AquaFlate Pre-Filled Sterile Water Syringe 10mL

K200556 · H R Pharmaceuticals, Inc. · EZL · Mar 31, 2020 · Gastroenterology, Urology

Device Facts

Record IDK200556
Device NameAquaFlate Pre-Filled Sterile Water Syringe 10mL
ApplicantH R Pharmaceuticals, Inc.
Product CodeEZL · Gastroenterology, Urology
Decision DateMar 31, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5130
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The AquaFlate Pre-Filled Sterile Water Syringe 10mL is intended to be used for inflating a Foley catheter balloon.

Device Story

Device is a 10mL syringe pre-filled with USP purified water; used for inflating indwelling urethral Foley catheter balloons. Healthcare professional removes cap, connects luer tip to catheter inflation port, and dispenses entire contents to inflate balloon; prevents catheter dislodgement. Syringe is then disconnected and discarded. Single-use only; not for injection. Provides sterile, biocompatible solution for balloon inflation in clinical settings.

Clinical Evidence

Bench testing only. Testing included small-bore connector testing (ISO 80369-7), manual syringe testing (ISO 7886-1), sterility testing (USP 42 <71>), biocompatibility (ISO 10993-5, -10), sterilization validation (ISO 11137), and packaging/shelf-life validation (ASTM F1980, ISTA 3A).

Technological Characteristics

Materials: Polypropylene barrel/plunger, latex-free synthetic rubber gasket/tip cap. Principle: Manual syringe inflation. Energy: None. Dimensions: 10mL. Sterilization: Gamma irradiation (25-40kGy). Connectivity: None. Standards: ISO 80369-7, ISO 7886-1, ISO 10993, ISO 11137, ASTM F1980.

Indications for Use

Indicated for inflating an indwelling urethral Foley catheter balloon in a healthcare facility setting.

Regulatory Classification

Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font in blue. March 31, 2020 HR Pharmaceuticals, Inc. % Prithul Bom Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, MN 55114 Re: K200556 Trade/Device Name: AquaFlate Pre-Filled Sterile Water Syringe 10mL Regulation Number: 21 CFR 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZL Dated: March 2, 2020 Received: March 3, 2020 Dear Prithul Bom: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jessica K. Nguyen, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K200556 #### Device Name AquaFlate Pre-filled Sterile Water Syringe 10 mL Indications for Use (Describe) The AquaFlate Pre-filled Sterile Water Syringe is intended to be used for inflating a Foley catheter balloon. | Type of Use (Select one or both, as applicable) | | | |-------------------------------------------------|--|--| | <table><tr><td></td><td></td></tr></table> | | | | | | | X Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ K200556 Page 1 of 6 510(k) SUMMARY 5-1 # 5.0 510(k) SUMMARY #### 510(k) SUMMARY for AquaFlate Pre-Filled Sterile Water Syringe 10mL #### I. SUBMITTER: HR Pharmaceuticals, Inc. 2600 Eastern Boulevard, Suite 201 York, PA 17402 Telephone Number:717-252-1110 x104 Fax Number:717-685-2590 Contact Person: Colby Wiesman Date Prepared: November 26, 2019 II. DEVICE: Name of Device: AquaFlate Pre-Filled Sterile Water Syringe 10mL Model Number: 250-400 Common or Usual Name: Catheter, Retention type, balloon Classification Description: Urological catheter and accessories (21 CFR 876.5130) Regulatory Class: II Product Code: EZL Review Panel: Gastroenterology/Urology ## III. PREDICATE DEVICES: AMSINO INTERNATIONAL, AMSure® Prefilled Syringe for Balloon Inflation with Sterile Water, K181814 No reference devices were used in this submission. ## IV. DEVICE DESCRIPTION: The proposed device, AquaFlate Pre-Filled Sterile Water Syringe 10mL (REF 250-400), is a 10mL syringe prefilled with USP purified water and sterilized via gamma irradiation. The syringe consists of a polypropylene barrel and plunger and latex free rubber for both the plunger gasket and syringe tip cap. The AquaFlate Pre-Filled Water Syringe 10mL is intended for inflating an indwelling. urethral Foley catheter balloon. In accordance with CDC Guidelines, healthcare facility protocols, and the Foley catheter manufacturer's instructions for use, prior to catheterization, the syringe cap of the AquaFlate Pre-Filled Sterile Water Syringe 10mL is Image /page/3/Picture/17 description: The image shows the logo for H-R Pharmaceuticals, Inc. The logo is in blue and features the letters "H-R" in a stylized font. To the right of the letters is the text "Pharmaceuticals, Inc." The logo is simple and professional, and it is likely used on the company's website, marketing materials, and products. {4}------------------------------------------------ removed and the luer tip of the syringe is connected to the inflation port of the catheter. Once the catheter is inserted into the bladder, the entire contents of the syringe are dispensed to inflate the catheter balloon to ensure the catheter does not slide out. A fter balloon inflation, the syringe is disconnected from the catheter inflation port and discarded. The AquaFlate Pre-Filled Sterile Water Syringe 10mL is not intended for injection and is for single use only. The shelf life of the AquaFlate Pre-Filled Sterile Water Syringe 10mL is one (1) year. HR Pharmaceuticals does not intend to market this device with accessories or as part of a system. ### V. INDICATIONS FOR USE: The AquaFlate Pre-Filled Sterile Water Syringe 10mL is intended to be used for inflating a Foley catheter balloon. ### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES: The subject and predicate devices have the same technological principle whereby, the prefilled sterile water syringe is intended for inflating an indwelling, urethral Foley catheter balloon. Once the catheter is inserted into the bladder, the entire contents of the syringe are dispensed to inflate the catheter balloon to ensure the catheter does not slide out. After balloon inflation, the syringe is disconnected from the catheter inflation port and discarded. Additionally, the prefilled sterile water syringe is not intended for injection and is for single use only. At a high level, the AquaFlate Pre-Filled Sterile Water Syringe 10mL and the predicate device are based on the following same elements: - Same Intended Use - Same Technological Characteristics ● - Same Materials - . Same Physical Characteristics - Same Performance Specifications ● - Packaged in the same quantity (10mL) ● - o Provided sterile - Biocompatible The only identified difference between the subject and predicate devices is the shelf life. The AquaFlate Pre-Filled Sterile Water Syringe 10mL has a one-year shelf life whereby, AMSure® Prefilled Syringe for Balloon Inflation with Sterile Water has a three-year shelf life. While the AquaFlate Pre-Filled Sterile Water Syringe 10mL is currently undergoing 2 year and 3-year shelf life testing, the one-year established shelf life for the subject device is acceptable for the intended use as demonstrated by successful stability Image /page/4/Picture/21 description: The image shows the logo for H-R Pharmaceuticals, Inc. The logo is in blue and features the letters "H" and "R" in a stylized font. To the right of the "R" is a symbol that looks like a water droplet inside of a flower. The words "Pharmaceuticals, Inc." are written in a smaller font to the right of the letters. {5}------------------------------------------------ and performance testing. Therefore, the difference in shelf life between the subject and predicate devices does not raise new questions of safety or effectiveness. A more detailed comparison of the predicate and subject devices is presented in the table below. | | Proposed Device<br>AquaFlate Pre-Filled Sterile<br>Water Syringe 10mL | Predicate Device<br>AMSure Prefilled<br>Syringe for Balloon<br>Inflation with Sterile<br>Water<br>(K181814) | Comparison | | |------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|------| | Manufacturer | HR Pharmaceuticals, Inc. | Amsino International, Inc. | N/A | | | Device Classification<br>Name | Catheter, Retention Type,<br>Balloon | Catheter, Retention Type,<br>Balloon | Same | | | Regulation Description | Urological catheter and<br>accessories | Urological catheter and<br>accessories | Same | | | Review Panel | Gastroenterology/Urology | Gastroenterology/Urology | Same | | | Product Code | EZL | EZL | Same | | | Reg Number | 876.5130 | 876.5130 | Same | | | Class | II | II | Same | | | Indications for Use | The AquaFlate Pre-Filled Sterile<br>Water Syringe 10mL is intended<br>to be used for inflating a Foley<br>catheter balloon. | Prefilled Syringe for<br>Balloon Inflation with<br>Sterile Water is intended<br>to be used in inflating<br>foley catheter balloon. | Same | | | Patient-contacting<br>material | Indirect patient contact of sterile<br>water used to inflate a catheter<br>balloon | Indirect patient contact of<br>sterile water used to<br>inflate a catheter balloon | Same | | | Sizes | 10mL | 10cc and 30cc | Same<br>(10mL) | | | Materials of<br>Main<br>Components | Barrel | Polypropylene | Polypropylene | Same | | | Plunger | Polypropylene | Polypropylene | Same | | | Plunger<br>Gasket | Isoprene (black synthetic rubber<br>that does not contain natural<br>rubber latex) | Black pharmaceutical<br>grade,<br>synthetic rubber (Latex<br>free) | Same | | | Tip<br>Cover | Isoprene (black synthetic rubber<br>that does not contain natural<br>rubber latex) | Same as gasket | Same | | | Solution | Purified water, USP | Purified water, USP | Same | | Principle of Operation | | Manual | Manual | Same | | Technical Performance | | Conforms with ISO7886-1,<br>ISO 80369-7 and USP 42<br>Chapter <71> | Conforms with ISO7886-<br>1,<br>ISO 80369-7 and USP 40 | Same | | Biocompatibility | | Meets ISO 10993-1:2009<br>requirements for irritation (ISO<br>10993-10), sensitization (ISO | Conforms with ISO<br>10993-5 and ISO 10993-<br>10 | Same | | | Proposed Device<br>AquaFlate Pre-Filled Sterile<br>Water Syringe 10mL | Predicate Device<br>AMSure Prefilled<br>Syringe for Balloon<br>Inflation with Sterile<br>Water<br>(K181814) | Comparison | | | | 10993-10), and cytotoxicity<br>(ISO 10993-5) | | | | | Utility | Single Use | Single Use | Same | | | Sterility (Method) | Sterile (Gamma Irradiation) | Sterile (Gamma Irradiation) | Same | | | Environment of Use | Healthcare Facility | Healthcare Facility | Same | | | Labeling | Prescription Use | Prescription Use | Same | | | Shelf Life | 1 year | 3 years | Different | | | Device Picture | Image: syringe | Image: syringes | N/A | | Image /page/5/Picture/6 description: The image shows the logo for H-R Pharmaceuticals, Inc. The logo is blue and features the letters "H" and "R" in a stylized font. To the right of the "R" is a symbol that looks like a water droplet inside of a flower. The words "Pharmaceuticals, Inc." are written in a smaller font to the right of the "R". {6}------------------------------------------------ *Values obtained from performance testing conducted per HR Pharmaceuticals' protocols. # VII. PERFORMANCE DATA There are no FDA device-specific guidance documents, special controls document, and/or requirements in a device-specific regulation for prefilled sterile water syringes. Additionally, no performance standards have been promulgated under Section 514 of the Food, Drug and Cosmetic Act for prefilled sterile water syringes. However, the following performance data were provided in support of the substantial equivalence determination. Bench Testing - Small-bore connector testing ISO 80369-7:2016. Small-bore connectors for ● liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications [FR Recognition No. 5-115] - Manual, single-use syringe testing - ISO 7886-1:2017, Sterile hypodermic syringes for single use - Part 1: Syringes for manual use [FR Recognition No. 6-404] - Sterility USP 42 Chapter <71> Sterility Tests [FR Recognition No. 14-532] ● ## Biocompatibility Testing The biocompatibility evaluation for the AquaFlate Pre-Filled Sterile Water Syringe 10mL was conducted in accordance with the FDA Final Guidance "Use of International Standard ISO-10993, 'Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process, " June 16, 2016, and International Standard ISO 10993-1:2009 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Image /page/6/Picture/13 description: The image shows the logo for H-R Pharmaceuticals, Inc. The logo is in blue and features the letters "H" and "R" in a stylized font. To the right of the "R" is a symbol that looks like a water droplet inside of a circle. The words "Pharmaceuticals, Inc." are written in a smaller font to the right of the letters. {7}------------------------------------------------ Testing Within a Risk Management Process," as recognized by FDA. The biocompatibility testing included the following and yielded acceptable results: - Cytotoxicity ANSI/AAMI/ISO 10993-5:2009 (R) 2014 Biological Evaluation of ● Medical Devices Part 5: Tests for In Vitro Cytotoxicity [FR Recognition No. 2-245] - Irritation ANSI/AAMI/ISO 10993-10:2013—Biological Evaluation of Medical . Devices-Part 10: Tests for Irritation and Skin Sensitization [FR Recognition No. 2-1741 - . Sensitization - ANSI/AAMI/ISO 10993-10:2013—Biological Evaluation of Medical Devices-Part 10: Tests for Irritation and Skin Sensitization [FR Recognition No. 2-174] The AquaFlate Pre-Filled Sterile Water Syringe 10mL falls into the category of limited (<24 hour) external communicating devices and is intended for indirect tissue contact. ## Sterilization The AquaFlate Pre-Filled Sterile Water Syringe 10mL was validated for sterilization using gamma irradiation at a minimum dose of 25kGy and a maximum dose of 40kGy to achieve a Sterility Assurance Level of 10th in accordance with the following recognized standards: - ANSI/ AAMI/ISO 11137-1:2006/(R) 2015 & A1:2013. Sterilization of health ● care products -Radiation-Part1: Requirements for development validation, and routine control of a sterilization process for medical devices. [FR Recognition No. 14-4281 - ANSI/AAMI/ISO 11137-2:2013. Sterilization of health care products-Radiation-Part 2: Establishing the sterilization dose. [FR Recognition No. 14-4091 - ANSI/AAMI/ISO 11137-3:2006 (R) 2010. Sterilization of healthcare products -Radiation-Part 3: Guidance on dosimetric aspects. [FR Recognition No. 14-510] - ANSI/AAMI/ISO 11737-1: 2018. Sterilization of health care products, Microbiological Methods-Part 1: Determination of a population of microorganisms on product. [FR Recognition No. 14-514] - ANSI/AAMI/ISO 11737-2: 2009 (R) 2014. Sterilization of Medical Devices-● Microbiological Methods-Part 2: Tests of sterility Performed in the definition, validation and maintenance of a sterilization Process. [FR Recognition No. 14-327] - AAMI TIR 17-2008. Compatibility of Materials subject to sterilization. ● - ANSI/AAMI/ISO TIR13004: 2013. Sterilization of health care products-● Radiation-Substantiation of a selected sterilization dose-Method VDmax® Image /page/7/Picture/17 description: The image shows the logo for H-R Pharmaceuticals, Inc. The logo is in blue and features the letters "H" and "R" in a stylized font. To the right of the "R" is a symbol that looks like a water droplet inside of a circle. The text "Pharmaceuticals, Inc." is to the right of the "R" and the water droplet symbol. {8}------------------------------------------------ #### Packaging Validation and Shelf Life The AquaFlate Pre-Filled Sterile Water Syringe 10mL underwent packaging validation testing and accelerated aging testing per the below standards to support a one-year shelf life. - ISTA Procedure 3A (2018) Series. General Simulation Performance Test Procedure Packaged-Products for Parcel Delivery System Shipment 150lb or less. - ASTM D4169-16. Performance Testing of Shipping Containers and Systems ● [FR Recognition No. 14-499] - ASTM F1980-16. Standard Guide for Accelerated Aging of Sterile Barrier ● Systems for Medical Devices [FR Recognition No. 14-497] # VIII. CONCLUSION All necessary testing has been performed on the AquaFlate Pre-Filled Sterile Water Syringe 10mL to support the safety and effectiveness of the device per its intended use. Additionally, the performance testing concluded that the subject device is substantially equivalent to the legally marketed predicate device. Image /page/8/Picture/10 description: The image shows the logo for H-R Pharmaceuticals, Inc. The logo is in blue and features the letters "H" and "R" in a stylized font. To the right of the "R" is a symbol that looks like a water droplet above a plant. The words "Pharmaceuticals, Inc." are written in a smaller font to the right of the "R".
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...