← Product Code [FKX](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FKX) · K171652

# Fresenius Liberty Select Cycler (K171652)

_Fresenius Medical Care Renal Therapies Group, LLC · FKX · Sep 15, 2017 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FKX/K171652

## Device Facts

- **Applicant:** Fresenius Medical Care Renal Therapies Group, LLC
- **Product Code:** [FKX](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FKX.md)
- **Decision Date:** Sep 15, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5630
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Fresenius Liberty Select Cycler is indicated for acute and chronic peritoneal dialysis.

## Device Story

The Fresenius Liberty Select Cycler is an electro-mechanical, table-top device for Automated Peritoneal Dialysis (APD) in professional and home settings. It delivers dialysis therapy by controlling fluid flow, heating, and monitoring via a cassette and tubing set. The device is programmed by a physician based on a prescription. It incorporates an IQdrive and an optional wireless modem for data connectivity. Key features include selectable drain exit criteria, STAT drain options, and 'Soft Alarms' to allow user intervention before full alarm activation. The device provides automated fluid management, ensuring fill and drain volumes are delivered and removed according to the prescription. By automating the dialysis process, it enables patients to perform therapy at home, improving treatment accessibility and management of ESRD.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including software verification and validation (per ANSI/AAMI/IEC 62304:2006), mechanical/acoustic testing (noise level ≤ 40 dBA), and human factors testing for new software features (drain exit criteria, STAT drain, and soft alarms).

## Technological Characteristics

Electro-mechanical table-top cycler. Materials: plastic housing, aluminum heater tray, aluminum cassette housing. Essential performance: Inflow 45-316 mL/min, Outflow 30-286 mL/min, Temperature 37°C ± 1°C, Fill accuracy ± 2%, Drain accuracy ± 3%. Connectivity: IQdrive and 3G/4G LTE wireless modem. Software version 2.8.7.

## Regulatory Identification

(1) A peritoneal dialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions, and that consists of a peritoneal access device, an administration set for peritoneal dialysis, a source of dialysate, and, in some cases, a water purification mechanism. After the dialysate is instilled into the patient's peritoneal cavity, it is allowed to dwell there so that undesirable substances from the patient's blood pass through the lining membrane of the peritoneal cavity into this dialysate. These substances are then removed when the dialysate is drained from the patient. The peritoneal dialysis system may regulate and monitor the dialysate temperature, volume, and delivery rate together with the time course of each cycle of filling, dwell time, and draining of the peritoneal cavity or manual controls may be used. This generic device includes the semiautomatic and the automatic peritoneal delivery system.(2) The peritoneal access device is a flexible tube that is implanted through the abdominal wall into the peritoneal cavity and that may have attached cuffs to provide anchoring and a skin seal. The device is either a single use peritioneal catheter, intended to remain in the peritoneal cavity for less than 30 days, or a long term peritoneal catheter. Accessories include stylets and trocars to aid in the insertion of the catheter and an obturator to maintain the patency of the surgical fistula in the abdominal wall between treatments. (3) The disposable administration set for peritoneal dialysis consists of tubing, an optional reservoir bag, and appropriate connectors. It may include a peritoneal dialysate filter to trap and remove contaminating particles. (4) The source of dialysate may be sterile prepackaged dialysate (for semiautomatic peritoneal dialysate delivery systems or “cycler systems”) or dialysate prepared from dialysate concentrate and sterile purified water (for automatic peritoneal dialysate delivery systems or “reverse osmosis” systems). Prepackaged dialysate intended for use with either of the peritoneal dialysate delivery systems is regulated by FDA as a drug.

## Special Controls

*Classification.* Class II (special controls). The following accessories are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9: A catheter finger grip that is non-patient contacting and intended for single use with a peritoneal catheter; a continuous ambulatory peritoneal dialysis (CAPD) belt; and a catheter stand that does not include weigh scales.

## Predicate Devices

- Liberty Cycler ([K123630](/device/K123630.md))
- Liberty PDx Cycler ([K141145](/device/K141145.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, resembling a bird-like shape.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 15, 2017

Fresenius Medical Care Renal Therapies Group. LLC Denise Oppermann Senior Director. Regulatory Affairs 920 Winter Street Waltham, MA 02451

Re: K171652

> Trade/Device Name: Fresenius Liberty Select Cycler Regulation Number: 21 CFR§ 876.5630 Regulation Name: Peritoneal Dialysis System and Accessories Regulatory Class: II Product Code: FKX Dated: June 2, 2017 Received: June 5, 2017

Dear Denise Oppermann:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Joyce M. Whang -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) Unknown K171652

Device Name Fresenius Liberty Select Cycler

Indications for Use (Describe) The Fresenius Liberty Select Cycler is indicated for acute and chronic peritoneal dialysis.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right, also in blue. The symbol is made up of three downward-pointing chevrons stacked on top of each other.

#### 5. 510(k) SUMMARY

This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR 807.92.

#### Submitter's Information 5.1.

| Name:             | Fresenius Medical Care Renal Therapies Group, LLC                 |
|-------------------|-------------------------------------------------------------------|
| Address:          | 920 Winter Street<br>Waltham, MA<br>02451-1457                    |
| Phone:            | (781) 699-4479                                                    |
| Fax:              | (781) 699-9635                                                    |
| Contact Person:   | Denise Oppermann, Senior Director<br>Regulatory Affairs – Devices |
| Preparation Date: | 2 June 2017                                                       |

#### 5.2. Device Name

| Trade Name:           | Fresenius Liberty Select Cycler        |
|-----------------------|----------------------------------------|
| Common Name:          | Peritoneal Dialysis Cycler             |
| Classification Name : | System, Peritoneal, Automatic Delivery |
| Regulatory Class:     | Class II per 21 CFR 876.5630           |
| Product Code:         | FKX                                    |
| Classification Panel: | Gastroenterology/Urology               |

#### 5.3. Legally Marketed Predicate Device

#### Primary Predicate – Liberty Cycler (K123630) 5.3.1.

The Liberty Cycler, K123630, will serve as the primary predicate for the Liberty Select Cycler as described in the 8 September 2016 pre-submission (Q161532).

#### Secondary Predicate – Liberty PDx Cycler (K141145) 5.3.2.

The Liberty PDx Cycler will serve as the secondary predicate for the Liberty Select Cycler.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the logo for Fresenius Medical Care. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three downward-pointing chevrons stacked on top of each other. The text is in a bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.

#### Device Description 5.4.

#### 5.4.1. Device Identification:

The Liberty Select Cycler will be available in the configuration described below.

| Product Line | Part Number | Description           |
|--------------|-------------|-----------------------|
| Liberty      | 180343      | Liberty Select Cycler |

#### 5.4.2. Device Characteristics

The Liberty Select Cycler is an electro-mechanical medical device. Software controls the functions of the machine during peritoneal dialysis treatment, including fluid flow, heating, and alarms.

#### 5.4.3. Environment of Use

The Liberty Select Cycler is designed as a table-top unit and is prescribed for use in both professional and home treatment settings.

#### 5.4.4. Brief Written Description of the Device

The Liberty Select Cycler is a software-controlled electro-mechanical medical device designed to deliver Automated Peritoneal Dialysis (APD) therapy for the treatment of end-stage renal disease (ESRD). The Liberty Select Cycler is used to perform continuous and intermittent peritoneal dialysis therapies. Treatment settings are programmed based on a physician's prescription. The Liberty Select Cycler accommodates three (3) accessory devices:

- Cassette and tubing set ●
- IQdrive
- Optional wireless modem ●

The cassette and tubing set and the IQdrive are the same accessory devices described in K123630. The peripheral wireless modem has been updated to support AT&T and Verizon 3G/4G LTE networks.

Performance modifications were made to the device. Selectable drain exit criteria and additional STAT Drain options were added. Alarm modifications introduce Soft Alarms that provide the user with an opportunity to correct drain and fill complications before issuing an alarm. The labeling has been modified to align with the software modifications and with the secondary predicate Liberty PDx Cycler.

#### 5.4.5. Materials of Use

There are no changes to the materials of use from the previous 510(k) submission (K123630).

The Liberty Select Cycler enclosure consists of the following materials:

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right, also in blue. The symbol is made up of three downward-pointing chevrons, stacked on top of each other.

- Plastic housing o
- o Aluminum heater tray
- Aluminum cassette housing ●

#### 5.4.6. Essential Performance Characteristics

The Liberty Select Cycler has the same essential performance characteristics as the predicate device (K123630) as listed in Table 1.

| Feature                | Specification                             |
|------------------------|-------------------------------------------|
| Inflow                 | 45-316 mL/min                             |
| Outflow                | Minimum: 30 mL/min<br>Maximum: 286 mL/min |
| Temperature            | 37°C ± 1°C                                |
| Volume Accuracy, Fill  | ± 2% of the fill volume                   |
| Volume Accuracy, Drain | ± 3% of the drain volume                  |

Table 1: Liberty Select Cycler Essential Performance Characteristics

#### Indications for Use 5.5.

The Fresenius Liberty Select Cycler is indicated for acute and chronic peritoneal dialysis.

### 5.6. Comparison of Technological Characteristics with the Predicate Device

The technological characteristics of the Liberty Select Cycler are equivalent to the predicate device. Although software modifications were made to improve the user experience, all other electrical components, mechanical components, pump module components, and safety components are identical to the predicate device.

The following technical specifications of the Liberty Select Cycler remain the same as those of the predicate device:

- Environmental requirements
- Accessories ●
- Transportation and storage specifications
- Hardware specifications
- Manufacturing location ●

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three downward-pointing chevrons stacked on top of each other. The text is in a bold, sans-serif font and is also blue.

#### 5.7. Performance Data

Performance testing requirements were determined through the application of a risk management process, applicable FDA guidance documents and performance standards (21 CFR 876.5630). Performance testing to support the determination of substantial equivalence included testing in compliance with ANSI/AAMI/IEC 62304:2006 and ANSI/AAMI/IEC 62366-1:2015.

#### 5.7.1. Software Verification and Validation Testing

Unit, integration and system level software verification testing was performed to demonstrate the effectiveness of the software modifications and to confirm operation of the machine.

#### 5.7.2. Mechanical and Acoustic Testing

Mechanical testing was conducted to verify that the Liberty Select Cycler noise level does not exceed 40 dBA while performing a treatment.

#### 5.7.3. Human Factors Testing

Testing was performed on the modifications found to impact Human Factors:

- Add selectable drain exit criteria (70% or 85%) 0
- Add selectable additional drain option with option for additional drain alert ●
- Add STAT drain at the end of the treatment ●
- Implement soft alarms

The remaining essential and critical user tasks were determined to be equivalent to the user tasks of the predicate devices.

FMCRTG concludes that the four (4) new features tested on the Liberty Select Cycler (software version 2.8.7) are safe and effective for the intended users, uses, and use environments.

#### 5.8. Conclusion

The indications for use, materials of construction, and technological characteristics of the Liberty Select Cycler are the same as the predicate devices. Differences between the Liberty Select Cycler and the predicates do not raise new concerns with regard to safety or efficacy. FMCRTG concludes that within the meaning of the Medical Device Amendments Act of 1976, the proposed Liberty Select Cycler is substantially equivalent to the primary predicate device, Liberty Cycler (K123630).

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FKX/K171652](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FKX/K171652)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
