← Product Code [KDN](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDN) · K243384

# Belzer UW® Cold Storage Solution (BTLBUW-001) (K243384)

_Bridge TO Life · KDN · Mar 31, 2025 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDN/K243384

## Device Facts

- **Applicant:** Bridge TO Life
- **Product Code:** [KDN](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDN.md)
- **Decision Date:** Mar 31, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5880
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

Belzer UW® Cold Storage Solution (BTLBUW-001) is indicated for the flushing and cold storage of kidney, liver and pancreas organs at the time of organ removal from the donor in preparation for storage, transportation and eventual transplantation into a recipient. Belzer UW® Cold Storage Solution (BTLBUW-001) is not indicated for continuous machine perfusion.

## Device Story

Belzer UW® Cold Storage Solution is a sterile, non-pyrogenic, hypothermic flushing and storage solution for donor organs (kidney, liver, pancreas). It functions by cooling organ vasculature to slow biological deterioration during storage and transport. The solution is supplied in non-PVC (EVA) bags (500 mL, 1 L, 2 L) and is intended for single use by trained healthcare staff. It is not for continuous machine perfusion. The solution is pre-cooled (2° to 6°C) before use. By maintaining hypothermic conditions, the device preserves organ viability for transplantation, benefiting patients awaiting organ grafts.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993-1/2/5/10/12), sterility validation, and chemical composition analysis.

## Technological Characteristics

Liquid solution; composition includes Pentafraction, Lactobionic Acid, Potassium Phosphate, Magnesium Sulfate, Raffinose, Adenosine, Allopurinol, Glutathione, Potassium Hydroxide, Sodium Hydroxide. Osmolarity 320 mOsmol/kg; pH ~7.4. Primary container: Ethylene Vinyl Acetate (EVA) bags, latex/PVC/DEHP-free. Sterilization: Aseptic fill with 0.2 µm and two 0.1 µm membrane filtration. Storage: 2° to 25°C.

## Regulatory Identification

An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.

## Predicate Devices

- CoStorSol® Solution ([K091245](/device/K091245.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 31, 2025

Bridge to Life Ltd Mark Harper Compliance Manager 707 Skokie Boulevard, Suite 340 Northbrook, Illinois 60062

Re: K243384

Trade/Device Name: Belzer UW® Cold Storage Solution (BTLBUW-001) Regulation Number: 21 CFR 876.5880 Regulation Name: Isolated kidney perfusion and transport system and accessories Regulatory Class: Class II Product Code: KDN, KDL Dated: February 19, 2025 Received: February 19, 2025

Dear Mark Harper:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.

{2}------------------------------------------------

See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Gema Gonzalez -S

Maura Rooney Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K243384

Device Name Belzer UW® Cold Storage Solution (BTLBUW-001)

Indications for Use (Describe)

Belzer UW® Cold Storage Solution (BTLBUW-001) is indicated for the flushing and cold storage of kidney, liver and pancreas organs at the time of organ removal from the donor in preparation for storage, transportation and eventual transplantation into a recipient. Belzer UW® Cold Storage Solution (BTLBUW-001) is not indicated for continuous machine perfusion.

Type of Use (Select one or both, as applicable)

| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)                 |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# K243384: 510(k) Summary

### 1.0 SUBMITTER

# Bridge to Life Ltd

707 Skokie Blvd. Ste. 340 Northbrook, IL 60062

| Contact Person | : | Mark Harper        |
|----------------|---|--------------------|
| Phone          | : | 847-796-3070       |
| Fax            | : | 803-753-9798       |
| Email          | : | m.harper@b2ll.com  |
| Date Prepared  | : | October 30th, 2024 |

### 2.0 DEVICE

| Common Name :           | Organ Cold Storage Solution                                    |
|-------------------------|----------------------------------------------------------------|
| Proprietary Name :      | Belzer UW® Cold Storage Solution (BTLBUW-001)                  |
| Classification Name :   | Isolated kidney perfusion and transport system and accessories |
| Device Classification : | Class II, 876.5880                                             |
| Product Code :          | KDN, KDL                                                       |

### 3.0 PREDICATE DEVICE

| Company                     | Device              | K Number | Date Cleared  |
|-----------------------------|---------------------|----------|---------------|
| Preservation Solutions, Inc | CoStorSol® Solution | K091245  | June 17, 2009 |

### 4.0 DEVICE DESCRIPTION

Belzer UW® Cold Storage Solution (BTLBUW-001) is a clear to light yellow, sterile, nonpyrogenic solution for hypothermic flushing and storage of organs. The solution has an approximate calculated osmolarity of 320 mosmol/kg, a sodium concentration of 29 mEq/L, a potassium concentration of 125 mEq/L, and a pH of approximately 7.4 at 20°C.

The solution should be used only by medical healthcare staff, adequately trained according to established operating protocols.

The Belzer UW® Cold Storage Solution (BTLBUW-001) is supplied in non-PVC bags: 500 mL bags, shelf carton of 6; 1000 mL in 1-liter bags, shelf carton of 10; and 2000 mL in 2-liter bags, shelf carton of 5. The solution is sterile and is intended for one single use.

The solution should be stored indoors at temperatures controlled between 2° and 25°C (36° and 77°F) until use. Excessive heat and freezing should be avoided. Belzer UW® Cold Storage Solution (BTLBUW-001) should not be used if discolored or if obvious particulate matter, precipitates, or contamination are evident in the solution.

Bridge to Life's Belzer UW® Cold Storage Solution (BTLBUW-001) is substantially CONFIDENTIAL Page 1 of 6 OCTOBER 2024

{5}------------------------------------------------

equivalent in composition, safety and efficacy to currently marketed CoStorSol® Solution, which was cleared by the US FDA in prior 510(k) submission K091245

Table 1 provides information about the composition of Bridge to Life's Belzer UW® Cold Storage Solution (BTLBUW-001).

| Component                      | g/L     | Mmol/L |
|--------------------------------|---------|--------|
| Pentafraction                  | 50      | N/A    |
| Lactobionic Acid (as Lactone)  | 35.83   | 105    |
| Potassium Phosphate monobasic  | 3.4     | 25     |
| Magnesium Sulfate heptahydrate | 1.23    | 5      |
| Raffinose pentahydrate         | 17.83   | 30     |
| Adenosine                      | 1.34    | 5      |
| Allopurinol                    | 0.136   | 1      |
| Total Glutathione              | 0.922   | 3      |
| Potassium Hydroxide            | 5.61    | 100    |
| Sodium Hydroxide, 5N           | 5.0mL/L | 25     |
| Water for Injection            |         | q.s.   |

# Table 1: Composition of Bridge to Life's Belzer UW® Cold Storage Solution (BTLBUW-001)

### 5.0 INDICATION FOR USE

Bridge to Life's Belzer UW® Cold Storage Solution (BTLBUW-001) is indicated for the flushing and cold storage of kidney, liver and pancreas organs at the time of organ removal from the donor in preparation for storage, transportation and eventual transplantation into a recipient. Belzer UW® Cold Storage Solution (BTLBUW-001) is not indicated for continuous machine perfusion.

### 6.0 COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The following table provides a comparison of attributes between the proposed device and the predicate device:

|                                                                          | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Predicate Device                                                                                                                                                                                                                                                                                                                                                                  | Comparison                                                                                                                                                                                                                                                                                          |
|--------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device                                                                   | BTL's Belzer UW® Cold<br>Storage Solution<br>(BTLBUW-001)                                                                                                                                                                                                                                                                                                                                                                                                                                       | CoStorSol® Solution                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                     |
| 510k Number                                                              | K243384                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | K091245                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                     |
| Manufacturer                                                             | Bridge to Life<br>707 Skokie Blvd, Ste. 340<br>Northbrook, IL 60062                                                                                                                                                                                                                                                                                                                                                                                                                             | Preservation Solutions, Inc.,<br>980 Proctor Drive, Elkhorn,<br>Wisconsin 32121                                                                                                                                                                                                                                                                                                   | -                                                                                                                                                                                                                                                                                                   |
| Classification &<br>Product Code                                         | 876.5880; KDN, KDL                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 876.5880; KDN                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                |
| Device Regulation<br>Description                                         | Isolated kidney perfusion and<br>transport system and accessories                                                                                                                                                                                                                                                                                                                                                                                                                               | Isolated kidney perfusion and<br>transport system and<br>accessories                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                                                                |
| Common Name                                                              | Organ Storage Solution                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Organ Storage Solution                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                                                |
| DESCRIPTION,<br>INDICATIONS,<br>& INTENDED<br>USE                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                     |
| Device Description                                                       | BTL's Belzer UW® Cold Storage CoStorSol® is a clear to light<br>Solution (BTLBUW-001) is a<br>clear to light yellow, sterile, non-pyrogenic solution for<br>hypothermic flushing and<br>storage of organs.<br><br>The solution has an approximate<br>calculated osmolarity of 320<br>mosmol/kg, a sodium<br>concentration of 29 mEq/L, a<br>potassium concentration of 125<br>mEq/L, and a pH of<br>approximately 7.4 at 20°C.<br><br>The solution is packaged in 500-<br>mL, 1-L and 2-L bags. | yellow, sterile, non-pyrogenic<br>solution for hypothermic<br>flushing and storage of organs.<br><br>The solution has an<br>approximate calculated<br>osmolarity of 320 mosmol/kg,<br>a sodium concentration of 29<br>mEq/L, a potassium<br>concentration of 125 mEq/L,<br>and a pH of approximately 7.4<br>at 20°C.<br><br>The solution is packaged in 1-<br>Liter bags          | Same<br>Subject and predicate<br>devices are both used as<br>organ preservation<br>solutions. Devices are<br>used in the organ<br>procurement and<br>transplant process per<br>standard organ<br>preservation practices.                                                                            |
| Indications for Use                                                      | Belzer UW® Cold Storage<br>Solution (BTLBUW-001) is<br>indicated for the flushing and<br>cold storage of kidney, liver<br>and pancreas organs at the time<br>of organ removal from the<br>donor in preparation for<br>storage, transportation and<br>eventual transplantation into a<br>recipient. Belzer UW® Cold<br>Storage Solution (BTLBUW-<br>001) is not indicated for<br>continuous machine perfusion.                                                                                   | CoStorSol® indicated for the<br>flushing and cold storage of<br>kidney, liver and pancreas<br>organs at the time of organ<br>removal from the donor in<br>preparation for storage,<br>transportation and eventual<br>transplantation into a<br>recipient.                                                                                                                         | Same<br>Subject and predicate<br>devices are both used as<br>organ preservation<br>solutions. All devices are<br>used at cold<br>temperatures, to slow<br>biological deterioration<br>in organs removed from<br>their physiological<br>environment per<br>standard organ<br>preservation practices. |
| Intended Use                                                             | Belzer UW® Cold Storage<br>Solution (BTLBUW-001) is<br>intended for the flushing and<br>cold storage of kidney, liver<br>and pancreas organs at the time<br>of organ removal from the<br>donor in preparation for<br>storage, transportation and<br>eventual transplantation into a<br>recipient.                                                                                                                                                                                               | Intended for the flushing and<br>cold storage of kidney, liver<br>and pancreas organs at the<br>time of organ removal from<br>the donor in preparation for<br>storage, transportation and<br>eventual transplantation into a<br>recipient.                                                                                                                                        | Same                                                                                                                                                                                                                                                                                                |
| TECNOLOGICAL<br>CHARACTERISTICS                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                     |
| Storage<br>Temperature                                                   | 2° to 25° C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 2° to 25° C                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                                                |
| Pre-Cooling                                                              | Pre-cool solution prior to use<br>(2° to 6° C)                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Pre-cool solution prior to use<br>(2° to 6° C)                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                |
| Maintain Cold<br>Organ<br>Temperature<br>During Storage<br>and Transport | Directly cools the organ<br>vasculature during hypothermic<br>storage and transportation (not<br>for continuous perfusion) to the<br>patient.                                                                                                                                                                                                                                                                                                                                                   | Directly cools the organ<br>vasculature during<br>hypothermic storage and<br>transportation (not for<br>continuous perfusion) to the<br>patient.                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                |
| Product State                                                            | Liquid - Solution                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Liquid - Solution                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                |
| Composition                                                              | 50 g/L Pentafraction<br>35.83 g/L Lactobionic Acid (as<br>Lactone)<br>3.4 g/L Potassium Phosphate<br>monobasic<br>1.23 g/L Magnesium Sulfate<br>heptahydrate<br>17.83 g/L Raffinose<br>pentahydrate<br>1.34 g/L Adenosine<br>0.136 g/L Allopurinol<br>0.922 g/L Total Glutathione<br>5.61 g/L Potassium Hydroxide<br>5.0 mL/L Sodium Hydroxide,<br>5N<br>Water for Injection q.s.                                                                                                               | 50 g/L Pentafraction<br>35.83 g/L Lactobionic Acid (as<br>Lactone)<br>3.4 g/L Potassium Phosphate<br>monobasic<br>1.23 g/L Magnesium Sulfate<br>heptahydrate<br>17.83 g/L Raffinose<br>pentahydrate<br>1.34 g/L Adenosine<br>0.136 g/L Allopurinol<br>0.922 g/L Total Glutathione<br>5.61 g/L Potassium Hydroxide<br>5.0 mL/L Sodium Hydroxide,<br>5N<br>Water for Injection q.s. | Same                                                                                                                                                                                                                                                                                                |
| Osmolality                                                               | 320 mOsmol/kg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 320 mOsmol/kg                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                |
| pH                                                                       | Approximately 7.4 at 20°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Approximately 7.4 at 20°C                                                                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                                                                                                                                                                                |
| Fluid Volume                                                             | 500 ml<br>1,000 ml<br>2,000 ml                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 1,000 ml                                                                                                                                                                                                                                                                                                                                                                          | Equivalent<br>Subject device has more<br>available size options<br>than the predicate<br>device.                                                                                                                                                                                                    |
| Single Use Only                                                          | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                                                                                                |
|                                                                          | Primary Container                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                     |
|                                                                          | Constructed from Ethylene<br>Vinyl Acetate (EVA). The film<br>is flexible, clear and durable.<br>The container bags are Latex-<br>Free, PVC-Free and DEHP-Free.<br>Sterile, non-pyrogenic fluid path.                                                                                                                                                                                                                                                                                           | Constructed from Ethylene<br>Vinyl Acetate (EVA). The film<br>is flexible, clear and durable.<br>The container bags are Latex-<br>Free, PVC-Free and DEHP-Free.<br>Sterile, non-pyrogenic fluid path.                                                                                                                                                                             | Equivalent                                                                                                                                                                                                                                                                                          |
| Protecting<br>Overwrap Bag                                               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                                                                                                |
| Shelf Life                                                               | 6 Months                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 24 Months                                                                                                                                                                                                                                                                                                                                                                         | Same. Shelf life stability<br>study is on-going to claim 24<br>months shelf life.                                                                                                                                                                                                                   |
| SAFETY                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                     |
| Pyrogenicity                                                             | Non-Pyrogenic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Non-Pyrogenic                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                |
| Sterility                                                                | Sterility assured via 0.2 µm and<br>two 0.1 µm membrane filtration                                                                                                                                                                                                                                                                                                                                                                                                                              | Sterility assured via 0.1 µ<br>membrane filtration                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                                                |
| Sterilization                                                            | Aseptic Fill                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Aseptic Fill                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                |
| Biocompatibility                                                         | Biocompatible in accordance<br>with ISO 10993-1/2/5/10/12.                                                                                                                                                                                                                                                                                                                                                                                                                                      | Biocompatible in accordance<br>with ISO 10993-1/2/5/10/12.                                                                                                                                                                                                                                                                                                                        | Equivalent                                                                                                                                                                                                                                                                                          |
| Sterile<br>Dispensing/<br>Administration                                 | Fluid is dispensed via sterile<br>port on bag.                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Fluid is dispensed via sterile<br>port on bag.                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                |
| Bag Connections                                                          | Fluid delivery and drug<br>administration ports                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Fluid delivery and drug<br>administration ports                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                                                |

CONFIDENTIAL

{6}------------------------------------------------

CONFIDENTIAL

{7}------------------------------------------------

CONFIDENTIAL

OCTOBER 2024

{8}------------------------------------------------

### 7.0 NON-CLINICAL PERFORMANCE DATA

Performance data is provided in support of the substantial equivalence determination. All tests performed are included in this submission.

# 7.1. Biocompatibility

Belzer UW® Cold Storage Solution (BTLBUW-001) has been demonstrated as substantially equivalent in biocompatibility to the predicate device via testing and biological risk assessment, such that it is safe for its intended use.

# 7.2 Sterilization

Belzer UW® Cold Storage Solution (BTLBUW-001) is aseptically filled and sterility assured via 0.2 µm and two 0.1 µm membrane filtration.

# 7.3 Shelf Life

Belzer UW® Cold Storage Solution (BTLBUW-001) has a shelf life of 6 months.

{9}------------------------------------------------

#### 8.0 DESIGN CONTROL SUMMARY

All procedural Design Controls were followed, and all design control elements were satisfied.

#### 9.0 SUMMARY OF VERIFICATION & VALIDATION

Verification and validation testing was done on the Belzer UW® Cold Storage Solution (BTLBUW-001). The subject device is substantially equivalent to the predicate device. Any risks to the overall system were addressed in the Risk Analysis. Concerns or risks identified were mitigated by the actual proper use and function Belzer UW® Cold Storage Solution (BTLBUW-001) indicated in the Instructions for Use. It is appropriate to conclude that Belzer UW® Cold Storage Solution (BTLBUW-001) is substantially equivalent to the predicate device K091245.

#### 10.0 CONCLUSION

The subject and predicate devices have the same intended use and are substantially equivalent in technological characteristics and performance (same composition). The Belzer UW® Cold Storage Solution (BTLBUW-001) is safe and effective as the predicate device. The non-clinical data supports and demonstrates the safety of the device.

The conclusion is that Belzer UW® Cold Storage Solution (BTLBUW-001) is substantially equivalent to the legally marketed CoStorSol® Solution and therefore should have clearance for premarket activities in the United States.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDN/K243384](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDN/K243384)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
