PLP · High Intensity Ultrasound System For Prostate Tissue Ablation
Gastroenterology, Urology · 21 CFR 876.4340 · Class 2
Overview
| Product Code | PLP |
|---|---|
| Device Name | High Intensity Ultrasound System For Prostate Tissue Ablation |
| Regulation | 21 CFR 876.4340 |
| Device Class | Class 2 |
| Review Panel | Gastroenterology, Urology |
Identification
A high intensity ultrasound system for prostate tissue ablation is a prescription device that transmits high intensity therapeutic ultrasound energy into the prostate to thermally ablate a defined, targeted volume of tissue, performed under imaging guidance. This classification does not include devices that are intended for the treatment of any specific prostate disease and does not include devices that are intended to ablate non-prostatic tissues/organs.
Classification Rationale
Class II (special controls). The special controls for this device are:
Special Controls
In combination with the general controls of the FD&C Act, the high intensity ultrasound system for prostate tissue ablation is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Characterization of acoustic pressure and power output at clinically relevant levels; (ii) Measurement of targeting accuracy and reproducibility of high intensity ultrasound output; (iii) Ultrasound-induced heating verification testing at target and non-target tissues; (iv) Electrical safety testing; and (v) Electromagnetic compatibility testing. (2) Software verification, validation, and hazard analysis must be performed. (3) The elements of the device that may contact the patient's mucosal tissue must be demonstrated to be biocompatible. (4) Performance data must demonstrate the sterility of the device components that contact the patient's mucosal tissue. (5) Performance data must support shelf life by demonstrating continued sterility of the device or the sterile components, package integrity, and device functionality over the identified shelf life. (6) Performance data must support the instructions for reprocessing all reusable components. (7) *In vivo* testing must demonstrate that the device thermally ablates targeted tissue in a controlled manner without thermal injury to adjacent, non-target tissues.(8) Clinical testing must document the adverse event profile, provide evidence of prostatic ablation, and demonstrate that the device performs as intended under anticipated conditions of use. (9) Training must be provided so that upon completion of the training program, the physician can: (i) Use all safety features of the device; (ii) Accurately target the high intensity ultrasound energy within the desired region of the prostate; and (iii) Perform the ablation procedure in a manner that minimizes damage to non-target tissues. (10) Labeling must include: (i) A section that summarizes the clinical testing results, including the adverse event profile and evidence of prostate ablation achieved; and (ii) An expiration date or shelf life for single use components.
Cleared Devices (13)
| Record | Device Name | Applicant | Decision Date | Decision |
|---|---|---|---|---|
| K251910 | Focal One® | Edap Technomed, Inc. | Nov 19, 2025 | SESE |
| K240296 | Tulsa Pro System (Pad-105) | Profound Medical, Inc. | May 9, 2024 | SESE |
| K231378 | Exablate Model 2100V1, Exablate Prostate System, Exablate MRgFUS, Exablate 2100V1 Type 3.0 | Insightec, Ltd. | Oct 30, 2023 | SESE |
| K230692 | TULSA-PRO System | Profound Medical, Inc. | Sep 20, 2023 | SESE |
| K211858 | TULSA-PRO System | Profound Medical, Inc. | Sep 6, 2022 | SESE |
| K212150 | Exablate Prostate System | Insightec, Inc. | Nov 23, 2021 | SESE |
| K202286 | Tulsa-Pro System | Profound Medical, Inc. | Sep 16, 2020 | SESE |
| K191200 | TULSA-PRO System | Profound Medical, Inc. | Aug 15, 2019 | SESE |
| K172721 | FOCAL ONE | Edap Technomed, Inc. | Jun 7, 2018 | SESE |
| K172285 | Ablatherm Fusion | Edap Technomed, Inc. | Oct 3, 2017 | SESE |
| K160942 | Sonablate | Sonacare Medical, LLC | Dec 21, 2016 | SESE |
| K153023 | Ablatherm Integrated Imaging | Edap Technomed, Inc. | Nov 6, 2015 | SESE |
| DEN150011 | Sonablate 450 | Sonacare Medical, LLC | Oct 9, 2015 | DENG |
Top Applicants
- Profound Medical, Inc. — 5 clearances
- Edap Technomed, Inc. — 4 clearances
- Sonacare Medical, LLC — 2 clearances
- Insightec, Ltd. — 1 clearance
- Insightec, Inc. — 1 clearance