← Product Code [KOC](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KOC) · K141997

# CRIT-LINE IV SYSTEM (K141997)

_Fresenius Medical Care · KOC · Apr 1, 2015 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KOC/K141997

## Device Facts

- **Applicant:** Fresenius Medical Care
- **Product Code:** [KOC](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KOC.md)
- **Decision Date:** Apr 1, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5820
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The intended use of the Crit-Line IV System is as a continuous real-time monitor of hematocrit, oxygen saturation, and percent change in blood volume.

## Device Story

Crit-Line IV System is a continuous, real-time blood monitoring system used in dialysis clinics. It consists of a monitor, a Crit-Line Clip (CLiC) sensor, and a single-use blood chamber. The CLiC sensor attaches to the blood chamber in the dialysis tubing set to perform non-invasive optical measurements of hematocrit, oxygen saturation, and percent change in blood volume (ABV%). Data is transmitted via USB serial signals to the monitor. The monitor features a touchscreen interface for data display, user ID entry, and language selection. It stores 30 days of session data, which can be transferred via USB or wireless ZigBee module. Clinicians use the real-time output to adjust fluid removal rates during dialysis, aiming to maximize fluid removal while minimizing patient complications like nausea, cramping, and vomiting. The system is designed for mounting on an IV pole.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance testing included software verification, electrical safety testing, electromagnetic compatibility (EMC) testing, intentional EM emissions immunity simulation, coexistence testing with RFID equipment, ship testing, and usability testing.

## Technological Characteristics

System comprises a dedicated monitor, CLiC sensor, and single-use blood chamber. Sensing principle: in-line non-invasive optical measurement. Connectivity: USB serial interface for sensor data; ZigBee wireless module for data download/printing. Form factor: compact unit mountable to IV pole. Software: touchscreen interface, file system for 30-day data storage. Power: not specified. Sterilization: not specified.

## Regulatory Identification

A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.

## Predicate Devices

- Crit-Line Clip Monitor (CLiC) ([K121599](/device/K121599.md))
- Crit-Line III Monitor ([K972470](/device/K972470.md))
- 2008T Hemodialysis Machine with optional Crit-Line Clip Monitor (CLiC) ([K131908](/device/K131908.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three human profiles facing right, stacked on top of each other, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 1, 2015

Fresenius Medical Care Denise Oppermann Senior Director, Regulatory Affairs, Devices 920 Winter Street Waltham, MA 02451

Re: K141997

> Trade/Device Name: Crit-Line IV System Regulation Number: 21 CFR§ 876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: KOC Dated: March 12, 2015 Received: March 13, 2015

Dear Denise Oppermann,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Herbert P. Lerner -S

Benjamin R. Fisher, Ph.D. for Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K141997

Device Name Crit-Line IV System

### Indications for Use (Describe)

The Crit-Line IV System is used to non-invasively to measure hematocrit, oxygen saturation and percent change in blood volume. The Crit-Line Clip (CLiC) measures hematocrit, percent change in blood volume and oxygen saturation in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dial. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e. increases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.

# Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

# FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 5. 510(K) SUMMARY

This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR § 807.92.

#### 5.1. Submitter's Information

| Name:                | Fresenius Medical Care Renal Therapies Group, LLC (FMC-RTG)        |
|----------------------|--------------------------------------------------------------------|
| Address:             | 920 Winter Street<br>Waltham, MA 02451-1457                        |
| Phone:               | (781) 699-4479                                                     |
| Fax:                 | (781) 699-9635                                                     |
| Contact Person:      | Denise Oppermann,<br>Senior Director, Regulatory Affairs - Devices |
| Date of Preparation: | 07-22-2014                                                         |

#### 5.2. Device Name

| Trade Name:                           | Crit-Line IV System                      |
|---------------------------------------|------------------------------------------|
| Common Name:                          | Accessories, blood circuit, hemodialysis |
| Classification Name:                  | Hemodialysis system and accessories      |
| Regulatory Class:                     | Class II per 21 CFR §876.5820            |
| Product Code/Classification<br>Panel: | KOC/ Gastroenterology - Urology          |

#### 5.3. Legally Marketed Predicate Device

Crit-Line Clip Monitor (CLiC), K121599

Crit-Line III Monitor, K972470

2008T Hemodialysis Machine with optional Crit-Line Clip Monitor (CLiC), K131908

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Fresenius Medical Care. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three downward-pointing chevrons stacked on top of each other. The text is in a bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.

#### 5.4. Device Description

The intended use of the Crit-Line IV System is as a continuous real-time blood monitoring system for displaying information measured by the CLiC sensor, including hematocrit, oxygen saturation, and percent change in blood volume. The Crit-Line IV System is comprised of the Crit-Line IV monitor (the subject of this submission), the Crit-Line Clip (CLiC) sensor, and the Crit-Line Clip Blood Chamber. The touchscreen interface and display of the Crit-Line IV monitor is used with the hardware and software of the CLiC sensor, which is a separate component of the Crit-Line Clip (CLiC) Monitor, K121599. Information displayed on the Crit-Line IV monitor may be viewed real-time during a dialysis treatment. There are additional display features incorporated into the Crit-Line IV System, which are similar to that of the Crit-Line III system (K972470) and the 2008T Hemodialysis Machine with (optional) CLiC (K131908). For this reason, the Crit-Line Clip Monitor (CLiC: K121599), is the predominant predicate and the 2008T Hemodialysis Machine with (optional) CLiC (K131908) and the Crit-Line III (K972470) are reference predicate devices.

#### 5.5. Indications for Use

The Crit Line IV System is used to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume. The Crit-Line Clip (CLiC) measures hematocrit, percent change in blood volume and oxygen saturation in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the clinician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e. increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common complications of dialysis which include nausea. cramping and vomiting.

#### 5.6. Intended Use

The intended use of the Crit-Line IV System is as a continuous real-time monitor of hematocrit, oxygen saturation, and percent change in blood volume.

#### 5.7. Technological Characteristics

Fundamental Scientific Technology/Operating Principle:

The Crit-Line IV System consists of a monitor running display software for data measured by the CLiC sensor. The data is sent as USB serial signals from the CLiC sensor to the monitor. The Crit-Line IV monitor connects with the CLiC sensor via the USB computer interface. The CLiC sensor connects to the external transparent section of a single-use Crit-Line Clip Blood Chamber. The Crit-Line Clip Blood Chamber provides in-line optic measurement access (non-invasive) to blood via connection to the dialysis tubing set.

The Crit-Line IV monitor displays and stores data generated by the CLiC sensor.

- The Crit-Line IV monitor is a dedicated compact display unit which can be

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Fresenius Medical Care. On the left side of the logo, there is a blue symbol that consists of three downward-pointing chevrons stacked on top of each other. To the right of the symbol, the words "FRESENIUS" and "MEDICAL CARE" are written in blue, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.

mounted to an IV pole. The monitor continuously displays volume (ABV%). hematocrit (Hct), estimated hemoglobin (Hb, Hct based estimate), and oxygen saturation (02 Sat).

- A ZigBee wireless module allows the user to wirelessly print or download data . collected during the monitoring session.
The display software of the Crit-Line monitor provides the following user interfaces:

- Touch screen. Virtual buttons on the screen allow the user to navigate through o the menu selections. The screen also allows for data entry, such as user ID, and for language selection.
- o File System. Data collected during the monitoring sessions will be maintained for a period of 30 days. Data stored in the patient run files are available for transfer to host computer systems or to a USB memory stick.

The above bulleted items represent the main differences between the function of the display software described in K121599 and K131908 as compared with the proposed monitor software of the Crit-Line IV System. Additional alerts, such as a "ΔΒΥ% greater than 8%/hr" alert, were added to enhance the user interface. The intended use and use environment are the same as that of the predicate monitoring devices. The algorithms used in the proposed monitor software for calculating △BV% and estimating hemoglobin (Hb) are the same algorithms used in the software driver provided with the predicate Crit-Line Clip (CLiC), K121599.

#### Performance Data 5.8.

Performance testing requirements were determined through the application of a risk management process and applicable performance standards. Performance testing included software verification, electrical safety testing, electromagnetic compatibility (EMC) testing, intentional EM emissions immunity simulation testing, coexistence testing with RFID equipment, ship testing, and usability testing.

#### 5.9. Conclusion

The information provided in this submission demonstrates the Crit-Line IV System functions as intended and is substantially equivalent to the predicate devices. The test results demonstrate that the proposed device does not raise any new concerns with regard to safety or effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KOC/K141997](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KOC/K141997)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
