K964937 · Boston Scientific Corp · LQC · Jan 24, 1997 · Gastroenterology, Urology
Device Facts
Record ID
K964937
Device Name
A LITHOTRIPTER-COMPATIBLE BASKET
Applicant
Boston Scientific Corp
Product Code
LQC · Gastroenterology, Urology
Decision Date
Jan 24, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Lithotripter-Compatible Basket is indicated for the retrieval of stones from the biliary system and endoscopic crushing (lithotripsy) of biliary calculi. The Lithotripter-Compatible Basket should not be used for any purpose other than its intended application.
Device Story
Device is a mechanical lithotripter and retrieval basket used during endoscopic procedures to capture and crush biliary calculi. Operated by a physician, the device consists of a basket (stainless steel, octagonal shape, 4-wire configuration) and a sheath (2.3 mm diameter, 180 cm length) with a detachable three-ring handle featuring an injection port. The device is inserted endoscopically into the biliary system to entrap stones; mechanical force is applied to crush the calculi for subsequent removal. The device facilitates stone clearance, potentially preventing complications associated with stone impaction, such as cholangitis or pancreatitis.
Clinical Evidence
Bench testing only. Biocompatibility assessment performed on patient- and fluid-contact materials with satisfactory results. Laboratory testing verified safety and performance characteristics.
Technological Characteristics
Stainless steel 4-wire octagonal basket; 2.3 mm diameter, 180 cm length sheath; detachable three-ring handle with injection port. Sterilized via ethylene oxide gas per AAMI protocol. Packaged in PVC tray within Tyvek-Mylar pouch.
Indications for Use
Indicated for patients requiring retrieval of biliary stones and endoscopic crushing (lithotripsy) of biliary calculi.
Regulatory Classification
Identification
A mechanical lithotriptor is a device with steel jaws that is inserted into the urinary bladder through the urethra to grasp and crush bladder stones.
Special Controls
*Classification.* Class II (special controls). The device, when it is a biliary mechanical lithotripter, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Biliary Helical and Mini-Helical Baskets (K925495)
Related Devices
K040447 — TRAPEZOID RX LITHOTRIPTER COMPATIBLE BASKET, MODELS 1086, 1087, 1088 · Boston Scientific Corp · Mar 12, 2004
K243807 — Single Use Retrieval Basket V FG-V421PR/V422PR/V431P/V432P · Olympus Medical Systems Corp. · Mar 18, 2025
K171969 — Web II Memory Extraction Basket or Memory II Double Lumen Extraction Basket, Memory 5 Fr. Soft Wire Baskets or Memory Helical Stone Extractor, Memory Eight Wire Baskets or Memory Hard Wire Baskets, Fusion Wire Guided Extraction Basket, Non-Lithotripsy Extraction Basket · Wilson-Cook Medical, Inc. · Mar 28, 2018
K170811 — Single Use Retrieval Nitinol Basket V · Olympus Medical Systems Corp. · Nov 27, 2017
K230598 — Stone Extraction Baskets · Jiangsu Vedkang Medical Science and Technology Co., Ltd. · Sep 12, 2023
Submission Summary (Full Text)
{0}
K964937
P193
SECTION 9
JAN 24 1997
510(K) SUMMARY
# FOI RELEASABLE
Pursuant to §513(i)(3)(A) of the Food, Drug, and Cosmetic Act, Boston Scientific Corporation is required to submit with this Premarket Notification “...adequate summary of any information respecting safety and effectiveness or state that such information will be made available upon request of any person.” Boston Scientific Corporation chooses to submit a summary of information respecting safety and effectiveness.
- **COMMON/USUAL NAMES:** Retrieval Basket, Mechanical Lithotripter
- **TRADE/PROPRIETARY NAME:** The tradename of the device has not been finalized.
- **CLASSIFICATION NAME & DEVICE CLASSIFICATION:** Class II
| Name | Number | 21 CFR Ref. |
| --- | --- | --- |
| Biliary Mechanical Lithotripter | 78 LQC | 876.4500 |
| Biliary Stone Dislodger | 78 LQR | 876.5010 |
- **DEVICE PANEL/BRANCH:** Gastroenterology-Urology (GU)<br/>Urology/Lithotripsy (ULDB)
- **OWNER/OPERATOR:** Boston Scientific Corporation<br/>One Boston Scientific Place<br/>Natick, MA 01760<br/>Owner/Operator No. 9912058
- **CONTACT PERSON:** Daniel J. Dillon, Senior Regulatory Affairs Specialist
## INDICATIONS FOR USE
The Lithotripter-Compatible Basket is indicated for the retrieval of stones from the biliary system and endoscopic crushing (lithotripsy) of biliary calculi. The Lithotripter-Compatible Basket should not be used for any purpose other than its intended application.
## CONTRAINDICATIONS
None known.
## COMPLICATIONS
Possible complications that may result from endoscopic crushing (lithotripsy) and retrieval of biliary calculi include, but may not be limited to: stone impaction; perforation; hemorrhage; hematoma; septicemia/infection; pancreatitis; cholangitis; allergic reaction to contrast medium.
Premarket Notification, Lithotripter-Compatible Basket, December 9, 1996
Proprietary and Confidential Information of Boston Scientific Corporation
57
{1}
K964937
p 2073
# DESCRIPTIVE AND TECHNOLOGICAL CHARACTERISTICS OF PROPOSED AND PREDICATE DEVICES
Boston Scientific Corporation believes that the Lithotripter-Compatible Basket is substantially equivalent to the currently-marketed Monolith Mechanical Lithotripter. Figure 9-1 compares the descriptive characteristics of these products.
## PERFORMANCE CHARACTERISTICS
Laboratory testing regarding characteristics was performed on Lithotripter-Compatible Basket to verify its safety and performance. A biocompatibility assessment was performed on the patient- and fluid-contact materials of the Lithotripter-Compatible Basket with satisfactory results.
## PACKAGING, STERILIZATION, AND PYROGENICITY
The Lithotripter-Compatible Basket will be packaged in a PVC tray and placed inside a Tyvek-Mylar pouch. The Lithotripter-Compatible Basket will be sterilized using ethylene oxide gas using the AAMI protocol for ethylene oxide sterilization. Pyrogenicity testing will be performed on a periodic basis to monitor bacterial endotoxin levels.
## CONCLUSION
Boston Scientific Corporation believes that Lithotripter-Compatible Basket is substantially equivalent to its currently-marketed Monolith Mechanical Lithotripter and Biliary Helical and Mini-Helical Baskets. Figure 9-1 compares the descriptive characteristics of these products. As demonstrated in Figure 9-1, the Lithotripter-Compatible Basket is equivalent in its indications for use, while being very similar in design and materials. In addition, Boston Scientific Corporation has presented laboratory testing and biocompatibility information. The information presented provides assurance that the Lithotripter-Compatible Basket will meet the minimum requirements that are considered acceptable for its intended use.
Premarket Notification, Lithotripter-Compatible Basket, December 9, 1996
Proprietary and Confidential Information of Boston Scientific Corporation
58
{2}
K964937
8293
FIGURE 9-1: SIMILARITIES AND DIFFERENCES BETWEEN LITHOTRIPTER-COMPATIBLE BASKET, MONOLITH MECHANICAL LITHOTRIPTER, AND BILIARY HELICAL AND MINI-HELICAL BASKETS
| | Lithotripter-Compatible Basket (This 510(k)) | Monolith Mechanical Lithotripter (K943191) | Biliary Helical and Mini-Helical Baskets (K925495) |
| --- | --- | --- | --- |
USE
| Indication | Biliary stone retrieval, crushing and removal | Biliary stone crushing and removal | Biliary stone entrapment and removal |
| --- | --- | --- | --- |
| Route of Administration | Endoscopic | «— Same | «— Same |
BASKET
| Dimensions | 5.5 x 2.5 cm
4 x 2 cm
3 x 1.5 cm | «— Same | 22 - 37 mm (Length)
8.5 - 20 mm (Diameter) |
| --- | --- | --- | --- |
| Number of Wires | 4 | «— Same | 3 - 5 |
| Shape | Octagonal | «— Same | Helical |
| Material | Stainless Steel | «— Same | «— Same |
SHEATH
| Diameter | 2.3 mm | «— Same | 0.63 - 1.6 mm |
| --- | --- | --- | --- |
| Length | 180 cm | «— Same | 35 -240 cm |
HANDLE
| Mechanism | Three-Ring, Detachable | Trigger, Non-Detachable | Three-Ring, Non-Detachable |
| --- | --- | --- | --- |
| Injection Port | Present | Present | Present |
Premarket Notification, Lithotripter-Compatible Basket, December 9, 1996
Proprietary and Confidential Information of Boston Scientific Corporation
59
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.