← Product Code [KNS](/submissions/GU/subpart-e%E2%80%94surgical-devices/KNS) · K232476

# Disposable Sphincterotome (K232476)

_Zhejiang Soudon Medical Technology Co., Ltd. · KNS · Nov 7, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/KNS/K232476

## Device Facts

- **Applicant:** Zhejiang Soudon Medical Technology Co., Ltd.
- **Product Code:** [KNS](/submissions/GU/subpart-e%E2%80%94surgical-devices/KNS.md)
- **Decision Date:** Nov 7, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.4300
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Disposable Sphincterotome is indicated for use in the cannulation of the biliary ducts and the transendoscopic sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi. The device is supplied sterile and intended for single use only.

## Device Story

Disposable Sphincterotome; sterile, single-use endoscopic accessory. Device consists of long plastic tube with internal wire; distal end features exposed wire segment. Used via endoscope working channel; physician-operated. Principle of operation: monopolar high-frequency current delivered through wire to tissue at duodenal papilla; thermal effect dehydrates/proteinizes tissue to perform incision. Models with injection capability allow delivery of contrast medium into biliary ducts. Output is physical incision of tissue; enables selective cannulation of biliary or pancreatic orifices. Benefits patient by facilitating minimally invasive access to biliary/pancreatic systems for diagnostic or therapeutic procedures.

## Clinical Evidence

No clinical data. Bench testing only, including appearance and physical properties testing. Biocompatibility testing performed per ISO 10993-1; sterilization validation per ISO 11135-1 and ISO 10993-7.

## Technological Characteristics

Materials: PTFE, ABS, SUS304. Energy: Monopolar radio frequency current. Form factor: Long tube with internal wire, 2000mm working length, 2.4mm minimum accessory channel. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Connectivity: None (mechanical/electrosurgical accessory).

## Regulatory Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

## Predicate Devices

- Sphincterotome, [K201121](/device/K201121.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 7, 2023

Zhejiang Soudon Medical Technology Co.,Ltd % Nick Wang RA Specialist Shanghai Vanhe Consulting Co., Ltd Shanghai, 201703 CHINA

Re: K232476

> Trade/Device Name: Disposable Sphincterotome Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: II Product Code: KNS Dated: July 18, 2023 Received: August 16, 2023

Dear Nick Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Glenn B. Bell -S

Glenn B. Bell, Ph.D. Director DHT3A: Division of Renal,Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known) K232476

Device Name Disposable Sphincterotome

Indications for Use (Describe)

The Disposable Sphincterotome is indicated for use in the biliary ducts and the transendoscopic sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi. The device is supplied sterile and intended for single use only.

| <div> <span> <b>                         ☑                     </b>                     Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span>                     ☐ Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|

 CONTINUE ON A SEPARATE PAGE IF NEEDED.

 This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** 

 The burden time for this collection of information is estimated to average 79 hours per response, including the
 time to review instructions, search existing data sources, gather and maintain the data needed and complete
 and review the collection of information. Send comments regarding this burden estimate or any other aspect
 of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services
   

 Food and Drug Administration
   

 Office of Chief Information Officer
   

 Paperwork Reduction Act (PRA) Staff
   

 PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Type of Use (Select one or both, as applicable)

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{4}------------------------------------------------

## 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

## 1. Submitter

| Submitted by:   | Zhejiang Soudon Medical Technology Co.,Ltd<br>Address:Room 302-1,Floor 3, Building 4, No.1 Nangonghe<br>Road, Donghu Street, Linping District, Hangzhou 311100<br>Zhejiang, China                                           |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Nick Wang<br>RA Specialist<br><br>Shanghai Vanhe Consulting Co., Ltd<br>Address: 2F, Building No.1, 3938 Huqingping Road, Qingpu<br>District, Shanghai, China.<br>Phone: 0086-13585860297<br>Email: vanheconsulting@126.com |
| Date Prepared:  | July 24, 2023                                                                                                                                                                                                               |

## Device

| Device Name:         | Disposable Sphincterotome                                       |
|----------------------|-----------------------------------------------------------------|
| Classification Name: | unit, electrosurgical, endoscopic (with or without accessories) |
| Regulatory Class:    | II                                                              |
| Regulation Number:   | 21 CFR 876.4300                                                 |
| Regulation Name:     | Endoscopic electrosurgical unit and accessories                 |
| Product Code:        | KNS                                                             |

## Predicate Device

| Device Name:         | Sphincterotome, K201121                                         |
|----------------------|-----------------------------------------------------------------|
| Manufacturer:        | Hangzhou AGS MedTech Co., Ltd.                                  |
| Classification Name: | unit, electrosurgical, endoscopic (with or without accessories) |
| Regulatory Class:    | II                                                              |
| Regulation Number:   | 21 CFR 876.4300                                                 |
| Regulation Name:     | Endoscopic electrosurgical unit and accessories                 |
| Product Code:        | KNS                                                             |

## 2. Device Description

The Disposable Sphincterotome described in this submission are a sterile, single use

{5}------------------------------------------------

devices compatible with the working channel of endoscope. The device consists of a long plastic tube with a wire running the length of its interior. A small portion of that wire is exposed at its distal end. The roof of the papilla is opened by passing high-frequency current through the wire, exposing the biliary or pancreatic orifices for selective cannulation.

The models with injection can inject contrast medium into the biliary ducts. The main materials of the proposed device include PTFE, ABS, SUS304. Type A, Type B and Type C are included in this submission.

The proposed devices are EO sterilized to achieve the Sterility Assurance Level (SAL) of 106 and placed in a sterility maintenance package to ensure a shelf life of 3 years.

## 3. Indication for Use:

The Disposable Sphincterotome is indicated for use in the cannulation of the biliary ducts and the transendoscopic sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi. The device is supplied sterile and intended for single use only.

## 4. Comparison of Technological Characteristics

The Disposable Sphincterotome has substantially equivalent device design, configuration, packaging fundamental technology, sterilization process and intended use as those featured in the predicate device Hangzhou AGS MedTech Co., Ltd.'s Sphincterotome, K201121. The differences between the proposed device and the predicate devices do not raise any questions regarding its safety and effectiveness. The differences are listed in the table below.

| Item                            | Disposable<br>Sphincterotome(Proposed<br>Device)                                                                                                                                                                                                                                                                                              | Sphincterotome,<br>K201121                                                                                                                                                                                                                                                                                                                                               | Discussion |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indication for<br>Use           | The Disposable<br>Sphincterotome is<br>indicated for use in the<br>cannulation of the biliary<br>ducts and the<br>transendoscopic<br>sphincterotomy of the<br>Papilla of Vater and/or the<br>Sphincter of Oddi. The<br>device is supplied sterile<br>and intended for single use<br>only.                                                     | The Sphincterotome is<br>indicated for use in the<br>cannulation of the<br>biliary ducts and the<br>transendoscopic<br>sphincterotomy of the<br>Papilla of Vater and/or<br>the Sphincter of Oddi.<br>The device is supplied<br>sterile and intended for<br>single use only.                                                                                              | Same       |
| Product Code                    | KNS                                                                                                                                                                                                                                                                                                                                           | KNS                                                                                                                                                                                                                                                                                                                                                                      |            |
| Regulation<br>Number            | 878.4300                                                                                                                                                                                                                                                                                                                                      | 878.4300                                                                                                                                                                                                                                                                                                                                                                 | Same       |
| Class                           | 2                                                                                                                                                                                                                                                                                                                                             | 2                                                                                                                                                                                                                                                                                                                                                                        | Same       |
| Supplied in                     | Yes                                                                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                                                                                                      | Same       |
| Item                            | Disposable<br>Sphincterotome(Proposed<br>Device)                                                                                                                                                                                                                                                                                              | Sphincterotome,<br>K201121                                                                                                                                                                                                                                                                                                                                               | Discussion |
| Sterile                         |                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                          |            |
| Principles of<br>Operation      | The product reaches the<br>duodenal papilla through<br>the working channel of the<br>duodenoscope, and the<br>cutting wire of the product<br>uses the thermal effect<br>produced by<br>high-frequency current to<br>dehydrate and proteinize<br>the tissue at the duodenal<br>papilla to achieve the<br>purpose of incision of the<br>tissue. | Sphincterotome<br>manufactured is an<br>applied part of<br>electrosurgical<br>generator,<br>using monopolar high<br>frequency current<br>delivered by the<br>electrosurgical<br>generator<br>for sphincterotomy<br>with<br>the electrode. The high<br>frequency electricity<br>flows from the active<br>electrode to the neutral<br>electrode placed on<br>patient skin. | Same       |
| Injection                       | Contrast Medium                                                                                                                                                                                                                                                                                                                               | Contrast Medium                                                                                                                                                                                                                                                                                                                                                          | Same       |
| Wire Guide<br>Diameter          | 0.89mm                                                                                                                                                                                                                                                                                                                                        | 0.63mm, 0.89mm                                                                                                                                                                                                                                                                                                                                                           | Similar    |
| Minimum<br>Accessory<br>Channel | 2.4mm                                                                                                                                                                                                                                                                                                                                         | 2.8mm                                                                                                                                                                                                                                                                                                                                                                    | Similar    |
| Working<br>Length               | 2000mm                                                                                                                                                                                                                                                                                                                                        | 1800mm                                                                                                                                                                                                                                                                                                                                                                   | Similar    |
| Energy<br>Used/Delivered        | Monopolar Radio<br>Frequency Current                                                                                                                                                                                                                                                                                                          | Monopolar Radio<br>Frequency Current                                                                                                                                                                                                                                                                                                                                     | Same       |
| Packaging                       | Single-use EO sterilized<br>pouch with one device per<br>pouch                                                                                                                                                                                                                                                                                | Single-use EO<br>sterilized pouch with<br>one device per pouch                                                                                                                                                                                                                                                                                                           | Same       |
| SAL                             | 10-6                                                                                                                                                                                                                                                                                                                                          | 10-6                                                                                                                                                                                                                                                                                                                                                                     | Same       |
| Sterilization<br>Method         | EO Sterilization                                                                                                                                                                                                                                                                                                                              | EO Sterilization                                                                                                                                                                                                                                                                                                                                                         | Same       |
| Biocompatibility                | Conform to ISO 10993-1                                                                                                                                                                                                                                                                                                                        | Conform to ISO<br>10993-1                                                                                                                                                                                                                                                                                                                                                | Same       |

{6}------------------------------------------------

## 5. Non-clinical Performance Data

{7}------------------------------------------------

The proposed device meets the requirements of ISO 10993 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing", ISO 11135-1 "Sterilization of Health Care products Ethylene Oxide - Part 1: Requirements for Development, Validation, and Routine Control of Sterilization processes for Medical Devices", and ISO 10993-7 "Biological evaluation of medical devices - Part 7: ethylene oxide sterilization residuals".

The following bench tests were performed on Disposable Sphincterotome: Appearance, Physical properties. The results of all testing were passing.

## 6. Clinical Test Data

No Clinical Study is included in this submission.

## 7. Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, Based on the information provided in this premarket notification, Zhejiang Soudon Medical Technology Co.,Ltd has demonstrated that proposed device Disposable Sphincterotome is substantially equivalent to Hangzhou AGS MedTech Co., Ltd.'s currently marketed Sphincterotome, K201121.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/KNS/K232476](https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/KNS/K232476)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).