header bag<br>5 Header bags are<br>placed in a shipping<br>carton<br><br>Translumbar tray<br>(Catheters greater than<br>33cm):<br>One unit gets a snap lid<br>One unit is sealed into a<br>header bag.<br>5 units are placed into a<br>shipping carton. | | | | Performance<br>Testing | Priming Volumes: | | | | | | Tip to Cuff<br>Length | Lumen<br>Volume | Tip to Cuff<br>Length | Lumen<br>Volume | | | 19cm | 2.1cc | 19cm | 2.1cc | | | 23cm | 2.3cc | 23cm | 2.3cc | | | 28cm | 2.5cc | 28cm | 2.5cc | | | 33cm | 2.6cc | 33cm | 2.6cc | | | 37cm | 2.8cc | 37cm | 2.8cc | | | 42cm | 3.2cc | 42cm | 3.2cc | | | Recirculation: | | Recirculation: | | Medcomp®: Symetrex™ Long Term Hemodialysis Catheter with Sideholes Section 6: 510 Summary 6-4 {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a red geometric shape that resembles a stylized letter "G" or a box with an open side. Below the geometric shape, the word "medCOMP" is written in a stylized font, with "med" in red and "COMP" in black. - 1499 Delp Drive Harleysville, PA 19438 Tel: 215-256-4201 Fax. 215-256-1787 www.medcompnet.com Medcomp®: Symetrex™ Long Term Hemodialysis Catheter with Sideholes | Symetrex™ Long Term<br>Hemodialysis Catheter<br>with Sideholes has less<br>than 1% recirculation in<br>forward and reverse flow<br>when tested in vitro | Symetrex™ Long Term<br>Hemodialysis Catheter<br>has less than 1%<br>recirculation in forward<br>and reverse flow when<br>tested in vitro | |---------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------| |---------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------| #### 8. Bench / Performance Data / Non-Clinical Testing: The results of performance testing, in conjunction with the substantial equivalence claims, effectively demonstrate the proposed device, Symetrex™ Long Term Hemodialysis Catheter with Sideholes, is equivalent to the predicate device, Symetrex™ Long Term Hemodialysis Catheter (K171618). The performance testing was performed in accordance with the following standards: #### Table 6.2: Performance Standards | Standard | Standard Title | Revision/ Date | Performance<br>Testing | |-------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|--------------------------------------------------------------------------------| | ISO 10555-1 | Intravascular<br>catheters --<br>sterile and single-<br>use<br>intravascular<br>catheters -- part<br>1: general<br>requirements. | Second Edition<br>2013-06-15 | Air Leak, Catheter<br>Leak, Extension-<br>Hub, Extrusion-<br>Hub, Gravity Flow | | ISO 11607-1 | Packaging for<br>terminally<br>sterilized medical<br>devices - part<br>1: requirements<br>for materials,<br>sterile barrier<br>systems and<br>packaging<br>systems<br>[including:<br>amendment 1<br>(2014)]. | First Edition<br>2006-04-15 | Shipping and Shelf<br>Life testing | | ISO 11607-2 | Packaging for<br>terminally<br>sterilized medical<br>devices - part<br>2: validation<br>requirements for | First Edition<br>2006-04-15 | Shipping and Shelf<br>Life testing | {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a red geometric shape that resembles a stylized letter "G" or an abstract building. Below the geometric shape, the word "medCOMP" is written in a stylized font, with "med" in red and "COMP" in black. Harleysville, PA 19438 Tel: 215-256-4201 Fax. 215-256-1787 www.medcompnet.com Medcomp®: Symetrex™ Long Term Hemodialysis Catheter with Sideholes | | forming, sealing<br>and assembly<br>processes<br>[including:<br>amendment 1<br>(2014)]. | | | |-----------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------| | ISO 594-1 | conical fittings<br>with a 6% (luer)<br>taper for<br>syringes, needles<br>and<br>certain other<br>medical<br>equipment - part<br>1: general<br>requirements. | First edition<br>1986-06-15, | Gauging | | ISO 594-2 | Conical fittings<br>with a 6% (luer)<br>taper for<br>syringes, needles<br>and<br>certain other<br>medical<br>equipment - part<br>2: lock fittings. | Second Edition<br>1998-09-01 | Liquid Leakage<br>Air Leakage<br>Separation Force<br>Unscrewing Torque<br>Ease of Assembly<br>Resistance to<br>Overriding<br>Stress Cracking | ## 9. Biocompatibility: Biocompatibility was performed for the Symetrex™ Long Term Hemodialysis Catheter with Sideholes per ISO 10993-1 for a blood implant device with permanent exposure (i.e. > 30 days). Biocompatibility was performed on the final, finished device. The biological end points include: - Hemocompatibility: 0 - · ISO 10993-4 Biological Evaluation Of Medical Devices Part 4: Selection Of Tests For Interaction With Blood - · ASTM F 756-08, Standard Practice for Assessment of Hemolytic Properties of Materials, 2008 - 0 Genotoxicity: - o ISO 10993-3, Biological Evaluation of Medical Devices Part 3: Tests for Genotoxicity, Carcinogenicity and Reproductive Toxicity. - 0 Cytotoxicity: - o AAMI / ANSI / ISO 10993-5: Biological evaluation of medical devices part 5: tests for in vitro cytotoxicity. {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a red geometric shape that resembles a stylized letter "M" or an abstract building. Below the geometric shape, the word "medCOMP" is written in a lowercase, italicized font, also in red. The overall design is simple and modern, with a focus on the geometric shape and the company name. Harlevsville, PA 19438 Tel: 215-256-4201 Fax. 215-256-1787 www.medcompnet.com Medcomp®: Symetrex™ Long Term Hemodialysis Catheter with Sideholes - Irritation/Intracutaneous: 0 - o ISO 10993-10 Biological Evaluation of Medical Devices Part 10: Tests for Irritation and Skin Sensitization. - Acute Systemic Toxicity: 0 - o ISO 10993-11 Biological Evaluation of Medical Devices Part 11: Tests for Systemic Toxicity - Implantation: 0 - o ISO 10993-6 Biological Evaluation of Medical Devices Part 6: Tests for Local Effects after Implantation. - ASTM F763-04 Standard Practice For Short-Term Screening of Implant O Materials. - ASTM F981-04 Standard Practice For Assessment of Compatibility of o Biomaterials for Surgical Implants With Respect To Effect of Materials on Muscle and Bone. - Additional Testing: 0 - o ISO 10993-18 Biological Evaluation of Medical Devices Part 18: Chemical Characterization of Materials - ASTM D4128-06 Standard Guide for Identification of Quantitation of O Organic Compounds in Water by Combined Gas Chromatography and Electron Impact Mass Spectrometry - ASTM D1971-11 Standard Practices for Digestion of Water Samples for O Determination of Metals by Flame Atomic Absorption, Graphite Furnace Atomic Absorption, Plasma Emission Spectroscopy, or Plasma Mass Spectroscopy - ISO 10993-17 Biological Evaluation of Medical Devices Part 17: O Establishment of Allowable Limits for Leachable Substances ## 10. Summary of Substantial Equivalence: In conclusion, the proposed device, Symetrex™ Long Term Hemodialysis Catheter with Sideholes, is considered substantially equivalent to the predicate device. Symetrex™ Long Term Hemodialysis (K171618) as demonstrated through non-clinical testing performed.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...