Cook Turbo-Flo HD Acute Hemodialysis Catheter Set/Tray

K161504 · Cook Incorporated · MPB · Jan 6, 2017 · Gastroenterology, Urology

Device Facts

Record IDK161504
Device NameCook Turbo-Flo HD Acute Hemodialysis Catheter Set/Tray
ApplicantCook Incorporated
Product CodeMPB · Gastroenterology, Urology
Decision DateJan 6, 2017
DecisionSEKD
Submission TypeSpecial
Regulation21 CFR 876.5540
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Cook Turbo-Flo HD Acute Hemodialysis Catheter Set/Tray is intended for acute hemodialysis, apheresis and hemofiltration via percutaneous insertion into the subclavian, jugular, or femoral veins. The device is a short-term use catheter (less than 30 days).

Device Story

Short-term, 12 Fr, radiopaque polyurethane central venous catheter; features two independent, noncommunicating lumens. Available in 15, 20, and 25 cm lengths. Used for acute hemodialysis, apheresis, and hemofiltration. Inserted percutaneously into subclavian, jugular, or femoral veins by clinicians. Device facilitates blood access for extracorporeal therapy. Modification from predicate involves removal of external hydrophilic coating. Benefits include reliable blood flow for acute renal replacement therapy.

Clinical Evidence

Bench testing and animal model validation performed. Animal study simulated clinical use, evaluating preparation, introduction, pushability, trackability, flexibility, radiopacity, blood draw, infusion, interaction, withdrawal, and inspection; all parameters rated 'adequate' or 'good'. Biocompatibility testing included cytotoxicity, direct/indirect hemolysis, complement activation, partial thromboplastin time, and in vitro hemocompatibility. Results confirmed modification does not impact biocompatibility.

Technological Characteristics

12 Fr radiopaque polyurethane central venous catheter; two independent, noncommunicating lumens; 15, 20, and 25 cm lengths. Non-implanted blood access device. No hydrophilic coating. Sterilization method not specified.

Indications for Use

Indicated for patients requiring acute hemodialysis, apheresis, or hemofiltration via percutaneous insertion into subclavian, jugular, or femoral veins for short-term use (<30 days).

Regulatory Classification

Identification

A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).

Special Controls

*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided. (ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated. (B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided. (C) Priming volumes must be established. (D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use. (E) Air leakage testing and liquid leakage testing must be conducted. (F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage. (G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern. (H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established. (iii) Performance data must demonstrate the sterility of the device. (iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing. (v) Labeling of implanted blood access devices for hemodialysis must include the following: (A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated. (B) Labeling must specify the forward and reverse recirculation rates. (C) Labeling must provide the arterial and venous priming volumes. (D) Labeling must specify an expiration date. (E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device. (F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient. (G) Labeling must include a patient implant card. (H) The labeling must contain comprehensive instructions for the following: ( *1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;( *2* ) Proper care and maintenance of the device and device exit site;( *3* ) Removal of the device;( *4* ) Anticoagulation;( *5* ) Management of obstruction and thrombus formation; and( *6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following: (A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port. (B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device. (C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use. (vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following: (A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied. (B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties. (C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens. (D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness. (viii) The following must be included for A-V shunt cannulae (with vessel tips): (A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply. (B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation. (C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use. (2) Class II (performance standards) for the nonimplanted blood access device. (3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section. (4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing right, representing health and human services. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 6, 2017 Cook Incorporated Steven Lawrie, MS, MA, RAC Regulatory Affairs Manager 750 Daniels Way Bloomington, IN 47404 Re: K161504 > Trade/Device Name: Cook Turbo-Flo HD Acute Hemodialysis Catheter Set/Tray Regulation Number: 21 CFR§ 876.5540 Regulation Name: Blood Access Device and Accessories Regulatory Class: II Product Code: MPB Dated: December 7, 2016 Received: December 8, 2016 Dear Steven Lawrie: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. In addition, we have determined that your device kit contains Lidocaine which is subject to regulation as a drug. Our substantially equivalent determination does not apply to the drug component of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug component. For information on applicable Agency requirements for marketing this drug, we suggest you contact: Director, Division of Drug Labeling Compliance Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0101 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. {2}------------------------------------------------ You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K161504 #### Device Name Cook Turbo-Flo® HD Acute Hemodialysis Catheter Set/Tray Indications for Use (Describe) The Cook Turbo-Flo® HD Acute Hemodialysis Catheter Set/Tray is intended for acute hemodialysis, apheresis and hemofiltration via percutaneous insertion into the subclavian, jugular, or femoral veins. The device is a short-term use catheter (less than 30 days). | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image is a logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in white, block letters in the upper portion of the logo. Below the word "COOK" is the word "MEDICAL" in white letters. ## 510(k) SUMMARY ## Cook Turbo-Flo® HD Acute Hemodialysis Catheter Set/Trav 21 CFR §880.5570 Date Prepared: 6 January 2017 | Submitted By: | Cook Incorporated | |-----------------------|------------------------------------------------------------| | Applicant: | Cook Incorporated | | Contact: | Steven Lawrie | | Applicant Address: | Cook Incorporated | | | 750 Daniels Way | | | Bloomington, IN 47404 | | Contact Phone Number: | (812) 335-3575 x104518 | | Contact Fax Number: | (812) 332-0281 | | Device Information: | | | Trade Name: | Cook Turbo-Flo® HD Acute Hemodialysis Catheter<br>Set/Tray | | Common Name: | Blood access device and accessories | | Classification Name: | Catheter, Hemodialysis, Non-implanted | | Regulation: | 21 CFR §876.5540 | | Product Code: | MPB | ## Predicate Device: - K122091 Cook Turbo-Flo® HD Acute Hemodialysis Catheter Set/Tray . #### Device Description: The Cook Turbo-Flo® HD Acute Hemodialysis Catheter Set/Tray is a short-term, 12 Fr, radiopaque polyurethane central venous catheter with two independent, noncommunicating lumens. Lengths of 15, 20, and 25 cm are available based on the anatomical needs of the patient. ## Intended Use: The Cook Turbo-Flo HD Acute Hemodialysis Catheter Set/Tray is intended for acute hemodialysis, apheresis and hemofiltration via percutaneous insertion into the subclavian, jugular, or femoral veins. The device is a short-term use catheter (less than 30 days). #### Comparison to Predicates: The Cook Turbo-Flo® HD Acute Hemodialysis Catheter Set/Tray is substantially equivalent to the predicate device, the Cook Turbo-Flo® HD Acute Hemodialysis Catheter Set/Tray (K122091) in that these devices have identical intended use, catheter materials, and technological characteristics. The only difference as compared to the {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, uppercase letters, stacked on top of the word "MEDICAL" in white, uppercase letters. Both words are set against a red background. The word "COOK" is in a square, while the word "MEDICAL" is in a rectangle with a diagonal cut on the left side. predicate is the removal of the hydrophilic coating from the external surface of the catheter shaft. # Technological Characteristics: The following tests have been conducted to ensure reliable design and performance under the specified design requirements: - . Validation testing - When tested in an animal model that simulates clinical use, the rating scale of the following performance parameters should be "adequate" or "good" in a three-point scale: preparation, introduction, pushability, trackability, flexibility, radiopacity, blood draw, infusion, interaction, withdrawal, and inspection. The pre-determined acceptance criterion was met. - Biocompatibility testing The results of testing (cytotoxicity, direct and indirect . hemolysis, complement activation, partial thromboplastin time, and in vitro hemocompatibility) indicate that the modification to the subject device does not impact its biocompatibility. ## Conclusion: After conducting appropriate design control activities and performing an appropriate validation test, it is concluded that the proposed Cook Turbo-Flo® HD Acute Hemodialysis Catheter Set/Tray has met the design input requirements based on the intended use and that the modification to the device does not raise new questions of safety or effectiveness as compared to the predicate device. Therefore, the Cook Turbo-Flo® HD Acute Hemodialysis Catheter Set/Tray has been demonstrated to be substantially equivalent to the predicate device, the Cook Turbo-Flo® HD Acute Hemodialysis Catheter Set/Tray (K122091).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...