← Product Code [KDJ](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDJ) · K022412

# FRESENIUS CAPD STAY.SAFE DISPOSABLE ADMINISTRATION SETS WITH STAY.SAFE CONNECTOR (K022412)

_Fresenius Medical Care North America · KDJ · Mar 5, 2003 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDJ/K022412

## Device Facts

- **Applicant:** Fresenius Medical Care North America
- **Product Code:** [KDJ](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDJ.md)
- **Decision Date:** Mar 5, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5630
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Fresenius CAPD stay•safe® Disposable Administration Sets with stay•safe® Connector are disposable devices for single use. The areas under protective guards and the fluid pathway are sterile and non-pyrogenic. The devices continue to be restricted to the sale by or on the order of a physician.

## Device Story

Disposable administration set for peritoneal dialysis; consists of circular housing with three outlet ports for patient, dialysis solution, and drain bag. Features manual knob to initiate treatment steps; outlet port plugs control solution flow. Used in clinical or home settings under physician prescription. Facilitates fluid delivery and drainage during peritoneal dialysis procedures.

## Clinical Evidence

No clinical data. Bench testing only. Materials tested to ISO 10993-1 for biological safety (indirect contact >30 days). Endotoxin testing to LAL specification of 0.01 EU/mL. Sterilization validated to SAL ≥ 10^-6.

## Technological Characteristics

Disposable administration set; circular housing with three ports. Materials tested to ISO 10993. Sterilized to SAL ≥ 10^-6. Non-pyrogenic fluid pathway. Manual operation via knob.

## Regulatory Identification

(1) A peritoneal dialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions, and that consists of a peritoneal access device, an administration set for peritoneal dialysis, a source of dialysate, and, in some cases, a water purification mechanism. After the dialysate is instilled into the patient's peritoneal cavity, it is allowed to dwell there so that undesirable substances from the patient's blood pass through the lining membrane of the peritoneal cavity into this dialysate. These substances are then removed when the dialysate is drained from the patient. The peritoneal dialysis system may regulate and monitor the dialysate temperature, volume, and delivery rate together with the time course of each cycle of filling, dwell time, and draining of the peritoneal cavity or manual controls may be used. This generic device includes the semiautomatic and the automatic peritoneal delivery system.(2) The peritoneal access device is a flexible tube that is implanted through the abdominal wall into the peritoneal cavity and that may have attached cuffs to provide anchoring and a skin seal. The device is either a single use peritioneal catheter, intended to remain in the peritoneal cavity for less than 30 days, or a long term peritoneal catheter. Accessories include stylets and trocars to aid in the insertion of the catheter and an obturator to maintain the patency of the surgical fistula in the abdominal wall between treatments. (3) The disposable administration set for peritoneal dialysis consists of tubing, an optional reservoir bag, and appropriate connectors. It may include a peritoneal dialysate filter to trap and remove contaminating particles. (4) The source of dialysate may be sterile prepackaged dialysate (for semiautomatic peritoneal dialysate delivery systems or “cycler systems”) or dialysate prepared from dialysate concentrate and sterile purified water (for automatic peritoneal dialysate delivery systems or “reverse osmosis” systems). Prepackaged dialysate intended for use with either of the peritoneal dialysate delivery systems is regulated by FDA as a drug.

## Special Controls

*Classification.* Class II (special controls). The following accessories are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9: A catheter finger grip that is non-patient contacting and intended for single use with a peritoneal catheter; a continuous ambulatory peritoneal dialysis (CAPD) belt; and a catheter stand that does not include weigh scales.

## Predicate Devices

- CAPD Safe-Lock™ Transfer Sets with Pre-filled Connector

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

page 1 of 2

K022412

# 510 (k) Summary

MAR 0 5 2003

| Applicant:                    | Fresenius Medical Care North America                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                              |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| Contact:                      | Arthur E. Eilinsfeld<br>Director, Regulatory Affairs<br>Fresenius Medical Care North America<br>95 Hayden Avenue<br>Lexington, MA 02420-9192                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Telephone: 781-402-9068<br>Fax: 781-402-9082 |
| Trade Name:                   | Fresenius CAPD stay•safe® Disposable Administration Sets<br>with stay•safe® Connector                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                              |
| Common Name:                  | Disposable Connector and Tubing Set for Solution Delivery<br>during Peritoneal Dialysis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                              |
| Classification Name:          | Peritoneal Dialysis System and Accessories per 21CFR<br>876.5630                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                              |
| Equivalence Predicate:        | CAPD Safe-Lock™ Transfer Sets with Pre-filled Connector                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                              |
| Device Description:           | The Fresenius CAPD stay•safe® Disposable Administration<br>Sets with stay•safe® Connector consists of a circular<br>housing designed with three, evenly spaced, outlet ports.<br>Individual ports specifically provide for attachment either to<br>the patient, to the peritoneal dialysis solution, or to a drain<br>bag. The outlet port plugs allow the start or stop of solution<br>flow; and prior to use the outlet port for the patient<br>connection is covered with a cap.<br><br>A knob is turned to different positions to initiate the various<br>treatment steps of the peritoneal dialysis procedure. |                                              |
| Statement of<br>Intended Use: | The Fresenius CAPD stay•safe® Disposable Administration<br>Sets with stay•safe® Connector are disposable devices for<br>single use. The areas under protective guards and the fluid<br>pathway are sterile and non-pyrogenic. The devices<br>continue to be restricted to the sale by or on the order of a<br>physician.                                                                                                                                                                                                                                                                                            |                                              |

Fresenius Medical Care North America

Corporate Headquarters: 95 Hayden Avenue Lexington, MA 02420-9192

Fresenius CAPD stay•safe® Disposable Administration Sets with stay•safe® Connector

{1}------------------------------------------------

Page 2 of 2

### 510 (k) Summary

It is indicated for use in patients with end-stage acute and Statement of Indications for Use: chronic renal disease.

- Summary of In general, the design and materials of the subject devices Technological are the same as the Fresenius predicate devices. The materials have been tested to ISO10993 and are shown not Differences: to raise any different issues regarding safety and effectiveness. A hazard analysis determination indicates that the design differences do not impact the safety or effectiveness of the device; and that the differences are not critical to the intended therapeutic use of the device.
Clinical Data: Not Applicable

Conclusions: Components of the proposed devices have met the ISO 10993-1 biological requirements for safety screening of materials in indirect contact for greater than 30 days. The sets are tested to a Limulus Amebocyte Lysate (LAL) specification of 0.01 EU/mL. Sets are sterilized by a method determined and verified to assure an SAL of ≥ 10°. Functional and physical testing is performed prior to product release.

Arthur E. Eilinsfeld

Director, Fresenius Regulatory Affairs

7/18/02
Date

K0224/2

---

Premarket Notification 510 (k) Number

Premarket Notification 510 (k) Number

AEE:dmk

Fresenius Medical Care North America

Corporate Headquarters: 95 Hayden Avenue Lexington, MA 02420-9192

Fresenius CAPD stay.safe® Disposable Administration Sets with stay.safe® Connector

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 0 5 2003

Mr. Arthur E. Eilinsfeld Director, Regulatory Affairs Fresenius Medical Care North America 95 Hayden Avenue LEXINGTON MA 02420-9192

Re: K022412

Trade/Device Name: Fresenius CAPD stay •safe® Disposable Administration Sets with stay safe® Connector Regulation Number: 21 CFR §876.5630 Regulation Name: Peritoneal dialysis system and accessories Regulatory Class: II Product Code: 78 KDJ Dated: December 3, 2002 Received: December 5, 2002

Dear Mr. Eilinsfeld:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy Crogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use Statement

#### Device Name:

Fresenius CAPD stay•safe® Disposable Administration Sets with stay•safe® Connector

#### Indications for Use:

The Fresenius CAPD stay.safe® Disposable Administration Sets with stay•safe® Connector is indicated for use in end stage acute and chronic renal disease.

Please Do Not Write Below This Line - Continue On Another Page If Needed

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109

OR

Over-The-Counter Use

Nancy Brogdon
(Division Sign-Off Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K022412

Fresenius Medical Care North America

Corporate Headquarters: 95 Hayden Avenue Lexington, MA 02420-9192

Fresenius CAPD stay.safe® Disposable Administration Sets with stay.safe® Connector

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDJ/K022412](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDJ/K022412)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
